Radiometric calibration of Michelson interferometric spectrophometers for the use in atmospheric sounders has been extensively studied, but its practical demonstration is not well reported. This paper describes one such demonstration for the InSb spectral region. The spectral resolution was 2 cm-1, and specific attention is given to the radiometric correctness of the fine scale spectral information. This work is paticularly applicable to the High-Resolution Interferometer Sounder (HIS) experiment, where the interferogram is used directly as the data element.
© (1983) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
D. D. LaPorte and R. Howitt "Ambient Temperature Absolute Radiometry Using Fourier Transform Spectrometers", Proc. SPIE 0364, Technologies of Cryogenically Cooled Sensors and Fourier Transform Spectrometers II, (16 August 1983); https://doi.org/10.1117/12.934175