17 January 2002 Intercomparison of UV-B radiometer
Shu Takeshita, Yukiharu Miyake, Masamitsu Ichihashi, Masako Sasaki
Author Affiliations +
Abstract
To evaluate regional differences of global ultraviolet-B (UVB: 290-315 nm) irradiance, eight band-spectral UVB radiometers (MS-210W, EKO), which were used in the UV monitoring network organized by five universities, were compared at the Shonan Campus of Tokai University, Japan from August 2-10, 1997. Reference irradiances were measured by a spectroradiometer (OL752, Optronic Lab.) and the reference band-spectral UVB radiometer (MS-210W, EKO). UVB radiometers participated were classified into three groups by the characteristics of the spectral response depending on the UVB irradiance. It suggests that characteristics of out of band is different from the reference radiometer. Furthermore, to access regional difference of UVB, it is necessary to use the calibration factor that is not constant but a function depending on output.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Shu Takeshita, Yukiharu Miyake, Masamitsu Ichihashi, and Masako Sasaki "Intercomparison of UV-B radiometer", Proc. SPIE 4482, Ultraviolet Ground- and Space-based Measurements, Models, and Effects, (17 January 2002); https://doi.org/10.1117/12.452949