5 September 2019 Absolute distance measurement with sub-fringe resolution
Author Affiliations +
Proceedings Volume 10565, International Conference on Space Optics — ICSO 2010; 105655W (2019) https://doi.org/10.1117/12.2552636
Event: International Conference on Space Optics—ICSO 2010, 2010, Rhodes Island, Greece
Abstract
A novel technique to measure absolute distances is presented. It is based on a Michelson interferometer where two tuneable lasers are superposed to create a synthetic wavelength. Relative and absolute interferometry theories are merged together. Its experimental realization allows absolute distance measurements with subfringe resolution. Preliminary results are presented in this work.

INTRODUCTION

Classical interferometers (mainly Michelson like) are relative interferometers and are known to be the best distance measurement instruments but suffer from the fact that are “incremental” instruments: i.e. since all optical fringes are identical, the only way to perform an “absolute” measurement is to count fringes while the target moves along the length to be measured. The issue of realizing an instrument capable of absolute distance measurement with low uncertainty has been afforded for several decades. Many technical solutions have been found, some of which have became popular commercial instruments. The two main approaches are the “time of flight” (TOF) and the “synthetic wavelength” (SW) methods, although most of the recently proposed solutions can be seen as a combination of the two [1]. TOF consists in sending a light pulse on a reflecting target and measuring time needed to the pulse to return back to the source. This method has been used since long for measuring very large distances (the most famous being the Heart to Moon distance) [2]. SW is basically a Michelson interferometer where the laser source is somehow modulated (amplitude, frequency, polarization etc.) with a given periodic signal and where the phase lag between the outgoing and the incoming modulating signal is measured. In other words, the system can be seen as a classical interferometer where the laser radiation acts as a carrier for an artificial wavelength (the synthetic wavelength) which in turn is used as the interferometer base wavelength. The difference with respect to classical interferometers is that the SW can be made as long as needed to solve the indetermination of integer number of wavelengths by analyzing the relationship between the SW phase and frequency. In practice the modulation of the carrier can be done in many different ways [3], [4], amongst which is the superimposition of two similar lasers [5], [6] and [7]. This technique consists in superimposing two lasers having same nominal wavelength where at least one of the two is tunable. In such a way, according to the interference theory, an amplitude modulated wave is obtained and the wavelength of the modulation can be changed from infinite (the two lasers having the same frequency) to a minimum limited by the maximum laser frequency detuning. With duplicated Nd:YAG lasers, it is possible to obtain tens of gigahertz detuning corresponding to few millimetres SW. Although SW interferometers allow absolute measurements, the resolution of these instruments is at present limited to at best a few micrometers: much worst of what can be obtained with relative incremental interferometry. This is mainly due to the scale difference between synthetic wavelength (tens of millimetres) and the “optical” wavelength (fractions of a micrometer). The realization of an absolute interferometer with nanometer resolution is a challenging task. A possible breakthrough could be the reduction of the absolute interferometer uncertainty down to less than an optical wavelength so that the optical fringes can be unambiguously numbered and the interferometer can be “switched” to the classical configuration eventually reaching nanometer resolution. In the present work we propose a simple interferometric set-up which allows to reach the aforementioned goal and experimentally demonstrate the feasibility of an absolute interferometer with subwavelength resolution.

PRINCIPLE OF THE EXPERIMENT

The experiment is based on the well known principle of feeding a Michelson interferometer with an amplitude modulated laser source made by superimposing two tunable lasers. If the two laser sources have equal intensity, the resulting radiation can be seen as a carrier having an optical frequency equal to the average of the two laser frequencies and modulated with a cosine function having frequency equal to the difference between the two laser frequencies.

Let us consider a simple Michelson interferometer with a 50:50 beamsplitter, two arms having length xm and xr and a single detector. The two laser sources have respectively amplitude E1 and E2 and wavenumber k1 and k2.

After the interferometer, we have four distinct fields: E1mcos(2k1xm), E1rcos(2k1xr), E2mcos(2k2xm) and E2rcos(2k2xr) that interfere to give a signal on the receiver limited in bandwidth:

00014_PSISDG10565_105655W_page_3_1.jpg

In the particular case the amplitudes are equal, the interferogram signal simplifies to:

00014_PSISDG10565_105655W_page_3_2.jpg

The resulting interferogram can be modelled as the product of the optical fringes (having frequency equal to the average of the two laser frequencies) and the synthetic fringe (having frequency equal to the difference between the two laser frequencies). The optical fringe is modulated by the synthetic wave and, every half a period, the latter reaches a zero amplitude point; at any of those points, the envelope of the optical fringes gets across the line of the average value of (2).

We can see the behaviour of the interferogram, by a numerical simulation, in Fig. 1:

Fig. 1.

Two laser beams theoretical interferogram

00014_PSISDG10565_105655W_page_3_3.jpg

Fig. 2 shows a detail of the interferogram close to the minimum amplitude region. On the left is the situation described in (2) with matched laser intensities. When the power of the two lasers is not balanced, we are in the case of a classical dual laser interferometer where the inversion point is not so clearly defined: in the right side we can see the simulated effect of a 0.1% power imbalance.

Fig. 2.

Detail of a two laser beams interferogram around a zero amplitude point. In the left, the power of the two laser power is equal in the two arms, in the right, the power of the two laser differs by a 0.1%

00014_PSISDG10565_105655W_page_3_4.jpg

When the power of the two lasers is well matched, it is easy to locate the inversion point with an uncertainty less than an optical fringe, so is possible to number the optical fringes contained in a synthetic fringe thus assigning an “absolute” value to the optical interferometer. This is the key point of the presented experiment.

EXPERIMENTAL SETUP

Let’s refer to the experimental setup depicted in Fig. 3.

Fig. 3.

Simplified experimental setup of the interferometer

00014_PSISDG10565_105655W_page_4_1.jpg

Two single-mode doubled Nd:YAG lasers emit green light at around λ = 532.4 nm. Laser “a” is frequency locked to the a10, R(56) 32-0, 127I2 transition line, at V1 = 563 260 223 513 kHz [8]. Laser “b” frequency can be tuned. The frequency difference is measured by beating the infrared output of the two lasers. Laser “b” is frequency controlled to keep the frequency difference constant. The beam power of laser “a” is trimmed by means of a half wave plate together with a polarizer. The two lasers are superposed with a beam splitter and coupled to a polarization maintaining fiber. A polarizer serves to avoid unwanted polarization to be coupled to the fiber. The output coupler of the fiber feeds the Michelson interferometer. M1 is a fixed plane mirror, M2 is the moving plane mirror which is placed on a piezocapacitive translation stage with a range of 25 μm, in turn, fixed on a motorized translation stage having 100 mm range. The interference signal is detected with a photodiode and sent to the acquisition system (a 16 bit 1 MHz sampling board).

PRELIMINARY EXPERIMENTAL RESULTS

The preliminary results shown here have been obtained with the following procedure. The motorized stage is moved in order to find a minimum amplitude point in the interferogram. The mirror is then moved back and forth by means of the piezo stage in order to see on the oscilloscope the shape of the minimum amplitude region as in figure 2. Here the power of the laser “a” is manually trimmed to obtain a good intensity balance by rotating the half wave plate. The fine power balancing is obtained by acting on the temperature control of the doubling crystal of laser “b”. Once the condition of laser power matching is reached, the piezo scan is stopped and the translation stage is moved while the interferogram is recorded. The frequency difference is recorded as well. The frequency difference ranges from 40 to 60 GHz corresponding to a distance between the minima of about 15 mm. Fig. 4 shows the two interferograms obtained while scanning the moving mirror at about 2 mm/s for two frequency differences.

Fig. 4.

Left: the two interferograms obtained while scanning the moving mirror at about 2 mm/s for two laser frequency differences. Right: position of the minima for two frequency differences. The zero optical path difference condition can be found around mm 78.

00014_PSISDG10565_105655W_page_4_2.jpg

Figure 5 shows a typical interferogram at the minimum amplitude region together with the fitting function (2). It is evident that the “inversion point” foreseen in (2) can be found and located with an uncertainty much less than one optical fringe. At present the experiment is made particularly difficult because of the non uniformity of the motorized translation stage which induces strong vibrations along motion axis. Indeed, in order to correctly fit the interferogram with function (2) it is necessary to have a constant speed scan. Improvements in the translation stage will be next step. Then a comparison with a reference laser interferometer will allow to complete the validation of the method.

Fig. 5.

Typical interferogram at the minimum amplitude region together with the fitting function (2)

00014_PSISDG10565_105655W_page_5_1.jpg

CONCLUSIONS

We have proposed a simple dual laser interferometer capable of absolute distance measurements with sub wavelength resolution. Although we have not performed absolute distance measurements yet, we have demonstrated the feasibility of the most critical point of the experiment: i.e. that it is possible to find the position of the synthetic wavelength (namely the minima) with an uncertainty better than an optical fringe. This possibility combined with an accurate knowledge of the absolute frequency of the two lasers will allow the accurate knowledge of the absolute optical path difference between the two mirrors.

REFERENCES

[1] 

J. Ye, “Absolute measurement of a long, arbitrary distance to less than an optical fringe,” Opt. Lett., 29 1153 –1155 (2004). https://doi.org/10.1364/OL.29.001153 Google Scholar

[2] 

J. O. Dickey, P. L. Bender, J. E. Faller, X. X. Newhall, R. L. Ricklefs, J. G. Ries, P. J. Shelus, C. Veillet, A. L. Whipple, J. R. Wiant, J. G. Williams, and C. F. Yoder, “Lunar laser ranging: a continuing legacy of the Apollo program,” Science, 265 482 –490 (1994). https://doi.org/10.1126/science.265.5171.482 Google Scholar

[3] 

I. Fujima, S. Iwasaki, and K. Seta, “High-resolution distance meter using optical intensity modulation at 28 GHz,” Measurement Science and Technology, 9 1049 –1052 (1998). https://doi.org/10.1088/0957-0233/9/7/007 Google Scholar

[4] 

K. Minoshima, and H. Matsumoto, “High-Accuracy Measurement of 240-m Distance in an Optical Tunnel by use of a Compact Femtosecond Laser,” Applied Optics, 39 5512 –5517 (2000). https://doi.org/10.1364/AO.39.005512 Google Scholar

[5] 

R. Dandliker, R. Thalmann, and D. Prongué, “Two-wavelength laser interferometry using superheterodyne detection,” Optics Letters, 13 339 –341 (1988). https://doi.org/10.1364/OL.13.000339 Google Scholar

[6] 

E. Gelmini, U. Minoni, and F. Docchio, “A tunable, double wavelength heterodyne detection interferometer with a frequency-locked diode-pumped Nd:YAG sources for absolute measurements,” Review of Scientific Instruments, 66 4073 –4080 (1995). https://doi.org/10.1063/1.1145419 Google Scholar

[7] 

V. Mahal, and A. Arie, “Distance measurements using two frequency-stabilized Nd:YAG lasers,” Applied Optics, 35 3010 –3015 (1996). https://doi.org/10.1364/AO.35.003010 Google Scholar

[8] 

T.J. Quinn, “Practical realization of the definition of the metre, including recommended radiations of other optical frequency standards,” Metrologia, 40 103 –133 (2003). https://doi.org/10.1088/0026-1394/40/2/316 Google Scholar
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
N. Antonietti, A. Giugni, and M. Pisani "Absolute distance measurement with sub-fringe resolution", Proc. SPIE 10565, International Conference on Space Optics — ICSO 2010, 105655W (5 September 2019); https://doi.org/10.1117/12.2552636