1INTRODUCTIONThe LISA mission is a joint ESA-NASA spaceborne project, aiming at detecting gravitational waves in the frequency range 10–4 – 1 Hz [1]. It consists of 3 spacecrafts in a nearly equilateral configuration orbiting around the sun, about 20 degrees behind the Earth. The satellites are separated by 5 × 106 km, constantly following free-flying masses located at their center. On each spacecraft, two laser beams are emitted towards the other satellites, resulting in six laser links. These interferometric measurements are used to precisely monitor the distance between the inertial masses and, hence, to detect the tiny variation due to the pass of a gravitational wave. The goal of LISA is to detect gravitational deformation as small as Nevertheless, this stability can be achieved for LISA thanks to three successive stabilization stages:
2IODINE STABILIZATION FOR LISAAs mentioned above, the pre-stabilization technique considered for LISA is based on a fixed-length, ultra stable optical resonator. This technique proved to meet the LISA requirement in terms of intrinsic frequency stability [2] by means of the well-known Pound-Drever-Hall stabilization method. Its tunability had been recently demonstrated using an additional electro-optic modulator and side band locking. Nevertheless, some limitations can be identified for this technique. First, the performance of the FP cavity is very sensitive to mechanical and mainly thermal disturbance. For example, the thermal stability of the tanks housing the cavities must of the order of 10 µK/ Locking to a molecular transition (hyperfine resonance line) can circumvent these issues. It offers an absolute frequency reference. The knowledge of the Doppler shift is therefore directly measured from the beat frequency. With a reasonably simple setup, the repeatability of the locking frequency has a long term drift of about a few kHz or less Therefore, a beat frequency well within the detector bandwidth will immediately be obtained while mixing two molecular stabilized lasers. The position of a molecular transition is also much less sensitive to thermal perturbations than an cavity, the constraints being typically relaxed by about 3 order of magnitudes. LISA makes use of Nd:YAG, NPRO lasers, known for their good intrinsic power and frequency stability. Unfortunately, there is no known strong molecular resonance around their wavelength of 1064 nm. However, iodine provides strong, narrow hyperfine transitions around 532 nm (2nd harmonic of Nd:YAG laser), that are often used on ground as frequency or length reference (e.g. see [7]). Previous experiments using iodine stabilization for LISA have already been done [8 9] and showed promising results. The goal of our experiment is to demonstrate that this technique can be used for space applications (i.e. robust, simple and compact) and meet the LISA requirements of 3EXPERIMENT AT THE APC3.1General layoutResonance transitions are usually broadened by the Doppler effect due to the thermal movements of the molecules. In order to get rid of this effect and lock on an hyperfine (roto-vibrational) line, two beams counter propagates within the iodine cell. One of this beam (the pump) is strong enough to ‘saturate’ the line (i.e. a large part of the molecules on its path are in a excited state). The other beam (the probe), of much weaker intensity, is used to scan the absorption profile. The molecules that have already been excited by the pump beam cannot absorb the probe and then cause a dip in the absorption profile [10]. The two beams being of the same wavelength (or shifted by a constant value see below), only molecules with a given velocity (projected on the light path) are simultaneously saturated by the pump and scanned by the probe beams. This technique is a standard method of Doppler-free spectroscopy and is used in our experiment. The general layout of the experiment is shown on fig. 1. Since there is no better frequency reference available in the laboratory, two identical systems have to be built, and their joint performance (i.e. the stability of the beat between the two infrared beams) measured. About 900 mW of infrared light is emitted from the laser source (Innolight Mephisto 1000). Its frequency is then doubled by passing through a non-linear crystal of PP:MgLNO (3 × 0.5 × 20 mm). This material had been chosen because of its good efficiency and it can stand high power density and space radiations [9]. These crystals are put in a oven and their temperature stabilized at 64.9 °C (system 1, 66.2 °C for system 2). For an input power of about 830 mW, about 10 mW of green light (at 532 nm) is produced. A dichroic mirror is then used to separate the two wavelength. The frequency noise of the beat between the infrared parts of the two systems is then assumed to be the sum of the frequency noises (quadratic sum) of each of them. The green beam is divided into a pump (about 5 mW) beam and a probe (about 0.5 mW) beam. The method used here is the modulation transfer spectroscopy (MTS), which already demonstrated very good results on similar experiments [11]. The pump beam is frequency shifted and modulated by an acousto-optic modulator. The modulation frequency is at 80 kHz with a constant frequency shift of 80 and 110 MHz. As the central frequency of the probe beam differs from the one of the pump beam, the position of the locking position is translated by half of the offset (i.e. 40 and 55 MHz). Since the carrier frequency of an AOM can be tuned up to about 20 % of its central value, the locked frequency could be tuned to accommodate the arm-locking requirements. By four waves mixing [12], the modulation is transfered on the probe beam. The carrier-side bands interference produce a beat on the probe beam whose amplitude is roughly proportional to the derivative of line profile and can then be used as an error signal for the feedback electronics. One of the main advantages of the MTS technique is that it is insensitive, at first order, to the residual amplitude noise of the pump (as long as it ‘saturates’ the molecules) that could be induced by the modulation process. We chose to lock on the a10 R(56)-32-0 line [13], which is intrinsically narrow and sufficiently isolated from other lines. The iodine cells are 20 cm long, 25 mm in diameter and have been provided by the BIPM (Bureau International des Poids et Mesures). The beam is folded three times in the cell, increasing the effective interaction length to 60 cm. T he cold finger of the cell is stabilized at a temperature of -15.1 °C (-15.7 °C) with a precision better than ± 0.01°C (see section 3.2.2). The iodine pressure is then regulated at 0.75 Pa (0.80 Pa) [14]. After demodulation with a lock-in amplifier, the error signal is fed through the feedback electronics, acting on the temperature and piezo-electric actuator of the laser crystal. The temperature is used for correction on long timescales (corner frequency at about 70 mHz). The piezo-electric actuator handles higher frequencies, up to about 1 kHz. 3.2Results3.2.1Frequency stabilityAssuming that the two stabilization setups have the same performance, the power spectrum of the infrared beat is twice the power spectrum of one system. The linear spectral density for a single laser is plotted on fig. 2. Above 70 mHz, the frequency noise falls as 1/f0.5 (1/f in terms of power), being at the level of about 15 Hz/ This setup already demonstrates good frequency stability, although about 3 times worse than the most conservative stability requirements for LISA. Taking advantages of the improved arm-locking techniques, the present stabilization is nevertheless an possible alternative to the FP cavity foreseen for LISA. Fig. 3 gives a typical time evolution of the beat frequency. The measured drift is of the order of 300 Hz over 9 hours. 3.2.2Temperature stability of the elementsIn the mHz region, one of the most prominent disturbance is due to thermal fluctuations. Two components requires a dedicated, efficient thermal control: the cold finger of the iodine cell and the doubling crystal. Within the cell, the vapor is in equilibrium with a solid grain of iodine, i.e. its pressure is equal to the normal vapor pressure of iodine, only depending on the temperature of solid iodine. These ‘saturated’ cells are used to compensate any loss of iodine vapor within the cell (e.g. due to adsorption on the glass). Therefore, the temperature of the cold finger must be tightly controlled. The linear variation of pressure as a function of temperature [14] is about 0.09 Pa/K at -15°C. In addition to the modification of linear absorption within the cell, a pressure change induces a shift of the position of the transition line; typical values of 3 kHz/Pa have been reported [11, 7]. Finally, we can estimate a shift of 278 Hz/K due to temperature fluctuations of the cold finger, our goal of 30 Hz/ Figure 4:Temperature stabilities of the cell cold finger and doubling crystal (improved feedback electronics). ![]() As explained in the next section, a good thermal stability was also needed for the non linear crystals. Fig. 4 (red curve) shows the typical achieved stability, almost limited by the instrument resolution. 3.2.3Effect of the doubling crystal on the frequency stabilityThe second harmonic generation (frequency doubling) is achieved by passing through a PP:MgLNO (non-linear) crystal. The (quasi-)phase matching requires a constant temperature of the crystal in order to match the length of the periodically poled regions with the laser wavelength. The doubling efficiency curve (as a function of temperature) peaks at the phase-matching temperature, with a width of a few tenths of degrees. Any temperature fluctuation, will therefore induce intensity fluctuations of the pump and probe beams. The reference frequency of our system was found to be sensitive to these power perturbations. Very different shifts (suggesting a role of the wavefront curvature) have already been reported in the literature, up to 2 kHz/mW [11]. Consequently, the temperature of the doubling crystals must be stabilized, at the optimum efficiency temperature, in order to minimize the influence of crystal temperature on the frequency stability. We first used a commercial temperature controller for the crystal oven. Unfortunately, strong temperature fluctuations, correlated with power stability of the green beam have been measured (see fig. 5). A dedicated temperature controller was then developed, whose performance are shown on fig. 4. With this improved controller, no effect of the fluctuations of the doubling crystal temperature could be measured on the frequency stability of the laser. 4FUTURE DEVELOPMENT: IODINE CELL IN A FP CAVITYThe best, long term, stabilization performance obtained with iodine, are of the order of a few 10–14 of relative frequency stability, but required very long interaction length [11, 15]. These systems are also quite sensitive to alignment and residual intensity noise (e.g. coming from fluctuations of the doubling efficiency). However, in order to be attractive for space application, any system would have to be compact and robust. Consequently, with collaborators from SYRTE (Obs. de Paris) and ARTEMIS (Obs. de la Côte d’Azur), we are currently designing a new stabilization technique based on iodine in a low finesse resonant cavity. Using a Fabry-Perot cavity increases the effective interaction length, while the geometry and curvature of the beams are fixed. This technique has already been used at others wavelength (and molecules) and demonstrated very good stability results [16, 17]. Experimentally, enclosing a molecular reference inside a Fabry-Perot cavity, seems to have been first demonstrated by Brillet et al. in the early 80’s [18]. Its possible application to gravity space missions was also underlined by Bender et al. [19]. Fig. 6 gives the general layout of the foreseen experimental setup. The whole system (excluding the laser source) will fit in a 30×50 cm2 box. As in the ‘standard’ experiment, the laser source is an Nd:YAG laser emitting at 1064 nm. The beam is then frequency doubled by passing through a non linear, PP:MgLNO crystal. The first AOM (acousto-optic modulator) is used to stabilize the intensity of the green beam. It can also be used to offset the reference frequency (of the green beam w.r.t the infrared one), and therefore is compatible with a slight tuning of the stabilization frequency, as needed by the arm-locking technique. The core of the system is based on a 10 cm long iodine cell (12 mm in diameter), with windows at Brewster angle to minimize losses. The cold finger of the cell will be maintained at -18°C (±1 mK), for which the width of the iodine line is estimated to be about 500 kHz. The cell will also be magnetically and thermally shielded and put into a small vacuum chamber (20×30 cm). The iodine is enclosed in a resonant cavity. The iodine absorption drives the losses of the cavity and consequently its low finesse (34). Two modulation stages of the pump are needed: one used to stabilize the laser frequency, the other one to control the length of the cavity. The first EOM (electro-optic modulator) has a modulation frequency equal to the free spectral rage of the cavity (540 MHz), according to the NICE-OHMS (Noise Immune Cavity-Enhanced Optical Heterodyne Modulation Spectroscopy) technique [17]. In that way, both the carrier and the side bands are transmitted by the cavity and their beat signal is used for frequency stabilization of the source. The second EOM provides a modulation at 80 MHz. The length of the cavity is then tuned so that it is resonant with the laser frequency, according to the Pound-Drever-Hall technique [20]. According to previous experiments and our own calculation, the achievable frequency stability could be as low as 2 Hz/ 5CONCLUSIONIodine laser stabilization had been investigated in the context of LISA. The standard stabilization scheme does not yet meet the most stringent requirements for LISA (30 Hz/ Nevertheless, it would be of great interest to relax the constraints of the second and third stages of stabilization and rely on a single, excellent absolute frequency reference. For this reason, our team developed a new concept of stabilization based on a iodine cell within a low finesse cavity. This setup, expected to achieve stability of the or der of 2 Hz/ ACKNOWLEDGMENTSWe acknowledge the French Space Agency (CNES) for its technical and financial support of this work, under contract R-S06/SU-0001-012. References,Danzmann K.,
“LISA Mission Overview,”
Adv. Space Res, 25 1129
(2000). Google Scholar
LISA Technology Status Report - DRAFT, ESA-NASA,
(2006). Google Scholar
Sheard B.S. et al.,
“Laser Frequency Stabilization by Locking to a LISA Arm,”
Physics Letters A, 320 9
(2003). Google Scholar
Tinto M. and Rakhmanov M.,
“On the Laser Frequency Stabilization by Locking to a LISA Arm,”
gr-qc/0408076v1,
(2007). Google Scholar
Sylvestre J.,
“Simulations of Laser Locking to a LISA Arm,”
Phys. Rev. D, 70 102002
(2004). Google Scholar
Tinto M. et al.,
“Implementation of Time-Delay Interferometry for LISA,”
Phys. Rev. D, 67 122003
(2003). Google Scholar
Picard S. et al.,
“Results from International Comparisons at the BIPM Providing a World-Wide Reference Network of 127I2 Stabilized Frequency-Doubled Nd:YAG Lasers,”
IEEE Trans. on Instr. and Meas, 52 236
(2003). Google Scholar
Mondin L.,
“Stabilisation de Fréquence de Laser Nd:YAG pour Applications Spatiales,”
Thse de l’Universit de Nice-Sophia Antipolis,
(2005). Google Scholar
Leonhardt V. and Camp J.B.,
“Space Interferometry Application of Laser Frequency Stabilization with Molecular Iodine,”
Applied Optics, 45 4142
(2006). Google Scholar
Demtröder W., Laser Spectroscopy, 3Springer-Verlag, Berlin Heidelberg
(2003). Google Scholar
Hall J.L. et al.,
“Stabilization and Frequency Measurement of the I2-Stabilized Nd:YAG Laser,”
IEEE Trans. on Instr. and Meas, 48 583
(1999). Google Scholar
Raj R.K. et al.,
“High-Frequency Optically Heterodyned Saturation Spectroscopy via Resonant Degenerate Four-Wave Mixing,”
Physical Review Letters, 44 1251
(1980). Google Scholar
Ye J. et al,
“Absolute Frequency Atlas of Molecular I2 Lines at 532 nm,”
IEEE Trans. on Instr. and Meas, 48 544
(1999). Google Scholar
Gillepsie L.J. and Fraser L.H.D.,
“Normal Vapor Pressure of Crystalline Iodine,”
Journal Of the American Chemical Society, 58 2260
(1936). Google Scholar
Zang E.J. et al.,
“ Realization of Four-Pass I2 Absorption Cell in 532-nm Optical Frequency Standard,”
IEEE Trans. on Instr. and Meas, 56 673
(2007). Google Scholar
Clairon A. et al.,
“State-of-the-Art for High Accuracy Frequency Standards in the 28 THz Range Using Saturated Absorption Resonances of OsO4 and CO2,”
Frequency Standards and Metrology, 212 Springer-Verlag, Heidelberg
(1989). Google Scholar
Sheng M.L. and Hall J.L.,
“Optical Heterodyne Spectroscopy Enhanced by an External Cavity: Toward Improved Working Standards,”
IEEE Journal of Quantum Electronics, 26
(1990). Google Scholar
Brillet A. et al.,
“Recent Work on 612 nm He-Ne Stabilized Lasers,”
in Proceedings of the Second International Conference,
(1981). Google Scholar
Bender P.L. et al.,
“Satellite-Satellite Laserlinks for Future Gravity Missions,”
Space Science Reviews, 108 377
(2002). Google Scholar
Drever R.W.P. et al.,
“Laser Phase and Frequency Stabilization using an Optical Resonator,”
Appl. Phys. B: Photophys. Laser Chem, 31 97
(1983). Google Scholar
|