9 March 2018 Intensity-modulated dental CBCT: noise and absorbed dose study
Author Affiliations +
Abstract
Dental computed tomography (CT) typically uses a cone-beam geometry with a flat-panel detector. Although the flat-panel detector normally covers the maxillary and mandibular jaws, the cone-beam scan can deliver the dose to organs that are located out of the direct beam path but sensitive to radiation damage, such as eye lens. For typical dentoalveolar cone-beam CT (CBCT) scans, this study investigates the absorbed dose distributions in the head and neck using the Monte Carlo technique, and quantifies specific organ doses. Then, we design an intensity-modulated CBCT scan protocol that can provide a higher tomographic image quality at a lesser patient dose. The beam-intensity modulation includes the changes of tube current (mA) and/or voltage (kVp) during circular scanning, and the modulation scenarios are designed considering the cervical spine through which x-ray beam attenuates largely. We assess the noise-to-dose performances for various intensity-modulation scenarios, and compare the results with that obtained for the conventional scan.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jinwoo Kim, Jonghee Yun, Seungwoo Ha, Hanbean Youn, and Ho Kyung Kim "Intensity-modulated dental CBCT: noise and absorbed dose study", Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 105734C (9 March 2018); https://doi.org/10.1117/12.2293396
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation

Monte Carlo methods

X-rays

Image quality

Tissues

Head

Particles

RELATED CONTENT

A Monte Carlo study on the effect of the orbital...
Proceedings of SPIE (March 18 2015)
A method for optimizing the x ray tube current in...
Proceedings of SPIE (February 15 2021)
Quality assessment of clinical computed tomography
Proceedings of SPIE (September 15 2008)
CT vs. VolumeScope: image quality and dose comparison
Proceedings of SPIE (December 06 2006)
X ray tube voltage and image quality in adult and...
Proceedings of SPIE (March 02 2006)

Back to Top