PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
To achieve an ultrahigh-resolution for soft X-ray beamlines, the slope error of a highly precise grating is required on the level of 0.1 μrad root-mean-square (RMS) under thermal loading. To realize the goal, a specially designed 25-actuator optical surface bender for the gratings and mirrors is developed and operated at Taiwan Photon Source (TPS) [1]. In this paper, the construction and operation of the in situ LTP measuring system is described[2]. This LTP consists of a switchable optical reflection system that let the LTP can switch to measure horizontal or vertical mounting mirrors/gratings in the beamline. The other is a low optical distortion and bakeable to 120 ˚C glass viewport which is used for the ultra-high vacuum[3,4] interface for the beamlines optics and LTP. The surface slope error being reduced down to 0.1 and 0.15 μrad (RMS) by the 25-actuator bender without/with the glass viewport as verified by the in situ LTP measurements in the beamline.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The alert did not successfully save. Please try again later.
Shang-Wei Lin, Duan-Jen Wang, Chih-Yu Hua, Hok-Sum Fung, Ming-Ying Hsu, Gung-Chian Yin, "In-situ metrology for adaptive x-ray optics with an in-position LTP system in the soft x-ray beamline," Proc. SPIE 11492, Advances in Metrology for X-Ray and EUV Optics IX, 114920G (21 August 2020); https://doi.org/10.1117/12.2570538