15 February 2021 Model-based inter- and intra-panel inconsistency correction for photon counting detector CT
Author Affiliations +
Abstract
Due to the subtle variations of energy response functions across photoconductive panels, photon counting detector CT are subject to severe banding artifacts. This work presents a physics based method to correct for these artifacts. It employs calibration objects with known thicknesses and composition to estimate the panel-specific response functions, which are used to concert the raw photon counting projection data of an arbitrary image object into acrylic- and aluminum-equivalent pathlengths. Experimental results show the method not only remove the banding artifacts but also address the beam hardening artifacts.
Conference Presentation
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mang Feng, Xu Ji, Ran Zhang, and Ke Li "Model-based inter- and intra-panel inconsistency correction for photon counting detector CT", Proc. SPIE 11595, Medical Imaging 2021: Physics of Medical Imaging, 115951V (15 February 2021); https://doi.org/10.1117/12.2580817