10 November 1998 Calibration and preliminary results on the performance of the XMM EPIC PN camera: timing modes
Eckhard Kendziorra, Edgar Bihler, Michael Colli, Marcus Kirsch, Markus Kuster, Bernhard Pflueger, Ruediger Staubert, Heinrich W. Braeuninger, Ulrich G. Briel, Konrad Dennerl, Frank Haberl, Gisela D. Hartner, Norbert Meidinger, Elmar Pfeffermann, Claus Reppin, Lothar Strueder, Joachim E. Truemper, Peter Holl, Josef Kemmer, Heike Soltau, Christoph von Zanthier
Author Affiliations +
Abstract
The pm-CCD camera is one of the three focal plane instruments of the European Photon Imaging Camera (EPIC) on board the x-ray multi mirror (XMM) mission scheduled for launch in August 1999. The detector consists of four quadrants of three pn-CCDs each, which are integrate don one 4 inch silicon wafer. Each CCD has 200 by 64 pixels with 280 micrometers depletion depth. One CCD of a quadrant is readout at a time, while the four quadrants can be processed independently of each other. Observations of point sources brighter than 11 mCrab in imaging mode will be effected by photon pile-up. However, special operating modes can be used to observe bright sources up to 150 mCrab in Timing Mode with 30 microsecond(s) time resolution and very bright sources up to several Crab in Burst Mode with 7 microsecond(s) time resolution. We have tested and calibrate the flight model FM of the EPIC pn-CCD camera at the long beam test facility Panter near Munich and at the synchrotron monochromators of the Institut d'Astrophysique Spatiale in Orsay, France. In this paper describe the calibration of the pn-CCD detector in high time resolution/bright source operating modes and present preliminary results on the performance in these modes.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Eckhard Kendziorra, Edgar Bihler, Michael Colli, Marcus Kirsch, Markus Kuster, Bernhard Pflueger, Ruediger Staubert, Heinrich W. Braeuninger, Ulrich G. Briel, Konrad Dennerl, Frank Haberl, Gisela D. Hartner, Norbert Meidinger, Elmar Pfeffermann, Claus Reppin, Lothar Strueder, Joachim E. Truemper, Peter Holl, Josef Kemmer, Heike Soltau, and Christoph von Zanthier "Calibration and preliminary results on the performance of the XMM EPIC PN camera: timing modes", Proc. SPIE 3445, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy IX, (10 November 1998); https://doi.org/10.1117/12.330303
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cameras

Calibration

Charge-coupled devices

Sensors

Chlorine

Point spread functions

Fermium

RELATED CONTENT

eROSITA on SRG
Proceedings of SPIE (July 18 2016)
PN CCD camera for XMM performance of full frame...
Proceedings of SPIE (October 15 1997)
XMM science simulator: EPIC mode performance issues
Proceedings of SPIE (November 10 1998)
eROSITA in orbit calibration strategy and plan from the...
Proceedings of SPIE (September 13 2012)
PN CCD camera for XMM performance of high time...
Proceedings of SPIE (October 15 1997)

Back to Top