Paper
22 June 1999 Improved tympanic thermometer based on a fiber optic infrared radiometer and an otoscope and its use as a new diagnostic tool for acute otitis media
Gadi Fishman, Ari DeRowe, Eyal Ophir, Vered Scharf, Abraham Shabtai, Dov Ophir, Abraham Katzir
Author Affiliations +
Proceedings Volume 3590, Lasers in Surgery: Advanced Characterization, Therapeutics, and Systems IX; (1999) https://doi.org/10.1117/12.350995
Event: BiOS '99 International Biomedical Optics Symposium, 1999, San Jose, CA, United States
Abstract
Clinical diagnosis of acute otitis media (AOM) in children is not easy. It was assumed that there is a difference ΔT between the Tympanic Membrane (TM) temperatures in the two ears in unilateral AOM and that an accurate measurement of ΔT may improve the diagnosis accuracy. An IR transmitting fiber, made of AgClBr, was coupled into a hand held otoscope and was used for the non-contact (radiometric) measurements of TT, the TM temperature. Experiments were carried out, first, on a laboratory model that simulated the human ear, including an artificial tympanic membrane and an artificial ear canal. Measurements carried out using commercially available tympanic thermometers shown that the temperature Tc of the ear canal affected the results. Tc did not affect the fiberoptic radiometer, and this device accurately measured the true temperature, TT of the tympanic membrane. A prospective blinded sampling of the TM temperature was then performed on 48 children with suspected AOM. The mean temperature difference between the ears, for children with unilateral AOM was ΔT = (0.68 ± 0.27)°C. For children with bilateral AOM it was ΔT = (0.14±0.10)°C (p<0.001). It was demonstrated that afor unilateral AOM the difference ΔT was proportional to the systemic temperature. In conclusion, the fiberoptic interferometric measurements of the TM can be a useful non-invasive diagnostic tool for AOM, when combined with other data.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gadi Fishman, Ari DeRowe, Eyal Ophir, Vered Scharf, Abraham Shabtai, Dov Ophir, and Abraham Katzir "Improved tympanic thermometer based on a fiber optic infrared radiometer and an otoscope and its use as a new diagnostic tool for acute otitis media", Proc. SPIE 3590, Lasers in Surgery: Advanced Characterization, Therapeutics, and Systems IX, (22 June 1999); https://doi.org/10.1117/12.350995
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Ear

Temperature metrology

Radiometry

Bragg cells

Fiber optics

Optical fibers

Fiber optics tests

Back to Top