Paper
6 December 2006 An absorbed dose due to examination on the VolumeScope x-ray device
V. N. Vasiliev, M. Yu. Zaytsev
Author Affiliations +
Proceedings Volume 5943, X-ray and Neutron Capillary Optics II; 59430X (2006) https://doi.org/10.1117/12.637994
Event: X-ray and Neutron Capillary Optics II, 2004, Zvenigorod, Russian Federation
Abstract
An absorbed dose in a patient due to an X-ray examination by the VôlumeScope Compton scatter tomographic scanner was calculated and measured. The calculation model takes into account the energy spectrum of incident photons, the attenuation of primary and scattered radiation in the surrounding medium, multiple scattering and detector/collimator efficiency. The measurements were performed in various parts of a tissue-equivalent human phantom by LiF thermoluminescent detectors. Typical measured dose per examination was 1-3 mGy, the calculated dose was 1.95-2.2 mGy.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
V. N. Vasiliev and M. Yu. Zaytsev "An absorbed dose due to examination on the VolumeScope x-ray device", Proc. SPIE 5943, X-ray and Neutron Capillary Optics II, 59430X (6 December 2006); https://doi.org/10.1117/12.637994