Paper
18 March 2008 Detective quantum efficiency of an energy resolving photon counting detector
Author Affiliations +
Abstract
The output response characteristics of an X-ray photon counting detector are measured experimentally and simulated using a Monte Carlo method in order to quantify the loss of statistical information due to pile-up. The analysis is applied to idealize counting detector models, but is adaptable to realistic event processing that is not amenable to analytic solution. In particular, the detective quantum efficiency (DQE) is calculated as a function of flux rate and shown to have an intermediate zero for the paralyzable case at the maximum periodic rate. The progressive degradation of the spectral response as a function of increasing flux rate is also modeled. Analogous metrics to DQE are defined in regards to the detector's ability to resolve atomic number and enhance image contrast based on atomic number differentiation. Analytic solutions are provided for the output and linearized response statistics and these interpolate well across the Monte Carlo and experimental results.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Erik Tkaczyk, Samit Basu, Daniel McDevitt, Wen Li, and Yanfeng Du "Detective quantum efficiency of an energy resolving photon counting detector", Proc. SPIE 6913, Medical Imaging 2008: Physics of Medical Imaging, 69130N (18 March 2008); https://doi.org/10.1117/12.770309
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Photon counting

Monte Carlo methods

Aluminum

Statistical analysis

Quantum efficiency

X-rays

Back to Top