Paper
6 May 2009 An OLED-based MRC measurement bench
Author Affiliations +
Abstract
We present a novel device for fast, reproducible and low contrast minimum resolvable contrast (MRC) measurements based on an OLED microdisplay. The high intensity resolution and luminosity of the employed display allows the generation of target contrasts well below 0.5 % at brightness levels of more than 500 nits. Using a 4 m collimator we were able to perform an MRC measurement over the full relevant spatial frequency range of a recently developed terrestrial TV camera system. The small pixel pitch of the display allowed us to use a short collimator length and to perform MRC measurements with high frequency resolution.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bernd Mohring, Martin Gangl, Hannes Böhm, and Alfred Fendt "An OLED-based MRC measurement bench", Proc. SPIE 7327, Display Technologies and Applications for Defense, Security, and Avionics III, 73270I (6 May 2009); https://doi.org/10.1117/12.818131
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Organic light emitting diodes

Cameras

Spatial frequencies

Imaging systems

Collimators

Sensors

Visualization