11 February 2011 Instantaneous coherence length measurement of a swept laser source using a Mach-Zehnder interferometer
Tim von Niederhäusern, Christoph Meier, Marcus Duelk, Philipp Vorreau
Author Affiliations +
Abstract
We present a measurement method which is capable of measuring the instantaneous coherence length as a function of the wavelength while the source is working at its full sweep rate. The measurement principle is based on the dynamic decrease of fringe contrast as a function of the optical path difference. The measurement setup consists of a free-space Mach-Zehnder interferometer with a variable optical path difference. We present results for instantaneous coherence lengths in a range from 0 mm to 50 mm with a mean standard deviation of 0.42 mm at sweep rates of up to 120 kHz.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tim von Niederhäusern, Christoph Meier, Marcus Duelk, and Philipp Vorreau "Instantaneous coherence length measurement of a swept laser source using a Mach-Zehnder interferometer", Proc. SPIE 7889, Optical Coherence Tomography and Coherence Domain Optical Methods in Biomedicine XV, 78892R (11 February 2011); https://doi.org/10.1117/12.873769