1 January 1987 Spectral And Radiometric Calibration Of The Airborne Visible/Infrared Imaging Spectrometer
Gregg Vane, Thomas G. Chrien, Edward A. Miller, John H. Reimer
Author Affiliations +
Abstract
The laboratory spectral and radiometric calibration of the Airborne Visible/Infrared Imaging Spectrometer (AVIRIS) used in the radiometric calibration of all AVIRIS science data collected in 1987 is described. The instrumentation and procedures used in the calibration are discussed and the calibration accuracy achieved in the laboratory as determined by measurement and calculation is compared with the calibration requirements. Instrument performance factors affecting radiometry are described. The paper concludes with a discussion of future plans.
© (1987) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gregg Vane, Thomas G. Chrien, Edward A. Miller, and John H. Reimer "Spectral And Radiometric Calibration Of The Airborne Visible/Infrared Imaging Spectrometer", Proc. SPIE 0834, Imaging Spectroscopy II, (1 January 1987); https://doi.org/10.1117/12.942288