6 March 2013 Intrinsic and total system performance evaluation for a newly developed Solid State X-ray Image Intensifier (SSXII) detector
Author Affiliations +
Proceedings Volume 8668, Medical Imaging 2013: Physics of Medical Imaging; 86680K (2013) https://doi.org/10.1117/12.2006451
Event: SPIE Medical Imaging, 2013, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
The new Solid State X-ray Image Intensifier (SSXII) is a high-resolution, high-sensitivity, real-time region-ofinterest (ROI) x-ray imaging detector. Evaluations were made of both standard linear systems metrics (MTF, DQE) and total system performance with generalized linear systems metrics (GMTF, GDQE) including scatter and geometric un-sharpness for simulated clinical conditions. The SSXII is based on a 1k x 1k EMCCD sensor coupled to a 300 μm thick CsI(Tl) phosphor through a 2.88:1 fiber optic taper resulting in a 37 μm effective pixel size and an effective 3.7 cm x 3.7 cm square field-of-view (FOV). Standard methods were used to calculate MTF, NNPS and DQE. Generalized metrics were calculated and compared for three different magnifications (1.03, 1.11 and 1.2) and three different focal spots (0.3 mm, 0.5 mm and 0.8 mm) for a scatter fraction of 0.28. For an RQA5 spectrum, at 5 cycles/mm the MTF was found to be 0.06 and DQE was 0.04, while the DQE(0) was 0.60. Focal spot un-sharpness and scatter significantly degrades the GMTF and GDQE performance of the detector. A low frequency drop is caused by scatter and is almost independent of focal spot size and magnification. The degradation for middle range frequencies is caused by geometric un-sharpness and increases with focal spot size and magnification. This degradation was least in the case of the small focal spot and almost independent of magnification. In spite of this degradation, the high resolution SSXII with a small FOV may have a significant impact on ROI image-guided neuro-interventions since it demonstrates far better performance than standard current detectors.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
V. Singh, S. N. Swetadri Vasan, A. Jain, P. Sharma, D. R. Bednarek, and S. Rudin "Intrinsic and total system performance evaluation for a newly developed Solid State X-ray Image Intensifier (SSXII) detector", Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86680K (6 March 2013); https://doi.org/10.1117/12.2006451
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Spatial frequencies

Modulation transfer functions

X-rays

X-ray imaging

X-ray detectors

Image intensifiers

Back to Top