10 October 2013 Optimization of BRDF measurement method using Spectralon white reflectance standard
Author Affiliations +
Proceedings Volume 8916, Sixth International Symposium on Precision Mechanical Measurements; 89160S (2013) https://doi.org/10.1117/12.2035690
Event: Sixth International Symposium on Precision Mechanical Measurements, 2013, Guiyang, China
Abstract
There are two kinds of methods to measure the bidirectional reflectance distribution function (BRDF), direct method and relatively method. The relatively method is widely used because of its convenience and small error. However, the BRDF of the Spectralon white reflectance standard in relatively method was found to deviate from the Lambertian BRDF. It was found to vary with the angle of incidence, the wavelength of incidence and the states of polarization of the incident and reflected light. This paper proposes an optimized measurement formula of BRDF. The result is corrected. Taking the BRDF of standard roughness specimens for example, experimental results show that the measurement accuracy is improved comparing with the original formula. Finally the error analysis is carried out on the accuracy of measurement. The research has certain reference significance for the study of target characteristics on non-cooperative target ranging.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hui-Rong Tao, Xing-Hua Qu, Fu-Min Zhang, and Yuan-Qing Zhu "Optimization of BRDF measurement method using Spectralon white reflectance standard", Proc. SPIE 8916, Sixth International Symposium on Precision Mechanical Measurements, 89160S (10 October 2013); https://doi.org/10.1117/12.2035690
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Bidirectional reflectance transmission function

Reflectivity

Error analysis

Sensors

Calibration

Light sources

Standards development

Back to Top