|
1.IntroductionThe presence of endogenous nitric oxide in exhaled breath of humans and animals was first reported in 1991.1 Since then, with more than 1000 publications in the field, it is becoming increasingly apparent that measurements of exhaled constitute a new way to monitor the inflammatory status in respiratory disorders, such as asthma and other pulmonary conditions. Exhaled nitric oxide as a measure of inflammation is suggested as providing the best combination of disease evaluation and practical implementation for improved asthma outcomes.2 Exhaled nitric oxide has been successfully employed in chronic asthma treatment monitoring to reduce the dose of inhaled corticosteroids, which have serious side effects, without compromising asthma control.3 In treating asthma, which affects 19 million Americans, exhaled may soon be incorporated into routine clinical care.4 Exhaled concentrations from the lower respiratory tract exhibit significant expiratory flow rate dependence.5 Because of this, exhaled is commonly collected using a single breath maneuver at a constant exhalation flow rate. Exhaled sharply rises and reaches a plateau, which can have a positive, negative, or near-zero slope. The plateau concentration is defined as the average concentration over a 3-sec window in the plateau region.6 The plateau level is reported as the exhaled value. Measurements of the plateau at multiple flow rates allow for determination of the origin of (proximal or distal lung) by estimating the flow-independent exchange parameters.7 Both offline and online techniques are used to study exhaled . In offline sampling, a portion of the exhaled breath is collected in a suitable reservoir for subsequent analysis. In online sampling, is measured over time during exhalation. Offline sampling has the advantages of remote collection, independence from instrument response time, and more efficient use of the analyzer, because breath from several patients may be rapidly analyzed in succession. Online sampling has the advantages of immediate identification of contamination from gas not derived from the lower airways, near-instantaneous test results, and no error introduced by sample storage and handling. Online analysis is generally preferred for the rapid test result. In both on- and offline collection, breath samples are typically collected at a standardized flow rate using a single exhalation.6 Sampling during tidal breath, either spontaneous or paced, is gaining momentum as a noninvasive assessment of lung inflammation in infants and children less than 5 years of age.8, 9 The American Thoracic Society and European Respiratory Society (ATS) jointly published requirements for a nitric oxide analyzer.6 According to the requirements, a analyzer should have a sensitivity of with a response time of . Plateau measurements during tidal breathing range from 1 to and occur over a short time at the end of a tidal exhalation.8, 9 Therefore, investigations of measurements during tidal breathing require a sensor with a -sec response time and an equal or better sensitivity than is required for sampling during a single breath maneuver. Simultaneous measurement of the exhaled profile can be used as a marker of end-exhaled air when collecting breath from mechanically ventilated patients and subjects who are breathing normally.10 A rapid rise in the inspiratory flow rate and a sharp decrease in end-tidal concentrations can be used to adjust and synchronize the and waveforms both during recording and data processing after and measurements.11 Additionally, end-tidal may be useful as an internal standard to verify correct breath collection and as a normalization procedure to reduce variation in calculating exhaled values, as proposed by Roller 12 This technique was used to measure exhaled in a diverse population.13 2.Available Technologies for Exhaled Nitric Oxide DetectionSeveral technologies for exhaled nitric oxide measurement have been reported and are summarized in Table 1 . In electrochemical analysis, a voltage potential is applied across two electrodes. sensing is achieved through oxidation of at the electrode surface and measuring the resulting current between the electrodes. Recently, a portable electrochemical analyzer was introduced14 and compared to chemiluminescence analyzers described later.15, 16 The electrochemical analyzer showed good agreement with the chemiluminescent analyzer in both studies. However, the 5-ppb sensitivity of the electrochemical analyzer prevents its use in determining the flow-independent exchange parameters and for concentration measurement during spontaneous breathing. Mass spectrometry relies on the nonselective ionization of molecules within a sample, requiring a mass-selective apparatus to enhance selectivity. One selective ionization technique is resonance-enhanced multiphoton ionization (REMPI). In this technique, a target species is photoionized and detected using time-of-flight mass spectrometry (TOF-MS). Recently, a system combining REMPI with TOF-MS techniques was reported capable of breath measurements.17 Short, Frey, and Berter report sub-ppbv detection sensitivity in less than . A merit of this technique, as well as the optical techniques discussed later, is that multiple target gas species can be detected. Another advantage is the capability to measure concentrations of isotopomers of , which may be useful in metabolic studies. The chemiluminescent technique is based on a gas-phase reaction between and ozone .18 Oxygen supplied to the instrument is converted to ozone via a high voltage discharge. reacts with ozone in a reaction chamber at a pressure of 4 to to produce nitrogen dioxide in the excited state, which releases its energy by emitting light in the visible spectrum. A photomultiplier converts the emitted light to an electrical current. The amount of light produced is proportional to the amount of nitric oxide in the sample. Current clinical and point-of-care chemiluminescent analyzers for exhaled nitric oxide breath analysis are based on instrumention developed primarily by Aerocrine, Incorporated,19 (Providence, New Jersey) and Sievers, Limited (GE Analytical Instruments, Boulder, Colorado). The chemiluminescent analyzer used in this work (model 280, Sievers) has a sensitivity of . However, the need for frequent calibration and high voltage operation has impacted the wide-scale use of chemiluminescent instruments for exhaled monitoring in clinics. Optical techniques have been employed for exhaled nitric oxide detection including photoacoustic, laser absorption, and cavity-enhanced spectroscopy. A sensor using photoacoustic spectroscopy was used to detect and quantify , but the reported detection sensitivity was .20 The sensitivity of such a sensor can be improved by use of distributed feedback (DFB)-QC lasers with higher output power, because the photoacoustic signal detected by the microphones is directly proportional to the incident optical power. Laser absorption spectroscopy allows sensitive, selective, and fast-response concentration measurements. For example, a lead salt laser-based system using a multipass cell has achieved a detection sensitivity of with a 4-sec integration time and is commercially available.21 More recently, two thermoelectrically cooled continuous-wave (cw) quantum cascade laser (QCL)-based spectrometers, which achieved sub-ppbv detection sensitivities22, 23 using a multipass cell, were reported. An innovative QCL-based sensor technology that exploits magnetic field modulation in a Faraday rotation configuration24 has been reported, but so far a minimum detection limit of was observed, which can be improved by using a QC laser targeting a ro-vibrational line in the Q-branch. Very sensitive trace gas measurements can be performed using cavity ring-down spectroscopy (CRDS). This technique uses a sample cell constructed of two ultra-low-loss dielectric mirrors (with reflectivities of ), which form a high finesse cavity. A CRDS measurement consists of coupling the laser radiation into the high finesse cavity, rapidly switching off the laser source, and measuring the decay rate of the cavity output radiation. The advantages of CRDS are high sensitivity at sub-ppbv levels25, 26 due to a long effective optical path, immunity to laser power fluctuations, and self calibration. Halmer used a cavity ring-down absorption spectrometer based on a continuous-wave laser to achieve a minimum detectable concentration for of in .27 Integrated cavity output spectroscopy (ICOS) is an alternative method, which also takes advantage of a high finesse cavity, but is less technically demanding than CRDS, as it does not require mode matching or high speed signal sampling electronics required for precise ring-down event measurements. In ICOS only the total laser intensity exiting the high finesse cavity is recorded and not its time dependence. Hence this method relies on exciting many cavity modes simultaneously.25, 28 Additional dithering of one of the cavity mirrors improves the cavity throughput and helps minimizing mode noise in the resulting absorption spectra. Off-axis injection of the laser beam into ICOS cell provides an increase of spectral density of cavity modes and thus further minimizes the mode noise.29, 30 Several ICOS-based detection systems have been reported.31, 32, 33 A mid-infrared, cw, thermoelectrically cooled, quantum cascade laser33, 34, 35 is an ideal spectroscopic source for ICOS-based sensor platforms for medical diagnostics because of its high power and narrow laser spectral width , necessary for efficient coupling of laser radiation into an ICOS cavity. In this work, we used an off-axis ICOS-based sensor employing a 50-cm-long optical cavity. Details of the performance characteristics of the ICOS-based sensor using a quantum cascade laser operating at were reported previously by the authors.36 In this work, a liquid nitrogen cooled quantum cascade laser operating at was used. We report what we believe to be the first quantum cascade laser-based sensor capable of combined tidal and concentration measurements. 3.Experimental Method3.1.Gas AnalysisFor simultaneous and analysis, a quantum cascade laser-based sensor system employing a cw, DFB quantum cascade laser operating at , and off-axis ICOS was used.33 The performance of the ICOS sensor was assessed by evaluating the agreement between the ICOS sensor and two commercial gas sensors. was measured with a chemiluminescent nitric oxide analyzer (model 280, Sievers). was measured with a capnograph (model 8100, Novametrix), which uses broadband infrared absorption spectroscopy [also referred to as nondispersive infrared absorption (NDIR)]. The ICOS sensor was always calibrated with a reference mixture of in before analysis of a breath sample. The sample flow rate into the ICOS system was set at . The scan rate was chosen to be . The data acquisition and processing were realized using a National Instruments LabView-based control platform. Concentration measurements were performed by means of a least-squares linear fitting algorithm of a preacquired spectrum of reference gas mixture. This eliminates the need of taking into account changes in optical pathlength in the ICOS QCL-based system, which is necessary to consider during the synthesis of a spectrum using a spectroscopic database. The Sievers model 280 was calibrated every two weeks with the manufacturer’s recommended 45-ppm calibration gas mixture and daily using air entrained via an filter (Sievers Instruments) as an -free gas. The Sievers was calibrated with the -free gas more frequently than specified by the manufacturer and at least were allowed to pass between the 45-ppm gas calibration and sample analysis or -free gas calibration to achieve the best accuracy with the chemiluminescent technique. Based on the results of the two previous studies showing the importance of calibration with a low concentration gas standard, the chemiluminescent analyzer was calibrated daily with a low concentration (76-ppb in ) gas mixture.37, 38 The sample flow rate into the Sievers analyzer was . The NDIR capnograph was also calibrated daily according to the manufacturer’s instructions by inserting reference and zero-gas cells provided by the manufacturer into the infrared beam path. The calibration was performed automatically by the capnograph software. 3.2.Breath CollectionA custom-built breath-collection device was used to perform exhaled breath measurements.36 The apparatus meets the American Thoracic Guidelines6 for collecting breath for lower airway measurement by: 1. providing a back pressure to prevent nasal contamination, and 2. allowing the subject to maintain a constant exhalation flow for online collection. Mouth pressure was maintained by an adjustable stopcock valve and was monitored with a pressure sensor (model 860, Autotran). A constant exhalation flow rate was maintained using feedback from an in-line mass flowmeter (model 4021, TSI Incorporated) displayed on a laptop personal computer (PC) using a Labview interface. The exhalation flow rates ranged from 0.5 to to provide a range of concentrations in the breath samples. Only one sample was collected at each flow rate from each subject. For offline collection, the plateau region was determined by first obtaining online profiles at each flow rate for each patient and estimating the plateau region based on the exhalation time for each flow rate. Breath was collected at least after the estimated start of the plateau region. 3.3.Study SubjectsSubjects were recruited from the Pulmonary Clinic at the Michael E. Debakey Veterans Affairs Medical Center (MED-VAMC) and from Rice University (Houston, Texas). The breath collection protocol was approved by the Institutional Review Boards of Baylor College of Medicine, MED-VAMC, and Rice University. Written informed consent was obtained from each enrolled patient prior to study entry. The inclusion criteria were age equal to or greater than 21. The exclusion criteria were the inability to adequately perform the breath collection maneuvers. MED-VAMC patients were included in the study to obtain breath samples from individuals with pulmonary diseases or conditions that alter levels (e.g., asthma, chronic obstructive pulminary disease (COPD), and smoking). The study subjects were enrolled to obtain biogenic in breath samples with varying concentrations to compare the analyzers. Factors affecting such as acute respiratory infection, diet, and medications were not taken into account, since no intra- or intersubject comparisons were made between subjects. 3.4.Offline Nitric Oxide and Carbon DioxideBreath samples were collected in 1-L Tedlar bags and measured with both the ICOS sensor and the chemiluminescent sensor. Breath was sampled over a range of exhalation flow rates (from 0.5 to ) to produce a range of concentrations. To assess the performance agreement of the three sensors, a statistical analysis was employed. We selected the Bland-Altman method39 rather than the Pearson product-moment correlation coefficient. The Pearson product-moment correlation coefficient, as an estimate of agreement between two techniques, can be misleading for several reasons39, 40: 1. two sets of data can have high correlation if the points lie along any straight line, not only unity; 2. correlation depends on the range of values; and 3. data with limited agreement can have high correlations. For the Bland-Altman method, a plot of the difference between a pair of measurements (in this case ICOS minus chemiluminescence and ICOS minus NDIR) versus their average is obtained. Subsequently, the limits of agreement between each pair of gas sensors can be determined as differences within standard deviations, provided that differences in this range are not clinically important. Offline measurements were made from the same 1-L Tedlar bags used in the previously mentioned offline concentration measurements. 28 breath samples were analyzed with the QCL-based ICOS sensor and the capnograph. The ICOS sensor was calibrated with a reference mixture of 5% in after analysis of five samples. The capnograph was calibrated using a reference cell at the start of each measurement session, as recommended by the manufacturer. The limits of agreement were determined using the Bland-Altman method. 3.5.Online and Tidal Nitric Oxide and Carbon DioxideFor online exhaled breath measurements, a single subject performed single breath maneuvers at six flow rates , 2, 3, 6, 9, and . As mentioned previously, measurement of the plateau for multiple flow rates allows an estimation of flow independent parameters, which may provide additional clinical insight as to the output from the alveolar and airway regions of the lung.7 Exhaled breath samples were measured by only one instrument, since the sensors were in two different geographic locations and could not be relocated. measurements were made using the ICOS sensor and the chemiluminescent sensor. measurements were made using the ICOS sensor and the NDIR capnograph. The average and standard deviation of the plateau and end-tidal were determined for each set of exhaled breaths at each flow rate. For tidal measurements, the flow restrictor was removed from the breathing circuit to allow comfortable spontaneous breathing. 4.Results4.1.Study SubjectsStudy subjects ranged from 29 to 72 years of age. The subject group included five healthy subjects, three patients with asthma, 14 exsmokers with chronic obstructive pulmonary disease, and eight current smokers with chronic obstructive pulmonary disease. 4.2.Offline Nitric Oxide and Carbon Dioxide Comparison72 breath samples were measured first with the ICOS-based sensor followed by the chemiluminescent analyzer, and 65 breath samples were measured in the reverse order, with several hours lapsing between measurements. For exhaled , no change in agreement of the results obtained with different sensors was noted during the entire period of this study. Offline measurements were performed with a 1.8-sec integration time with both sensors. The best to-date minimum detectable concentration achieved with the ICOS sensor is with a 1-sec integration time, as determined by of the Voigt fit residual using a 76-ppb calibration gas in as a balance gas. The average difference in concentration measurements between the ICOS and chemiluminescent analyzers (ICOS minus chemiluminescence) was , and the limits of agreement (as ) were and . It was observed that the concentration gradually decreases with time in the collected breath samples. To remove systematic errors, the offline measurements were separated into two groups—those measured with ICOS and then chemiluminescence, and those measured in the reverse order. Figure 1 shows Bland-Altman plots for offline concentration measurements for the sequence of ICOS followed by chemiluminescence [Fig. 1a] and chemiluminescence followed by ICOS [Fig. 1b]. concentration measurements from samples measured by ICOS first had an average difference (ICOS minus chemiluminescence) of and limits of precision of ( - and 4-ppb ). concentration measurements from samples measured by the chemiluminescent analyzer first had an average difference (ICOS minus chemiluminescence) of and limits of precision of ( - and 2.1-ppb ). For exhaled , 28 breath samples were collected during a period of two weeks. Offline measurements were made using a 1-sec averaging time. The average difference between ICOS and NDIR shown in Fig. 2 was , and the limits of agreement were and . 4.3.Online Nitric Oxide and Carbon Dioxide ComparisonFigure 3 shows exhaled at measured with ICOS with a 2.4-sec averaging time and computation time, and chemiluminescence with a 2.4-sec averaging time. The solid line represents the average of 15 breath measurements with the chemiluminescent analyzer at a 16 samples/sec rate after applying a 50-point Savitzky-Golay smoothing routine with a first-order polynomial. The squares represent the average concentration measured by the chemiluminescent analyzer after a 2.4-sec averaging (wide cap error bars are 1 standard deviation for the 15 chemiluminescent breath measurements). The solid circles connected with a dashed line represent average measurements with ICOS using a 2.4-sec averaging and computation time (narrow cap error bars are 1 standard deviation for the 15 ICOS breath measurements). The end-tidal did not vary significantly with exhalation flow rate. The standard deviation of end-tidal values was equivalent between sensors ( for ICOS and NDIR). Figure 4 shows exhaled at measured with off axis ICOS (1.8-sec data averaging and computation time) [Fig. 4a] and NDIR (1.8-sec data averaging) [Fig. 4b]. A time delay between the start of the NDIR waveform and the ICOS waveform was observed for all exhalation flow rates. 4.4.Tidal Nitric Oxide and Carbon Dioxide ComparisonAn waveform was obtained with both the ICOS and chemiluminescence analyzers. Figure 5 shows exhaled data collected from the ICOS with a 0.33-sec data averaging and computation [Fig. 5a] and chemiluminescence with a rate during tidal breathing [Fig. 5b]. The chemiluminescent data were not averaged to preserve rapid changes in over time. For tidal , the in-line NDIR capnograph detected rapid changes in , whereas the ICOS sensor had a response delay due to sample cell filling. Figures 6a and 6b show exhaled data collected from the ICOS (0.15-sec averaging and computation) and NDIR (0.15-sec averaging time), respectively. The NDIR capnograph sensor measured a typical expiratory waveform41 consisting of three phases: phase 1, a portion of zero ; phase 2, a rapid rise in ; and phase 3, a plateau region with a small positive slope. The ICOS sensor measured a ramp waveform with end-tidal values in agreement with NDIR. Due to the slower gas delivery system, the standard phases visible in the waveform acquired by the NDIR capnograph were strongly distorted in the ICOS tidal waveform. Table 1Performance summary of five exhaled nitric oxide measurement techniques.
5.DiscussionThe ICOS offline measurements of exhaled and are in good agreement with commercially available sensors. In this work, the current gold standards are the chemiluminescence and NDIR methods. The purpose of the Bland-Altman method is to assess the agreement of two methods, where one method is the clinical gold standard and the true value is not known or cannot be easily determined. The Bland-Altman method does not compare the accuracy of the two methods. Instead, the Bland-Altman method is used to determine if the new method has readings sufficiently similar to the standard technique to allow the new technique to replace the standard technique in clinical decision making. For the offline concentration measurements, a narrower limit of precision versus the entire dataset was obtained by dividing the offline samples into two groups according to the order of analysis: 1. ICOS first and chemiluminescence second, and 2. vice versa. To verify the 2- limits of precision determined using the Bland Altman method, the standard deviations of both instruments were determined at the time of measurement, which yielded for the chemiluminescent analyzer and for the ICOS analyzer. The two standard deviations were added quadratically, yielding , which is in reasonably good agreement with the 1- limits of precision estimated using the Bland-Altman method. The standard deviations of the chemiluminescent and ICOS analyzers were determined using the 76-ppb calibration gas. The ICOS analyzer was undergoing continuous development during the reported studies, and therefore our best to date minimum detectable concentration was better than the minimum detectable concentration measured at the time of intercomparison investigations for the offline measurements. The limit of precision for each group individually had remarkably similar limits of precision , suggesting that a different systematic error was introduced into each group of data. The systematic error in the offline comparison appeared to be due to a decreasing concentration over the 4 to between consecutive concentration measurements. In the group measured by ICOS first [Fig. 1a], the average difference (ICOS minus chemiluminescence) was greater than 0 , whereas in the group measured by chemiluminescence first [Fig. 1b], the average difference (ICOS minus chemiluminescence) was less than 0 ( -ppbv ). This finding suggests that is chemically unstable and decreases in Tedlar bags during a 4- to 6-h period. The ICOS offline measurements were in good agreement with the NDIR capnograph. The average difference was and the limits of agreement were . The percent differences are not large enough to be clinically important, and ICOS measurements can be used clinically. For online analysis, the plateau concentrations were estimated for breath measurements with both the ICOS and chemiluminescent analyzers over a range of exhalation flow rates (data not shown). The average of concentration profiles at a exhalation rate are shown in Fig. 3. The 3-l/min exhalation flow rate was chosen, as it is recommended by the lower airway guidelines.6 The plateau regions for ICOS and chemiluminescence were in good agreement, with all data points within 1 standard deviation of the 15 measurements. The ICOS measurements had a larger standard deviation for each averaged concentration. Tidal breath measurements require subsecond time resolution with a detection sensitivity of . From Fig. 5, it is evident that the chemiluminescent sensors had adequate sensitivity to detect the plateau from the waveform. For tidal measurements, better detection sensitivity of the ICOS sensor is required to improve clinical decision making. The ICOS sensor was found to be in good agreement with the NDIR capnograph for online end-tidal measurements. End-tidal varies with ventilation, where increased ventilation produces a lower end-tidal . Variation in ventilation was minimized in this study by using paced breathing during online and tidal measurements. For online measurements, the standard deviation of end-tidal values was similar for the ICOS and NDIR sensors. The ICOS online waveforms began after the NDIR waveforms (see Fig. 4). The delay was due to the difference in the position of the sensor in the breath collection system. The NDIR sensor was positioned near the mouth in-line with the exhalation flow, whereas the ICOS sensor drew air from the exhalation flow into the sensor at a rate of . The time lag is estimated to be based on the sample flow rate, an total gas cell volume (at ), and a 1.8-sec integration time. A time lag was not observed between the chemiluminescent and ICOS sensors because the chemiluminescent sensor also drew air from the exhalation flow. As with online analysis, the different positions of the NDIR and ICOS sensors accounted for the faster response time of the NDIR sensor. The NDIR sensor was able to rapidly detected changes in during tidal breathing, allowing visualization of the three phases of exhaled , labeled in Fig. 6. The gas handling system of the ICOS sensor did not have sufficient time response to resolve the three phases of expiratory .41 The maximum sample flow rate of into the ICOS sample cell restricted the system gas exchange capability and was the limiting factor in the time resolution of the measurement. At the flow rate of , the time required for a complete gas exchange within the sampling cell is . The volume of the ICOS sensor system includes the sample cell and tubing connecting the breath collection system, the sample cell, and the pump. A higher gas flow and smaller total volume would improve the time response of the ICOS sensor. In the case of breath measurement, in which the expected concentrations are at levels between 2 and 5%, it is important that the absorption lines are strong enough to have high signal-to-noise ratio but at the same time are not saturated. In this work we were able to target optimal absorption lines in the region that allowed for rapid and sensitive exhaled measurements. Simultaneous and measurements can be made with the ICOS sensor by measuring the absorption line at and the absorption line at , accessible within a single frequency scan of the QCL. The linewidth of the absorption feature is sufficiently low to prevent spectral interference with the neighboring line. No spectral interferences from other molecules are present in this wavelength region. Several improvements can be made to the ICOS sensor to achieve better sensitivity of exhaled during tidal breathing and to improve time resolution of measurement during tidal breathing. Mirrors with higher reflectivity are available and would improve the sensitivity of both gases and allow a smaller ICOS gas sample cell to be used. Quantum cascade lasers are available at , where the strongest, interference-free line in the R-branch of the fundamental ro-vibrational band is located. The line is times stronger than the line used in this work. Wavelength modulation spectroscopy, where the QCL current is modulated at a high frequency and a lock-in amplifier is used to detect the second harmonic frequency, can also be used to improve the signal-to-noise ratio.42 Off-axis ICOS sensors have benefited by applying this technique.32, 33, 43 Furthermore, a concentration determination method applied in this work, based on a fitting of the measurement data with the tabulated reference spectrum, can benefit from faster data acquisition by increasing the number of points per laser frequency scan, and thereby reducing the fitting uncertainty. 6.SummaryIn this work an exhaled sensor for breath analysis using mid-infrared QCL-based integrated cavity output spectroscopy is compared to a commercial chemiluminescence sensor and nondispersive infrared absorption sensor. The ICOS sensor shows good agreement with the two commercial sensors using offline, online, and tidal sampling techniques. The ICOS sensor has a noise-equivalent sensitivity for of with a 1-sec averaging time. Potential improvements to the ICOS sensor include incorporating higher reflectivity mirrors and utilizing the absorption line at . A key advantage of the ICOS-QCL sensor is its ability to simultaneously detect multiple clinically relevant exhaled molecules, in this case and . AcknowledgmentsThis work was supported by the National Aeronautics and Space Administration, National Science Foundation (ERC MIRTHE), and the Welch Foundation. The authors wish to thank Stuart Abramson and David Walding of Texas Children’s Hospital (Houston, Texas) for generously allowing the use of the Sievers model 280 and Novametrix model 8100. ReferencesL. E. Gustafsson,
A. M. Leone,
M. G. Persson,
N. P. Wiklund, and
S. Moncada,
“Endogenous nitric oxide is present in the exhaledair of rabbits, guinea pigs and humans,”
Biochem. Biophys. Res. Commun., 181 852
–857
(1991). 0006-291X Google Scholar
D. Bukstein,
M. Kraft,
A. H. Liu, and
S. P. Peters,
“Asthma endpoints and outcomes: What have we learned?,”
J. Allergy Clin. Immunol., 118 S1
–S15
(2006). 0091-6749 Google Scholar
A. D. Smith,
J. O. Cowan,
K. P. Brassett,
G. P. Herbison, and
D. R. Taylor,
“Use of exhaled nitric oxide measurements to guide treatment in chronic asthma,”
N. Engl. J. Med., 352 2163
–2173
(2005). https://doi.org/10.1056/NEJMoa043596 0028-4793 Google Scholar
D. R. Taylor,
M. W. Pijnenburg,
A. D. Smith, and
J. C. De Jongste,
“Exhaled nitric oxide measurements: clinical application and interpretation,”
Thorax, 61 817
–827
(2006). 0040-6376 Google Scholar
P. E. Silkoff,
P. A. McClean,
A. S. Slutsky,
H. G. Furlott,
E. Hoffstein,
S. Wakita,
K. R. Chapman,
J. P. Szalai, and
N. Zamel,
“Marked flow-dependence of exhaled nitric oxide using a new technique to exclude nasal nitric oxide,”
Am. J. Respir. Crit. Care Med., 155 260
–267
(1997). 1073-449X Google Scholar
American Thoracic Society and European Respiratory Society,
“ATS/ERS recommendations for standardized procedures for the online and offline measurement of exhaled lower respiratory nitric oxide and nasal nitric oxide, 2005,”
Am. J. Respir. Crit. Care Med., 171 912
–930
(2005). 1073-449X Google Scholar
S. C. George,
M. Hogman,
S. Permutt, and
P. E. Silkoff,
“Modeling pulmonary nitric oxide exchange,”
J. Appl. Physiol., 96 831
–839
(2004). https://doi.org/10.1152/japplphysiol.00950.2003 8750-7587 Google Scholar
A. Artlich,
B. Jonsson,
M. Bhiladvala,
P. A. Lonnqvist, and
L. E. Gustafsson,
“Single breath analysis of endogenous nitric oxide in the newborn,”
Biol. Neonate, 79 21
–26
(2001). 0006-3126 Google Scholar
E. Baraldi,
C. Dario,
R. Ongaro,
M. Scollo,
N. M. Azzolin,
N. Panza,
N. Paganini, and
F. Zacchello,
“Exhaled nitric oxide concentrations during treatment of wheezing exacerbation ininfants and young children,”
Am. J. Respir. Crit. Care Med., 159 1284
–1288
(1999). 1073-449X Google Scholar
T. Birken,
J. Schubert,
W. Miekisch, and
G. Noldge-Schomburg,
“A novel visually controlled alveolar breath sampling technique,”
Technol. Health Care, 14 499
–506
(2006). 0928-7329 Google Scholar
G. L. Hall,
B. Reinmann,
J. H. Wildhaber, and
U. Frey,
“Tidal exhaled nitric oxide in healthy, unsedated newborn infants with prenatal tobacco exposure,”
J. Appl. Physiol., 92 59
–66
(2002). 8750-7587 Google Scholar
C. B. Roller,
K. Namjou,
J. D. Jeffers,
W. Potter,
P. J. McCann, and
J. Grego,
“Simultaneous and measurement in human breath with a single IV-VI mid-infrared laser,”
Opt. Lett., 27 107
–109
(2002). 0146-9592 Google Scholar
K. Namjou,
C. B. Roller,
T. E. Reich,
J. D. Jeffers,
G. L. McMillen,
P. J. McCann, and
M. A. Camp,
“Determination of exhaled nitric oxide distributions in a diverse sample population using tunable diode laser absorption spectroscopy,”
Appl. Phys. B, 85 427
–435
(2006). https://doi.org/10.1007/s00340-006-2301-3 0946-2171 Google Scholar
T. Hemmingsson,
D. Linnarsson, and
R. Gambert,
“Novel hand-held device for exhaled nitric oxide-analysis in research and clinical applications,”
J. Clin. Monit Comput., 18 379
–387
(2004). 1387-1307 Google Scholar
K. Alving,
C. Janson, and
L. Nordvall,
“Performance of a new hand-held device for exhaled nitric oxide measurement in adults and children,”
Respir. Res., 20 57
–67
(2006). Google Scholar
C. McGill,
G. Malik, and
S. W. Turner,
“Validation of a hand-held exhaled nitric oxide analyzer for use in children,”
Pediatr. Pulmonol, 41 1053
–1057
(2006). 8755-6863 Google Scholar
L. C. Short,
R. Frey, and
T. Benter,
“Real-time analysis of exhaled breath via resonance-enhanced multiphoton ionization-mass spectrometry with a medium pressure laser ionization source: observed nitric oxide profile,”
Appl. Spectrosc., 60 217
–222
(2006). https://doi.org/10.1366/000370206776023241 0003-7028 Google Scholar
A. Fontijn,
A. Sabadell, and
R. Ronco,
“Homogenous chemiluminescent measurement of nitric oxide with ozone. Implications for continuous selective monitoring of gaseous air pollutants,”
Anal. Chem., 42 575
–579
(1970). https://doi.org/10.1021/ac60288a034 0003-2700 Google Scholar
P. Silkoff,
M. Carlson,
T. Bourke,
R. Katial,
E. Ogren, and
S. Szefler,
“The aerocrine exhaled nitric oxide monitoring system is cleared by the Food and Drug Administration for monitoring therapy in asthma,”
J. Allergy Clin. Immunol., 114 1241
–1256
(2004). 0091-6749 Google Scholar
A. Elia,
P. M. Lugara, and
C. Giancaspro,
“Photoacoustic detection of nitric oxide by use of a quantum-cascade laser,”
Opt. Lett., 30 988
–990
(2005). https://doi.org/10.1364/OL.30.000988 0146-9592 Google Scholar
C. B. Roller,
K. Namjou,
J. D. Jeffers,
M. Camp,
A. Mock,
P. J. McCann, and
J. Grego,
“Nitric oxide breath testing by tunable-diode laser absorption spectroscopy: application in monitoring respiratory inflammation,”
Appl. Opt., 41 6018
–6029
(2002). 0003-6935 Google Scholar
D. D. Nelson,
J. B. McManus,
S. C. Herndon,
J. H. Shorter,
M. S. Zahniser,
S. Blaser,
L. Hvozdara,
A. Muller,
M. Giovannini, and
J. Faist,
“Characterization of a near-room-temperature, continuous-wave quantum cascade laser for long-term, unattended monitoring of nitric oxide in the atmosphere,”
Opt. Lett., 31 2012
–2014
(2006). https://doi.org/10.1364/OL.31.002012 0146-9592 Google Scholar
B. W. M. Moeskops,
S. M. Cstescu, and
F. J. M. Harren,
“Sub-part-per billion monitoring of nitric oxide using wavelength modulation spectroscopy in combination with a thermoelectrically cooled, continuous wave quantum cascade laser,”
Opt. Lett., 31 823
–825
(2006). https://doi.org/10.1364/OL.31.000823 0146-9592 Google Scholar
H. Ganser,
W. Urban, and
J. Brown,
“The sensitive detection of by Faraday modulation spectroscopy with a quantum cascade laser,”
Mol. Phys., 101 545
–550
(2003). https://doi.org/10.1080/00268970210159460 0026-8976 Google Scholar
B. A. Paldus and
A. A. Kachanov,
“An historical overview of cavity-enhanced methods,”
Can. J. Phys., 83 975
–999
(2005). https://doi.org/10.1139/p05-054 0008-4204 Google Scholar
A. A. Kosterev,
A. L. Malinovsky,
F. K. Tittel,
C. Gmachl,
F. Capasso,
D. L. Sivco,
J. N. Baillargeon,
A. L. Hutchinson, and
A. Y. Cho,
“Cavity ring down spectroscopic detection of nitricoxide with a continuous-wave quantum-cascade laser,”
Appl. Opt., 40 5522
–5529
(2001). 0003-6935 Google Scholar
D. Halmer,
G. von Basum,
M. Horstjann,
P. Herring, and
M. Murtz,
“Time resolved simultaneous detection of and via mid-infrared cavity leak-out spectroscopy,”
Isotopes Environ. Health Stud., 41 303
–311
(2005). https://doi.org/10.1080/10256010500384408 Google Scholar
A. O’Keefe,
J. J. Scherer, and
J. B. Paul,
“CW integrated cavity output spectroscopy,”
Chem. Phys. Lett., 307 343
–349
(1999). https://doi.org/10.1016/S0009-2614(99)00547-3 0009-2614 Google Scholar
J. B. Paul,
L. Larson, and
J. G. Anderson,
“Ultrasensitive absorption spectroscopy with a high-finesse optical cavity and off-axis alignment,”
Appl. Opt., 40 4904
–4910
(2001). 0003-6935 Google Scholar
D. S. Baer,
J. B. Paul,
M. Gupta, and
A. O’Keefe,
“Sensitive absorption measurements in the near-infrared region using off-axis integrated-cavity-output spectroscopy,”
Appl. Phys. B, 75 261
–265
(2002). https://doi.org/10.1007/s00340-002-0971-z 0946-2171 Google Scholar
M. L. Silva,
D. M. Sonnenfroh,
D. I. Rosen,
M. G. Allen, and
A. O’Keefe,
“Integrated cavity output spectroscopy measurements of levels in breath with a pulsed room-temperature QCL,”
Appl. Phys. B, 81 705
–710
(2004). https://doi.org/10.1007/s00340-005-1922-2 0946-2171 Google Scholar
Y. A. Bakhirkin,
A. A. Kosterev,
C. Roller,
R. F. Curl, and
F. K. Tittel,
“Mid-infrared quantum cascade laser based off-axis integrated cavity output spectroscopy for biogenic nitric oxide detection,”
Appl. Opt., 43 2257
–2266
(2004). https://doi.org/10.1364/AO.43.002257 0003-6935 Google Scholar
Y. A. Bakhirkin,
A. A. Kosterev,
R. F. Curl,
F. K. Tittel,
D. A. Yarekha,
L. Hvozdara,
M. Giovannini, and
J. Faist,
“Sub-ppbv nitric oxide concentration measurements using cw thermoelectrically cooled quantum cascade laser-based integrated cavity output spectroscopy,”
Appl. Phys. B, 82 149
–154
(2006). https://doi.org/10.1007/s00340-005-2058-0 0946-2171 Google Scholar
S. Blaser,
D. A. Yarekha,
L. Hvozdara,
Y. Bonetti,
A. Muller,
M. Giovannini, and
J. Faist,
“Room-temperature, continuous-wave, single-mode quantum cascade lasers at ,”
Appl. Phys. Lett., 86 041109
(2005). https://doi.org/10.1063/1.1853511 0003-6951 Google Scholar
L. Diehl,
D. Bour,
S. Corzine,
J. Zhu,
G. Hofler,
M. Loncar,
M. Troccoli, and
F. Capasso,
“High-temperature continuous wave operation of strain-balanced quantum cascade lasers grown by metal organic vapor-phase epitaxy,”
Appl. Phys. Lett., 89 081101
(2006). https://doi.org/10.1063/1.2337284 0003-6951 Google Scholar
M. R. McCurdy,
Y. A. Bakhirkin, and
F. K. Tittel,
“Quantum cascade laser-based integrated cavity output spectroscopy of exhaled nitric oxide,”
Appl. Phys. B, 85 445
–452
(2006). https://doi.org/10.1007/s00340-006-2365-0 0946-2171 Google Scholar
Z. Borrill,
D. Clough,
N. Truman,
J. Morris,
S. Langley, and
D. Singh,
“A comparison of exhaled nitric oxide measurements performed using three different analysers,”
Respir. Med., 100 1392
–1396
(2006). 0954-6111 Google Scholar
K. C. Muller,
R. A. Jorres,
H. Magnussen, and
O. Holz,
“Comparison of exhaled nitric oxide analysers,”
Respir. Med., 99 631
–637
(2005). 0954-6111 Google Scholar
J. M. Bland and
D. G. Altman,
“Statistical methods for assessing agreement between two methods of linical measurement,”
Lancet, 8 307
–310
(1986). 0140-6736 Google Scholar
J. Ludbrook,
“Statistical techniques for comparing measurers and methods of measurement: a critical review,”
Clin. Exp. Pharmacol. Physiol., 29 527
–536
(2002). https://doi.org/10.1046/j.1440-1681.2002.03686.x 0305-1870 Google Scholar
C. T. Anderson and
P. H. Breen,
“Carbon dioxide kinetics and capnography during critical care,”
Crit. Care, 4 207
–215
(2000). Google Scholar
J. Reid and
D. Labrie,
“Second-harmonic detection with tunabled iode lasers—comparison of experiment and theory,”
Appl. Phys. B, 26 203
–210
(1981). https://doi.org/10.1007/BF00692448 0946-2171 Google Scholar
A. Zybin,
Y. A. Kuritsyn,
V. R. Mironenko, and
K. Niemax,
“Cavity enhanced wavelength modulation spectrometry for application in chemical analysis,”
Appl. Phys. B, 78 103
–109
(2004). https://doi.org/10.1007/s00340-003-1342-0 0946-2171 Google Scholar
|