20 May 2022 Methods and applications of full-field optical coherence tomography: a review
Ling Wang, Rongzhen Fu, Chen Xu, Mingen Xu
Author Affiliations +
Abstract

Significance: Full-field optical coherence tomography (FF-OCT) enables en face views of scattering samples at a given depth with subcellular resolution, similar to biopsy without the need of sample slicing or other complex preparation. This noninvasive, high-resolution, three-dimensional (3D) imaging method has the potential to become a powerful tool in biomedical research, clinical applications, and other microscopic detection.

Aim: Our review provides an overview of the disruptive innovations and key technologies to further improve FF-OCT performance, promoting FF-OCT technology in biomedical and other application scenarios.

Approach: A comprehensive review of state-of-the-art accomplishments in OCT has been performed. Methods to improve performance of FF-OCT systems are reviewed, including advanced phase-shift approaches for imaging speed improvement, methods of denoising, artifact reduction, and aberration correction for imaging quality optimization, innovations for imaging flux expansion (field-of-view enlargement and imaging-depth-limit extension), new implementations for multimodality systems, and deep learning enhanced FF-OCT for information mining, etc. Finally, we summarize the application status and prospects of FF-OCT in the fields of biomedicine, materials science, security, and identification.

Results: The most worth-expecting FF-OCT innovations include combining the technique of spatial modulation of optical field and computational optical imaging technology to obtain greater penetration depth, as well as exploiting endogenous contrast for functional imaging, e.g., dynamic FF-OCT, which enables noninvasive visualization of tissue dynamic properties or intracellular motility. Different dynamic imaging algorithms are compared using the same OCT data of the colorectal cancer organoid, which helps to understand the disadvantages and advantages of each. In addition, deep learning enhanced FF-OCT provides more valuable characteristic information, which is of great significance for auxiliary diagnosis and organoid detection.

Conclusions: FF-OCT has not been completely exploited and has substantial growth potential. By elaborating the key technologies, performance optimization methods, and application status of FF-OCT, we expect to accelerate the development of FF-OCT in both academic and industry fields. This renewed perspective on FF-OCT may also serve as a road map for future development of invasive 3D super-resolution imaging techniques to solve the problems of microscopic visualization detection.

1.

Introduction

From the perspective of resolution, radiation damage, price–performance ratio, and other aggregate indicators, optical imaging technology featuring high spatiotemporal resolution, high information capacity, and nonionizing properties have garnered significant research attention in the fields of biomedical and material science.1,2 Currently, high-resolution imaging technologies, such as fluorescence microscopy (FM),35 confocal microscopy (CM),6,7 two-photon microcopy (TPM),811 optical coherence tomography (OCT),12,13 and full-field optical coherence tomography (FF-OCT),14,15 have been widely adopted. Each technology has its distinct features. Combining fluorescence probe and optical microscopy, FM can perform micron-scale high-resolution imaging of tissues, organoids, cells, and other samples. However, FM cannot provide depth information and exhibits phototoxicity to tissues.16,17 Based on single-photon or two-photon excitation, CM and TPM render high imaging resolution and images with good contrast, but their imaging depth is mostly limited to a few hundred μm.9 Based on the principle of low coherent interference, OCT can provide histology-like images of a sample noninvasively; the transverse resolution and axial resolution of OCT can be controlled, respectively, which can reach 10  μm or even 1  μm; OCT imaging depth can reach 1 mm or higher, according to the scattering characteristics of samples, whereas the system cost is considerably lower than that of TPI. Therefore, OCT has attracted increased attention since it was first proposed in 1991.12,13,18

FF-OCT, which is developed from OCT, adopts a broadband light source with low spatial coherence to evenly irradiate the whole surface of the imaging sample and collects interferograms of different phases at the focusing depth of the sample by the planar detector, such as CMOS or CCD camera. Without performing transverse scanning, FF-OCT can reconstruct the optical slice image at the layer of the focal plane and the coherence plane as well as perform axial scanning similar to time-domain optical coherence tomography (TD-OCT) or time- and space-sharing parallel detection similar to Fourier-domain optical coherence tomography (FD-OCT).14,19 Therefore, optical slices with different focusing depths can be obtained to eventually reconstruct three-dimensional (3D) images of the sample. FF-OCT adopts lens with a high numerical aperture (NA) for sample imaging, which can significantly enhance its transverse resolution.20,21 The use of the broadband light source ensures high axial resolution, making the overall spatial imaging resolution of FF-OCT reach subcellular level, and the imaging results are comparable to biopsy.15,22 Based on full-field illumination, FF-OCT can conduct microscopy on two-dimensional (2D) slices at any depth within the imageable depth range of the sample without requiring transverse point scanning, thus correlating with conventional microscopy, CM, and TPM.2325 In addition, FF-OCT can significantly suppress stray light using the coherent gate, given these advantages, FF-OCT has attracted extensive attention from researchers worldwide and demonstrated its applications in biomedicine, material detection, security identification, and other scenarios. As a type of optical microscopy, FF-OCT is advantageous in terms of resolution, penetration depth, imaging speed, noninvasiveness, and cost, but its application depends on the performance optimization of this technology, such as enhancing image reconstruction efficiency, optimizing image quality, extending imaging depth and field, and mining image information. In recent years, with the advancement of multimodal FF-OCT, particularly dynamic full-field optical coherence tomography (D-FFOCT), researchers can obtain the dynamic information of biological tissues and cells nondestructively, which further expands the application range of FF-OCT.23,2632 It is of significance to review the research progress and application of FF-OCT to further promote the development of 3D high-resolution imaging technology and solve the problem of nondestructive detection in microscopic scale.

Beginning with an overview of the working principle and basic interference structure of FF-OCT, this work analyzes the possible optimization schemes of the system design based on the theoretical formula of its performance parameters. The latest research development of FF-OCT is reviewed in terms of imaging speed increase, imaging quality optimization, imaging flux enhancement, functional imaging extension, target information mining, and other performance improvements. The advantages and disadvantages of each specific imaging scheme are discussed along with its applicable scenarios. This work also provides examples of FF-OCT applications in biomedicine, materials science, security, and identification; further, it concludes by stating the technical bottlenecks in the development of FF-OCT.

2.

Summary of FF-OCT Imaging

2.1.

Working Principle and Basic Structure of FF-OCT

FF-OCT has simplified the system structure of conventional point scanning TD-OCT and line scanning FD-OCT by performing plane detection based on full-field illumination. Meanwhile, its acquisition speed is enhanced by full-field imaging via the CCD or CMOS camera.14 The imaging principle of the FF-OCT system is shown in Fig. 1. Its basic structure is a low coherence interference microscope system with a Linnik interference structure [Fig. 1(a)], spatial incoherent broadband light source. The plane array detector acquires the en face 2D image of the sample during one exposure.20 The frequently used interference structures in FF-OCT are shown in Fig. 1(b), including Michelson, Linnik, and Mirau interference structures.14,20,3335 The Linnik interference structure is the most common type used in FF-OCT, whose two identical high NA microscope objectives are placed on the sample arm and the reference arm, respectively. The optical path of the two arms can be flexibly adjusted, and the optical path length as well as the focus can be modified independently. No other optical path exists between the objectives of the two arms and the imaging target, thereby extending the selectable range of NA and working distance of the objectives. Using microscope objectives with high NA, FF-OCT can achieve ultrahigh spatial resolution (axial and transverse) at the subcellular level and obtain real-time noninvasive images of the sample tissues [Fig. 1(c)], whose detection results are comparable to those of tissue biopsy. Besides the interference type, different overall structures of FF-OCT systems have been proposed, for example, external spectral modulation FF-OCT system, endoscopic FF-OCT system, and compact miniaturized FF-OCT system. In FF-OCT devices with external spectral modulation, imaging utilizing spectrally modulated, spatially incoherent illumination and a static Linnik interferometer. The first demonstration of spectrally modulated imaging through a multimode fiber optic bundle represents an important step toward achieving endoscopic FFOCM.36 In a compact miniaturized FF-OCT device, the interferometer part of the setup was designed to be as compact and light as possible (11×11×5  cm3, 210 g), while keeping a Linnik interferometer design, which lays the foundations for a handheld device.37

Fig. 1

(a) Schematic diagram an FF-OCT. (b) Different optical set-ups of FF-OCT: Michelson-type, Linnik-type, and Mirau-type. Adapted from Refs. 19, 20, and 33. (c) FF-OCT images and tissue biopsy images of human forearm skin. Adapted from Ref. 22. (d) Tomographic reconstruction of simulated interference fringe with SPA and FPA, respectively. (e) Tomographic reconstruction of CaF2 crystal with SPA and FPA, respectively. (f) Tomographic reconstruction of embryo sample with SPA and FPA, respectively. Adapted from Ref. 19.

JBO_27_5_050901_f001.png

The tomographic images of FF-OCT are reconstructed from the interference signal acquired by its detector. The reconstruction scheme varies based on different light sources and signal detection methods. According to the light sources and signal detection methods, FF-OCT is categorized into two types, i.e., time-domain full-field optical coherence tomography (TD-FFOCT, developed from TD-OCT) and Fourier-domain full-field optical coherence tomography (FD-FFOCT, developed from swept-source OCT).

TD-FFOCT adopts the spatial incoherent broadband light source for illumination and can obtain the tomographic interference spectrum data of the sample front without transverse scanning, thereby enabling rapid 3D imaging. In TD-FFOCT, the interference signal captured by plane array detector, contains sample information and serves as the basis of tomographic image reconstruction. The signal of each pixel in the interferogram can be expressed as

Eq. (1)

I=I¯(x,y)+A(x,y)·cos[ϕ(x,y)+α],
where I¯(x,y) is the DC signal, α is the single phase shift, ϕ(x,y) is the initial phase, and A(x,y) is the intensity signal containing the information of light reflectivity of the tissue at the focal depth. Equation (1) contains three unknowns, i.e., I¯(x,y), A(x,y), and ϕ(x,y). To analytically solve A(x,y), which is the intensity of the tomography, at least three interferograms with different phases are required. The results of obtained tomography would be better if more interferograms are involved during reconstruction.19 Nevertheless, the acquisition time and the computational load of interferograms would also increase accordingly. Commonly used reconstruction methods include three-step, four-step, and five-step phase-shifting algorithms as well as single-step phase-shifting algorithms (SPAs) based on the Hilbert transform.3843

In general, to balance the acquisition time and reconstruction quality, the four-step phase-shifting algorithm (FPA) is the most adopted method for reconstruct tomography; the signals of four interferograms with a phase interval of π/2 are assigned as I1, I2, I3, and I4, corresponding to α as 0, π/2, π, and 3π/2, respectively, in Eq. (1). The intensity A(x,y) and phase ϕ(x,y) maps are calculated using Eqs. (2) and (3):

Eq. (2)

A(x,y)=(I3I1)2+(I4I2)2,

Eq. (3)

ϕ(x,y)=arctan(I4I2I3I1).

The five-step method is similar to FPA, except that it performs reconstruction based on five interferograms with different phases; therefore, the reconstruction speed of the five-step method is slower than that of FPA; its intensity and phase maps can be calculated using Eqs. (4) and (5):

Eq. (4)

A(x,y)=4*(I2I4)2+(I1+2*I3I5)2,

Eq. (5)

ϕ(x,y)=arctan(2*(I2I4)2*I3I5I1).
The comparison of the imaging results of the FPA and SPA based on Hilbert transform is shown in Figs. 1(d)1(f), in which tomographic reconstruction of simulated interference fringe and embryo sample was performed.19

Based on the Hilbert transform, SPA can reconstruct tomography with only two interferograms, thereby accelerating the imaging speed.40 The signals of interferograms Ia and Ib with a phase difference of θ are described using Eqs. (6) and (7):

Eq. (6)

Ia=I¯(x,y)+A(x,y)·cos[ϕ(x,y)],

Eq. (7)

Ib=I¯(x,y)+A(x,y)·cos[ϕ(x,y)θ].
Here, θ is the phase shift, I(x,y) is the background signal, A(x,y) is the intensity signal required to be solved, and ϕ(x,y) is the initial phase signal. The analytical solution of intensity signal can be expressed by Eq. (8). The contrast of the tomography generated by this method is impacted by the value of θ, and the best contrast effect can be achieved when θ equals π; nonetheless, when θ reaches 0.7π, the contrast effect can already achieve 90% of the optimum:42

Eq. (8)

A(x,y)=12sinθ2·(IaIb)2+[Hilbert(IaIb)]2.

Notably, as shown in Figs. 1(d) and 1(e), the fringes cannot be completely eliminated in the reconstruction of FF-OCT tomographic interference images via SPA based on Hilbert transform. However, when SPA based on the Hilbert transform is adopted for the tomographic reconstruction of biological samples, almost no issue of residual fringes exists because the randomly distributed micro/nanostructures in biological samples make the fringes in the reconstructed image invisible, as shown in Fig. 1(f).19 A simplified algorithm for digital fringe analysis in two-wave interferometry with sinusoidal phase modulation was proposed in 2017. The algorithm consists of simple mathematical combinations of four frames obtained by integration by a camera of the time-varying intensity in an interference pattern during the four successive quarters of the modulation period. The algorithm is invariant by circular permutation of the four image frames. Any set of four consecutive frames can be used for the calculations, which simplifies the practical implementation of the method.44 In addition to the above-mentioned algorithms with sinusoidal phase modulation, there is also the “Larkin” algorithm for digital envelope detection, which is a simple but highly effective nonlinear algorithm for envelope detection in white light correlograms. The method is applicable to any bandpass signals where either the envelope or the phase, or both, need to be detected.45

FD-FFOCT adopts spatial coherent swept light source combining planar array detector to perform time-sharing detection, which can achieve parallel detection in both frequency and spatial domain. The system structure is shown in Fig. 2(a); the axial information of spatial domain can be obtained from the frequency-domain information acquired by a pixel on the plane array detector.46,47 The image reconstruction aims to analyze the frequency-domain information of a plane (such as 512(x)×512(y)] then obtain the 3D volumetric data of spatial domain (as 512(x)×512(y)×1024(z)]; the processing flow is shown in Fig. 2(b). FD-FFOCT can conduct rapid 3D imaging without axial or transverse scanning, thereby overcoming motion artifacts caused by mechanical dither and repeated motion showing higher sensitivity and imaging speed. However, swept source is a type of spatial coherent light, and the backscattered light information of the corresponding sample points will be mapped to the wrong pixels during parallel detection, resulting in optical crosstalk and affecting the imaging results. The schematics of crosstalk generation are shown in Fig. 2(c). Reducing the spatial coherence of the light source or using multiangle illumination can diminish crosstalk and improve the imaging quality to a certain extent. Figure 2(d) shows the cross section of the 3D reconstruction of human forearm skin at different depths with reduced crosstalk.46

Fig. 2

Schematic diagram of an FD-FFOCT system that can suppress crosstalk and speckle. (a) Optical set-up. (b) Reconstruction process of the FD-FFOCT system. (c) Crosstalk caused by incorrect backscattered ray mapping. (d) Human forearm skin images acquired by the cross-talk-free FD-FFOCT. (a) and (c) Reprinted from Ref. 46.

JBO_27_5_050901_f002.png

2.2.

Key Performance Parameters of FF-OCT

The main performance parameters of conventional OCT and FF-OCT include axial resolution, transverse resolution, penetration depth, imaging field, imaging speed, detection sensitivity, and motion artifacts. Each performance parameter usually is mutually restricted.

The axial resolution of FF-OCT mainly depends on the coherence length of the light source, whose theoretical value is calculated using Eq. (9):

Eq. (9)

δz=lc22ln2nπ·λ02Δλ,
where lc is the coherence length, Δλ and λ0 are the spectral bandwidth and central wavelength of the light source, respectively, and n is the refractive index of the medium, which equals 1 in air and is between 1.3 and 1.5 in biological tissue. Equation (9) can serve as a criterion for selecting light source.48

The transverse resolution of FF-OCT is determined by the half peak width of the transverse point spread function, whose theoretical value is calculated using Eq. (10):

Eq. (10)

δxλ02NA.
Here, NA is the numerical aperture of microscope objective; λ0 is the central wavelength of the light source.

Imaging field, as a significant parameter of the system, is related to the NA of the microscope objective and the central wavelength of the light source; its theoretical value can be estimated using Eq. (11):

Eq. (11)

FOV0.3592λ0NA2.
A large NA can increase the transverse resolution while reducing the imaging field. Therefore, it is inadvisable to excessively pursue high transverse resolution and neglect its adverse effects. In fact, FF-OCT can achieve subcellular level by just choosing an appropriate NA (0.3 to 0.5).

Light source parameters, such as wavelength, bandwidth, spatial incoherence, and power will directly impact the imaging performance of the system. Figure 3(a) shows the absorption curves in each spectral band of several common molecules or substances in biological tissues, including water, aorta, oxygenated hemoglobin (HBO), and melanin. Considering the spectral response range and tissue imaging window of the plane array detector, the central wavelength of common light sources for FF-OCT is mainly 500 to 1000 nm.44 In this spectral band range and the NA range of 0.1 to 1.2, the transverse and axial resolutions of FF-OCT system are visualized, and the results are demonstrated in Figs. 3(b) and 3(c). The axial/transverse resolution data of typical light sources and microscope objectives are given here, as shown in Table 1. The types of light sources include light-emitting diode (LED), tungsten halogen lamp (THL), superluminescent diode (SLD), THL was first applied to the FF-OCT system in 2002.50 In the recent decade, due to the development of LED technology, the use of LED light sources became more and more common. These light sources are covered in Table 1. The central wavelength of the light source and NA of the microscope objectives can be preliminarily determined based on these results.

Fig. 3

(a) Spectral absorption curve of common substances or molecules in biological tissues. Adapted from Ref. 48. (b) The influence of NA and center wavelength on transverse resolution. (c) The influence of spectral bandwidth and central wavelength on axial resolution.

JBO_27_5_050901_f003.png

Table 1

Axial/transverse resolution data for common light sources and microscope objectives.

Parameters of the light sourceParameters of water immersion objectivesδxδz
λ0ΔλTypeNAMagnification
565 nm49104 nmLED0.25 NA10×1.5  μm1.35  μm
565 nm41104 nmLED0.8 NA10×0.35  μm1.35  μm
660 nm3120 nmLED1.05 NA30×0.41
660 nm2720 nmLED0.8 NA40×0.5  μm1.7  μm
800 nm50235 nmTHL0.3 NA10×1.3  μm1.2  μm
840 nm1430 nmLED0.25NA10×2  μm8  μm
850 nm3140 nmLED0.3 NA10×1.7  μm7.7  μm
1309 nm3926 nmSLD**70  μm29  μm
Note: “*” indicates that no relevant data is provided in the paper.

The detection sensitivity of the system can be expressed by the minimum intensity signal that can be detected by the detector, which is one of the key parameters affecting the contrast and imaging depth of the tomogram. Its theoretical value can be calculated using Eq. (12):

Eq. (12)

R=K(ξ+χ2M*ξ)(Rinc+Rr)2Rr.

The detection sensitivity is defined as 10×log(R), where K is a fixed constant, ξ is the full-well capacity of the camera, M is the superposition times of interference signals, χ is the electrical noise signal during data collection, Rinc is the light reflectivity of the sample, and Rr is the light reflectivity of the reference mirror in the reference arm. The detection sensitivity of FF-OCT system can be enhanced by increasing the full-well capacity of the planar array detector, using a higher-power light source, and increasing the superposition times of interference signals.27 The performance of the camera affects that of the FF-OCT system; Professor W. Y. Oh’s team first applied an indium-gallium-arsenide (InGaAs) camera to the FF-OCT system in 2006. InGaAs area array cameras with a good responsivity in the 0.9 to 1.7  μm wavelength range may enable imaging at longer wavelengths, thereby increasing the penetration depth of FF-OCM.51 Moreover, the use of water or oil immersion on microscope objectives not only can improve the matching degree of the microscope objective and sample refractive index but also enhances the system sensitivity to a certain extent.19,41

3.

FF-OCT Performance Optimization Methods

Compared with conventional OCT, the overall performance of FF-OCT is superior in terms of faster imaging speed and better resolution. However, in the practical application events of FF-OCT, constraints of the system implementation and insufficiency of image information extraction cannot be neglected. The overall performance and applicable scope of FF-OCT can be enhanced via technological innovation. The ways to improve performance can be divided into five directions: (1) quickening the imaging speed, (2) optimizing the imaging quality by improving the contrast and decreasing the dispersion, aberration, and motion artifacts, (3) extending the imaging depth and field of view (FOV), (4) additional contrast via multimodal imaging for providing complementary information in addition to structure; and (5) combining deep learning for image information mining. Optimization of the specific implementation scheme of FF-OCT can improve its application effect in the field of microscopic detection, thus enabling different application scenarios and requirements. The review of FF-OCT performance optimization methods can help researchers to fully understand this submicron high-resolution imaging technology and overcome the issues in system implementation under the requirements of innovative applications.

3.1.

Increasing the Imaging Speed

The area array sensor and high NA microscope objectives adopted by FF-OCT can achieve submicron-level resolution while ensuring fast imaging speed. However, realizing real-time in vivo imaging remains challenging, and enhancing the overall imaging speed of the system is still a research focus. Based on the imaging principle of FF-OCT, reconstruction algorithms involving fewer interferograms as well as phase shifts with higher speed and accuracy can enhance the imaging speed of the system. The specific reconstruction algorithm of FF-OCT tomography is detailed in Sec. 2.1. The acquisition of interferograms with different phases is achieved by introducing a phase shift, which can be categorized into two methods depending on whether mechanical displacement is involved: mechanical phase shifting using the piezoelectric ceramic (PZT) phase shifters and instantaneous phase shifting without mechanical displacement using wave plates.

PZT phase shifting is the most common approach, which uses the piezoelectric effect to produce accurate small displacement to change the position of the reference mirror for phase shifting. PZT phase shifting has two methods, i.e., fixed-step phase shifting and carrier frequency phase shifting. The phase of the fixed-step approach changes discretely between intensity measurements, and the data acquisition is conducted after phase shifting. The carrier frequency method uses sine waves to enable the reciprocating motion of PZT and uses a square wave with a frequency four or five times the sine wave to trigger the detector and collect interferograms, which can accelerate the imaging speed but remains in the category of discrete phase shift.50 By combining the PZT phase shift with the instantaneous phase shift using wave plate, two movements of PZT can achieve four-step phase shift results by constructing the optical Hilbert transform.52 This dual-channel structure with two detectors can enhance the overall phase shifting speed. In the genuine instantaneous phase shift without mechanical displacement, the combination of a half-wave plate and linear polarizer introduces a phase difference of π between the optical paths of two arms; two interferograms with a phase difference of π are simultaneously acquired by two cameras. The SPA based on the Hilbert transform is then used to reconstruct the tomography, the construction of which is shown in Fig. 4(a).

Fig. 4

(a) Schematic diagram of dual-channel instantaneous phase-shifting FF-OCT set-up. Adapted from Ref. 40. (b) Instantaneous phase-shifting FF-OCT set-up of a single channel with two divisions. Adapted from Ref. 39. (c) Instantaneous phase-shifting FF-OCT set-up of a single channel quarter, and four interferograms collected at the same time with a phase interval of π/2. Scale bar: 100  μm. Reprinted from Ref. 41.

JBO_27_5_050901_f004.png

Industrial cameras with high well capacity and high frame rate are expensive; thus, using two cameras would increase the system cost. The dichotomy of a single channel can be achieved using a Wollaston prism, then different sensor regions of one camera can simultaneously detect two quadrature polarized light fields, as shown in Fig. 4(b).39,43 Currently, the instantaneous phase-shifting structure with a quartered single channel has also been achieved, as shown in Fig. 4(c), which can simultaneously acquire four phase-shifting images using a single camera. Compared with the PZT four-step phase shift, this method can significantly enhance the imaging speed. Instantaneous phase-shifting approaches with a dichotomized or quartered single channel can reduce the system cost and enhance the imaging speed, but it would also decrease the number of pixels available for single imaging on an industrial camera, hence rendering a certain impact on the transverse imaging range of the system.

Each phase shifting approach varies in phase shifting speed and quality, economic cost, system complexity, and imaging field. Several phase shifting implementation schemes are listed in Table 2, which can serve as the reference for selecting phase-shifting approach during system design. The number of asterisks in Table 1 represents the degree of excellence regarding the performance, and more asterisks indicate a better performance.

Table 2

Features of different FF-OCT structures.

Reconstruction algorithm + the way of phase shiftingA channel divided areaReconstruction speedReconstruction qualityCCDStructure complexity
Single-step algorithm + PZT421(***)(*)1(*)
Three-step algorithm + PZT191(***)(*)1(*)
Four-step algorithm + PZT141(**)(***)1(*)
Five-step algorithm + PZT191(*)(***)1(*)
Single-step algorithm + wave plate + Wollaston prism39,402(****)(**)1(**)
Single-step algorithm + wave plate522(****)(**)2(**)
Four-step algorithm + wave plate + PZT432(***)(***)2(**)
Four-step algorithm + wave plate38,414(****)(****)1(***)

3.2.

Enhancing Imaging Quality

To enhance imaging quality, the following aspects can be considered: reducing external noise to improve the accuracy and sensitivity of signal acquisition, reducing aberration to improve resolution, and reducing image noise to increase contrast. The FF-OCT image quality can be enhanced by either hardware or by algorithm and software.

Phase shifting and sample dithering during FF-OCT imaging may render motion artifacts and confuse image signals and noise, thereby impairing signal acquisition accuracy.53,54 The instantaneous phase shift without mechanical displacement mentioned in Sec. 3.1 is an approach to enhance the speed and accuracy of phase shifting as well as to reduce motion artifacts.55 And methods with two frames from the two outputs of the interferometer, en face tomographic images are calculated as a combination of two phase-opposed interferometric images acquired simultaneously by two CCD cameras placed at both outputs of the interferometer, with a spatial resolution of 0.8  μm×1.6  μm (axial × transverse). This new FF-OCT system, which uses instantaneous phase shifting with nonpolarizing optics and pulsed illumination, can suppress most of the artifacts that arise due to sample motion.53 Motion artifacts can also be reduced via image processing; as shown in Fig. 5(a). The method based on singular value decomposition (SVD) can effectively filter the motion artifacts in collagen fibers, rendering a clearer display of cells.56,60,61 When imaging biological tissue, an interface would occur between cover glass and tissues. If the refractive indexes of the two materials do not match, a strong “specular reflection” would occur, which can severely interfere with the recognition of backscattered signals of biological tissue. The dark-field FF-OCT can block specular reflections by the help of an opaque disk in the pupil-conjugated plane; the reference mirror is replaced by a blazed grating, which eliminates a walk-off between the sample and the reference beams on a camera that otherwise limits the imaging FOV, which suppresses specular reflection by at least two orders of magnitude.49,62 Therefore, high-contrast tomography can be captured at the cover glass–tissue interface. The results of dark-field detection are shown in Fig. 5(b), exhibiting a significantly increased contrast.

Fig. 5

Image enhancement methods for FF-OCT. (a) The method based on SVD can effectively filter the motion artifacts in collagen fibers, rendering a clearer display of cells, pointed by the red arrow in the figure; (b) FF-OCT image and dark-field FF-OCT image recorded at the interface of cover glass and sheep brain tissue slice. Adapted from Refs. 49 and 56. (c) The process of deep learning network used to remove noise signals in FF-OCT images. Adapted from Ref. 57. (d) Comparison of FF-OCT images of a Ficus leaf before and after sample self-induced aberration was corrected when imaging at a depth of 75  μm. Adapted from Ref. 58. (e) The difference in refractive index between the microscope objective lens and air causes the coherence plane and the focal plane to separate. Adapted from Ref. 59.

JBO_27_5_050901_f005.png

The coherence plane and the focus plane of the FF-OCT system are usually in the same plane. However, when imaging biological tissues with a dry objective, the difference between the refractive indexes of light in air (nim) and biological tissues (n) would separate the coherence plane and the focus plane, leading to defocus. As shown in Fig. 5(e), defocus would blur images reconstructed by FF-OCT. The differences in refractive index between the sample and the immersion medium of the microscope objectives may degrade the FFOCM image quality because of focus defect and optical dispersion mismatch. Dubois59 studied the effects of focusing defects and dispersion mismatch in full-field optical coherence microscopy. Performing water immersion or oil immersion objectives can ensure that nim and n are close, for example, water immersion objectives can be used to mitigate defocus for imaging macrophages in Petri dishes.19,41 The multiscale aberration structure of biological tissues would diminish the imaging resolution of different depths. The adaptive optics (AO) system and OCT can be combined to correct the inherent aberration of biological tissues, in which AO devices can compensate for the signal loss caused by tissue wavefront distortion as well as improve the image clarity. More structure information can be observed due to AO correction, as the image shown in Fig. 5(d).58,6365

Inherent speckle noise exists in FF-OCT, which would generate a grainy pattern and impair image quality. When it cannot be eliminated by hardware, certain image-processing algorithms can be adopted to reduce speckle noise, such as Gaussian filtering, median filtering, deep learning denoising, and other algorithms.66,67 In recent years, deep learning algorithms have achieved satisfying results in the noise reduction of FF-OCT images and paved a new way to improve FF-OCT image quality. The fundamental approach for noise elimination based on deep learning is to distinguish noise from signals by combining spatial mixing and noise spectrum prediction technology.57 As shown in Fig. 5(c), a neural network of two-phase image-spectrum conversion is adopted to reduce the noise of the FF-OCT image. By averaging adjacent B-scan images, speckles were reduced and the denoise database was obtained. A denoising convolution neural network based on the spatial-compounding (SC-DnCNN) was trained for noise recognition in the denoise database. After training, SC-DnCNN is used to eliminate the speckle noise of FF-OCT and output clean images.

3.3.

Improving Imaging Throughput (FOV and Depth)

Large imaging range and high spatial resolution are the two goals expected to be simultaneously achieved for biomedical imaging, but mutual restraint exists between these two goals. To expand the imaging FOV of the system on the premise of ensuring a high spatial resolution of FF-OCT, a scanning mechanism can be added to the detection unit of the system for transverse stitching of tomograms. Alternatively, a multiconjugate AO structure can also be used to obtain a wider imaging FOV.63,65,6870 The disadvantages of these approaches are as follows: transverse stitching of multiple elongates the imaging time, and the accuracy of image stitching is easily influenced by motion artifacts. The multiconjugate AO device increases the system complexity and shrinks the actual effective imaging field of a single AO device. Currently, coherent shaping is adopted to expand the effective imaging FOV of the FF-OCT system: the eye serving as the objectives breaks the symmetry of the Linnik interferometer in the FF-OCT system during retinal imaging, forming a curved coherent gate. Therefore, the effective imaging FOV of the FF-OCT system becomes smaller.7173 The reference arm is equipped with an electric displacement platform to adjust the position of the coherent gate, and the sample arm is equipped with an optical window element to correct the geometry of the coherent gate. As shown in Fig. 6(a), the geometry of the coherent gate matches the curvature of the imaging sample, thereby expanding the effective imaging FOV. In retinal imaging, the effective imaging FOV after coherent gate shaping is three to four times larger than that before shaping [Fig. 6(b)].

Fig. 6

(a) Schematic showing different degrees of curvature for the coherence gate and their consequence on the FOV of the FF-OCT image, the left picture shows the coherence gate curvature does not perfectly match the retinal curvature, the middle picture shows the coherence gate is reshaped, although the curvature does not completely match the retinal curvature, a full FOV is still generated, as the coherence gate has a given thickness defined by the axial resolution. Adapted from Ref. 72. (b) The retina FF-OCT images before and after correcting for the coherence gate curvature. Adapted from Ref. 72. (c) FF-OCT imaging system based on coherent reflection matrix. Adapted from Ref. 74. (d) Samples with high scattering coefficient to verify deepening effect of imaging depth. Adapted from Ref. 74.

JBO_27_5_050901_f006.png

Owing to the existence of aberration and multiple scattering, the imaging depth of FF-OCT for biological tissues is mostly limited to 1  mm, whereas the optimal imaging depth is limited to a few hundred microns. There are currently two approaches to increase the imaging depth of the FF-OCT in the scattering medium as much as possible. One approach utilizes optical clearing or other hardware, such as a light source with appropriate central wavelength based on the scattering properties of the sample or optical devices to reduce the imaging aberration. Another approach adopts a computational matrix, such as ballistic wave identification combining matrix calculation and iterative time reversal, which theoretically and experimentally confirmed that the imaging depth can be increased by at least two times.75 Chen’s group at University of California, Irvine (UCI) proposed a calculation method based on coherent reflection matrix measurement, which confirmed that the imaging depth can be extended. The reflection matrix is obtained by point-by-point scanning and phase shift, then SVD is conducted to retrieve the single backscattered light.74 Combining the reflection matrix measurement and wide-field heterodyne detection, satisfactory tomography of the microbeads through highly scattering medium is obtained with a doubled imaging depth; the system principle and imaging results are shown in Figs. 6(c) and 6(d).

3.4.

Functional Imaging Expansion

Like OCT with spot scanning, FF-OCT can also develop multimodal composite imaging, which can be combined with one or more technologies such as dynamic imaging,30 fluorescence microscopic imaging,24,25 elastography,76 polarization-sensitive,77 and photothermal imaging78 and hyperspectral imaging.79 Therefore, multimodal images of one sample can be collected at the same position, which can not only provide structure information with subcellular-level resolution but also render data regarding tissue function.23,25,31

As an extension of FF-OCT, dynamic functional imaging detects intracellular motility by analyzing the signal fluctuation. It provides a method of label-free cell imaging, which is a mostly a nondestructive detection method for tissue or cell dynamics compared with FM. As shown in Fig. 7(a) shows the signal acquisition process of dynamic FF-OCT, where the camera collects the interference images of the sample in temporal sequence and obtains a 3D data matrix. Temporal signal on each pixel is collected, and it is converted from the time domain to the frequency domain for power spectrum analysis. Dynamic FF-OCT uses cell motility as the endogenous contrast agent and exploits metabolically driven dynamic scattering changes of cellular structures to enhance contrast.30 Dynamic FF-OCT images serve as a useful supplement to static FF-OCT images, as static FF-OCT has high imaging contrast for static tissues such as fibers, blood vessels, and collagen, while dynamic FF-OCT is sensitive to cell membrane floating and cell surface reconstruction. Therefore, the superposition of static and dynamic FF-OCT images can enhance the overall visibility of tissue morphology. Due to the motion discrepancy between cytoplasm and connective tissue, dynamic FF-OCT images can clearly display the cell boundaries and demonstrate the hidden cells in static FF-OCT with high contrast. As shown in Fig. 7(b), the dynamic image of liver tissue exhibited clearer cell structures and boundaries than those in static images. The superposition of dynamic and static FF-OCT images of mouse brain tissue are shown in Fig. 7(c), which has significantly enriched the information capacity and contrast of the images.30

Fig. 7

(a) Signal acquisition process of dynamic FF-OCT. Adapted from Ref. 27. (b) Static and dynamic FF-OCT images and partial enlarged images of rat liver. Adapted from Ref. 30. (c) Static/dynamic composite FF-OCT images of mouse brain cortex. Adapted from Ref. 30. (d) Dynamic images are computed in the HSV color space; hue is computed with the mean frequency, from blue (low temporal frequencies) to red (high temporal frequencies); saturation is computed as the inverse of the frequency bandwidth. The value is computed as the running STD. Adapted from Ref. 27. (e) Computation of the mean frequency (hue); frequency bandwidth (saturation); dynamic (value) grayscale image; dynamic FF-OCT of retinal organoid in the HSV color space. Adapted from Ref. 27.

JBO_27_5_050901_f007.png

Dynamic FF-OCT imaging is calculated based on the standard deviation (STD) to obtain the interference signal fluctuation over time.30 The time-lapse sequence signal of the intensity I(x,y,ti) is acquired and processed using Eq. (13):

Eq. (13)

DSTD(x,y)=1Ni=1N(I(x,y,ti)I¯(x,y))2,
where I(x,y,ti) is the signal intensity and I¯(x,y) is the average of all signals in time dimension. To achieve better results of dynamic images, the time-lapse sequence signal I(x,y,ti) is divided into multiple subdatasets; their STDs are first calculated separately, then several values are averaged. If the acquisition time interval between successive subdatasets is short, faster changes of movement can be captured while slower changes would be filtered; if the acquisition time interval is long, slower changes of movement can be captured while faster changes would be filtered.

To further enhance the display of dynamic images, the signal obtained by the STD method is analyzed by power spectrum to generate “frequency-power spectrum” data. The analysis results of STD power spectrum are displayed in different color spaces, reflecting cell motility. For instance, the cell dynamics can be divided into three normalized frequency bands by RGB color space, in which R, G, and B represent low, medium, and high frequency cell movements, respectively.80,81 Alternatively, the hue–saturation–value (HSV) color space can also be used, and the image brightness reflects the amplitude of cell movement while the image color is related to the cell motion speed.27,29 The HSV color display approach and the results of dynamic FF-OCT imaging of retinal organoids are demonstrated in Figs. 7(d) and 7(e). Colored dynamic images can demonstrate the overall activity distribution and local abnormal activity distribution of the sample.

In addition, dynamic imaging signals can also be analyzed using autocorrelation (AUTO),82 logarithmic intensity variance (LIV),8285 and other methods. The specific calculation principles are provided in Eq. (14) and Eq. (15):

Eq. (14)

DAUCO(x,y)=(1N1i=1N1I(x,y,ti)I(x,y,ti+1))I¯(x,y)2,

Eq. (15)

DLIV(x,y)=1Ni=0N1(10logI(x,y,ti)10logI¯(x,y))2.
By comparing the imaging results of various methods with data from the same sample, a more detailed intuitive understanding can be developed for each method. A data set of the colorectal cancer organoid (Organoid-O, Regenovo Ltd., Hangzhou, China) were acquired by exploiting the self-developed OCT equipment, and three dynamic imaging methods were adopted to process the same OCT data. The results are shown in Fig. 8. For the LIV method, the variance was calculated using the logarithm of intensity signals, and the static and dynamic components in the intensity signal were transformed into an additive form. Therefore, the static component can be easily subtracted by separating the moving tissue sample from the background [Fig. 8(b)]. For the STD method, the STD was directly calculated based on the intensity; thus, the images reflecting the motion frequency of samples were obtained. However, the speckle noise in the background cannot be well separated from organoids [Fig. 8(c)], which should be further processed, i.e., assigning color in the HSV channel according to the motion frequency [Fig. 8(e)] and eliminating the background noise [Fig. 8(f)]. The AUTO method makes the organoids with a low physical position in the image (far away from the focal position with low signal-to-noise ratio) obtain the signal intensity close to the sample at the focal plane and increases the overall signal-to-noise ratio of the image [Fig. 8(d)]. It must be highlighted that it is difficult to directly detect the dynamic discrepancy of different tissue layers using FF-OCT and D-FFOCT because only the en face tomography is obtained. Dynamic OCT imaging can directly visualize depth-related cell changes and structural maturity with time-spectroscopic analysis of cross-sectional tomography, which is particularly effective for the physiological and pathological analysis of epithelial tissue.80,81

Fig. 8

Cancer organoid imaging using different dynamic OCT method. (a) SD-OCT image of colorectal cancer organoids which are cultured in Matrigel. (b) Dynamic OCT image with LIV method. (c) Dynamic OCT image with STD method. (d) Dynamic OCT image with AUTO method. (d) Dynamic images in the HSV color space with STD method (256 fps SD-OCT images is used for calculation), where the orange-red background in the picture is due to the rapid movement of the inherent speckle noise of SD-OCT. (f) Dynamic images in the HSV color space with STD method with background noise removed using a mask. Scale bar: 125  μm.

JBO_27_5_050901_f008.png

3.5.

Combination of FF-OCT Images and Deep Learning

Intelligent recognition of the important features or regions in FF-OCT images as well as the pathological or physiological information contained in images is a research focus in this field, which will bring revolutionary improvement to the clinical diagnostic application of FF-OCT.86

In the latest research, deep learning has been exploited in the target recognition of FF-OCT images. For instance, U-net neural networks are used to identify red blood cells in skin microvessels from FF-OCT images with a subcellular-level resolution;87 multidirectional CNN is applied for predicting the possible pixels of epidermis in 3D data to enhance the accuracy of distinguishing dermal–epidermal junction (DEJ);88 the OCT image data of enhanced capillary network generated by automatic simulation algorithm are combined with the U-net neural network to achieve automatic segmentation of skin capillaries.89 The recognition of specific targets, particularly the recognition of structural features related to pathology, is crucial for the early diagnosis of severe diseases (e.g., cancer) and the in-depth understanding of tissue physiology and pathology. The combination of FF-OCT and deep learning can also enhance the accuracy of tissue segmentation [Fig. 9(c)]. A Pruned-ResNet-18 classifier based on the CNN network can be combined with FF-OCT images with subcellular-level resolution to detect skin squamous cell carcinoma, which has an accuracy of 80% in the early detection of this disease; the recognition results are shown in Fig. 9(f).90 Moreover, the automatic classification algorithm based on a support vector machine is exploited to distinguish the adipose tissue in the hollow structure of breast tissue; combined with the texture features from ultrahigh-resolution FF-OCT image, the invasive ductal carcinoma (IDC) tissue, and normal fibrous matrix can be further distinguished.91

Fig. 9

FF-OCT image processing using deep-learning techniques. (a) FF-OCT image of the DEJ of human skin. Adapted from Ref. 88. (b) Hematoxylin–eosin staining (H&E) image of DEJ of human skin. Adapted from Ref. 88. (c) Deep learning algorithm performs DEJ detection on FF-OCT images of human skin, where the red line is the DEJ manually marked by the doctor, and the green one is the automatic recognition result by the deep learning algorithm. Adapted from Ref. 88. (d) FF-OCT image of mouse squamous cell carcinoma skin, where the red region is the stratum corneum (SC), the purple region is the epidermis (EPI), and the green region is the nucleus (NUC). Adapted from Ref. 90. (e) H&E image of mouse squamous cell carcinoma skin; adapted from Ref. 90. (f) The result of deep learning tissue classification algorithm for squamous cell carcinoma detection in FFOCT images of mouse skin, where the yellow circles represent the identified cancerous cells. Adapted from Ref. 90.

JBO_27_5_050901_f009.png

The combination of high-resolution FF-OCT images and deep learning has especially revolutionized histopathological diagnosis, but it still has the shortcoming of manual marking, which is time- and labor-intensive, and the marking accuracy is influenced by precise identification of tissue characteristic margins. The colored images of cells and tissues with high resolution and contrast from dynamic FF-OCT can directly serve as characteristic marking images to train the deep learning network. Currently, based on feather engineering and the automatic diagnosis algorithm of CNN, some researchers performed the auxiliary diagnosis of cancer with an accuracy of nearly 100% using the high-resolution colored image of D-FFOCT at both small scale (2  mm2) and macroscale (2  cm2). This has verified the huge potential of combining D-FFOCT and deep-learning for fast, automatic, and accurate histopathological diagnosis.92

4.

Applications of FF-OCT

Compared with FF-OCT, OCT has a larger imaging FOV and imaging depth and has been widely used in materials science, in situ process monitoring, and tissue engineering.8083,9395 But FF-OCT has isotropic micrometric resolution and has the ability of producing a histology-like image from a thick sample within a few milliseconds without physical slicing. Therefore, this technology is also known as optical biopsy. Benefiting from the advancement of the overall performance of the FF-OCT system, FF-OCT has appeared in many fields, such as biomedicine, materials science, security, and identification.

4.1.

Biomedical Imaging Applications

FF-OCT can perform feature recognition from the microstructural level to cell level in ophthalmology, dermatology, oncology, and other fields. OCT has been most widely used in ophthalmology; compared with currently commercialized FD-OCT, FF-OCT can characterize cells and tissues as well as present pathological structure details.31,54,65,71,72,96100 In the early stage of FF-OCT development, high-quality, high-resolution tomographic images can be obtained in some fields. For example, for retinal imaging, as early as 2004, some researchers used FF-OCT detected high-resolution retinal images and achieved the effect of distinguishing different structures.15 In addition, in other early applications, it also included tissue engineering scaffolds, etc.101 Even though high-quality imaging cannot be achieved due to hardware limits, FF-OCT opened a door for unique imaging performance, which made an important contribution to the development and application of FF-OCT. In recent years, imaging quality of FF-OCT has observed a significant improvement, Fig. 10(a) depicts the nerve fiber layer, ganglion cell layer (GCL), inner plexiform layer, outer plexiform layer (OPL), and outer nuclear layer (ONL) of macaque retina. Besides, due to its subcellular resolution, FF-OCT can detect the cellular characteristics of each layer within the retina such as the details of ganglion cell axons and cell bodies (yellow box) demonstrated in Figs. 10(a) and 10(b). Wide-field images are important for the early diagnosis of retinal diseases in clinical settings. Therefore, FF-OCT with a large imaging FOV, high resolution, and fast imaging speed has a prospective future in the field of ophthalmic detection. As shown in Fig. 10(c), the AO technology was adopted to obtain the retinal image of 5deg×5deg FOV, then five images were stitched into a 12deg×12deg FOV via a certain stitching method, in which the color map represents the calculated photoreceptor density. Both coherent gate shaping and AO technology can be used to expand the imaging field of FF-OCT during retinal detection. Moreover, optical phase modulation via movement of the eye itself to retrieve tomographic images of TD-FFOCT has also been achieved. The combination of this method with coherent gate shaping technology has the potential to facilitate the clinical application of FF-OCT in the ophthalmology field.54

Fig. 10

(a) Merged static (red) and dynamic (green) FF-OCT images on the macaque retina to identify stationary (fiber, blood vessel, collagen) and dynamic (cell) structures. Adapted from Ref. 28. (b) The FF-OCT image shows the details of the cell axons and cell bodies of the human retinal GCL, which can be used to check the details of the damage. Adapted from Ref. 96. (c) Combine AO and stitching methods to get the FF-OCT image of retina with a larger FOV. Adapted from Ref. 71.

JBO_27_5_050901_f010.png

Combining FF-OCT with other imaging technologies to perform multimodal image acquisition at the same position of the sample can significantly enrich imaging contrast and provide functional information besides structure. Therefore, FF-OCT performance in medical diagnosis can be greatly enhanced, exhibiting clinical potential in early cancer diagnosis and serving as guidance in tumor resection. As shown in Figs. 11(a)11(d), the reported applications include, but are not limited to, the discrimination of benign and malignant tissues of skin cancer,35,103,105 breast invasive ductal cancer,102,106 head and neck cancer,104 as well as optical biopsy of intestinal tissues,107,108 detection of esophageal mucosa, liver, and gastric mucosa.86,102,106,109113 Besides, FF-OCT incorporating artificial intelligence can benefit deep mining of the abundant image data, which can reduce the difficulty of human eye recognition, improve the accuracy of information recognition, and promote the efficiency of FF-OCT histological diagnosis.

Fig. 11

(a) D-FFOCT image of esophageal mucosa at a 10  μm depth, where the epithelial cell cytoplasm (white arrows) and nuclei (blue arrows), pericellular lining (blue arrowheads), lamina propria papillae (white arrowheads). Adapted from Ref. 109. (b) D-FFOCT image of invasive ductal carcinoma of breast tissue, in which nested cells with malignant features have local low intensity. Adapted from Ref. 102. (c) FF-OCT image of mouse skin squamous cell carcinoma where the cancerous tissue in the black box. Adapted from Ref. 103. (d) FF-OCT image of head and neck cancer, where the yellow dotted arrow points to the stromal reaction, the yellow cross mark points to the cancerous tissue, and the yellow asterisks points to the keratin pearls. Adapted from Ref. 104. (e) In the first line, D-FFOCT monitors the dynamic profile (the boundary between the two cells) changes during cell differentiation; the second line, the very active area in the organoid, is composed of cells that exhibit high dynamics, and may be undergoing apoptosis. Adapted from Ref. 27. (f) D-FFOCT monitors the activity and morphology of Hela cells over time. Adapted from Ref. 32.

JBO_27_5_050901_f011.png

FF-OCT can provide high-resolution structural and functional information of 2D or 3D cell cultures and 3D-printed cell scaffolds under a normal culture environment.15,34,114 Researchers have used dynamic FF-OCT to conduct long-term longitudinal comparative studies on organoid samples, including long-term observation of cell apoptosis, proliferation, or migration as well as observation of the dynamic contour changes of tissues during organoid cell differentiation [Fig. 11(e)]. Moreover, the dynamic and static dual-mode of FF-OCT detection has also been exploited to monitor the changes in cell viability in cancer cells (e.g., Hela cells) and complex culture systems [Fig. 11(f)]. The ability of FF-OCT to track the growth and development of cells in vitro can monitor the cell viability change in disease modeling or artificial tissue transplantation as well as reveal the mechanism of structural and functional reconstruction of 3D cultured tissues such as organoids and tumor microspheres, thereby facilitating research on the mechanism of human development and disease.15,26,27,32,34,114 The functional imaging extension of FF-OCT is also applied in spectroscopic FF-OCT, with different pseudocolors showing melanocytes in ex vivo Xenopus laevis tadpole embryos at high resolution.115,116 In addition, it also includes superimposing the fluorescence microscopic image of the mouse colon and the FF-OCT image, so more information can be displayed in one image;117 the images obtained at the same time at the center wavelength of 800 and 1200 nm are given different colors, respectively. Superimposed multimodal imaging was applied to detect rabbit trachea, which can increase the contrast of different tissue structures, significant improvement in imaging penetration depth was demonstrated at 1200 nm compared to 800 nm, and a color image representation was applied to fuse both images into a single image for better comparison of wavelength effects.118

4.2.

Expansion of FF-OCT Applications

Benefiting from the enhancement of the overall performance of FF-OCT system, its potential in scientific research and application of materials science, security, and identification has gradually been valued by researchers. FF-OCT can nondestructively acquire 2D tomography of a certain depth with high axial and transverse resolution.119121 Therefore, this technology can detect the specific spatial location and detailed characteristics of material damage to analyze the degree and cause of the damage. For instance, FF-OCT can be used to assess the subsurface (dozens of microns below the surface) damage of optical glass [Fig. 12(a)] and observe the internal and external damage of hair [Fig. 12(b)].122,123

Fig. 12

(a) Subsurface damage of optical glass using FF-OCT. Adapted from Ref. 122. (b) Human hair damage detection. Adapted from Ref. 123. (c) FF-OCT cross-sectional views of different depths of banknotes, used for authenticity identification. Adapted from Ref. 124. (d) FF-OCT fingerprint patterns at different depths, from top to bottom are external fingerprints (0  μm), internal fingerprints (300  μm), and sweat pores (200  μm). Adapted from Ref. 125.

JBO_27_5_050901_f012.png

The unique internal characteristics of security objects can be adopted as new indicators of anticounterfeiting measures. Noninvasive detection of the internal details of banknotes by FF-OCT can be used to identify their authenticity [Fig. 12(c)].124 The internal fingerprint detection can function as identity recognition or a password [Fig. 12(d)], which is more secure than the surface fingerprint. Modern forgery is highly specialized, and the appearance of imitations may be mistaken as genuine.124127 Therefore, FF-OCT is attracting increased attention as an effective tool for authenticity identification of certificates, paintings, and porcelain, exhibiting a broad range of application prospects in anticounterfeiting science.

5.

Summary and Discussion

This work explicitly introduced the working principle and performance parameters of the FF-OCT system and summarized the latest advances and applications of FF-OCT performance optimization. Relevant technologies for increasing the imaging speed (acquisition, phase shift, and reconstruction), imaging quality (optimization of contrast and signal-to-noise ratio), and imaging throughput (extension of imaging FOV and depth), as well as adding complementary contrast to structural images and information mining have also been detailed in this work.

Although FF-OCT has undergone significant progress, the following aspects require in-depth investigation, which could become the focus of innovative research in this field:

(1) Combining optical field control technology and computational optics to extend the actual detection depth of FF-OCT. The transverse and axial resolution of FF-OCT can reach submicron levels, but the effective imaging depth is limited to 1 mm due to multiple scattering and aberration of the sample itself. The FF-OCT imaging method, which is based on measuring of the time-gated reflection matrix of a modulated optical field and matrix SVD theory, can eliminate a majority of the multiple scattering and effectively extend the imaging penetration depth.

(2) Enhancing the information capacity and information extraction efficiency of FF-OTC images by mathematical methods or signal processing. Multimodal OCT is a conventional method to achieve structure–function composite imaging; however, owing to the difference in acquisition methods, the FF-OCT optical system must be redesigned when incorporating laser confocal, two-photon, and other imaging technologies. If the information in the existing spectral interference signals can be mined from the perspective of signal processing, the contrast of the original image can be increased and new functional information can be added. For instance, dynamic FF-OCT images can more clearly display cell boundaries and cell activity states with high contrast. Further, the combination of deep learning and super-resolution FF-OCT images with high information capacity can efficiently mine the valuable information of images, thereby exhibiting remarkable application prospects in early cancer diagnosis and histological diagnosis.

(3) Expanding the applications of FF-OCT. FF-OCT imaging extends the detection depth of ultrahigh-resolution imaging and can be utilized in nondestructive continuous detection for organoids, cell spheres, and artificial tissues in subcellular level. Moreover, it provides a new tool for in-depth investigation of the growth and development regularity of tissues and organs in vitro. In addition, FF-OCT can detect the characteristics of samples at a specific depth, thereby functioning as a rapid detection approach for flaw detection of material subsurfaces, safe-guarding, anticounterfeiting, art identification, and other fields.

Disclosures

The authors declare that they have no conflict of interest.

Acknowledgments

This work was supported by the Nature Science Foundation of China under Grant Nos. 61675059 and 31927801. We thank Regenovo Ltd. (Hangzhou, China) for providing colorectal cancer organoid samples.

References

1. 

M. Leahy et al., “Functional imaging for regenerative medicine,” Stem Cell Res. Ther., 7 57 (2016). https://doi.org/10.1186/s13287-016-0315-2 Google Scholar

2. 

P. Sarder et al., “All-near-infrared multiphoton microscopy interrogates intact tissues at deeper imaging depths than conventional single- and two-photon near-infrared excitation microscopes,” J. Biomed. Opt., 18 (10), 106012 (2013). https://doi.org/10.1117/1.JBO.18.10.106012 JBOPFO 1083-3668 Google Scholar

3. 

C. Li et al., “Advanced fluorescence imaging technology in the near-infrared-II window for biomedical applications,” J. Am. Chem. Soc., 142 (35), 14789 –14804 (2020). https://doi.org/10.1021/jacs.0c07022 JACSAT 0002-7863 Google Scholar

4. 

E. H. K. Stelzer, “Light-sheet fluorescence microscopy for quantitative biology,” Nat. Methods, 12 (1), 23 –26 (2015). https://doi.org/10.1038/nmeth.3219 1548-7091 Google Scholar

5. 

D. Rupsa et al., “Fluorescence lifetime imaging microscopy: fundamentals and advances in instrumentation, analysis, and applications,” J. Biomed. Opt., 25 (7), 071203 (2020). https://doi.org/10.1117/1.JBO.25.7.071203 JBOPFO 1083-3668 Google Scholar

6. 

M. Pascolini et al., “Non-contact laser confocal microscope for corneal imaging,” Invest. Ophth. Vis. Sci., 60 (9), 2123 (2019). IOVSDA 0146-0404 Google Scholar

7. 

L. D. Jessica et al., “Feasibility of confocal fluorescence microscopy for real-time evaluation of neoplasia in fresh human breast tissue,” J. Biomed. Opt., 18 (10), 106016 (2013). https://doi.org/10.1117/1.JBO.18.10.106016 JBOPFO 1083-3668 Google Scholar

8. 

A. Darine and O. Thomas, “Confocal reflectance and two-photon microscopy studies of a songbird skull for preparation of transcranial imaging,” J. Biomed. Opt., 14 (3), 034038 (2009). https://doi.org/10.1117/1.3155522 JBOPFO 1083-3668 Google Scholar

9. 

F. Helmchen and W. Denk, “Deep tissue two-photon microscopy,” Nat. Methods, 2 (12), 932 –940 (2005). https://doi.org/10.1038/nmeth818 1548-7091 Google Scholar

10. 

W. Song et al., “Fully integrated reflection-mode photoacoustic, two-photon and second harmonic generation microscopy in vivo,” Sci. Rep., 6 32240 (2016). https://doi.org/10.1038/srep32240 SRCEC3 2045-2322 Google Scholar

11. 

L. Po-Yen et al., “Two-photon scanned light sheet fluorescence microscopy with axicon imaging for fast volumetric imaging,” J. Biomed. Opt., 26 (11), 116503 (2021). https://doi.org/10.1117/1.JBO.26.11.116503 JBOPFO 1083-3668 Google Scholar

12. 

D. Huang et al., “Optical coherence tomography,” Science, 254 (5035), 1178 –1181 (1991). https://doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar

13. 

A. F. Fercher et al., “A thermal light source technique for optical coherence tomography,” Opt. Commun., 185 (1), 57 –64 (2000). https://doi.org/10.1016/S0030-4018(00)00986-X OPCOB8 0030-4018 Google Scholar

14. 

E. Beaurepaire et al., “Full-field optical coherence microscopy,” Opt. Lett., 23 (4), 244 –246 (1998). https://doi.org/10.1364/OL.23.000244 OPLEDP 0146-9592 Google Scholar

15. 

A. Dubois et al., “Three-dimensional cellular-level imaging using full-field optical coherence tomography,” Phys. Med. Biol., 49 (7), 1227 –1234 (2004). https://doi.org/10.1088/0031-9155/49/7/010 Google Scholar

16. 

P. P. Laissue et al., “Assessing phototoxicity in live fluorescence imaging,” Nat. Methods, 14 (7), 657 –661 (2017). https://doi.org/10.1038/nmeth.4344 1548-7091 Google Scholar

17. 

V. Magidson, A. Khodjakov, “Circumventing photodamage in live-cell microscopy,” Methods in Cell Biology, 545 –560 Academic Press, Boston (2013). Google Scholar

18. 

A. L. Rainer et al., “Enhanced medical diagnosis for dOCTors: a perspective of optical coherence tomography,” J. Biomed. Opt., 26 (10), 100601 (2021). https://doi.org/10.1117/1.JBO.26.10.100601 JBOPFO 1083-3668 Google Scholar

19. 

A. Dubois, Handbook of Full-Field Optical Coherence Microscopy: Technology and Applications, CRC Press, Singapore (2016). Google Scholar

20. 

A. Dubois et al., “High-resolution full-field optical coherence tomography with a Linnik microscope,” Appl. Opt., 41 (4), 805 –812 (2002). https://doi.org/10.1364/AO.41.000805 APOPAI 0003-6935 Google Scholar

21. 

A. Federici and A. Dubois, “Full-field optical coherence microscopy with optimized ultrahigh spatial resolution,” Opt. Lett., 40 (22), 5347 –5350 (2015). https://doi.org/10.1364/OL.40.005347 OPLEDP 0146-9592 Google Scholar

22. 

E. Dalimier and D. Salomon, “Full-field optical coherence tomography: a new technology for 3D high-resolution skin imaging,” Dermatology, 224 (1), 84 –92 (2012). https://doi.org/10.1159/000337423 DERMEI 0742-3217 Google Scholar

23. 

O. Thouvenin et al., “Full-field optical coherence tomography as a diagnosis tool: recent progress with multimodal imaging,” Appl. Sci., 7 (3), 236 (2017). https://doi.org/10.3390/app7030236 Google Scholar

24. 

E. Auksorius et al., “Dual-modality fluorescence and full-field optical coherence microscopy for biomedical imaging applications,” Biomed. Opt. Express, 3 (3), 661 –666 (2012). https://doi.org/10.1364/BOE.3.000661 BOEICL 2156-7085 Google Scholar

25. 

O. Thouvenin et al., “Dynamic multimodal full-field optical coherence tomography and fluorescence structured illumination microscopy,” J. Biomed. Opt., 22 (2), 026004 (2017). https://doi.org/10.1117/1.JBO.22.2.026004 JBOPFO 1083-3668 Google Scholar

26. 

V. Gorti et al., “D-FFOCT as a tool to detect cellular activity,” Biophotonics Congress: Biomedical Optics (Translational, Microscopy, OCT, OTS, BRAIN), W1E –W5E Optical Society of America, Washington, DC (2020). Google Scholar

27. 

J. Scholler et al., “Dynamic full-field optical coherence tomography: 3D live-imaging of retinal organoids,” Light Sci. Appl., 9 (1), 140 (2020). https://doi.org/10.1038/s41377-020-00375-8 Google Scholar

28. 

O. Thouvenin et al., “Cell motility as contrast agent in retinal explant imaging with full-field optical coherence tomography,” Invest. Ophth. Vis. Sci., 58 (11), 4605 –4615 (2017). https://doi.org/10.1167/iovs.17-22375 IOVSDA 0146-0404 Google Scholar

29. 

G. Kassandra et al., “Dynamic full-field optical coherence tomography of retinal pigment epithelium cell cultures to model degenerative diseases,” Proc. SPIE, 11630 1163014 (2021). https://doi.org/10.1117/12.2581048 PSISDG 0277-786X Google Scholar

30. 

C. Apelian et al., “Dynamic full field optical coherence tomography: subcellular metabolic contrast revealed in tissues by interferometric signals temporal analysis,” Biomed. Opt. Express, 7 (4), 1511 –1524 (2016). https://doi.org/10.1364/BOE.7.001511 BOEICL 2156-7085 Google Scholar

31. 

J. Scholler et al., “Probing dynamic processes in the eye at multiple spatial and temporal scales with multimodal full field OCT,” Biomed. Opt. Express, 10 (2), 731 –746 (2019). https://doi.org/10.1364/BOE.10.000731 BOEICL 2156-7085 Google Scholar

32. 

S. Park et al., “Quantitative evaluation of the dynamic activity of HeLa cells in different viability states using dynamic full-field optical coherence microscopy,” Biomed. Opt. Express, 12 (10), 6431 –6441 (2021). https://doi.org/10.1364/BOE.436330 BOEICL 2156-7085 Google Scholar

33. 

T. Ho et al., “Mirau-type full-field optical coherence tomography with switchable partially spatially coherent illumination modes,” Biomed. Opt. Express, 12 (5), 2670 –2683 (2021). https://doi.org/10.1364/BOE.422622 BOEICL 2156-7085 Google Scholar

34. 

C. Tsai et al., “Submicron spatial resolution optical coherence tomography for visualising the 3D structures of cells cultivated in complex culture systems,” Sci. Rep., 11 3492 (2021). https://doi.org/10.1038/s41598-021-82178-4 SRCEC3 2045-2322 Google Scholar

35. 

T. Ming-Rung et al., “In vivo dual-mode full-field optical coherence tomography for differentiation of different types of melanocytic nevus,” Proc. SPIE, 11618 116180C (2021). https://doi.org/10.1117/12.2582891 PSISDG 0277-786X Google Scholar

36. 

W. Oh et al., “Spectrally-modulated full-field optical coherence microscopy for ultrahigh-resolution endoscopic imaging,” Opt. Express, 14 (19), 8675 –8684 (2006). https://doi.org/10.1364/OE.14.008675 OPEXFF 1094-4087 Google Scholar

37. 

J. Ogien and A. Dubois, “A compact high-speed full-field optical coherence microscope for high-resolution in vivo skin imaging,” J. Biophotonics, 12 (2), e201800208 (2019). https://doi.org/10.1002/jbio.201800208 Google Scholar

38. 

C. Dunsby et al., “Single-shot phase-stepped wide-field coherence-gated imaging,” Opt. Express, 11 (2), 105 –115 (2003). https://doi.org/10.1364/OE.11.000105 OPEXFF 1094-4087 Google Scholar

39. 

M. Sato et al., “Quadrature fringes wide-field optical coherence tomography and its applications to biological tissues,” Opt. Commun., 271 (2), 573 –580 (2007). https://doi.org/10.1016/j.optcom.2006.10.029 OPCOB8 0030-4018 Google Scholar

40. 

Y. Zhu and W. Gao, “Single-shot wavelength-independent phase-shifting method for full-field optical coherence tomography,” Appl. Opt., 58 (4), 806 –813 (2019). https://doi.org/10.1364/AO.58.000806 APOPAI 0003-6935 Google Scholar

41. 

M. Žurauskas et al., “Simultaneous 4-phase-shifted full-field optical coherence microscopy,” Biomed. Opt. Express, 12 (2), 981 –992 (2021). https://doi.org/10.1364/BOE.417183 BOEICL 2156-7085 Google Scholar

42. 

J. Na et al., “Image restoration method based on Hilbert transform for full-field optical coherence tomography,” Appl. Opt., 47 (3), 459 –466 (2008). https://doi.org/10.1364/AO.47.000459 APOPAI 0003-6935 Google Scholar

43. 

H. M. Subhash, “Full-field and single-shot full-field optical coherence tomography: a novel technique for biomedical imaging applications,” Adv. Opt. Technol., 2012 1 –26 (2012). https://doi.org/10.1155/2012/435408 1687-6393 Google Scholar

44. 

A. Dubois, “A simplified algorithm for digital fringe analysis in two-wave interferometry with sinusoidal phase modulation,” Opt. Commun., 391 128 –134 (2017). https://doi.org/10.1016/j.optcom.2017.01.004 OPCOB8 0030-4018 Google Scholar

45. 

K. G. Larkin, “Efficient nonlinear algorithm for envelope detection in white light interferometry,” J. Opt. Soc. Am. A, 13 (4), 832 –843 (1996). https://doi.org/10.1364/JOSAA.13.000832 JOAOD6 0740-3232 Google Scholar

46. 

P. Stremplewski et al., “In vivo volumetric imaging by crosstalk-free full-field OCT,” Optica, 6 (5), 608 –617 (2019). https://doi.org/10.1364/OPTICA.6.000608 Google Scholar

47. 

E. Auksorius et al., “Multimode fiber enables control of spatial coherence in Fourier-domain full-field optical coherence tomography for in vivo corneal imaging,” Opt. Lett., 46 (6), 1413 –1416 (2021). https://doi.org/10.1364/OL.417178 OPLEDP 0146-9592 Google Scholar

48. 

A. F. Fercher et al., “Optical coherence tomography—principles and applications,” Rep. Prog. Phys., 66 (2), 239 (2003). https://doi.org/10.1088/0034-4885/66/2/204 RPPHAG 0034-4885 Google Scholar

49. 

E. Auksorius and A. C. Boccara, “Dark-field full-field optical coherence tomography,” Opt. Lett., 40 (14), 3272 –3275 (2015). https://doi.org/10.1364/OL.40.003272 OPLEDP 0146-9592 Google Scholar

50. 

L. Vabre, A. Dubois and A. C. Boccara, “Thermal-light full-field optical coherence tomography,” Opt. Lett., 27 (7), 530 –532 (2002). https://doi.org/10.1364/OL.27.000530 OPLEDP 0146-9592 Google Scholar

51. 

W. Oh et al., “Ultrahigh-resolution full-field optical coherence microscopy using InGaAs camera,” Opt. Express, 14 (2), 726 –735 (2006). https://doi.org/10.1364/OPEX.14.000726 OPEXFF 1094-4087 Google Scholar

52. 

A. Dubois, “Full-field optical coherence microscopy,” Selected Topics in Optical Coherence Tomography, 3 –20 InTech, Rijeka, Croatia (2012). Google Scholar

53. 

D. Sacchet et al., “Motion artifact suppression in full-field optical coherence tomography,” Appl. Opt., 49 (9), 1480 –1488 (2010). https://doi.org/10.1364/AO.49.001480 APOPAI 0003-6935 Google Scholar

54. 

M. Viacheslav et al., “Optical phase modulation by natural eye movements: application to time-domain FF-OCT image retrieval,” (2021). Google Scholar

55. 

M. Ning et al., “Image contrast correction method in full-field optical coherence tomography,” Opt. Eng., 58 (7), 074102 (2019). https://doi.org/10.1117/1.OE.58.7.074102 Google Scholar

56. 

J. Scholler, “Motion artifact removal and signal enhancement to achieve in vivo dynamic full field OCT,” Opt. Express, 27 (14), 19562 –19572 (2019). https://doi.org/10.1364/OE.27.019562 OPEXFF 1094-4087 Google Scholar

57. 

I. Chen et al., “Full field optical coherence tomography image denoising using deep learning with spatial compounding,” in IEEE 17th Int. Symp. Biomed. Imaging, 1975 –1978 (2020). https://doi.org/10.1109/ISBI45749.2020.9098452 Google Scholar

58. 

P. Xiao et al., “Adaptive optics full-field optical coherence tomography,” J. Biomed. Opt., 21 (12), 121505 (2016). https://doi.org/10.1117/1.JBO.21.12.121505 JBOPFO 1083-3668 Google Scholar

59. 

A. Dubois, “Focus defect and dispersion mismatch in full-field optical coherence microscopy,” Appl. Opt., 56 (9), D142 –D150 (2017). https://doi.org/10.1364/AO.56.00D142 APOPAI 0003-6935 Google Scholar

60. 

H. Ammari et al., “A signal separation technique for sub-cellular imaging using dynamic optical coherence tomography,” Multiscale Model. Simul., 15 (3), 1155 –1175 (2017). https://doi.org/10.1137/16M1090387 Google Scholar

61. 

C. Demene et al., “Spatiotemporal clutter filtering of ultrafast ultrasound data highly increases Doppler and fUltrasound sensitivity,” IEEE Trans. Med. Imaging, 34 (11), 2271 –2285 (2015). https://doi.org/10.1109/TMI.2015.2428634 ITMID4 0278-0062 Google Scholar

62. 

E. Auksorius and A. C. Boccara, “High-throughput dark-field full-field optical coherence tomography,” Opt. Lett., 45 (2), 455 –458 (2020). https://doi.org/10.1364/OL.381888 OPLEDP 0146-9592 Google Scholar

63. 

M. Laslandes et al., “Increasing the field of view of adaptive optics scanning laser ophthalmoscopy,” Biomed. Opt. Express, 8 (11), 4811 –4826 (2017). https://doi.org/10.1364/BOE.8.004811 BOEICL 2156-7085 Google Scholar

64. 

V. Barolle et al., “Manifestation of aberrations in full-field optical coherence tomography,” Opt. Express, 29 (14), 22044 –22065 (2021). https://doi.org/10.1364/OE.419963 OPEXFF 1094-4087 Google Scholar

65. 

M. Pedro et al., “Adaptive glasses wavefront sensorless full-field OCT for high-resolution retinal imaging over a wide field-of-view,” Proc. SPIE, 11623 1162306 (2021). https://doi.org/10.1117/12.2576689 PSISDG 0277-786X Google Scholar

66. 

Z. Dong et al., “Optical coherence tomography image denoising using a generative adversarial network with speckle modulation,” J. Biophotonics, 13 (4), e201960135 (2020). https://doi.org/10.1002/jbio.201960135 Google Scholar

67. 

C. C. Po et al., “Nonlocally adaptive image enhancement system for full-field optical coherence tomography,” Opt. Eng., 60 (7), 073103 (2021). https://doi.org/10.1117/1.OE.60.7.073103 Google Scholar

68. 

M. Serge et al., “Optical incoherence tomography: a method to generate tomographic retinal cross-sections with non-interferometric adaptive optics ophthalmoscopes,” Proc. SPIE, 11623 1162308 (2021). https://doi.org/10.1117/12.2576684 PSISDG 0277-786X Google Scholar

69. 

G. Recher et al., “Remote scanning for ultra-large field of view in wide-field microscopy and full-field OCT,” Biomed. Opt. Express, 11 (5), 2578 –2590 (2020). https://doi.org/10.1364/BOE.383329 BOEICL 2156-7085 Google Scholar

70. 

“Adaptive-glasses time-domain FF-OCT for wide-field high-resolution retinal imaging with increased SNR,” Opt. Lett., 45 (21), 5901 –5904 (2020). https://doi.org/10.1364/OL.403135 OPLEDP 0146-9592 Google Scholar

71. 

M. Pedro et al., “Retinal curvature correction with coherence gate shaping for full-field OCT high-resolution retinal imaging over a wide field-of-view,” Proc. SPIE, 11630 116300C (2021). https://doi.org/10.1117/12.2576687 PSISDG 0277-786X Google Scholar

72. 

P. Mecê et al., “Coherence gate shaping for wide field high-resolution in vivo retinal imaging with full-field OCT,” Biomed. Opt. Express, 11 (9), 4928 –4941 (2020). https://doi.org/10.1364/BOE.400522 BOEICL 2156-7085 Google Scholar

73. 

P. Mecê et al., “Curved-full-field OCT for high-resolution imaging of living human retina over a large field-of-view,” (2020). Google Scholar

74. 

Q. Yang et al., “Extended imaging depth of en-face optical coherence tomography based on fast measurement of a reflection matrix by wide-field heterodyne detection,” Opt. Lett., 45 (4), 828 –831 (2020). https://doi.org/10.1364/OL.382898 OPLEDP 0146-9592 Google Scholar

75. 

A. Badon et al., “Smart optical coherence tomography for ultra-deep imaging through highly scattering media,” Sci. Adv., 2 (11), e1600370 (2016). https://doi.org/10.1126/sciadv.1600370 STAMCV 1468-6996 Google Scholar

76. 

A. Nahas et al., “3D static elastography at the micrometer scale using full field OCT,” Biomed. Opt. Express, 4 (10), 2138 –2149 (2013). https://doi.org/10.1364/BOE.4.002138 BOEICL 2156-7085 Google Scholar

77. 

G. Moneron et al., “Polarization-sensitive full-field optical coherence tomography,” Opt. Lett., 32 (14), 2058 –2060 (2007). https://doi.org/10.1364/OL.32.002058 OPLEDP 0146-9592 Google Scholar

78. 

A. Nahas et al., “Detection of plasmonic nanoparticles with full field-OCT: optical and photothermal detection,” Biomed. Opt. Express, 5 (10), 3541 –3546 (2014). https://doi.org/10.1364/BOE.5.003541 BOEICL 2156-7085 Google Scholar

79. 

A. Federici and A. Dubois, “Three-band, 1.9-μm axial resolution full-field optical coherence microscopy over a 530-1700 nm wavelength range using a single camera,” Opt. Lett., 39 (6), 1374 –1377 (2014). https://doi.org/10.1364/OL.39.001374 OPLEDP 0146-9592 Google Scholar

80. 

H. M. Leung et al., “Imaging intracellular motion with dynamic micro-optical coherence tomography,” Biomed. Opt. Express, 11 (5), 2768 –2778 (2020). https://doi.org/10.1364/BOE.390782 BOEICL 2156-7085 Google Scholar

81. 

M. Munter et al., “Dynamic contrast in scanning microscopic OCT,” Opt. Lett., 45 (17), 4766 –4769 (2020). https://doi.org/10.1364/OL.396134 OPLEDP 0146-9592 Google Scholar

82. 

A. L. Oldenburg et al., “Inverse-power-law behavior of cellular motility reveals stromal-epithelial cell interactions in 3D co-culture by OCT fluctuation spectroscopy,” Optica, 2 (10), 877 –885 (2015). https://doi.org/10.1364/OPTICA.2.000877 Google Scholar

83. 

I. A. El-Sadek et al., “Optical coherence tomography-based tissue dynamics imaging for longitudinal and drug response evaluation of tumor spheroids,” Biomed. Opt. Express, 11 (11), 6231 –6248 (2020). https://doi.org/10.1364/BOE.404336 BOEICL 2156-7085 Google Scholar

84. 

I. A. El-Sadek et al., “Three-dimensional dynamics optical coherence tomography for tumor spheroid evaluation,” Biomed. Opt. Express, 12 (11), 6844 –6863 (2021). https://doi.org/10.1364/BOE.440444 BOEICL 2156-7085 Google Scholar

85. 

M. Pradipta et al., “Three-dimensional imaging of mouse liver dynamics by polarization-sensitive optical coherence tomography,” Proc. SPIE, 11655 116550O (2021). https://doi.org/10.1117/12.2577974 PSISDG 0277-786X Google Scholar

86. 

S. Nandy et al., “Classification and analysis of human ovarian tissue using full field optical coherence tomography,” Biomed. Opt. Express, 7 (12), 5182 –5187 (2016). https://doi.org/10.1364/BOE.7.005182 BOEICL 2156-7085 Google Scholar

87. 

B. K. Mekonnen et al., “Deep learning approach for red blood cell segmentation from full-field OCT data of human skin,” in IEEE Int. Conf. BioPhotonics, 1 –2 (2019). https://doi.org/10.1109/BioPhotonics.2019.8896748 Google Scholar

88. 

H. Chou et al., “Dermal epidermal junction detection for full-field optical coherence tomography data of human skin by deep learning,” Comput. Med. Imaging Graph., 87 101833 (2021). https://doi.org/10.1016/j.compmedimag.2020.101833 Google Scholar

89. 

B. K. Mekonnen et al., “Generation of augmented capillary network optical coherence tomography image data of human skin for deep learning and capillary segmentation,” Diagnostics, 11 (4), 685 (2021). https://doi.org/10.3390/diagnostics11040685 Google Scholar

90. 

C. J. Ho et al., “Detecting mouse squamous cell carcinoma from submicron full-field optical coherence tomography images by deep learning,” J. Biophotonics, 14 (1), e202000271 (2021). https://doi.org/10.1002/jbio.202000271 Google Scholar

91. 

X. Yao et al., “Visualization and tissue classification of human breast cancer images using ultrahigh-resolution OCT,” Laser. Surg. Med., 49 (3), 258 –269 (2017). https://doi.org/10.1002/lsm.22654 Google Scholar

92. 

O. Thouvenin et al., “Automatic diagnosis and biopsy classification with dynamic full-field OCT and machine learning,” (2021). Google Scholar

93. 

S. Yang et al., “In situ process monitoring and automated multi-parameter evaluation using optical coherence tomography during extrusion-based bioprinting,” Addit. Manuf., 47 102251 (2021). https://doi.org/10.1016/j.addma.2021.102251 Google Scholar

94. 

R. Zhao et al., “Automatic quantitative analysis of structure parameters in the growth cycle of artificial skin using optical coherence tomography,” J. Biomed. Opt., 26 (9), 095001 (2021). https://doi.org/10.1117/1.JBO.26.9.095001 JBOPFO 1083-3668 Google Scholar

95. 

L. Wang et al., “Automated quantitative assessment of three-dimensional bioprinted hydrogel scaffolds using optical coherence tomography,” Biomed. Opt. Express, 7 (3), 894 –910 (2016). https://doi.org/10.1364/BOE.7.000894 BOEICL 2156-7085 Google Scholar

96. 

K. Grieve et al., “Appearance of the retina with full-field optical coherence tomography,” Invest. Ophth. Vis. Sci., 57 (9), T96 –T104 (2016). https://doi.org/10.1167/iovs.15-18856 IOVSDA 0146-0404 Google Scholar

97. 

P. Xiao et al., “In vivo wavefront-correctionless full-field OCT human retinal imaging,” Invest. Ophth. Vis. Sci., 60 (9), 1274 –1274 (2019). IOVSDA 0146-0404 Google Scholar

98. 

K. Grieve et al., “Ocular tissue imaging using ultrahigh-resolution, full-field optical coherence tomography,” Invest. Ophth Vis. Sci., 45 (11), 4126 –4131 (2004). https://doi.org/10.1167/iovs.04-0584 Google Scholar

99. 

V. Mazlin et al., “In vivo high resolution human corneal imaging using full-field optical coherence tomography,” Biomed. Opt. Express, 9 (2), 557 –568 (2018). https://doi.org/10.1364/BOE.9.000557 BOEICL 2156-7085 Google Scholar

100. 

P. Xiao et al., and M. Wojtkowski, “Combining FF-OCT with SD-OCT for retinal imaging,” Optical Coherence Imaging Techniques and Imaging in Scattering Media II, 104160K Optical Society of America, p. 2017). Google Scholar

101. 

Y. Yang et al., “Investigation of optical coherence tomography as an imaging modality in tissue engineering,” Phys. Med. Biol., 51 (7), 1649 (2006). https://doi.org/10.1088/0031-9155/51/7/001 PHMBA7 0031-9155 Google Scholar

102. 

H. Yang et al., “Use of high-resolution full-field optical coherence tomography and dynamic cell imaging for rapid intraoperative diagnosis during breast cancer surgery,” Cancer, 126 (S16), 3847 –3856 (2020). https://doi.org/10.1002/cncr.32838 CANCAR 0008-543X Google Scholar

103. 

C.-K. Chang et al., “Skin cancer detection using cellular resolution optical coherence tomography,” in 24th Microoptics Conf., 70 –71 (2019). https://doi.org/10.23919/MOC46630.2019.8982810 Google Scholar

104. 

F. De Leeuw et al., “Value of full-field optical coherence tomography imaging for the histological assessment of head and neck cancer,” Laser. Surg. Med., 52 (8), 768 –778 (2020). https://doi.org/10.1002/lsm.23223 Google Scholar

105. 

Y. Wang et al., “Application of cellular resolution full-field optical coherence tomography in vivo for the diagnosis of skin tumours and inflammatory skin diseases: a pilot study,” Dermatology, 238 (1), 121 –131 (2021). https://doi.org/10.1159/000514686 DERMEI 0742-3217 Google Scholar

106. 

O. Assayag et al., “Large field, high resolution full-field optical coherence tomography: a pre-clinical study of human breast tissue and cancer assessment,” Technol. Cancer Res. Treat., 13 (5), 455 –468 (2014). https://doi.org/10.7785/tcrtexpress.2013.600254 Google Scholar

107. 

M. Camus et al., “Full-field optical coherence tomography: a new imaging modality for rapid on-site evaluation of resected polyps during colonoscopy,” Gastroenterology, 155 (6), 1692 –1694 (2018). https://doi.org/10.1053/j.gastro.2018.07.054 GASTAB 0016-5085 Google Scholar

108. 

L. Quénéhervé et al., “Full-field optical coherence tomography: novel imaging technique for extemporaneous high-resolution analysis of mucosal architecture in human gut biopsies,” Gut, 70 (1), 6 –8 (2020). https://doi.org/10.1136/gutjnl-2020-321228 GUTTAK 0017-5749 Google Scholar

109. 

P. Kumar et al., “Full-field optical coherence tomography for optical biopsy applications,” in Eur. Conf. Biomed. Opt., (2021). Google Scholar

110. 

M. Jain et al., “Rapid evaluation of fresh ex vivo kidney tissue with full-field optical coherence tomography,” J. Pathol. Inf., 6 53 (2015). https://doi.org/10.4103/2153-3539.166014 Google Scholar

111. 

L. van Manen et al., “Validation of full-field optical coherence tomography in distinguishing malignant and benign tissue in resected pancreatic cancer specimens,” PLoS One, 12 (4), e0175862 (2017). https://doi.org/10.1371/journal.pone.0175862 POLNCL 1932-6203 Google Scholar

112. 

T. Maldiney et al., “Full-field optical coherence tomography for the diagnosis of giant cell arteritis,” PLoS One, 15 (8), e0234165 (2020). https://doi.org/10.1371/journal.pone.0234165 POLNCL 1932-6203 Google Scholar

113. 

Z. Yue et al., “Rapid and high-resolution imaging of human liver specimens by full-field optical coherence tomography,” J. Biomed. Opt., 20 (11), 116010 (2015). https://doi.org/10.1117/1.JBO.20.11.116010 JBOPFO 1083-3668 Google Scholar

114. 

Z. Livia et al., “Label-free characterization of vitrification-induced morphology changes in single-cell embryos with full-field optical coherence tomography,” J. Biomed. Opt., 20 (9), 096004 (2015). https://doi.org/10.1117/1.JBO.20.9.096004 JBOPFO 1083-3668 Google Scholar

115. 

D. Sacchet et al., “Simultaneous dual-band ultra-high resolution full-field optical coherence tomography,” Opt. Express, 16 (24), 19434 –19446 (2008). https://doi.org/10.1364/OE.16.019434 OPEXFF 1094-4087 Google Scholar

116. 

A. Dubois, “Spectroscopic polarization-sensitive full-field optical coherence tomography,” Opt. Express, 20 (9), 9962 –9977 (2012). https://doi.org/10.1364/OE.20.009962 OPEXFF 1094-4087 Google Scholar

117. 

H. Makhlouf et al., “Simultaneous optically sectioned fluorescence and optical coherence microscopy with full-field illumination,” Opt. Lett., 37 (10), 1613 –1615 (2012). https://doi.org/10.1364/OL.37.001613 OPLEDP 0146-9592 Google Scholar

118. 

A. Dubois et al., “Spectroscopic ultrahigh-resolution full-field optical coherence microscopy,” Opt. Express, 16 (21), 17082 –17091 (2008). https://doi.org/10.1364/OE.16.017082 OPEXFF 1094-4087 Google Scholar

119. 

V. Srivastava et al., “Tomographic and volumetric reconstruction of composite materials using full-field swept-source optical coherence tomography,” Meas. Sci. Technol., 23 (5), 055203 (2012). https://doi.org/10.1088/0957-0233/23/5/055203 MSTCEP 0957-0233 Google Scholar

120. 

J. Zhang et al., “Non-destructive analysis of flake properties in automotive paints with full-field optical coherence tomography and 3D segmentation,” Opt. Express, 25 (16), 18614 –18628 (2017). https://doi.org/10.1364/OE.25.018614 OPEXFF 1094-4087 Google Scholar

121. 

Z. Zhang et al., “Sub-surface imaging of soiled cotton fabric using full-field optical coherence tomography,” Opt. Express, 27 (10), 13951 –13964 (2019). https://doi.org/10.1364/OE.27.013951 OPEXFF 1094-4087 Google Scholar

122. 

S. Frank et al., “Three-dimensional evaluation of subsurface damage in optical glasses with ground and polished surfaces using FF-OCT,” Appl. Opt., 60 (8), 2118 –2126 (2021). https://doi.org/10.1364/AO.413090 APOPAI 0003-6935 Google Scholar

123. 

W. J. Choi et al., “Qualitative investigation of fresh human scalp hair with full-field optical coherence tomography,” J. Biomed. Opt., 17 (3), 036010 (2012). https://doi.org/10.1117/1.JBO.17.3.036010 JBOPFO 1083-3668 Google Scholar

124. 

W. Choi et al., “Counterfeit detection using characterization of safety feature on banknote with full-field optical coherence tomography,” J. Opt. Soc. Korea, 14 (4), 316 –320 (2010). https://doi.org/10.3807/JOSK.2010.14.4.316 1226-4776 Google Scholar

125. 

K. B. Raja et al., “Robust verification with subsurface fingerprint recognition using full field optical coherence tomography,” in Proc. IEEE Conf. Comput. Vision and Pattern Recognit. Workshops, 144 –152 (2017). https://doi.org/10.1109/CVPRW.2017.93 Google Scholar

126. 

E. Auksorius and A. C. Boccara, “Fingerprint imaging from the inside of a finger with full-field optical coherence tomography,” Biomed. Opt. Express, 6 (11), 4465 –4471 (2015). https://doi.org/10.1364/BOE.6.004465 BOEICL 2156-7085 Google Scholar

127. 

A. Egidijus and A. C. Boccara, “Fast subsurface fingerprint imaging with full-field optical coherence tomography system equipped with a silicon camera,” J. Biomed. Opt., 22 (9), 096002 (2017). https://doi.org/10.1117/1.JBO.22.9.096002 JBOPFO 1083-3668 Google Scholar

Biography

Ling Wang received her PhD in measuring and testing technologies and instruments from Zhejiang University, China, in 2008, and has been a visiting scholar at the University of California, Irvine. Currently, she works as a full professor at Hangzhou Dianzi University and serves as a deputy director of Key Laboratory of Medical Information and Biological 3D Printing of Zhejiang Province. Her research is mainly focuses on the biomedical imaging.

Rongzhen Fu is a graduate student at Hangzhou Dianzi University, China. His current research interests include multimodal biomedical imaging, including full-field optical coherence tomography and dynamic contrast optical coherence tomography, as well as building the high-resolution full-field optical coherence tomography system used in the laboratory.

Chen Xu received his BS degree in optoelectronics and PhD in optical engineering from Beijing Institute of Technology, Beijing, China, in 2012 and 2018, respectively. He is currently with Hangzhou Dianzi University, China. His research interests include optical design, freeform optics, and biomedical optics. He is a member of SPIE.

Mingen Xu received his PhD in biomedical engineering from Tsinghua University, China. He is a full professor and head of the Bio-3D Printing and Medical Device Research Institute, Key Laboratory of Medical Information and 3D Biological of Zhejiang Province, and has presided over a total of more than 20 national major and key projects at all levels. His research interests include 3D bioprinting, biomedical optical imaging, etc.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Ling Wang, Rongzhen Fu, Chen Xu, and Mingen Xu "Methods and applications of full-field optical coherence tomography: a review," Journal of Biomedical Optics 27(5), 050901 (20 May 2022). https://doi.org/10.1117/1.JBO.27.5.050901