1.IntroductionContinuous and accurate monitoring of physiological changes is critical in medical settings, making peripheral oxygen saturation () a crucial vital sign in patient treatment and care.1 serves as an indicator of the overall functioning of the body’s main organ systems, and its continuous observation is key for early problem detection and timely delivery of treatment and care. The gold standard for measuring arterial blood oxygen saturation () is arterial blood gas (ABG) sampling, a technique performed using a blood gas analyzer. This method is, however, intermittent, expensive, and time-consuming and involves invasive procedures. By contrast, pulse oximetry offers an alternative technology that is a continuous, inexpensive, pain-free, and non-invasive way of monitoring . This technique is realized using a pulse oximeter (PO), which operates by transmitting light into the tissue and measuring photoplethysmography (PPG) signals that represent light modulated by cardiac-synchronized variations in the arterial blood volume.2 Pulse oximeters (POs) commonly utilize red (660 nm) and infrared (IR, 940 nm) light wavelengths to measure the absorption of deoxyhemoglobin (Hbb) and oxyhemoglobin () in the blood, respectively. The basis of pulse oximetry is the “ratio of ratios” R of the pulsatile (AC)/non-pulsatile (DC) detected light components at red and IR wavelengths, obtained from PPG signals.3 The relationship between R and is usually derived via calibration with ABG during controlled desaturation events.4 It is a simple, yet powerful technique in estimating continuous blood oxygen saturation readings in real time that has gained significant use in various clinical and remote patient monitoring settings. For instance, during the COVID-19 pandemic, POs played a pivotal role in guiding clinicians and doctors in making timely clinical decisions, especially during patient triaging or when adjusting supplemental oxygen levels.5 Silent hypoxia, also known as “happy hypoxia,” was a common symptom experienced by people infected with this disease, wherein there was a sudden drop in levels without any signs of deterioration in health.6 As a result, remote use of POs became widespread globally among COVID-19 patients or those at high risk to monitor their levels regularly.5 Simultaneously, there was a rise in studies documenting POs inaccuracies, particularly in individuals with non-White skin tones.7–10 One large-scale retrospective study by Sjoding et al.7 highlighted that, when POs are utilized, Black patients were three times more likely to experience occult hypoxemia compared with those with White pigmentation. Occult hypoxemia is defined as a condition during which PO measures , despite (i.e., overestimation). This discrepancy has since raised considerable alarm within the medical and scientific communities as overestimating levels could hinder the timely and appropriate escalation of care delivered for those affected. Another extensive study conducted by Crooks et al.11 on patients at the time of transfer to the intensive care unit (ICU) confirmed that skin-tone-related measurement errors in pulse oximetry could lead to health and healthcare disparities across ethnic groups worldwide. Non-White subjects recorded similar levels, but reduced readings compared with White subjects and, as a result, were found to be more severely ill at the time of transfer to the ICU. This is consistent with the overestimation and delayed treatment hypothesis for heavily pigmented patients. With the growing demand for improved PO reliability and accuracy across diverse skin tones, understanding how skin color influences readings becomes crucial. Dark skin has a high melanin concentration that attenuates light greatly. As the basis of pulse oximetry is the ratio of ratios , differences in melanin concentration should be compensated. However, the melanin absorbs and scatters light differently (red attenuation greater than IR),12 which may induce differences in for different skin tones and therefore errors in (as is inversely proportional to ).13 In our recent study, a laboratory benchtop simulator was constructed to generate PPG signals in a controlled manner, and the performance of commercially available home and hospital-based POs under the simulated effects of different levels of melanin in the skin were assessed.14,15 We hypothesized that high melanin concentration in the skin has two principal effects: (a) a reduced signal-to-noise ratio (SNR) and (b) preferential attenuation of red over IR light. These effects were simulated using neutral density and synthetic melanin filters. We discovered that none of the tested POs exhibited overestimation under the effects of varying melanin attenuation. Although this study provided an opportunity to assess POs without the more complex ABG procedures that require testing on human volunteers, the use of a simplified simulator introduced limitations by not accounting for variations in optical properties that may occur in real-world situations (e.g., scattering). To further investigate this issue, numerical simulations such as Monte Carlo (MC) can be used to model both the scattering and absorption properties of melanin. MC simulations have been extensively used to investigate light–tissue interactions across a wide range of PPG applications.16,17 The impact of skin tone on these interactions has been investigated in previous MC studies and revealed that increased melanin concentration reduces light intensity, resulting in a lower SNR.18,19 A study led by Arefin et al.20 showed that overestimation in highly pigmented cohorts was minimized when the calibration equation was derived from a population cohort with a higher number of dark-pigmented models. Despite these efforts, a comprehensive physical understanding of the root cause of overestimation due to skin color remains unknown. The primary aim of this paper is to address this by offering insights into the specific factors that may drive overestimation in POs due to skin color variations, paying particular attention to the trajectories of the light paths at different wavelengths. 2.MethodsMCmatlab, an open-source software tool developed by Marti et al.,21 was utilized for carrying out MC simulations on a four-layer tissue model. These simulations aimed to investigate the impact of melanin’s optical properties on light–tissue interactions, ratio and . This section outlines the geometry, parameters, and optical properties of the tissue model and the simulation methodology employed in this study. 2.1.Geometry of the Finger ModelA finger tissue model was constructed; it contained four layers: epidermis, dermis, two arteries, and a bone. The geometry of this four-layer tissue model is presented in Fig. 1. The overall structure of the geometry was modeled by a three-dimensional, semi-infinite cuboid model. The dimensions of the cuboid geometry were , and it was divided into bins in the direction, with the finger thickness similar to that used in other work.22 The thickness of the various tissue layers included in the model is provided in Table 1. Table 1Thickness of tissue layers in the finger model.22,23
2.2.Tissue Parameters and Optical PropertiesThe parameters used to simulate the layers in Table 1, namely water volume (), melanin volume (), arterial blood volume (), and venous blood volume (), are all presented as fractions in Table 2. Two skin tones were simulated by adjusting in the epidermal layer to 0.05 (light skin) and 0.25 (dark skin), which correspond to Fitzpatrick type 2 (light) and type 5 (dark), respectively.24 The pulse was simulated by increasing during systole twice as high as in diastole. was maintained equal to the diastolic throughout simulations.16 Table 2Parameters utilized for modeling tissue layers in the simulation. These values were adapted from the literature.16,21
The two principal optical properties governing light attenuation through tissue are the absorption coefficient () and the scattering coefficient (). These properties differ across tissue layers, depending on the type and concentration of chromophores present. To simulate variations in skin tone, both and values within the epidermal layer were altered. The of any tissue layer (denoted as the ’th layer) at a particular wavelength () was determined as where was defined by Eq. (2) and accounts for the absorption due to chromophores in the tissue other than the melanin, water, and blood.16Parameters , , , and are the absorption coefficients () of the arterial blood, venous blood, melanin, and water, respectively. Oxygen saturation is defined by the ratio of concentration to the total Hb concentration present in the blood. Based on this definition, and are calculated using Eqs. (3) and (4), respectively.16 Absorption coefficients () of water, melanin, , and Hbb at 660 and 940 nm are presented in Table 3. was varied from 60% to 100% in increments of 10%, and venous oxygen saturation () was maintained at 10% lower than the corresponding values.16 where at 660 and 940 nm for the epidermis and bone layer16 are summarized in Table 4. Although values for the dermal and arterial layers are not listed in the table due to their dependency on oxygen saturation levels, they have been included in the simulation (with the values derived according to the steps outlined above). In addition, this table includes for all tissue layers at both wavelengths. Scattering is known to vary with an increase in melanin concentration,27 as a result, an estimate of the scattering coefficient was introduced in the epidermis to simulate different scattering properties associated with variations in skin tone.23Table 3Absorption coefficients of chromophores at specific pulse oximetry wavelengths.25
Table 4Optical properties of tissue layers.16,23,26
For calculating the for the dermal and arterial layers, at first, reduced scattering coefficient () values that accounts for the anisotropy of scattering (g) were calculated as26 where and correspond to Rayleigh and Mie scattering coefficients, respectively, and denotes a scaling factor. These parameters were obtained from existing literature for the dermis and arterial layer and are summarized in Table 5. Finally, was estimated by scaling by a factor of (1-g).Table 5Parameters used to calculate reduced scattering coefficient for the dermal and arterial layers.26
For simplification, all tissue layers in the model were assigned a uniform refractive index () and anisotropy factor (), so reflections at the air-tissue and tissue-tissue boundaries are not taken into account. The primary aim was to study overestimation due to skin tone by isolating the effect of epidermal absorption and scattering properties on light–tissue interactions from other complex interactions, so a simple model was desirable. 2.3.Simulation ProcedureTransmission and reflection modes were simulated with the source and detector were either positioned on opposite sides of the tissue (transmission) or adjacent to each other on the same side of the tissue (reflection). The position and placement of the light source and collector in transmission and reflection modes are depicted in Figs. 2(a) and 2(b), respectively. The simulations were run on a remote desktop with Windows 10 Enterprise, an Intel Xeon CPU E7 with 40 cores and 1 TB of main memory. A Gaussian beam with a radial width of 0.2 cm simulated the illumination at the tissue surface. In commercial POs, the detector is usually a photodiode. The most effective way of implementing this in MCmatlab was by simulating this as a fiber optic collector with a diameter of 0.3 cm and a numerical aperture of 1. For each simulation run, a total of were simulated, and this process was repeated three times to ensure repeatability and consistency in the measured results. After the execution of the MC simulation, various outputs could be extracted during systolic and diastolic states at 660 and 940 nm such as the normalized DC light intensities from the light collector ( and ), normalized fluence rate distribution of the detected photons within the tissue, optical path length (OPL), and penetration depth (PD, only in reflection mode). The AC pulsatile intensity was determined from the difference in DC intensities between diastolic and systolic states. The perfusion index, denoted by (), was defined as the ratio of the AC pulsatile intensity to the DC diastolic intensity: Finally, was computed by taking the ratio of PIs at red and IR wavelengths, given as To aid the comparison, we also defined attenuation factors that describe the relative responses of light and dark skin, where is the relative AC attenuation of dark skin to light skin at the red wavelength, is the relative DC attenuation of dark skin to light skin at the red wavelength, is the relative AC attenuation of dark skin to light skin at the IR wavelength, and is the relative DC attenuation of dark skin to light skin at the IR wavelength. This leads to a ratio of ratios for dark skin, given as The median PD of the detected photons was defined as the median of the depths (i.e., the maximum distance traveled along the -axis). Similarly, the median OPL of the detected photons was defined as the median of the total OPL (i.e., the sum of all step sizes corresponding to the number of scattering and absorption events). It should be noted that the PD and OPL data were tested for normality using the Kolmogorov-Smirnov (KS) test in MATLAB. The findings revealed that the data was not normally distributed for both cases; hence the median (IQR) was chosen as the appropriate measure to describe the OPL and PD in this article. 3.ResultsThis section presents the results investigating the influence of skin color on versus curves, using both transmission and reflection mode configurations. In addition, it examines the impact of source–detector (SD) separation on light and dark skin tissue models. To simulate the effect of skin color in Secs. 3.1–3.3, the absorption and scattering coefficients for the epidermal layer, as presented in Table 4, were varied. In addition, Table 4 contains the optical properties used for modeling different tissue layers. The oxygen saturation was varied from 60% to 100% in increments of 10% for each skin type in both the dermal and arterial layers. However, the pulse was only simulated in the arterial layer by doubling the blood volume in the systolic state compared with the diastolic state. The absorption coefficient for both the dermal and arterial layers was determined using the values in Table 3 in Eq. (1), and the scattering coefficient was calculated from Eq. (5) using the values in Table 5. A total of were simulated during the diastolic and systolic phases at red and IR wavelengths, and this process was repeated three times. 3.1.Transmission ModeIn transmission mode, the light source was positioned at the origin, and the collector was placed at a distance of 0.41 cm, on the opposite side of the tissue surface [see Fig. 2(a)]. A two-dimensional normalized fluence rate distribution that illustrates the magnitude of detected photon distribution within the tissue at an oxygen saturation value of 90% for both light and dark skin at different wavelengths is shown in Fig. 3. IR light, being the longer wavelength, exhibits a higher fluence rate distribution within the tissue (due to relatively lower light absorption) compared with the shorter red light in both skin types [seen from Figs. 3(b) and 3(d)]. Furthermore, the absorption of red light is more pronounced in dark skin, affecting its light distribution (lower fluence rate) within the tissue [observed in Figs. 3(a) and 3(c)]. The horizontal and vertical line profiles, drawn at and , respectively, for both skin types corresponding to the two wavelengths of interest, are presented in Figs. 3(e) and 3(f). The maximum fluence rate () within the tissue is found in the following order: IR, light skin (0.44) > red, light skin (0.18) > IR, dark skin (0.16) > red, dark skin (). From Fig. 3(f), the fluence rate decreased with increasing tissue depth. The highest fluence rate () at the collector facing the tissue boundary () is found in the following order: IR, light skin , dark skin , light skin , dark skin (). This reveals that the intensity of light reaching the detector in transmission mode is highest for the IR wavelength in both skin types compared with the intensity at the red wavelength. Figure 4 presents the detected light intensities, PIs, and curves for light and dark skin tissue models in transmission mode. The overall transmitted light intensity was found to be higher for light skin compared with dark skin at both wavelengths [Figs. 4(a) and 4(b)]. However, in dark skin, the relative attenuation at the red wavelength was greater compared with the IR wavelength. For instance, at , the relative attenuation in dark skin at diastolic red was 89%, whereas at the diastolic IR, it was only 50% compared with light skin. Throughout this paper, trends at were selected as an illustrative example as it represents a clinically significant threshold for diagnosing hypoxemia28 and may indicate an urgent need for critical medical attention. Further, from Figs. 4(a) and 4(b), it is observed that the red-light intensity gradually increased with an increase in , whereas the IR light intensity showed a gradual decrease with an increase in saturation. This is expected as the molar extinction coefficient of Hbb is higher than in the red, and in the IR, the opposite is true. For both light and dark skin, the diastolic light intensities were consistently higher than their systolic counterparts due to lower volumes of absorbing blood. The calculated PI values at red and IR for light and dark skin are displayed in Figs. 4(c) and 4(d), respectively. To quantify the overall difference or similarity in PI curves between skin types, the pairwise Euclidean distance was calculated; it quantifies the absolute difference in the magnitude for each wavelength. This method is consistently applied throughout the analysis. The difference in PIs at each wavelength between light and dark skin types was similar, as indicated by their Euclidean distances: for red PI and for IR PI. For example, at an , the mean (standard deviation, std) of PI in red is () in light skin and () in dark skin. Similarly, at the same saturation, the mean (std) of PI at IR is () in light skin and () in dark skin. The ratio for both skin types as a function of is illustrated in Fig. 4(e). To evaluate the statistical significance between curves in this paper, the Wilcoxon signed rank test, a non-parametric statistical test for paired data, was utilized with a significance level set at . No measurable difference in was observed between the two skin types under this configuration (). Figure 5 displays a histogram of OPL distributions based on a sample of 10,000 detected photons at both red and IR diastolic wavelengths for light and dark skin with an of 90%. The OPL quantifies the total distance traveled by each detected photon within the tissue. The plot reveals a slight reduction in OPL for dark skin at the red wavelength, with the median (IQR) decreasing from 1.65 (1.29) to 1.47 (1.02) cm. At the IR, the median (IQR) OPL for dark skin is 1.26 (1.0) cm, which is comparable to the median (IQR) OPL for light skin at 1.24 (0.95) cm. 3.2.Reflection ModeIn reflection mode, the light source was located at the origin, and the detector was slightly offset from the origin at (0, 0, ) to simulate a case of central illumination and detection [see Fig. 2(b)]. Figure 6 shows normalized fluence rate plots of the detected photons in reflection mode at an oxygen saturation of 90%, revealing similar observations to those seen in transmission mode (Fig. 3). Here, the increased absorption of red light in darker skin leads to reduced light penetration within the tissue. As seen in the transmission mode case, horizontal and vertical line profiles drawn at and , respectively, for both skin types corresponding at the two wavelengths are presented in Figs. 6(e) and 6(f). The magnitude of the fluence rate () is higher for the IR wavelength compared with the red wavelength. For instance, at a depth of 0.1 cm, the maximum fluence rate is as follows: IR, light skin , light skin , dark skin , dark skin (0.06). This implies that the tissue penetration depth is greatest in IR for light skin, followed by red for light skin, IR for dark skin, and red for dark skin. The detected light intensities, PIs, and curves for light and dark skin tissue models in reflection mode are presented in Fig. 7. Although the detected light intensities in reflection mode were significantly higher compared with those in transmission mode (36 times higher for IR and 18 times higher for red at an of 90%), the observed trends with varying oxygen saturation for both light and dark skin, as illustrated in Figs. 7(a) and 7(b), show similar characteristics to the transmission mode plots [Figs. 4(a) and 4(b)]. Interestingly, as observed from Figs. 7(c) and 7(d), variations in skin tone did not have a large effect on the PI at IR wavelengths but caused a significant reduction in the red PI for darker skin tones across varying levels. This is confirmed by the very small Euclidean distance between the PIs of light and dark skin at IR () compared with the larger Euclidean distance at red (). As an example, at , the mean (std) of PI at red is () in light skin, which decreased to () in dark skin. At the same saturation, the mean (std) of the PI at IR is () in light skin and () in dark skin. Contrary to the curves seen in transmission mode, the reflection mode shows significant variations between light and dark skin, as seen in Fig. 7(e) (). For instance, at an value of 0.25 (dashed horizontal line), the actual measured for dark skin is 90%, which is lower compared with a value exceeding 90% shown for light skin. This indicates overestimation in dark skin if a calibration curve originally derived from light skin is used. 3.3.Varying Source–Detector (SD) Separation—Reflection ModeTo further explore the effects of varying SD separation on light and dark skin tissue models, simulations were conducted at three different separations: small (0.1 cm), medium (0.25 cm), and large (0.45 cm). As the SD separation increased, the PI values at IR and red also rose in both light and dark skin types due to greater DC light attenuation with larger SD separations, as depicted in Figs. 8(a), 8(b), 9(a), 9(b), 10(a), and 10(b). The IR PI values showed minimal differences between light and dark skin types for all SD separations considered (Euclidean distance: small , medium , and large ). However, slightly larger variations in red PI values were seen between light to dark skin due to the change in SD separation (Euclidean distance: small , medium , and large ). At a smaller SD distance, dark skin exhibited a relatively lower PI with a mean (std) of () at red compared with light skin PI of () at 90% saturation, as shown in Figs. 8(a) and 8(b). At the largest SD separation, the red and IR PIs in both light and dark skin became comparable [as shown in Figs. 10(a) and 10(b)] with Euclidean distances of and , respectively, at 90% saturation. Further, the variations in curves between light and dark skin became less pronounced as the SD separation increased, as illustrated in Figs. 8(c), 9(c), and 10(c). This trend is statistically confirmed by the increasing -values found with greater SD separations (-value: small , medium , and large ). The decrease in intensity in dark skin relative to light skin, expressed as a fraction, was calculated at a particular value of 90% to compare relative attenuation factors under different simulated configurations. This is summarized in Table 6. Mathematically, attenuation in dark skin with respect to light skin is expressed as shown in Eq. (8). Table 6Summary of AC and DC relative attenuation factors in dark skin with respect to light skin at red and IR wavelengths for SpO2=90% across different configurations.
Figure 11 displays the median (IQR) OPL and PD at for both light and dark skin, measured at diastolic red and IR. The -axis shows varying SD separations, and the -axis represents the median (IQR) of the OPL and PD in Figs. 11(a) and 11(b), respectively. Both the median OPL and median PD at red and IR increase with an increase in SD separation. At the IR wavelength, variations in median OPL between the two skin types are similar in all configurations (Euclidean distance: zero , small , medium , and large ), whereas at red, differences in median OPL was found to be increasing with increase in SD separation (Euclidean distance: zero , small , medium , and large ). At small SD separation, the median OPL in light skin at red was similar to light skin at IR (Euclidean distance: ); however, as the separation increased, the median OPL at red exceeded that of IR (Euclidean distance: medium and large ). With respect to PD, IR consistently penetrated deeper than red across all SD separations as seen in Fig. 11(b). The median PD for IR was similar between light and dark skin across varying SD separations (Euclidean distance: small , medium , and large ). In contrast to IR, the median PD at red light in dark skin decreased due to relatively greater absorption compared with light skin (Euclidean distance: small , medium , and large ). 4.DiscussionThe effect of skin tones on curves has been investigated using a four-layer finger tissue Monte Carlo simulation. The main aim of this paper was to gain insights into the origin of the problem of overestimation that is frequently reported in highly pigmented subjects. Inaccurate high readings can pose significant risks, resulting in delayed or denied disease detection and clinical treatment. Hence, comprehending the factors contributing to this issue is vital to enable solutions that could be devised to ensure the equitable performance of POs regardless of skin color. Initially, a transmission mode PO configuration was studied under the effects of varying melanin in the epidermal layer. As the level increased, there was a gradual decrease and increase in the detected intensities at IR and red wavelengths, respectively [Figs. 4(a) and 4(b)]. This trend was attributed to higher concentrations of at elevated levels, resulting in increased and decreased absorption of IR and red, respectively, thereby affecting the detected light levels. Furthermore, diastolic intensities were observed to be higher than systolic intensities at both wavelengths due to the reduced blood volume during diastole, which led to the reduction in light absorption. These trends observed in detected light intensities as a function of are consistent with the findings from previous MC studies.16,19 As melanin’s absorption and scattering properties increased, there was a decrease in absolute light intensity at both wavelengths. However, the relative attenuation at AC red (denoted by ) and DC red (denoted by ) was greater than at AC IR (denoted by ) and DC IR (denoted by ), respectively. This indicates that melanin preferentially absorbs red light more than IR (as is observed in Fig. 3). The OPL distributions from a sample of 10,000 detected photons showed that median OPL at red decreased in dark skin, whereas in IR, it was similar to that of light skin (Fig. 5). Decrease in the median OPL at the red wavelength implies that the effect of absorption dominates the effect of scattering in dark skin. Moreover, it was observed that both the AC (difference between diastole and systole intensities) and the DC components at red and IR were relatively attenuated by the same factor ( and ) in dark skin (Table 6, transmission mode). Consequently, this uniform attenuation resulted in no change in PI for red and IR [Figs. 4(c) and 4(d)], leading to similar values for and [Fig. 4(e)]. Contrary to the results in transmission mode, in reflection mode simulation geometry, was overestimated in darker skin as the optical properties in the epidermal layer varied. The detected light intensities [Figs. 7(a) and 7(b)] as a function of exhibited a trend similar to that observed in transmission mode [Figs. 4(a) and 4(b)], whereas the absolute detected intensities in this geometry were significantly higher for both light and dark skin compared with those in transmission mode. This difference can be attributed to the central illumination and detection geometry, which results in shallower penetration of light within the tissue. Consequently, a greater amount of reflected light is detected. The simulations revealed that the PI at IR remained consistent between light and dark skin, whereas the PI at red exhibited variations between the two skin types [Figs. 7(c) and 7(d)]. This can be explained due to the equal/similar attenuation of AC and DC at IR () and unequal attenuation at the red in dark skin, with the AC attenuation factor exceeding the DC attenuation factor (). This differential attenuation at the red wavelength (Table 6, central illumination and detection) resulted in reduced PI for dark skin compared with light skin, causing overestimation [Fig. 7(e)]. To understand further the effects of SD separation in reflection mode on varying skin tones, three different SD configurations (small SD, medium SD, and large SD) were employed. The simulations illustrated that, with an increase in SD distance, the disparities in curves due to skin color became less pronounced [Figs. 8(c), 9(c), and 10(c)]. The largest SD separation distance [Fig. 10(c)] yielded results that aligned closely with those observed in transmission mode [Fig. 4(e)]. At smaller SD distances, light penetration into the tissue is limited, causing photons to interact longer with the epidermal layer. This increased interaction resulted in relatively higher absorption and scattering of DC light at both wavelengths compared with longer SD separations. The increased scattering led to a higher amount of DC light being reflected, whereas increased absorption contributed to reduced light interaction with the arterial layer, thereby increasing AC attenuation. This effect is more pronounced at the red wavelength due to its higher optical properties of melanin, which resulted in unequal attenuation between red AC and DC (). As the SD separation distance increased, so did the PD, leading to a decrease in DC light reflection (DC attenuation factor increases) due to greater interactions along the photons’ paths to the detector, i.e., longer OPL. This phenomenon resulted in equal attenuation for both DC and AC components at red wavelengths (); hence, there is no change in PI for red in dark skin relative to light skin [Figs. 10(a) and 10(b)]. It can be inferred that both the absorption and scattering effects of the epidermal layer contribute to the overestimation of , with absorption having a greater effect than scattering in dark skin. This is evident from the observation that median OPL and median PD decrease in dark skin relatively at the red wavelength by a greater amount across all SD separation distances compared with light skin [Figs. 11(a) and 11(b)]. If scattering dominated, then the median OPL in dark skin would be higher than in light skin at the red wavelength. In this study, overestimation in dark skin was not observed when using a transmission-based tissue model. This does not imply that melanin affects PO readings in reflection mode sensors only. Unfortunately, retrospective studies that highlighted overestimation in dark-skinned individuals did not specify the PO manufacturer or whether transmission or reflection mode POs were used in their research, although it is likely that the majority would be transmission mode.7,11 Reflection mode sensors theoretically can be applied to any vascular area, whereas transmission mode devices are generally restricted to smaller extremities, such as toes, fingers, or earlobes.29 In medical settings, transmission mode POs are commonly used devices due to their high accuracy and stability compared with reflection mode.30 More recently, a prospective study involving healthy adult volunteers with desaturation, which used two transmission finger pulse oximeters (Masimo Radical 7 and Nellcor N-595) under low perfusion conditions, reported a missed diagnosis of hypoxemia in dark-pigmented skin types.31 The discrepancies between our findings and those of other published clinical studies in transmission mode geometry could be attributed to limitations in both sets of studies. In previously published clinical studies, variations in experimental protocols, differences in skin color between subjects, and the absence of information on the PO used and its type may have contributed to this difference. Alternatively, differences could be attributed to some of the following limitations.
5.ConclusionA Monte Carlo-based finger tissue model that includes four layers has been developed to investigate the effect of varying optical properties of melanin on values. The simulations were carried out in two different PO configurations: transmission mode and reflection mode. In transmission mode, uniform attenuation of AC and DC components at red and infrared wavelengths resulted in no change in the “ratio of ratios” or . This does not conclude that melanin does not affect oxygen saturation readings in transmission mode. Further complex MC models are essential to confirm this result. By contrast, in reflection mode, significant variations in curves were observed between dark and light skin tissue models. Differential attenuation of red AC and DC components and uniform attenuation of IR AC and DC components led to reduced red perfusion index and unaltered infrared perfusion index in dark skin relative to light skin. This caused to decrease and to increase. Finally, with a change in source–detector separation distance, variations in curves due to skin tone were more significant at shorter distances. As the SD separation distance increased, the effect of skin tone on curves diminished. Code and Data AvailabilityThe material support for this study was obtained through the open-source Monte Carlo simulation platform MCmatlab, accessible at https://github.com/ankrh/MCmatlab. All data generated during this study are available from the University of Nottingham data repository at https://doi.org/10.17639/nott.7429. AcknowledgmentsThis work was supported by the Medical Research Council (United Kingdom) (Grant No. MR/T025638/1). Suvvi K. Narayana Swamy is funded by a PhD studentship from the Engineering and Physical Sciences Research Council (United Kingdom) (Grant No. EP/T517902/1) and a studentship award from SurePulse Medical Ltd. We extend our sincere gratitude to Dr. Anders Kragh Hansen, co-creator and owner of MCmatlab, for his invaluable assistance in this open-source software platform. ReferencesT. A. Neff,
“Routine oximetry: a fifth vital sign?,”
Chest, 94
(2), 227 https://doi.org/10.1378/chest.94.2.227a CHETBF 0012-3692
(1988).
Google Scholar
J. Allen,
“Photoplethysmography and its application in clinical physiological measurement,”
Physiol. Meas., 28
(3), R1 https://doi.org/10.1088/0967-3334/28/3/R01 PMEAE3 0967-3334
(2007).
Google Scholar
E. D. Chan, M. M. Chan, M. M. Chan,
“Pulse oximetry: understanding its basic principles facilitates appreciation of its limitations,”
Respir. Med., 107
(6), 789
–799 https://doi.org/10.1016/j.rmed.2013.02.004 RMEDEY 0954-6111
(2013).
Google Scholar
, “Medical electrical equipment. Particular requirements for basic safety and essential performance of pulse oximeter equipment,”
https://bsol.bsigroup.com
().
Google Scholar
A. Alboksmaty et al.,
“Effectiveness and safety of pulse oximetry in remote patient monitoring of patients with COVID-19: a systematic review,”
Lancet Digit. Health, 4
(4), e279
–e289 https://doi.org/10.1016/S2589-7500(21)00276-4
(2022).
Google Scholar
A. Rahman et al.,
“Silent hypoxia in COVID-19: pathomechanism and possible management strategy,”
Mol. Biol. Rep., 48
(4), 3863
–3869 https://doi.org/10.1007/s11033-021-06358-1 MLBRBU 0301-4851
(2021).
Google Scholar
M. W. Sjoding et al.,
“Racial bias in pulse oximetry measurement,”
N. Engl. J. Med., 383
(25), 2477
–2478 https://doi.org/10.1056/NEJMc2029240 NEJMAG 0028-4793
(2020).
Google Scholar
G. W. Burnett et al.,
“Self-reported race/ethnicity and intraoperative occult hypoxemia: a retrospective cohort study,”
Anesthesiology, 136
(5), 688
–696 https://doi.org/10.1097/ALN.0000000000004153 ANESAV 0003-3022
(2022).
Google Scholar
A. Fawzy et al.,
“Racial and ethnic discrepancy in pulse oximetry and delayed identification of treatment eligibility among patients with COVID-19,”
JAMA Internal Med., 182
(7), 730
–738 https://doi.org/10.1001/jamainternmed.2022.1906
(2022).
Google Scholar
E. R. Gottlieb et al.,
“Assessment of racial and ethnic differences in oxygen supplementation among patients in the intensive care unit,”
JAMA Internal Med., 182
(8), 849
–858 https://doi.org/10.1001/jamainternmed.2022.2587
(2022).
Google Scholar
C. J. Crooks et al.,
“Differential pulse oximetry readings between ethnic groups and delayed transfer to intensive care units,”
QJM Int. J. Med., 116
(1), 63
–67 https://doi.org/10.1093/qjmed/hcac218
(2022).
Google Scholar
S. K. Alla, J. F. Clark and F. R. Beyette,
“Signal processing system to extract serum bilirubin concentration from diffuse reflectance spectrum of human skin,”
in Annu. Int. Conf. IEEE Eng. in Med. and Biol. Soc.,
1290
–1293
(2009). https://doi.org/10.1109/IEMBS.2009.5333236. Google Scholar
D. He et al.,
“A single-chip CMOS pulse oximeter with on-chip lock-in detection,”
Sensors, 15
(7), 17076
–17088 https://doi.org/10.3390/s150717076 SNSRES 0746-9462
(2015).
Google Scholar
S. K. N. Swamy et al.,
“Evaluation of pulse oximeter performance under varying skin tones in bench tests and using Monte Carlo simulations,”
Proc. SPIE, 12833 1283304 https://doi.org/10.1117/12.3000705 PSISDG 0277-786X
(2024).
Google Scholar
S. K. N. Swamy et al.,
“Pulse oximeter bench tests under different simulated skin tones,”
Med. Biol. Eng. Comput., https://doi.org/10.1007/s11517-024-03091-2 MBECDY 0140-0118
(2024).
Google Scholar
S. Chatterjee and P. A. Kyriacou,
“Monte Carlo analysis of optical interactions in reflectance and transmittance finger photoplethysmography,”
Sensors, 19
(4), 789 https://doi.org/10.3390/s19040789 SNSRES 0746-9462
(2019).
Google Scholar
S. Chatterjee, K. Budidha and P. A. Kyriacou,
“Investigating the origin of photoplethysmography using a multiwavelength Monte Carlo model,”
Physiol. Meas., 41
(8), 084001 https://doi.org/10.1088/1361-6579/aba008 PMEAE3 0967-3334
(2020).
Google Scholar
R. Al-Halawani, S. Chatterjee and P. A. Kyriacou,
“Monte Carlo simulation of the effect of human skin melanin in light-tissue interactions,”
in 44th Annu. Int. Conf. IEEE Eng. Med. & Biol. Soc. (EMBC),
1598
–1601
(2022). https://doi.org/10.1109/EMBC48229.2022.9871350 Google Scholar
M. Bolic,
“Simulating the effects of melanin and air gap depth on the accuracy of reflectance pulse oximeters,”
in BIODEVICES 2023 - 16th Int. Conf. Biomed. Electron. and Devices,
(2023). https://doi.org/10.5220/0011749300003414 Google Scholar
M. S. Arefin, A. P. Dumont and C. A. Patil,
“Monte Carlo based simulations of racial bias in pulse oximetry,”
Proc. SPIE, 11951 1195103 https://doi.org/10.1117/12.2610483 PSISDG 0277-786X
(2022).
Google Scholar
D. Marti et al.,
“MCmatlab: an open-source, user-friendly, MATLAB-integrated three-dimensional Monte Carlo light transport solver with heat diffusion and tissue damage,”
J. Biomed. Opt., 23
(12), 121622 https://doi.org/10.1117/1.JBO.23.12.121622 JBOPFO 1083-3668
(2018).
Google Scholar
L. Hernández-Quintanar et al.,
“Fiber-optic pulse oximeter for local oxygen saturation determination using a Monte Carlo multi-layer model for calibration,”
Comput. Methods Programs Biomed., 187 105237 https://doi.org/10.1016/j.cmpb.2019.105237 CMPBEK 0169-2607
(2020).
Google Scholar
K. Tichauer et al.,
“50 years after pulse oximetry was invented: pulse oximetry revisited,”
https://www.spiedigitallibrary.org/jbo-hot-topics-webinar-series
(2022).
Google Scholar
T. Boonya-Ananta et al.,
“Monte Carlo analysis of optical heart rate sensors in commercial wearables: the effect of skin tone and obesity on the photoplethysmography (PPG) signal,”
Biomed. Opt. Express, 12
(12), 7445
–7457 https://doi.org/10.1364/BOE.439893 BOEICL 2156-7085
(2021).
Google Scholar
S. Prahl and S. Jacques,
“mcxyz.c: how to use make TissueList.m to create a library of tissue optical properties,”
https://omlc.org/software/mc/mcxyz/index.html
().
Google Scholar
S. L. Jacques,
“Optical properties of biological tissues: a review,”
Phys. Med. Biol., 58
(11), R37 https://doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155
(2013).
Google Scholar
X. U. Zhang et al.,
“Multidiameter single-fiber reflectance spectroscopy of heavily pigmented skin: modeling the inhomogeneous distribution of melanin,”
J. Biomed. Opt., 24
(12), 127001 https://doi.org/10.1117/1.JBO.24.12.127001 JBOPFO 1083-3668
(2019).
Google Scholar
A. Herbst et al.,
“Oxygen saturation targets for adults with acute hypoxemia in low and lower-middle income countries: a scoping review with analysis of contextual factors,”
Front. Med., 10 1148334 https://doi.org/10.3389/fmed.2023.1148334 FMBEEQ
(2023).
Google Scholar
T. Leppänen et al.,
“Pulse oximetry: the working principle, signal formation, and applications,”
Advances in the Diagnosis and Treatment of Sleep Apnea: Filling the Gap between Physicians and Engineers, 205
–218 Springer(
(2022). Google Scholar
H. Lee, H. Ko and J. Lee,
“Reflectance pulse oximetry: practical issues and limitations,”
ICT Express, 2
(4), 195
–198 https://doi.org/10.1016/j.icte.2016.10.004
(2016).
Google Scholar
M. K. Gudelunas et al.,
“Low perfusion and missed diagnosis of hypoxemia by pulse oximetry in darkly pigmented skin: a prospective study,”
Anesth. Analg., 138
(3), 552
–561 https://doi.org/10.1213/ANE.0000000000006755
(2022).
Google Scholar
“Biomonitoring sensor: BIOFY SFH7072 data sheet,”
https://ams-osram.com/products/multi-chips/optical-modules/osram-biofy-sfh-7072
(2022).
Google Scholar
BiographySuvvi K. Narayana Swamy is a third year PhD student in the Department of Electrical and Electronic Engineering at the University of Nottingham (UoN). She received her Bachelor of Technology degree in electronics and communications in 2018 from Alliance University, India, and her master’s in advanced electrical and electronic engineering with an extended research project from the UoN in 2021. Her current research interests include medical devices, fiber optics, biosensors, and signal processing. Chong Liu is a post-doctoral researcher affiliated with the Faculty of Engineering at the University of Nottingham. He mainly focuses on researching optical-electronic medical/healthcare sensors and physiological signal applications. Ricardo Correia is an assistant professor in optical fiber sensing at the University of Nottingham. As an instrumentation engineer specializing in the development of fiber optic sensors, his research focuses on the development and application of fiber optic sensors for a wide range of applications that range from healthcare to civil engineering and aerospace. His main interest is in the development and application of fiber optic sensors for physical measurements. Barrie R. Hayes-Gill is a professor of medical devices and electronic systems at the University of Nottingham. His main research interests include optoelectronic design, signal processing, technology transfer, and medical device regulations. Stephen P. Morgan is a professor of biomedical engineering at the University of Nottingham. His research involves the development of medical devices with sensing capability. For example, he is currently developing a novel endotracheal tube that can monitor the microcirculation at the cuff/trachea interface and a smart wound dressing for monitoring the healing rate and infection. His work has involved close collaboration with industry partners facilitated by a Royal Society Industry Fellowship. He is the director of the Centre for Healthcare Technologies, which aims to bring together key stakeholders, capabilities, and expertise to support the rapid translation of scientific discoveries into healthcare adoption. |
Skin
Tissues
Reflection
Oxygen
Signal attenuation
Monte Carlo methods
Simulations