28 December 2023 Deep learning-based breast region segmentation in raw and processed digital mammograms: generalization across views and vendors
Sarah D. Verboom, Marco Caballo, Jim Peters, Jessie Gommers, Daan van den Oever, Mireille J. M. Broeders, Jonas Teuwen, Ioannis Sechopoulos
Author Affiliations +
Abstract

Purpose

We developed a segmentation method suited for both raw (for processing) and processed (for presentation) digital mammograms (DMs) that is designed to generalize across images acquired with systems from different vendors and across the two standard screening views.

Approach

A U-Net was trained to segment mammograms into background, breast, and pectoral muscle. Eight different datasets, including two previously published public sets and six sets of DMs from as many different vendors, were used, totaling 322 screen film mammograms (SFMs) and 4251 DMs (2821 raw/processed pairs and 1430 only processed) from 1077 different women. Three experiments were done: first training on all SFM and processed images, second also including all raw images in training, and finally testing vendor generalization by leaving one dataset out at a time.

Results

The model trained on SFM and processed mammograms achieved a good overall performance regardless of projection and vendor, with a mean (±std. dev.) dice score of 0.96±0.06 for all datasets combined. When raw images were included in training, the mean (±std. dev.) dice score for the raw images was 0.95±0.05 and for the processed images was 0.96±0.04. Testing on a dataset with processed DMs from a vendor that was excluded from training resulted in a difference in mean dice varying between 0.23 to +0.02 from that of the fully trained model.

Conclusions

The proposed segmentation method yields accurate overall segmentation results for both raw and processed mammograms independent of view and vendor. The code and model weights are made available.

1.

Introduction

Automatic image processing and analysis of digital mammograms (DMs), such as computer-aided detection (CAD), CAD diagnosis, and breast density estimation, have become an important part of breast care.13 Most of these methods require a prior segmentation of the mammogram into background, breast, and pectoral muscle areas. The segmentation of the breast is a necessary step to extract image features only from the breast, avoiding the background and, if depicted, the pectoral muscle. If not correctly removed, the pectoral muscle could introduce important biases in any automated system aimed at the analysis of the breast tissue.1,4

Several methods have been proposed to segment mammograms. Most of these methods are designed for digitized screen-film mammograms or raw (for processing) DMs58 and mainly focus on only the mediolateral oblique (MLO) view.5,7 Screen-film mammography has widely been replaced by digital mammography, making methods that work solely for digitized screen-film mammography less relevant. Raw (for processing) DMs are often used for physics-based analysis, such as for breast density quantization, but they are not saved in clinical and screening practice, making segmentation methods dedicated to raw mammograms of limited use (especially in large, retrospective studies). Therefore, segmentation methods able to analyze processed (for presentation) DMs, instead of raw data, are needed. However, this shift in data type introduces a new challenge in generalizing across different manufacturers. This is because different vendors use different postprocessing algorithms to convert the raw x-ray projection data to DMs for display, resulting in vendor specific differences in processed images with respect to gray levels, contrast, and texture.9 Therefore, a segmentation method that is intended for processed DMs should be designed to handle these differences to generalize well across DMs of different vendors. Finally, although most previously developed methods have focused on only the MLO view, the cranio-caudal (CC) view is also acquired during a standard mammographic examination. In contrast to the MLO view, this view does not always depict the pectoral muscle, and when it is present, it is often small. This makes the automated segmentation of the pectoral muscle in CC images challenging. However, there are cases in which the pectoral muscle is clearly depicted in the CC view and can therefore influence image analysis. For these cases, therefore, it would be beneficial to automatically detect the pectoral muscle also in CC images.

Therefore, considering the limitations in existing methods, the aim of this study is to develop a model to segment the breast region and the pectoral muscle that is applicable to both raw and processed DMs, designed to generalize across models of different vendors and across both CC and MLO views. The code and trained model weights of the model are available at https://github.com/radboud-axti/maseg.

2.

Methods

A convolutional neural network (CNN) was trained in a supervised fashion to segment mammograms into background, breast, and pectoral muscle areas. For this, eight different datasets were used to train and test the model on mammograms from different vendors. During training, several augmentations were used to increase the model generalization across the contrast differences among vendors. Furthermore, a hyperparameter search was done with two different losses to compensate for the pectoral muscle class imbalance introduced by training a single model for both CC and MLO views. To evaluate the appropriateness of our model, several experiments were performed; these aimed at evaluating the model performance on generalization across views and vendors, the effect of including raw images, and the potential performance of generalization to new vendors.

2.1.

Datasets

For training, validation, and testing, eight different datasets of mammographic images were used: two consisted of public datasets, mini-MIAS10 and INbreast,11 and six were datasets of different vendors. An overview of these datasets is shown in Table 1. One to 36 images from each woman are included in the datasets with a median of four images, and in the Hologic, GE, and Siemens datasets, some women had multiple exams acquired on different dates. All datasets consist of 50% MLO images, except the INbreast dataset, which consists of only MLO images. Of the eight datasets, four also include the corresponding raw images. All images were linearly resampled to have a pixel spacing of 400  μm.

Table 1

Description of the eight datasets that were used for training, validation, and testing. INbreast and mini-MIAS are publicly available datasets. The mini-MIAS dataset contains only CC images.

No. imagesNo. womenDM/SFMNo. raw imagesPectoral muscle visiblePixel spacing (μm)YearCountry
MLO (%)CC (%)
Hologic1012195DM1012993670 or 65.52004 to 2009The Netherlands
IMS Giotto821203DM821992682.92022Italy
Fuji756200DMNo9918502022Germany
GE720104DM7209841100 or 942000 to 2019The Netherlands
INbreast11 (Siemens)410108DMNo998702008 to 2010Portugal
Mini-MIAS10322161SFMNo982001994United Kingdom
Siemens26840DM26810040852001 to 2019The Netherlands
PlanMed26466DMNo10021832018 to 2021Finland
Total4573107728219928
DM, digital mammography; SFM, screen film mammography; MLO, medio-lateral oblique; and CC, cranio-caudal.

2.1.1.

Ground truth

A segmentation mask with pixel-wise ground truth for background, breast, and pectoral muscle (when present) was made using the processed mammograms and screen film mammograms (SFMs). If present, the labels were then applied to the corresponding raw images because no positional shifts are performed by any of the image processing algorithms. Segmentation masks were created in two or three steps: first, initialization of the breast boundary by Otsu thresholding;12 second, a pectoral muscle initialization with Otsu thresholding only for MLO images; and finally, a manual adjustment of the mask, as shown in Fig. 1. The pectoral muscle initialization in MLO images was done by reapplying the Otsu thresholding method after excluding the background for seven of the eight datasets. For the MLO view of the INbreast dataset,11 instead, the already publicly available annotations of the pectoral muscle were used as the initialization. Finally, each segmentation mask was checked visually and adjusted manually using ITK-SNAP 3.6.013 by one of four medical imaging scientists with experience in mammography.

Fig. 1

Annotation process for processed images with one or two initialization steps and manual adjustment.

JMI_11_1_014001_f001.png

2.1.2.

Data split

Each dataset was split randomly on a patient level into at least 100 test images, 80 validation images, and the rest in training. In total, this led to a training set of 3116 images, a validation set of 645 images, and a test set of 812 images, as shown in Table 2. The same split was used for all experiments.

Table 2

Split of each of the eight datasets into a training, validation, and test sets.

TrainValidationTest
No. womenNo. imagesNo. womenNo. imagesNo. womenNo. images
Hologic155832188022100
IMS Giotto158638208125102
Fuji152576238025100
GE75540178012100
INbreast62224228024106
mini-MIAS71142408050100
Siemens1280138415104
PlanMed2184208025100
Total7063116173645198812

2.2.

Network

2.2.1.

Architecture

The multiclass segmentation was performed using a U-Net14 with four downsampling blocks of two convolutional layers each with a kernel size of three, and a max-pooling layer with a kernel size of two and stride of two. The initial number of convolutional filters was set to 64 and doubled at each scale. The four upsampling steps consisted of a bilinear upsampling, a convolutional layer with a kernel size of unity to reduce the number of filters by a factor of two, and two convolutional layers with a kernel size of three. Before the final SoftMax layer, a convolutional layer (three filters, unity kernel size) was included. Dropout regularization (probability of 0.5) was applied in this layer to allow for an improved generalization and to improve the training optimization of the deeper layers. All activation functions were rectified linear units, and weights were initialized with Kaiming uniform initialization.

2.2.2.

Augmentations

Processed DMs of different vendors have different intensity and contrast properties mainly due to differences in processing of the raw images. To generalize over these different properties, several types of data augmentation were used as a regularizer during training, as shown in Table 3. For each training image, a random look-up table was chosen from the choices provided in its DICOM header, with a jitter varying the window center and window width slightly. This was followed by a random horizontal flip, a random rotation, and an elastic deformation.15 After these augmentations, a random cropping of the image of 512×512  pixels was performed. The cropping was allowed to go at most 64 (12.5%) pixels outside of the original image boundary and, where necessary, the images was padded with zeros. Finally, a random gamma transform was applied, and Gaussian noise was added. Two examples of the result of these augmentations on mammograms are shown in Fig. 2.

Table 3

Augmentations that were performed during training with their corresponding parameter ranges.

AugmentationParameters
Random lookup table±5% of window center and window width
Random horizontal flipProbability = 0.5
Random rotation−5 deg to +5 deg
Elastic deformation15α=1, σ=100 and αaffine=20
Random cropping512 × 512 pixels
Random gamma transformγ(0.5,1.5)
Random Gaussian noiseStandard deviation 5% of image range

Fig. 2

Examples of two training images after all data augmentation steps. The examples show the variety in gray level intensity and contrast between the images.

JMI_11_1_014001_f002.png

2.2.3.

Losses

If present in the CC view, the pectoral muscle region is often small, comprising a very small area of the overall image and of the breast. To account for this class imbalance, two different weighted losses were used to train the network: weighted cross entropy and weighted focal loss.

The weighted cross-entropy loss is defined as

Eq. (1)

cross-entropy loss(pt)=c=13αclogpt,c,

Eq. (2)

pt,c={pc,y=c1pc,otherwise,
where pc denotes the probability of a pixel belonging to class c and y is the ground truth of that pixel. Hyperparameters αc weigh the classes, giving a higher or lower weight to a certain class c.

The focal loss16 is a variation of the weighted cross-entropy loss, in which high confidence predictions are weighted lower than low confidence predictions by tuning a hyperparameter (γ). The focal loss is defined as

Eq. (3)

focal loss(pt)=c=13αc(1pt,c)γlogpt,c,
with hyperparameters αc and γ (γ=1 corresponds to cross-entropy loss). In this study, we used γ=2, as was suggested in the original paper.14

The model was trained with the entire training set with both losses and with α=[11αpectoral] such that αpectoral{1,1.5,2,2.5}. Training was stopped when the validation loss did not improve for 10 epochs. The final loss and αpectoral were chosen based on the highest segmentation dice coefficient on the entire validation set for the pectoral class and were not modified further when assessing the model performance on the test set.

2.3.

Experiments

2.3.1.

Comparing views and vendors

The model was trained with all eight training datasets including only the processed (and digitized screen film) images, totaling 3116 images using the selected loss and hyperparameters. The segmentation performance of the network was tested on the processed test images of all eight datasets, with the look-up table provided in the DICOM header as the “normal” setting, with no further test augmentation. The performance was quantified by the dice coefficient for each class c present in the image, defined as

Eq. (4)

DICE(c)=2·|y^cyc||y^c|+|yc|,
which compares the class segmentation estimate y^c to the class ground truth yc. An overall dice coefficient was computed for each image by averaging the dice coefficients for the breast and pectoral areas weighted by the number of pixels of the corresponding class in the ground truth.

Dice coefficients for CC and MLO view test set images were compared with an unpaired t-test. The class and average dice coefficients of the eight datasets were compared using a one-way ANOVA test and Tukey’s method as a post hoc test. All tests were performed with SciPy, and a significance level of 0.05 was used.

As discussed, not all mammograms depict the pectoral muscle. This could lead to the model segmenting the pectoral muscle when it is not present, a false positive, or not segmenting a pectoral muscle when it is present, a false negative. Therefore, in addition to the dice coefficient, the performance in pectoral muscle segmentation was also quantified in terms of the number of false positive and false negative cases. A false positive was defined as an output in which some pixels are classified as pectoral muscle when there is no pectoral muscle present in the ground truth. A false negative was defined as an output in which no pectoral muscle is segmented when there is a pectoral muscle present in the ground truth.

2.3.2.

Including raw images in training

In the first experiment, the model was trained only on processed and SFM images. To expand the applicability of the model to raw images also, these were included during training in the second experiment. For this, the model was retrained on the same training set, but additionally including all 2090 available raw images of the training set. The dice coefficients of the 812 processed and 406 raw test images for both models (processed-only training and processed + raw training) were compared with a paired t-test with a significance level of 0.05.

2.3.3.

Test for vendor generalization

A third experiment was performed to assess the potential ability of the model to generalize to images from a possible new, unseen vendor or images processed with a new vendor processing algorithm. For this, eight different models were trained, each time including all processed-image-only training sets except for one, the excluded dataset. The resulting model was then tested on the training and test sets of the excluded dataset. For each model, the resulting dice coefficients were compared with the dice coefficients of the test set of the datasets that were used for training with an unpaired t-test. The significance level of 0.05 was adjusted for multiple comparisons with a Bonferroni correction to 0.05/8.

2.3.4.

Comparison with Volpara commercial segmentation algorithm

The research version of a commercially available breast density quantification software (Volpara version 3.4, Volpara Health, Wellington, New Zealand) was used to generate a segmentation of all raw images. Volpara returns a segmentation with four classes: background, pectoral muscle, breast tissue with a constant thickness, and breast tissue with a varying thickness. To compare these segmentations to the ground truth masks, both breast tissue classes were combined into one breast tissue class. The resulting dice coefficients were compared with the performance of the model that was trained on both processed and raw images applied to the raw images and corresponding processed images with a dependent t-test.

3.

Results

The cross-entropy loss with αpectoral=2.0 resulted in the highest pectoral and overall dice coefficient in the validation set (as shown in Table 4) and was thus used for all following models. Higher values of αpectoral resulted in over-segmentation of the pectoral muscle, whereas lower values resulted in under-segmentation. All following models were trained with a cross-entropy loss with αpectoral=2.0.

Table 4

Dice coefficients for the two losses with different values of αpectoral for the validation set. The pectoral muscle dice was calculated only for the 445/645 images in which the pectoral muscle was visible. Results are given in mean ± standard deviation.

LossαpectoralBreastPectoral
Cross-entropy loss1.00.94 ± 0.070.56 ± 0.32
1.50.93 ± 0.050.58 ± 0.27
2.0a0.93 ± 0.050.63 ± 0.26
2.50.94 ± 0.060.60 ± 0.29
Focal loss1.00.92 ± 0.080.23 ± 0.29
1.50.96 ± 0.050.60 ± 0.34
2.00.93 ± 0.100.59 ± 0.31
2.50.93 ± 0.070.45 ± 0.26

aSelected hyperparameter for training.

3.1.

Comparison of Views and Vendors

Table 5 shows the overall and view-specific results from the first experiment. Figures 3(a) and 3(b) show representative examples of segmentation outputs for both the MLO and CC views, respectively. Figure 3(d) shows an example from the mini-MIAS dataset in which the model mistakes the white edges for breast tissue and pectoral muscle, resulting in a low dice coefficient. Figure 3(e) shows a CC example with a low dice coefficient.

Table 5

Dice coefficients on the test set for different views. Results are given in mean ± standard deviation. The number of images that contain the pectoral muscle class in the ground truth is given.

ViewnBreastPectoral (n)Overall
MLO4560.96 ± 0.040.83 ± 0.20 (451)0.94 ± 0.07
CC3560.98 ± 0.020.50 ± 0.35 (93)0.98 ± 0.02
All8120.97 ± 0.030.78 ± 0.26 (544)0.96 ± 0.06
MLO, medio-lateral oblique and CC, cranio-caudal.

Fig. 3

Example of segmentation outputs (a) of a CC view with a high overall dice (0.98) from GE, (b) MLO view with a high overall dice (0.99) from Hologic, (c) MLO view with a median overall dice (0.95) and split pectoral segmentation from Inbreast, (d) MLO view with a low overall dice (0.80) from mini-MIAS CC view with a low overall dice (0.94) from GE, and (f) CC view with a median overall dice (0.99) but false positive pectoral muscle from Hologic. The red line indicates the background-breast edge, and the blue line indicates the breast-pectoral muscle edge.

JMI_11_1_014001_f003.png

The overall dice coefficient differed significantly between MLO (0.94±0.07) and CC (0.98±0.02) views, p<0.001, as shown in Table 5. The segmentation of the breast class yielded higher dice coefficients for CC compared with MLO views (p<0.001), and the pectoral muscle dice coefficient was lower for CC than for MLO views (p<0.001).

When the pectoral muscle was present, it was missed in 25/93 (26.9%) CC images, as shown in Table 6. When the pectoral muscle was not missed, the dice of the pectoral muscle in the CC views was 0.68±0.21 (n=68). The pectoral muscle was never missed in MLO images. False positive pectoral muscle segmentation larger than 1% of the breast area occurred in 28/263 (10.6%) of CC images without a pectoral muscle.

Table 6

False negative and false positive rates for the pectoral muscle in different views. The size of the pectoral muscle is given in mean percentage of the breast area ± standard deviation.

ViewnPectoral presentPectoral sizeFalse negativesFalse positives
MLO45698.9% (451/456)21.1% ± 13.6%0.0% (0/451)100.0% (5/5)
CC35626.1% (93/356)4.1% ± 3.2%26.9% (25/93)44.5% (117/263)
MLO, medio-lateral oblique and CC, cranio-caudal.

The mean overall dice coefficient of the different datasets differed significantly (p<0.001) and ranged from 0.86 to 0.98, as shown in Table 7. The post hoc analysis showed that mini-MIAS was significantly different from all others (p<0.001) with the lowest overall dice coefficient (mean ± std. dev) of 0.86±0.09. When only including DMs, the overall dice coefficient (mean ± std. dev) was 0.97±0.03. Furthermore, the post hoc analysis also showed that the overall dice coefficients of Siemens and INbreast were statistically different (p=0.017).

Table 7

Dice coefficients of the test set for each of the eight datasets, the total test set, and all DMs from the test set, which include all datasets excluding mini-MIAS. Results are given in mean ± standard deviation. The pectoral muscle dice was calculated only for images in which the pectoral muscle was visible. The number of images that contain the pectoral muscle class in the ground truth is given in (n).

DatasetnBreastPectoral (n)Overall
Hologic1000.98 ± 0.010.81 ± 0.30 (68)0.98 ± 0.02
IMS Giotto1020.98 ± 0.020.79 ± 0.31 (64)0.97 ± 0.04
Fuji1000.98 ± 0.020.87 ± 0.17 (58)0.98 ± 0.03
GE1000.97 ± 0.020.81 ± 0.29 (65)0.97 ± 0.03
INbreast1060.99 ± 0.010.81 ± 0.24 (57)0.98 ± 0.02
Mini-MIAS1000.91 ± 0.050.55 ± 0.15 (97)0.86 ± 0.09
Siemens1040.97 ± 0.020.83 ± 0.22 (76)0.96 ± 0.03
PlanMed1000.97 ± 0.020.87 ± 0.21 (59)0.97 ± 0.05
All8120.97 ± 0.030.78 ± 0.26 (544)0.96 ± 0.06
All DM7120.98 ± 0.020.83 ± 0.26 (447)0.97 ± 0.03
DM, digital mammography.

3.2.

Including Raw Images in Training

When the raw images were included in training, the overall segmentation dice for raw images improved, from an overall dice of 0.68±0.25 (mean ± std. dev.) with the processed-only-training model to an overall dice of 0.95±0.05 (mean ± std. dev.) (p<0.001), as shown in Table 8. As can also be seen, adding the raw images in training had no effect on the overall performance of the segmentation algorithm on the processed images (p=0.11).

Table 8

Dice coefficients of the test set for processed and raw images of each of the eight datasets, the total test set, and the four datasets that have raw images. Results are given in mean ± standard deviation and with the number of images (n).

DatasetnTrained with processed onlyTrained with processed and raw
ProcessedRawProcessedRaw
Hologic1000.98 ± 0.020.46 ± 0.140.97 ± 0.030.95 ± 0.04
IMS Giotto1020.97 ± 0.040.59 ± 0.270.96 ± 0.050.95 ± 0.07
Fuji1000.98 ± 0.030.97 ± 0.03
GE1000.97 ± 0.030.90 ± 0.080.96 ± 0.040.96 ± 0.03
INbreast1060.98 ± 0.020.96 ± 0.04
Mini-MIAS1000.86 ± 0.090.90 ± 0.08
Siemens1040.96 ± 0.030.76 ± 0.210.96 ± 0.050.96 ± 0.04
PlanMed1000.97 ± 0.050.96 ± 0.06
All8120.96 ± 0.060.95 ± 0.05
Datasets containing processed and raw4060.97 ± 0.030.68 ± 0.250.96 ± 0.040.95 ± 0.05

3.3.

Vendor Generalization

Of the eight models that were trained, three performed significantly better on the excluded dataset than on the included datasets, two had no significant difference, and three performed significantly worse on the excluded set, as shown in Table 9. The differences in mean overall dice were small, varying between 0.23 to +0.02. The largest decrease in performance was found for the model that had to segment the mini-MIAS dataset without having seen it during training.

Table 9

Dice coefficients when the model is trained on all processed training images of all datasets, excluding one dataset, and tested on (1) the complete test set excluding the same dataset and (2) the training and test set of the excluded dataset. Results are given in mean ± standard deviation. A Bonferroni corrected significance level of 0.05/8 = 0.00625.

TrainingTest set included datasetsTrain and Test set excluded datasetDifferencep
Excluded datasetnDicenDicen
Hologic22840.95 ± 0.077120.95 ± 0.03932+0.000.202
IMS Giotto24780.95 ± 0.077100.88 ± 0.15740−0.07a<0.001
Fuji25400.94 ± 0.077120.97 ± 0.03676+0.02a<0.001
GE25760.94 ± 0.077120.89 ± 0.10640−0.04a<0.001
INbreast28920.95 ± 0.067060.95 ± 0.07330+0.000.701
Mini-MIAS2,9740.97 ± 0.057120.74 ± 0.13242−0.23a<0.001
Siemens30360.96 ± 0.057080.97 ± 0.04184+0.01a<0.001
PlanMed30320.94 ± 0.077120.97 ± 0.04184+0.02a<0.001

aStatistically significant difference.

3.4.

Comparing with Volpara Segmentation

The mean overall dice coefficient of Volpara applied to the raw images was 0.98±0.03, which was higher than the overall dice coefficient of the model trained with both processed and raw images for the corresponding processed images (0.96±0.04) and raw images (0.95±0.05, both p<0.001).

4.

Discussion

The aim of this study was to propose a segmentation method suited for both raw and processed DMs that can generalize across CC and MLO views and across different vendors. The model provided a good overall performance regardless of raw/processed image, view, and vendor, achieving an overall dice of >0.95 on average across all of these conditions. When excluding the mini-MIAS dataset with SFMs and evaluating on only processed DMs, the mean overall dice surpassed 0.97.

Although the segmentation performance is satisfactory for both views, there was a difference between segmenting CC and MLO views, especially in the segmentation of the pectoral muscle. The performance of pectoral muscle segmentation was lower and more widely distributed in CC images compared with that of MLO images. There are three main possible explanations for this. First, the pectoral muscle was not always present in the CC views, being in only 28% of the images compared with 99% of the MLO images. Therefore, the model was trained on a much larger dataset when segmenting pectoral muscles in MLO images compared with CC images. Second, the area of the pectoral muscle, with respect to that of the breast region, was much smaller in CC views (4.7%±3.0%) compared with MLO views (21.4%±13.5%). As seen in Eq. (4), a smaller area in the annotation can make the dice coefficient more sensitive to small errors in segmentation. Third, 26.9% of the CC views with an annotated pectoral muscle were false negatives. These images have a pectoral dice of 0, greatly decreasing the mean pectoral dice. If the pectoral muscle is found in a CC view image, it is segmented with a lower, but satisfactory, dice score compared with the MLO view. Of course, given the small size of the pectoral muscle in the CC view, the impact of a subpar pectoral segmentation on misclassified number of pixels and on the overall dice is minor.

The mean overall dice coefficients across the different DM datasets varied from 0.86 to 0.98. These differences are most likely not clinically relevant. An exception to this is the mini-MIAS dataset. With 0.86, the mean overall dice for this dataset was considerably lower than that for the other datasets. This relatively low performance can be explained by the underrepresentation of SFMs in the training set. The public dataset mini-MIAS is the only included dataset that consists of digitized SFMs instead of DMs.

In the second experiment, raw images were included in training to expand the applicability of the segmentation model from only processed DMs to processed and raw DMs. In clinical practice, raw images are often not saved. However, some physics-based analysis, such as breast density quantization, is performed in real-time on raw images, whereas in research, raw images can play an important role due to their pixel values representing actual attenuation information of different tissues. Without including raw images in training, the performance of the segmentation on raw images was much lower than for processed images and was not sufficient. Including raw images during training, instead, increased the performance on raw images and almost closed the gap in performance between processed and raw images. Moreover, including raw DMs in training also did not significantly decrease the performance on processed DMs. In short, this experiment shows that the model trained on processed and raw mammograms can be applied to both processed and raw DMs without the need to specify the type of image.

The final experiment illustrates how the model might perform when presented with mammograms from a different, unseen, vendor. When the mini-MIAS dataset was excluded, the model was trained only on DMs and, therefore, did not generalize well to the SFMs in the mini-MIAS dataset. This replicates what was also found during the first experiment. Figure 3 shows the clear differences of the SFMs in the mini-MIAS dataset (d) to the DMs in the other datasets. The SFMs in the mini-MIAS dataset have a dark breast edge and include bright labels and edges of the screen film. However, for the other seven datasets, all containing DMs, the difference in mean overall dice between included and excluded datasets varied between 0.07 and +0.02. Excluding IMS Giotto data seems to lead to the poorest performance of the DMs sets for the excluded set. Overall, this experiment shows that the model is likely to generalize well to DMs of other vendors or of new image processing algorithms from the vendors included in this study.

Comparing the segmentation results with those by Volpara showed a minor difference in performance. In contrast to the proposed segmentation method, however, Volpara runs on only raw images. There are several other studies that proposed a method for segmenting mammograms, based on rule-based and machine learning approaches. A rule-based approach by Pawar et al.17 used an adaptive threshold for segmenting the background and a region growing method with an automatically selected starting point for pectoral muscle segmentation. The method was tested on only scanned MLO SFMs and had a dice of 0.955 for pectoral muscle segmentation. This is higher than that found in this study for pectoral muscle segmentation. The difference can be explained by the heterogeneity of the images in this study as CC images and MLO images without a pectoral muscle were also included. Zebari et al.5 introduced a segmentation method for both scanned MLO screen-film mammograms and digital MLO mammograms. The background segmentation was also done with an adaptive threshold, and local feature extraction and a two-layer neural network were used to segment the pectoral muscle. The average dice coefficient for breast segmentation with this method was 0.991, which is in the range of the dice coefficient of breast segmentation of this study.

Other studies using a CNN to segment mammograms found lower or comparable segmentation performance. Two articles both used a U-Net with a ResNet encoder; one designed for raw DM MLO images18 and the other for both scanned MLO film mammograms and digital MLO mammograms19 had an overall dice coefficient of 0.949±0.019 and 0.92, respectively, which is lower than the overall dice found in the current study. Rampun et al.7 introduced a CNN inspired by holistically nested edge detection for finding the pectoral boundary in both scanned MLO film mammograms and digital MLO mammograms. The pectoral segmentation had a dice coefficient of 0.975±0.063, which is higher than the average dice coefficients for MLO images in our study of 0.835±0.196. Note that Rampun et al.7 corrected the output by selecting only the pectoral muscle segmentation that had the longest edge. The method proposed in this study has no such correction. It would be expected that a correction of selecting only the largest connected area could increase the segmentation performance. Overall, the segmentation results of this study fall within the performance range of published segmentation methods and, in contrast to existing methods, can be applied to both raw and processed DMs of different vendors and to both MLO and CC views.

Our study has some limitations, mainly related to the datasets used. One of the limitations of this study is the limited information on the imaged women. All eight datasets were acquired in Europe, and their population varies from screening to diagnostic. Furthermore, every image has only one annotation made by one of four researchers. Therefore, there is no information on the interreader variability to compare the model segmentation performance to. Testing on raw images was also limited because raw images were also available for only four datasets (Hologic, IMS Giotto, GE, and Siemens). Therefore, the performance on raw images could be evaluated only on these four datasets. However, the results of this study suggest good generalization across processed images of different vendors; therefore, it is likely that the model will also generalize well across raw images of different vendors. Finally, the segmentations were evaluated only by the dice coefficients. Although this is an objective measure of segmentation performance, it is not a guarantee for applicability for all downstream tasks.

5.

Conclusion

The proposed deep learning-based mammogram segmentation method yielded accurate overall segmentation results for raw and processed mammograms for both standard views and was able to generalize well over mammograms from different vendors. The model should perform well on processed DMs of vendors that were not included in this study, whereas its performance on unseen digitized screen-film mammograms might be more limited.

Disclosures

This study is part of the aiREAD Project (No. 17912), which is financed by the Dutch Research Council, Dutch Cancer Society, and Health Holland. The authors declare no potential conflicts of interest associated with this work.

Code and Data Availability

The code and trained weights for the model trained with processed and raw images (experiment 2) is available at https://github.com/radboud-axti/maseg. Manually annotated masks of the publicly available datasets INbreast and mini-MIAS are available at https://zenodo.org/records/10171732 and https://zenodo.org/records/10149914.

Acknowledgments

This paper would not have been possible without the datasets kindly provided by Saara Muhli from PlanMed in Finland, Monica Sordi and Sara Vecchio from IMS Giotto S.p.A. in Italy, and Caroline Behr from the Klinik für Radiologie und Neuroradiologie at Kliniken Maria Hilf GMbH in Germany. This paper is an extension of the proceedings for SPIE Medical Imaging 2022.20 We would also like to thank Volpara Health (Volpara Health, Wellington, New Zealand) for providing a comparison with their latest software version.

References

1. 

O. Haji Maghsoudi et al., “Deep-LIBRA: an artificial-intelligence method for robust quantification of breast density with independent validation in breast cancer risk assessment,” Med. Image Anal., 73 102138 https://doi.org/10.1016/j.media.2021.102138 (2021). Google Scholar

2. 

N. Saffari et al., “Fully automated breast density segmentation and classification using deep learning,” Diagnostics, 10 (11), 988 https://doi.org/10.3390/diagnostics10110988 (2020). Google Scholar

3. 

F. J. Pérez-Benito et al., “A deep learning system to obtain the optimal parameters for a threshold-based breast and dense tissue segmentation,” Comput. Methods Prog. Biomed., 195 105668 https://doi.org/10.1016/j.cmpb.2020.105668 (2020). Google Scholar

4. 

Y. Zheng et al., “Parenchymal texture analysis in digital mammography: a fully automated pipeline for breast cancer risk assessment,” Med. Phys., 42 (7), 4149 –4160 https://doi.org/10.1118/1.4921996 MPHYA6 0094-2405 (2015). Google Scholar

5. 

D. A. Zebari et al., “Improved threshold based and trainable fully automated segmentation for breast cancer boundary and pectoral muscle in mammogram images,” IEEE Access, 8 203097 –203116 https://doi.org/10.1109/ACCESS.2020.3036072 (2020). Google Scholar

6. 

O. H. Maghsoudi et al., “Deep-LIBRA: artificial intelligence method for robust quantification of breast density with independent validation in breast cancer risk assessment,” Med. Image Anal., 73 102138 https://doi.org/10.1016/j.media.2021.102138 (2020). Google Scholar

7. 

A. Rampun et al., “Breast pectoral muscle segmentation in mammograms using a modified holistically-nested edge detection network,” Med. Image Anal., 57 1 –17 https://doi.org/10.1016/j.media.2019.06.007 (2019). Google Scholar

8. 

X. Ma et al., “Automated Pectoral muscle identification on MLO-view mammograms: comparison of deep neural network to conventional computer vision,” Med. Phys., 46 (5), 2103 –2114 https://doi.org/10.1002/mp.13451 MPHYA6 0094-2405 (2019). Google Scholar

9. 

A. Gastounioti et al., “Breast parenchymal patterns in processed versus raw digital mammograms: a large population study toward assessing differences in quantitative measures across image representations,” Med. Phys., 43 (11), 5862 –5877 https://doi.org/10.1118/1.4963810 MPHYA6 0094-2405 (2016). Google Scholar

10. 

J. Suckling et al., “The mammographic image analysis society digital mammogram database,” Expert. Med. Int. Congr. Ser., 1069 (Jan.), 375 –378 https://doi.org/10.1016/j.acra.2011.09.014 (1994). Google Scholar

11. 

I. C. Moreira et al., “INbreast: toward a full-field digital mammographic database,” Acad. Radiol., 19 (2), 236 –248 https://doi.org/10.1016/j.acra.2011.09.014 (2012). Google Scholar

12. 

N. Otsu, “A Threshold selection method from gray-level histograms,” IEEE Trans. Syst. Man. Cybern., 9 (1), 62 –66 https://doi.org/10.1109/TSMC.1979.4310076 ISYMAW 0018-9472 (1979). Google Scholar

13. 

P. A. Yushkevich et al., “User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability,” Neuroimage, 31 (3), 1116 –1128 https://doi.org/10.1016/j.neuroimage.2006.01.015 NEIMEF 1053-8119 (2006). Google Scholar

14. 

O. Ronneberger, P. Fischer and T. Brox, “U-Net: convolutional networks for biomedical image segmentation,” Lect. Notes Comput. Sci., 9351 234 –241 https://doi.org/10.1007/978-3-319-24574-4_28 LNCSD9 0302-9743 (2015). Google Scholar

15. 

A. Buslaev et al., “Albumentations: fast and flexible image augmentations,” Information, 11 (2), 125 https://doi.org/10.3390/info11020125 (2020). Google Scholar

16. 

T. Y. Lin et al., “Focal loss for dense object detection,” IEEE Trans. Pattern Anal. Mach. Intell., 42 (2), 318 –327 https://doi.org/10.1109/TPAMI.2018.2858826 ITPIDJ 0162-8828 (2018). Google Scholar

17. 

S. D. Pawar et al., “Segmentation of pectoral muscle from digital mammograms with depth-first search algorithm towards breast density classification,” Biocybern. Biomed. Eng., 41 1224 –1241 https://doi.org/10.1016/j.bbe.2021.08.005 (2021). Google Scholar

18. 

O. Haji Maghsoudi et al., “Automatic breast segmentation in digital mammography using a convolutional neural network,” Proc. SPIE, 11513 1151322 https://doi.org/10.1117/12.2564235 PSISDG 0277-786X (2020). Google Scholar

19. 

M. J. Ali et al., “Enhancing breast pectoral muscle segmentation performance by using skip connections in fully convolutional network,” Int. J. Imaging Syst. Technol., 30 (4), 1108 –1118 https://doi.org/10.1002/ima.22410 IJITEG 0899-9457 (2020). Google Scholar

20. 

S. D. Verboom et al., “Deep learning-based breast tissue segmentation in digital mammography: generalization across views and vendors,” Proc. SPIE, 12032 1203236 https://doi.org/10.1117/12.2611437 PSISDG 0277-786X (2022). Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Sarah D. Verboom, Marco Caballo, Jim Peters, Jessie Gommers, Daan van den Oever, Mireille J. M. Broeders, Jonas Teuwen, and Ioannis Sechopoulos "Deep learning-based breast region segmentation in raw and processed digital mammograms: generalization across views and vendors," Journal of Medical Imaging 11(1), 014001 (28 December 2023). https://doi.org/10.1117/1.JMI.11.1.014001
Received: 14 June 2023; Accepted: 11 December 2023; Published: 28 December 2023
Advertisement
Advertisement
KEYWORDS
Image segmentation

Education and training

Muscles

Image processing

Breast

Mammography

Data modeling

Back to Top