Image guidance capability is an important feature of modern radiotherapy linacs, and future cobalt-60 units will be
expected to have similar capabilities. Imaging with the treatment beam is an appealing option, for reasons of simplicity
and cost, but the dose needed to produce cone beam CT (CBCT) images in a Co-60 treatment beam is too high for this
modality to be clinically useful. Digital tomosynthesis (DT) offers a quasi-3D image, of sufficient quality to identify
bony anatomy or fiducial markers, while delivering a much lower dose than CBCT.
A series of experiments were conducted on a prototype Co-60 cone beam imaging system to quantify the resolution,
selectivity, geometric accuracy and contrast sensitivity of Co-60 DT. Although the resolution is severely limited by the
penumbra cast by the ~2 cm diameter source, it is possible to identify high contrast objects on the order of 1 mm in
width, and bony anatomy in anthropomorphic phantoms is clearly recognizable. Low contrast sensitivity down to
electron density differences of 3% is obtained, for uniform features of similar thickness. The conventional shift-and-add
reconstruction algorithm was compared to several variants of the Feldkamp-Davis-Kress filtered backprojection
algorithm result. The Co-60 DT images were obtained with a total dose of 5 to 15 cGy each.
We conclude that Co-60 radiotherapy units upgraded for modern conformal therapy could also incorporate imaging
using filtered backprojection DT in the treatment beam. DT is a versatile and promising modality that would be well
suited to image guidance requirements.
Megavoltage computed tomography (MVCT) has been an active area of research and development in image guided radiation therapy. We have been investigating a particular implementation of MVCT in conjunction with studies of the potential for tomotherapy with a Cobalt-60 radiation source. In this paper, we present results comparing MVCT using a Co-60 source and a 4 MV linear accelerator to conventional kVCT imaging. The Co-60 and linac MVCT measurements were obtained with a first generation benchtop CT imager; the KVCT measurements were obtained using a Philips AcQSim CT Simulator). Phantoms containing various inserts ranging in density from air, through lung, soft tissue and bone equivalent materials and extending to high atomic number metals were imaged with the three modalities. The results enable characterization of image artifacts, CT number linearity and beam hardening. The MVCT images have sufficient contrast that soft tissue regions with 2.8% difference in electron density can be visualized. In MVCT, a linear relationship between CT numbers and electron densities extends to materials with Z ≈ 60. In the 4MV CT imaging there is a position dependence of the CT numbers within a uniform water phantom, which is absent in Co-60 CT images, indicating the presence of beam hardening artifacts in the linac MVCT images. The differences between kVCT and MVCT will be discussed considering the variation of the photon interactions dominating the images. Our investigations indicate that MVCT has properties that may potentially extend its utility beyond radiation therapy.
There has been considerable interest in megavoltage CT (MVCT) imaging associated with the development of image guided radiation therapy. It is clear that MVCT can provide good image quality for patient setup verification with soft tissue contrast much better than noted in conventional megavoltage portal imaging. In addition, it has been observed that MVCT images exhibit considerably reduced artifacts surrounding metal implants (e.g., surgical clips, hip implants, dental fillings) compared to conventional diagnostic CT images (kVCT). When encountered, these artifacts greatly limit the usefulness of kVCT images, and a variety of solutions have been proposed to remove the artifacts, but these have met with only partial success. In this paper, we investigate the potential for CT imaging in regions surrounding metal implants using high-energy photons from a Cobalt-60 source and from a 4 MV linear accelerator. MVCT and kVCT images of contrast phantoms and a phantom containing a hip prosthesis are compared and analysed. We show that MVCT scans provide good fidelity for CT number quantification in the high-density regions of the images, and in the regions immediately adjacent to the metal implants. They also provide structural details within the high-density inserts and implants. Calculations will show that practical clinical MVCT imaging, able to detect 3% contrast objects, should be achievable with doses of about 2.5cGy. This suggests that MVCT not only has a role in radiotherapy treatment planning and guidance, but may also be indicated for surgical guidance and follow-up in regions where metal implants cannot be avoided.
Access to the requested content is limited to institutions that have purchased or subscribe to SPIE eBooks.
You are receiving this notice because your organization may not have SPIE eBooks access.*
*Shibboleth/Open Athens users─please
sign in
to access your institution's subscriptions.
To obtain this item, you may purchase the complete book in print or electronic format on
SPIE.org.
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.