In computed tomography the Hounsfield Units (HU) are used as an indicator of the tissue type based on the linear attenuation coefficients of the tissue. HU accuracy is essential when this metric is used in any form to support diagnosis. In hybrid imaging, such as SPECT/CT and PET/CT, the information is used for attenuation correction (AC) of the emission images. This work investigates the HU accuracy of nodules of known size and HU, comparing diagnostic quality (DQ) images with images used for AC.
Ultrasound imaging of the breast is highly operator dependent. The amount of pressure applied with the transducer has a direct impact on the lesion visibility in breast ultrasound. The conspicuity index is a quantitative measure of lesion visibility, taking into account more parameters than standard measures that impact on lesion detection. This study assessed the conspicuity of lesions within a breast phantom using increased transducer compression in breast ultrasound.
Methods
A phantom was constructed of gelatine to represent adipose tissue, steel wool for glandular/blood vessels and silicone spheres to represent lesions, this meant that the lesions were also compressible, but less than the surrounding tissue. The phantom was imaged under increasing transducer compression. The conspicuity index was measured using the Conspicuity Index Software. The distance between the transducer surface and lesion surface was measured as an indication of increased compression.
Results
When moderate compression (17mm) was applied, the conspicuity index increased resulting in better visualisation of the silicone lesions. However, with increased compression the conspicuity index decreased.
New work to be presented
The conspicuity index has never been demonstrated in ultrasound imaging before. This is preliminary phantom work to demonstrate the impact of increased transducer compression on quantitative lesion visibility assessment.
Conclusion
The compression applied should be considered for optimum visualisation, as excessive pressure decreases conspicuity. However, further work needs to be conducted in order to consider other factors, such as density of the breast and lesion location, for a better understanding of the effect of compression on the visualisation of the lesion. A human study is planned.
Tube current modulation is a method employed in the use of CT in an attempt to optimize radiation dose to the patient. The acceptable noise (noise index) can be varied, based on the level of optimization required; higher accepted noise reduces the patient dose. Recent research [1] suggests that measuring the conspicuity index (C.I.) of focal lesions within an image is more reflective of a clinical reader's ability to perceive focal lesions than traditional physical measures such as contrast to noise (CNR) and signal to noise ratio (SNR). Software has been developed and validated to calculate the C.I. in DICOM images. The aim of this work is assess the impact of tube current modulation on conspicuity index and CTDIvol, to indicate the benefits and limitations of tube current modulation on lesion detectability. Method An anthropomorphic chest phantom was used “Lungman” with inserted lesions of varying size and HU (see table below) a range of Hounsfield units and sizes were used to represent the variation in lesion Hounsfield units found. This meant some lesions had negative Hounsfield unit values.
A novel software programme and associated Excel spreadsheet has been developed to provide an objective measure of the expected visual detectability of focal abnormalities within DICOM images. ROIs are drawn around the abnormality, the software then fits the lesion using a least squares method to recognize the edges of the lesion based on the full width half maximum. 180 line profiles are then plotted around the lesion, giving 360 edge profiles.
Purpose: To investigate the dose saving potential of iterative reconstruction (IR) in a computed tomography (CT) examination of the thorax.
Materials and Methods: An anthropomorphic chest phantom containing various configurations of simulated lesions (5, 8, 10 and 12mm; +100, -630 and -800 Hounsfield Units, HU) was imaged on a modern CT system over a tube current range (20, 40, 60 and 80mA). Images were reconstructed with (IR) and filtered back projection (FBP). An ATOM 701D (CIRS, Norfolk, VA) dosimetry phantom was used to measure organ dose. Effective dose was calculated. Eleven observers (15.11±8.75 years of experience) completed a free response study, localizing lesions in 544 single CT image slices. A modified jackknife alternative free-response receiver operating characteristic (JAFROC) analysis was completed to look for a significant effect of two factors: reconstruction method and tube current. Alpha was set at 0.05 to control the Type I error in this study.
Results: For modified JAFROC analysis of reconstruction method there was no statistically significant difference in lesion detection performance between FBP and IR when figures-of-merit were averaged over tube current (F(1,10)=0.08, p = 0.789). For tube current analysis, significant differences were revealed between multiple pairs of tube current settings (F(3,10) = 16.96, p<0.001) when averaged over image reconstruction method.
Conclusion: The free-response study suggests that lesion detection can be optimized at 40mA in this phantom model, a measured effective dose of 0.97mSv. In high-contrast regions the diagnostic value of IR, compared to FBP, is less clear.
Access to the requested content is limited to institutions that have purchased or subscribe to SPIE eBooks.
You are receiving this notice because your organization may not have SPIE eBooks access.*
*Shibboleth/Open Athens users─please
sign in
to access your institution's subscriptions.
To obtain this item, you may purchase the complete book in print or electronic format on
SPIE.org.
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.