|
1.INTRODUCTIONSignificant advances in the general area of minimally invasive and noninvasive surgery have been made in the last few years. This can be attributed to advances in key enabling technologies in medical imaging, computers, and instrumentation. Thermal ablative and coagulative therapies utilizing various sources of heat have received significant attention due to the inherent safety and absence of harmful side effects when the energy source is controlled properly. Several research groups and industrial entities are developing image-guided minimally invasive or noninvasive techniques for thermal surgery with applications ranging from cardiac ablation to cancer treatments. In terms of energy sources, there are many forms of applicators with various degrees of sophistication and emplying a variety of modalities for tissue coagulation. Most of the available applicators have been optimized for specific tissue targets and desired coagulated volume based on reasonable, but imperfect, theoretical models. Multi-element applicators are becoming common place today and offer improved control over the coagulated volume. Without adequate temperature feedback, however, these additional degrees of freedom are rendered useless. Therefore, this brings us to the second major development area for the future of minimally invasive surgery. In particular, the development of noninvasive or minimally invasive methods for guidance of the energy source and monitoring the tissue response during thermal surgery. 1.1.Image GuidanceImage guidance for noninvasive thermal therapy is now a reality. This may be based on well established imaging modalities like MRI,15 CT,19 or ultrasound.16,18 Other imaging modalities are also being developed and may become available in the foeseeable future.22 The main component of an image guidance system for thermal surgery appear to be the temperature imaging which has received significant attention during the last 10 - 15 years. Several modalities are being used for noninvasive temperature estimation such as MR,24-26 CT,19 ultrasound.28,1,17,21,27,11 The specific imaging modality used for image guidance will depend on many factors such as doctor preference, target type and site, economics, etc. Nevertheless, some form of image guidance will definitely be available in the operating room in the near fututre. 1.2.Temperature ImagingAmong other forms of feedback, feedback information on the temperature evolution of the region being treated is needed.15,18 MRI,25,26 CT,24,19 impedance tomography,22 microwave radiation23 and ultrasound28,27,20,1,21,18,11 imaging have been studied for this purpose. Each of these modalities has its particular advantages and shortcomings. In this paper, we focus on the use of pulse-echo ultrasound in non-invasive temperature estimation. While the other methods, especially MRI, offer important advantages as means of noninvasive temperature estimation, ultrasound is very attractive due to several key advantages:
Of course, temperature imaging with pulse-echo ultrasound has its limitations which are discussed below. More research is needed to overcome these limitations to allow for practical use of this temperature imaging method in clinical situations or in in-vivo animal testing. Some of these limitations are discussed in detail in earlier publications.12,13,11 They are enumerated in the discussion section below with appropriate citation. 1.3.Image-Based Methods for Damage AssessmentBased on the current state of the art, MRI and CT are expected to offer superior performance when it comes to damage assessment. However, these may be very expensive and complex. Ultrasound offers several key advantages for image guidance such as portability, compatibility, real-time operation, ease of use, etc. Unfortunately, however, gray-scale ultrasonic imaging does not offer quantitative measurements that relate to the state of the tissue. Therefore, ultrasound has not been considered a serious tool for noninvasive damage assessment of treated tissue. This view of ultrasonic imaging misses the fact that ultrasound backscattered pulses are a result of interaction between the basic ultrasound wavelet with the microstructure of the tissue. Therefore, with appropriate signal processing, ultrasound sholud provide some useful parameters that correlate with tissue state. 1.4.Self-Guided Systems for Noninvasive SurgeryAnother area unique to ultrasound that could revolutionize the field of image guided surgery is the development of a new generation of dual-mode high-power phased array systems capable of both imaging and therapy.14 These piezocomposite transducers can produce focal intensity levels needed for ablative and coagulative theramal surgery with high precision. Furthermore, the operating bandwidth of such transducers allows for imaging the treatment region with adequate image quality to delineate important landmarks within and around the target volume. With these capabilities, it is possible to operate these arrays in a self-guided mode whereby the array’s imaging capabilities are utilized to improve the quality of the therapeutic foci, guide the therapeutic foci to the (possibly moving) target, and even perform temperature imaging around the therapeutic foci. 1.5.Organization and Scope of PaperThe rest of the paper is organized as follows: Section 2 describes the theoretical bases, formulation, and the implementation of noninvasive two-dimensional methods for temperature estimation with pulse-echo ultrasound. Section 3 addresses the use of diagnostic ultrasound for damage assessment and characterization of thermal lesions. In Section 4 we describe a phased array system capable of performing both imaging and therapy and, therefore, offers the potential for self-guided noninvasive method for noninvasive surgery. Section 5 offers a discussion on the current status and future directions in the development of ultrasonic methods for noninvasive surgery. In Section 6 we offer some concluding remarks. The reader will undoubtedly realize that much of the material included in this paper is based largely on the author's work in this area for the last few years. This was done primarily for convenience given some time constraints. It should not be interpreted as a judgement on the author's part as to the worthiness of other related work in this area. However, we do stress the fact that the material included herein is represensative of the state of the art in the three major areas identified in this paper. We hope that the citations given will give the reader a more complete perspective of the scope of ongoing research in this area. 2.TEMPERATURE IMAGING WITH PULSE-ECHO ULTRASOUNDThe temperature dependence of some of the acoustic propagation parameters, especially the speed of sound, has long been recognized and reasonably well understood.8,9 However, early attempts at imaging temperature noninvasively were primarily based on transmission mode9 or tomographic reconstruction techniques.8 Unfortunately, these techniques have many limitations that hampered any significant progress in this area beyond simple laboratory demonstrations. Recently, however, several groups have proposed new methods for temperature estimation using pulse-echo ultrasound using spectral methods20,27 and echo shift methods.1,21,11 2.1.Theoretical Formulation2.1.1.ModelThe temperature-change estimation method described in this paper is based on the thermal dependence of the ultrasound echo that accounts for two different physical phenomena: (1) local change in speed of sound due to changes in temperature and (2) thermal expansion of the propagating medium. The former produces an apparent shift in scatterer location and the latter leads to a physical shift. Along an A-line, however, the two effects lead to echo time-shifts that can be estimated and are shown to be related to local change in temperature in the medium. In many situations of pratical interest, these effects are typically small, so that a linearized approach can be used in the analysis that follows. It should be mentioned, however, that for large temperature rise and/or large size of the heated region, the linearized analysis breaks down. Different formulation is needed to handle these cases. This formulation will be introduced in a future report. The baseline temperature in the sample (tissue) medium is assumed to be constant and equal to θ0 at the initial time T0. We start by considering the time delay of the echo from a scatterer at axial depth z as being exclusively a function of the temperature-dependent speed of sound in the propagating medium: where θ(ξ) = θ0 + δθ(ξ) is the temperature at depth ξ, c(ξ,θ(ξ)) represents the speed of sound at depth ξ and temperature θ(ξ), and the subscript c indicates that only the thermal dependence of speed of sound has been considered. In order to account for thermal expansion in the propagating medium we substitute dξ by (1 + α(ξ)δθ(ξ))dξ in Eq. (1), where α(ξ) is the linear coefficient of thermal expansion of the medium at axial depth ξ. We get: At the initial time T0 = 0, the echo time delay is . Once the temperature in the medium changes, a time-shift will be observed on the echo from scatterer at axial depth z: Differentiating Eq. (3) with respect to the scatterer location z, we get: The thermal dependence of the speed of sound in tissue can be assumed to be approximately linear in the temperature range of interest9: where , and c0(z) = c(z, θ0). From Eqs. (4), (5), and using the fact that |β(z)δθ(z)| ≪ 1, solving for δθ(z) = (θ(z) – θ0) we get: The term is a medium (material) dependent parameter. For a homogeneous medium this parameter can be experimentally determined. For the remainder of this paper we assume k(z) and c0(z) to be invartiant with respect to axial depth z, i.e., we assume k(z) = k and c0(z) = c0. In practice, these parameters should be allowed to vary in space and as a function with temperature. For the purposes of this paper, however, we will limit our attention to homogeneous media and small changes in temperature so that these assumptions are approximately valid. It should be mentioned that the linearized approach is valid for temperature changes on the order of 10 - 15 °, even in inhomogeneous media. Therefore, the temperature estimation algorithm described below will be of practical for temperature estimation in hyperthermia and, more importantly, for image guidance of thermal therapy devices. 2.2.CharacterizationIn30,11 we have presented experimental data comparing noninvasive temperature estimation data from a tissue mimicking phantom with thermocouple measurements. The results of this test can be summarized as follows:
2.3.ApplicationsIn this section, we describe three applications of the temperature estimation algorithm as it stands at this point. The limitations noted below not withstanding, these applications demonstate the enormous potential and some of the distinct advantages of using ultrasound for noninvasive temperature feedback. 2.3.1.Image Guidance for Noninvasive SurgeryFigure 1 shows an experimental setup used in image guidance based on the temperature imaging method described in this paper. In this experiment, a thick piece of bovine muscle was subject to a sequence of heating patterns due to the therapeutic array. Eight 2-second duration foci were used for simulating a coagulative thermal necrosis in tissue. The individual foci were spaced by two mm and deivered to the target in sequence with the foci approaching the imaging transducer. Ample wait time between shots was given to allow temperature to return to baseline before the next shot is delivered. Upon completion of the experiment, the tissue sample was cut and it was confirmed that tissue destruction (as indicated by visible change of color) at the desired location. In order to demonstrate the capability of the temperature estimation algorithm for guiding the therapy, the sequence of shots were initially delivered at low power (causing only 2 °C rise at the focal point. That is, the therapy array was operated well below the threshold for tissue damage and was run in a sequence that resembles the planned therapy session. While this was being done, temperature images were collected and processed to determine if the locations and quality of the foci are consistent with the treatment objectives. This allows for characterizing the quality of the heating foci in-situ at subthreshold (nondestructive) levels just before the theraputic pulses are to be delivered. Figure 2 shows the spatio-temporal temperature distribution along one image line going through all eight heating spots. The grayscale temperature image is obtained from an M-mode image consisting of 128 A-lines. For each shot, 16 A-lines were acquired and processed for temperature estimation, 2 before, 4 during, 5 after (fast rate), and 5 after (slow rate). Please note that the time axis is simply labeled A-line Number. For the temperature estimation experiment shown, each 16 lines represent 1 minute, but the A-lines were not acquired uniformly in time. Figure 2 shows clearly that the temperature estimation algorithm has the spatial contrast and sensitivity needed to characterize the quality of the therapeutic transducer. The peaks of the temperature rise profiles are spaced by 2 mm as planned. Also, the extent of each heated region is between 3-5 mm axially which also corresponds to the expected temperature profile of a 2 mm wide beam heating for two seconds. While the image clearly demonstrate the potential of the temperature estimation algorithm, it should be noted that some level of ripple artifact exists. However, this ripple is not a limiting factor in this case. It should also be emphasized that the maximum temperature rise needed to produce this result is approximately 2 °C. This is a major advantage for ultrasound in that the quality assurance of focusing and energy deposition can be performed energy levels well below the therapeutic thresholds. Furthermore, since these images can be acquired and processed within a fraction of a second, the initial heating rate can be evaluated two dimensionally. Therefore, it is possible to visualize the specific absorption rate (SAR) of the energy source in situ. These are perhaps the major strengths of using ultrasound in image guidance of energy sources for noninvasive or minimally invasive thermal surgery. 3.PULSE-ECHO IMAGING METHODS FOR DAMAGE ASSESSMENTAt present, the issue of image-based damage assessment remains largely unsolved. To the best of our knowledge, no research group has made significant progress on this issue with any imaging modality, including MRI and CT. These modalities hold the most promise due to their quantitative nature. Ultrasound, on the other hand, suffers from some significant shortcomings due to nonquantitative nature of gray scale imaging and lack of 3D imaging in most clinical scanners. Nonetheless, ultrasound continues to be attractive due to its real-time imaging capability and high flexibility. In addition, with appropriate signal processing, quantitative tissue parameters can be extracted from pulse-echo data. While these tissue chacterization techniques have not been widely successful in identifying and classifying tissue pathology in normal clinical scenarios, it is expected that they will have a role to play in chacterization of thermal lesions. This is due to the drastic change of some of the acoustic properties within the ablated or coagulated volume, e.g., attenuation. Imaging of attenuation in reflection mode has been investigated by many investigators for diagnostic purposes. One of the most successful algorithms for estimation of this parameter is based on low-order autoregressive (AR) modeling of backscattered ultrasonic signals. In what follows, we describe the use of this technique in an in-vitro experiment. While the results shown are encouraging, they are very preliminary and should be interpreted as an indicator of potential rather than true performance. Much more work needs to be done in this area for real prgress to be made. 3.1.Theoretical Formulation3.1.1.Backscattered Signal ModelA widely accepted model for ultrasound backscattered signal from tissue media expresses the received signal as a superposition of reflections from N scatterers within the resolution cell of the transducer: where Ai is the strength of the ith scatterer assumed to be independent of frequency, pz(t)wo-way propagation del is the basic ultrasound wavelet emitted from the transducer with z (range) dependence to account for attenuation, diffraction, etc., and τi is the two-way propagation delay to the ith scatterer . A simplified version of the spectrum of the bacscattered signal can be given by: where we have explicitly accounted for the attenuation (a is the attenuation coefficient in dB/cm). 3.1.2.Autoregressive ModelingAutoregressive (AR) modeling of ultrasound backscattered data has proved to be extremely beneficial in many applications, e.g., Doppler, mean scatterer spacing estimation, attenuation measurements, etc. The pth order AR model for the received signal given in Equation 7 is given by: where x(n) is the sampled version of x(t) at t = nT, ai are the AR filter coefficients to be determined from the time series, and ϵ(n) is white Gaussian noise sequence with zero mean and variance . A typical use of the AR model calls for a model order p such that ϵ is truly a white noise sequence. However, for many applications, a low-order AR model is preferred for computational simplicity and robustness. Furthermore, for values of p equal to 2 and 3 (AR2 and AR3), the coefiicients of the AR model take on physical significance with intuitive appeal. Regardless, once the model order has been determined the filter coefficients can be determined by many techniques based on second oder or higher order statistics. Furthermore, algorithms for determining the optimal filter coefficients have been developed for cases when the underlying model is stationary or nonstationary. The interested reader is referred to several textbooks on the subject. 3.1.3.AR Parameter EstimationIn an AR2 model, the coefficients a1 and a2 are direcly related to the reflection coefficients as well as the center frequency of the spectrum of the backscatter. The latter is an important parameter in determining the attenuation coefficients. The attenuation coefficient is known to change dramatically when the tissue is subjected to coagulative necrosis.29 Assuming that α is a linear function of frequency, α = α0f, then it can be shown that where σs is related to the transducer bandwidth and fo is the center frequency of the backscattered signal. It can also be shown that where fs is the sampling frequency. 3.2.Experimental ResultsThe system shown in Figure 1 was used for producing a 1 cm3 lesion in an in-vitro bovine muscle. The lesion was formed by electronically scanning the therapy transducer to produce 10 contiguous HIFU spots in one plane. The therapy array was mechanically scanned to produce a three dimensional lesion in 10 plane. Ultrasound B-scan imaages where acquired before and after the lesion formation. Examples of these B-scan gray scale images are shown in Figure 3. The lesion boundaries are shown by the box in Figure 3b). However, examining the B-scan image does not reveal any useful information about the lesion. This result is typical and it is the reason behind the perception that B-scan ultrasonic imaging does not provide quantitative information that can be utilized for quantitative analysis of tissue damage due to coagulative necrosis. As seen from Section 3.1.2, one can extract the attenuation coefficient from the center frequency though curve fitting algorithm. Since it is known that attenuation changes dramatically with changes in the tissue state due to thermal coagulation, it is expected that extracting this parameter should be a good indicator of the change in tissue state. Figure 4 is a subtraction image of the center frequency images obtained from the before and after images shown in Figure 3. Significant change in the center frequency can be seen in a region extending from 35 - 45 mm in the range direction and 13 - 26 mm in the lateral direction. To the best of our knowledge, this is the first such image obtained in reflection mode for this type of application. This result strongly suggest that it will be possible to obtain quantitative images from beamformed RF data that correlate with the damage to the tissue. 4.SELF-GUIDED PHASED ARRAY SYSTEM FOR NONINVASIVE SURGERYThe system described in Figure 1 provides an example of image-guided therapeutic phased array. In that case, the image guidance is provided by an independent diagnostic imaging system. Other implementations of image-guided phased arrays with ultrasound,16 MRI,15 and CT19 have been proposed. In all of these implementations, some form of transformation from the imaging coordinate system to the therapeutic array coordinate system is needed. This may be accomplished with various degrees of difficulty depending on the specific imaging modality and imaging geometry. An attractive alternative for the above paradigm is the use of the same therapeutic applicator to perform imaging. This was facilitated by recent advances in piezocomposite transducer technology. The main advantage of this proposed approach is the use of of the same coordinate system for both therapy and imaging. Therefore, refocusing information obtained for improving image quality are directly applicable to improving the heat localization to the target. Furthermore, acoustic obstacles and critical structures can be directly identified and methods for incorporating them in optimization of focusing are easily obtainable using imaging techniques and are directly applicable to the therapeutic array. In this section, we describe our recent efforts on developing such phased array systems and appropriate algorithms for their use as imaging transducers. 4.1.TransducerA spherical-shape piezocomposite 64-element 1-D array therapeutic transducer (Imasonic, France) operating at 1.1 MHz was used. The echo response received by one of the central elements, when pulsing a neighbor element (Panametrics Pulser Receiver 5052-PR), a thick flat Plexiglas reflector being at the focal distance, is given in Fig. 5a. The corresponding frequency response (Fig. 5b) indicates that the transducer has two predominant resonance frequencies, at 1.1 and 2.1 MHz, respectively. When the transducer operates in therapeutic mode, a high-Q resonant matching circuit is used to drive the individual transducer elements at 1.1 MHz. When the transducer operates in imaging mode, low-Q series inductors were used to select the 1.1 MHz operating frequency, providing an echo with relatively short duration (Fig. 5c). The -6dB full width bandwidth of the pulse-echo system in this case was measured to be 37%, the frequency response being shown in Fig. 5d. 4.2.Therapeutic CharacterizationA custom designed phased array driving system was used to independently drive each of the 64 elements of the array with control of amplitude and phase. Series inductors are used to obtain a high-Q resonant peak at the operating frequency. The output acoustic power, when driving the transducer with the current system, can be varied from 0 to 510 W, depending on the amplitude of the sinusoids used to drive the elements. The acoustic output power was measured in water using a force balance approach while focusing the transducer at its geometric center (all elements driven in phase). In order to demonstrate the therapeutic capability of this system, a series of lesions were created in an in vitro beef muscle HIFU experiment.6 Eleven single focus sonications were delivered 30 mm deep within the tissue, and laterally spaced by 5 mm, being distributed from -25 mm to +25 mm from the geometric center of the transducer. The total acoustic power was fixed at 200 W and the duration of each sonication was linearly varied from 0.25 sec to 2.75 sec. The tissue was allowed to cool down for a sufficiently long time (3 minutes) between successive sonications, therefore avoiding temperature build up between sonications. The tissue was cut along the scanned plane and lesions were observed for all sonications with duration longer than 1.25 sec, as shown in Fig. 6. 4.3.Pulse-Echo Imaging4.3.1.Data collection systemImaging data was obtained by driving the transducer in pulse-echo mode. The transmitted pulses were generated by a pulser/receiver (Panametrics 5052-PR). The received echoes were first amplified by 30dB using an ultrasound signal amplifier (NTR pfs-005a). The transmit and receive circuitry were connected to the array elements through a VXI programmable 64×4 relay matrix (Tektronix VX4380). Two different data acquisition systems were used to digitize the data (Fig. 7). The first digitizer consists of a digital signal analyzer (Tektronix DSA-601), the captured data being transfered to the controlling workstation through a GPIB interface. The duration of the data collection in this case is limited by the speed of the digitizer, a complete set of 64×64 A-lines taking about 3 hours. In order to reduce the total duration of the data collection, a gated digitizer was implemented. The data is digitized using a Burr Brown ADS-801 12-bit fast-A/D operated at 16 MHz (data sampling frequency). Gating is implemented using conventional fast TTL logic. The digitized data is stored in a dual port 128 Mb memory buffer, and transfered at slower rate to the controlling workstation. The duration of the acquisition is limited in this case by the dead time of the relay matrix after each switching (approximately 40 ms), a complete set of 64×64 A-lines being acquired in about 3 minutes. The current prototype implementation of this circuit provides data with lower SNR than the data obtained using the Tektronix digital signal analyzer. This increases the level of the background noise in the reconstructed image, ultimately limiting the image dynamic range. Nevertheless, the system is still capable of providing good image quality, as seen in Fig. 12. 4.3.2.Synthetic aperture imagingImages were obtained using a full synthetic aperture technique.31 In this case the system is in focus at every pixel of the image, in transmit and receive. Furthermore, in order to reduce side-lobes, apodization of the transmit and receive apertures was applied by weighting the echo responses with a Kaiser window34 (β = 3.25). The image pixel at coordinates (xp, zp) was therefore computed by (Fig. 8): where i is the transmit element index, j is the receive element index, Ai is the transmit apodization weight at element i, Bj is the receive apodization weight at element j, Rip and Rjp are the distances from the transmit and receive elements, respectively, from the image pixel, c is the speed of sound in the medium being imaged, and Si,j(t) is the base-band converted echo acquired when transmitting with element i and receiving with element j. The images were first sampled in the R × sin(θ) domain, and then scan converted into a Cartesian spatial sampling. 4.4.Results4.4.1.Multiple-wire phantomA multiple-wire phantom consisting of thirteen nylon wires spaced by 10 mm along a horizontal line was used to characterize the point spread function of the imaging system described above (Fig. 9). The phantom was aligned so that the middle wire was located at the geometric center of the spherically-shaped therapeutic transducer. The phantom and transducer were immersed in water to provide acoustic coupling. A full 64×64 data-set was collected using the setup represented in Fig. 7, the data being digitized by the Tektronix digital signal analyzer. The pulse-echo synthetic aperture reconstructed image is shown in Fig. 10 at dynamic ranges of 20, 30, 40, and 50 dB, where all images are referenced to the peak value in the reconstructed region. At 20 dB seven central wires are seen. As this transducer is highly focused, and no compensation for element directivity was applied during reconstruction, objects closer to the geometric center of the transducer appear to be brighter than objects farther away. Furthermore, as the aperture is undersampled grating lobes become visible when imaging off-axis (image at 40 and 50dB). At 50 dB eleven of the thirteen wires are seen. The results in Fig. 10 indicate that this system provides good image reconstruction of objects close to the transducer geometric center. However, the quality of the image degrades as objects off-center are considered. Although this image degradation is certainly not appropriate for diagnostic systems, in the case of therapeutic guidance the most relevant information is contained in the region undergoing treatment. Typically, this corresponds to the region close to the center of the therapeutic transducer, where the image quality is better. 4.4.2.Speckle-generating phantomIn order to characterize the performance of this system with speckle media, a tissue-mimicking phantom was designed, its dimensions being given in Fig. 11. The phantom is made of rubber material (plastic for making plastic worms, M-F Manufacturing Co., Fort Worth, TX), with embedded polystyrene scatterers (Amberlite IR-120plus, Sigma Chemical Co., St. Louis, MO). The scatterers are homogeneously distributed within the phantom, except for a 12.7 mm diameter cyst region, that was filled with water. The phantom and the transducer were immersed in water to provide acoustic coupling. The phantom was aligned so that the cyst region corresponded to the geometric center of the therapeutic transducer, as indicated in Fig. 11. A full 64×64 data set was collected using the experimental setup shown in Fig. 7, the custom gated digitizer being used in this case. The synthetic aperture image of the phantom is shown at four different dynamic ranges, namely 20, 30, 40 and 50 dB, referenced to the peak value detected within the imaged region. The image shows the cyst region, illustrating the performance of the system in the vicinity of the focal region. The strong reflection at the top and bottom edges of the cyst are caused by the acoustic impedance mismatch between the rubber material and water. 5.DISCUSSION5.1.Temperature ImagingThe algorithm described in this paper was developed based on an approximate theoretical model and was verified in tissue mimicking phantoms under conditions similar to those assumed in this paper. We have also performed many experiments in tissue media that have conclusively shown that the time shifts produced even by small to moderate temperature rise are measurable and correspond to expected spatiotemporal distribution of temperatures due to localized heating sources. However, there are several limitations of this temperature imaging algorithm that require further development before this method finds clinical application in in-vivo clinical and animal testing experiments. In the following, we enumerare these limitations and describe efforts to overcome them. Citations of relevant work in this regards are given where appropriate.
5.2.Damage AssessmentThe result shown in Figure 4 clearly demonstrate that the center frequency is very sensitive to changes in the tissue state after thermal coagulative necrosis. However, this image is still not of sufficient quality to provide easy visual characterization of thissue damage. At this point, it appears very likely that the center frequency will be one of several parameters extracted and processed (on a statistical basis) to characterize the state of the tissue. However, the importance of this result cannot be overemphasized. While MRI and X-ray CT are ahead of ultrasound in terms of quantitative damage assessment, ultrasound is superior in terms of portability, realtime imaging capability, cost, and complexity. If this and many of our in-vitro results can be demonstrated in in-vivo tissues, then a major limitation towards ultrasound guided noninvasive surgery will be removed. Given other advances in noninvasive temperature imaging11 and dual phased arrays for imaging and therapy,13 it is now possible to envision a self-guided ultrasound phased array that is capable of delivering the HIFU pulses, monitor the progression of the treatment, and perform a post treatment evaluation. This can save time and resources We should note that the image shown in this paper is a subtraction image. While subtraction images are very useful and help identify before and after type changes as we have in this experiment, they are not always possible. In practice, high temperature procedures cause tissue deformations. These, in addition to tissue motion during the procedure, make it very hard if not impossible to produce such subtraction images reliably. It is therefore necessary to develop practical algorithms for motion compensation. In addition, image registration techniques with deformable models will be needed. We are currently investigating such tools. It is hoped that more research effort will be directed to this area in the near future. 5.3.Self-Guided Ultrasound Phased Arrays for Noninvasive SurgeryPoor acoustic penetration in tissue usually limits the applicability of synthetic aperture imaging techniques.31 However, for the system presented in this paper the transducer elements are designed to have very high efficiency in transmit. Furthermore, the received echo signal is relatively strong due to the element large surface area as well as its focusing along the elevation direction. Moreover, in the case of diagnostic systems, the maximum acoustic intensity is limited to a relatively low value, in order to avoid potential for damage or risk to the patient. In a high-intensity focused ultrasound therapeutic context, however, the acoustic intensities used in the treatment are much higher than those generated by diagnostic systems. Therefore, when operating the therapeutic transducer in imaging mode to provide guidance for therapeutic procedures, as suggested in this paper, it is potentially acceptable to use acoustic intensities that exceed the limits specified for conventional diagnostic systems. Motion of the object being imaged also imposes practical limitations to the applicability of synthetic aperture imaging techniques. In the case of the system presented in this paper, it should be possible to acquire a full 64×64 data set within 12 ms, assuming that the echoes received at the 64 elements are collected simultaneously, and a maximum imaged depth of 150 mm is used. With the exception of cardiac applications, one can assume that most tissues will not move faster than 10 mm/s. In this case, the maximum error due to motion within the collection of a single frame (12 ms) will be limited to 0.16 μs, or λ/6 (λ=wavelength) for an imaging frequency of 1 MHz. 6.CONCLUSIONSWe have addressed three aspects of the use of ultrasound phased array technology in noninvasive thermal surgery. These are temperature imaging, quantitative imaging for damage assessment, and dual-mode self-guided array systems. The latter is unique to ultrasound and could prove to be key to the overall development of noninvasive surgery. However, at the very least, this approach may be the only viable approach for noninvasive surgery with ultrasound arrays. Given the complexity of ultrasound wave propagation, the issue of guiding ultrasound with a separate imaging modality may prove to be quite difficult. Self-guided phased arrays will be the only viable alternative. Much more research is needed in this area, but the early attempts are quite promising. With the availability of high-power arrays with improved bandwidths over what we presented in this paper, we can only expect significant improvements in image quality. Temperature imaging with ultrasound is also at an early development stage and will require significant effort before it can be used reliably in clinical situations. However, even at this early stage, one can argue that ultrasound stands the best chance of being used for localizing the heat source in multidimensional spatio-temporal maps in realtime and with very high spatial resolution. The main challenge at this point is the development of quantitative calibration-free temperature imaging algorithms. We predict that semiquantitative and calibration-aided quantitative temperature imaging with ultrasound will become available within the next few years, especially with the introduction of 3D imaging systems by several manufacturerers. This feature will eliminate one of the major limitations of ultrasound temperature imaging. In particular, out of plane motion will be easily tracked and, therefore, speckle tracking for temperature estimation will improve in robustness and accuracy. This is likely to prove to be a very fruitful research area for ultrasound imaging in general, but for temperature imaging in particular. In the area of damage assessment, ultrasound currently stands at a relative disadvantage. The results shown in this paper are representative of how little was achieved in this area. It is fair to say that no serious attempt at developing ultrasound imaging tools for damage assessment has been taken. Similar to what we have shown in this paper, several researchers have shown before and after type images of certain tissue parameters that reflect changes due to thermal lesion formation. However, these results are still not quantitative and, to the best of our knowledge, no specific imaging algorithm for damage characterization has been proposed in the literature. However, ultrasound wave propagation in tissue media is fundamentally quantitative and is sensitive to tissue structure as well as scattering characteristics. Ultrasound continues to hold significant promise in this regard. Several recent developments in harmonic imaging, elastography, and vibroacoustic emission imaging appear to offer the most promise for damage characterization. At the time of writing this article, some preliminary work has been presented in these areas. More work is needed to corroborate these results and establish repeatibility. In summary, the promise of ultrasound phased array technology in the area of noninvasive surgery has never looked better. However, much more work is needed to bring this promise to fruitful clinical application. We note that the interest in therapeutic ultrasound is quite high in both industry and at academic institutions. It is hoped that the collective effort of this community will move this research forward and bring us closer to true noninvasive surgery early in the new millenium. AcknowledgementsThis work was funded by Grant CA 66602 from the National Institutes of Health. Many thanks to former students P. VanBaren, C. Simon, and K. Oweiss for help with the data collection and storage. REFERENCESR. Seip, P. VanBaren, C. Simon, and E. Ebbini,
“Non-invasive spatio-temporal temperature change estimation using diagnostic ultrasound,”
IEEE Ultrason. Symp. Proc., 1613
–1616
(1995). Google Scholar
P. VanBaren and E. Ebbini,
“Multipoint temperature control during hyperthermia treatments: Theory and simulations,”
IEEE TRans. BME, 42
(8), 818
–827
(1995). https://doi.org/10.1109/10.398643 Google Scholar
R. Seip, P. VanBaren, C. Cain and E. Ebbini,
“Noninvasive real-time multipoint temperature control for ultrasound phased array treatments,”
IEEE Trans. UFFC, 43
(6), 1063
–1073
(1996). https://doi.org/10.1109/58.542050 Google Scholar
T. Sheljaskov, R. Lerch, M. Bechtold, K. Newerla, and U. Schätzle,
“A phased array antenna for simultaneous HIFU therapy and sonography,”
IEEE Ultrason. Symp. Proc., 1527
–1530
(1996). Google Scholar
T. Sheljaskov, R. Lerch, A. Fuchs, and U. Schätzle,
“A phased array antenna for simultaneous thermotherapy and sonography,”
IEEE Ultrason. Symp. Proc.,
(1997). Google Scholar
P. VanBaren, C. Simon, R. Seip, T. Solf, C. Cain, and E. Ebbini,
“Image-guided phased array system for ultrasound thermotherapy,”
IEEE Ultrason. Symp., 1269
–1272
(1996). Google Scholar
T. Loupas, J. Powers, and R. Gill,
“An axial velocity estimator for ultrasound blood flow imaging, based on a full evaluation of the Doppler equation by means of a two-dimensional autocorrelation approach,”
IEEE Trans. UFFC, 42
(4), 672
–688
(1995). https://doi.org/10.1109/58.393110 Google Scholar
J. Greenleaf,
“Computerized transmission tomography,”
Methods of Experimental Physics, 19 Ultrasonics, Academic Press, London
(1981). https://doi.org/10.1016/S0076-695X(08)60343-9 Google Scholar
R. Nasoni, T. Bowen, W. Connor, R. Sholes,
“In vivo temperature dependence of ultrasound speed in tissue and its application to noninvasive temperature monitoring,”
Ultrasonic Imaging, 1
(1), 34
–43
(1979). https://doi.org/10.1177/016173467900100103 Google Scholar
C. Simon,
“Two-Dimensional Temperature Estimation,”
Ph.D. Dissertation, Ann Arbor
(1998). Google Scholar
C. Simon, P. VanBaren, and E. Ebbini,
“Two-dimensional temperature estimation for ultrasound thermotherapy using diagnostic ultrasound,”
IEEE Trans. Ultrason., Ferroelec., Freq. Contr., 45
(4), 1088
–1099
(1988). https://doi.org/10.1109/58.710592 Google Scholar
C. Simon, P. VanBaren, and E. Ebbini,
“Motion compensation algorithm for non-invasive two-dimensional temperature estimation using diagnostic pulseecho ultrasound,”
SPIE Bios: Surgical Applications of Energy,
(1998). Google Scholar
C. Simon, J. Shen, T. Hall, and E. Ebbini,
“Combined ultrasound image guidance and therapy using a therapeutic phased array,”
SPIE: Medical Imaging 1998, 3341 San Diego,1998). Google Scholar
E. Ebbini and C. Simon,
“Combined imaging and therapy with piezocomposite phased arrays,”
1998 IEEE Ultrasonics Symposium, 2 1555
–1558
(1998). https://doi.org/10.1109/ULTSYM.1998.765240 Google Scholar
K. Hynynen,
“Review of ultrasound therapy,”
1997 IEEE Ultrasonics Symposium, 2 1305
–1313
(1997). https://doi.org/10.1109/ULTSYM.1997.661817 Google Scholar
N. Sanghvi,
“Noninvasive surgery of prostate tissue by high-intensity focused ultrasound,”
IEEE Trans. UFFC, 43
(6), 1099
–1110
(1996). https://doi.org/10.1109/58.542053 Google Scholar
R. Seip,
“Feedback for ultrasound thermotherapy,”
Ph.D. Dissertation,
(1996). Google Scholar
E. Ebbini, P. VanBaren, and C. Simon,
“Image-guided noninvasive surgery with ultrasound phased arrays,”
SPIE Bios: Surgical Applications of Energy,
(1998). Google Scholar
J. Jenne, M. Bahner, J. Spoo, P. Huber, R. Rastert, I. Simiantonakis, W. Lorenz, and J. Debus,
“CT on-line monitoring of HIFU therapy,”
IEEE Ultrason. Symp.,
(1997). Google Scholar
R. Seip and E. Ebbini,
“Non-invasive estimation of tissue temperature response to heating fields using diagnostic ultrasound,”
IEEE Trans. Biomed. Eng., 42
(8), 828
–839
(1995). https://doi.org/10.1109/10.398644 Google Scholar
R. Moreno, C. Damianou, and N. Sanghvi,
“Tissue temperature estimation in-vivo with pulse-echo,”
IEEE Ultrason. Symp., 1225
–1229
(1995). Google Scholar
K. Paulsen, M. Moskowitz, T. Ryan, S. Mitchell, and P. Hoopes,
“Initial in vivo experience with EIT as a thermal estimator during hyperthermia,”
Int. J. Hyperthermia, 12
(5), 573
–591
(1996). https://doi.org/10.3109/02656739609027666 Google Scholar
P. Meaney, K. Paulsen, A. Hartov, and R. Crane,
“Microwave imaging for tissue assessment: Initial evaluation in multitarget tissue-equivalent phantoms,”
IEEE Trans. Biomed Eng., 43
(9), 878
–890
(1996). https://doi.org/10.1109/10.532122 Google Scholar
B. Fallone, P. Moran, and E. Podgorsak,
“Noninvasive thermometry with a clinical X-ray scanner,”
Med. Phys., 9
(5), 715
–721
(1982). https://doi.org/10.1118/1.595117 Google Scholar
D. Parker,
“Applications of NMR imaging in hyperthermia: An evaluation of the potential for localized tissue heating and noninvasive temperature monitoring,”
IEEE Trans. Biomed. Eng., 31
(1), 161
–167
(1984). https://doi.org/10.1109/TBME.1984.325382 Google Scholar
J. Poorter, C. Wagter, Y. Deene, C. Thomsen, F. Stahlberg, and E. Achten,
“Noninvasive MRI thermometry with the proton resonance frequency (PRF) method: In vivo results in human muscle,”
Magn. Reson. Med., 33 74
–81
(1995). https://doi.org/10.1002/(ISSN)1522-2594 Google Scholar
W. Straube and R. Arthur,
“Theoretical estimation of the temperature dependence of backscattered ultrasonic power for noninvasive thermometry,”
Ultrasound in Med. & Biol., 20
(9), 915
–922
(1994). https://doi.org/10.1016/0301-5629(94)90051-5 Google Scholar
S. Ueno, M. Hashimoto, H. Fukukita, and T. Yano,
“Ultrasound thermometry in hyperthermia,”
IEEE Ultrason. Symp., 1645
–1652
(1990). https://doi.org/10.1109/ULTSYM.1990.171648 Google Scholar
A. Malcom, and G. ter Haar,
“Ablation of tissue volumes using high intensity focused ultrasound,”
Ultrasound in Med. and Biol., 22 659
–699
(1996). https://doi.org/10.1016/0301-5629(96)00020-8 Google Scholar
C. Simon, P. VanBaren, and E. Ebbini,
“Quantitative analysis and applications of non-invasive temperature estimation using diagnostic ultrasound,”
1997 IEEE Ultrasonics Symposium, 1319
–1322
(1997). https://doi.org/10.1109/ULTSYM.1997.661819 Google Scholar
K. Thomenius,
“Evolution of ultrasound beamformers,”
IEEE Ultrason. Symp., 1615
–1622
(1996). Google Scholar
I. Hein, and W. O’Brien Jr.,
“Current time-domain methods for assessing tissue motion by analysis from reflected ultrasound echoes — a review,”
IEEE Trans. Ultrason., Ferroelec., Freq. Contr., 40
(2), 84
–102
(1993). https://doi.org/10.1109/58.212556 Google Scholar
R. Nasoni and T. Bowen,
“Ultrasonic speed as a parameter for non-invasive thermometry,”
Non-invasive temperature measurement, 1 95
–107 Gordon and Breach Science Publishers,1989). Google Scholar
A. Oppenheim, and R. Schafer,
“Discrete-Time Signal Processing,”
95
–107 Prentice-HallGordon and Breach Science Publishers, Englewood Cliffs
(1989). Google Scholar
|