X-ray computed tomography (XCT) is an important clinical diagnostic tool which is also used in a range of biological
imaging applications in research. The increasing prevalence of metallic implants in medical and dental radiography and
tomography has driven the demand for new approaches to solving the issue of metal artefacts in XCT. Metal artefacts
occur when a highly absorbing material is imaged which is in boundary contact with one or more weakly absorbing
components, such as soft-tissue. The resulting ‘streaking’ in the reconstructed images creates significant challenges for
X-ray analysis due to the non-linear dependence on the absorption properties of the sample. In this paper we introduce a
new approach to removing metal artefacts which exploits the capabilities of the recently available, photon-counting
PiXirad detector. Our approach works for standard lab-based polychromatic X-ray tubes and does not rely on any postprocessing
of the data. The method is demonstrated using both simulated data from a test phantom and experimental data
collected from a cochlear implant. The results show that by combining the individual images, which are simultaneously
generated for each different energy threshold, artefact -free segmentation of the implant from the surrounding biological
tissue is achieved.
Viona S. K. Yokhana,Benedicta D. Arhatari,Timur E. Gureyev, andBrian Abbey
"Lab-based x-ray tomography of a cochlear implant using energy discriminating detectors for metal artefact reduction", Proc. SPIE 10456, Nanophotonics Australasia 2017, 104565S (2 January 2018); https://doi.org/10.1117/12.2283320
ACCESS THE FULL ARTICLE
INSTITUTIONAL Select your institution to access the SPIE Digital Library.
PERSONAL Sign in with your SPIE account to access your personal subscriptions or to use specific features such as save to my library, sign up for alerts, save searches, etc.
The alert did not successfully save. Please try again later.
Viona S. K. Yokhana, Benedicta D. Arhatari, Timur E. Gureyev, Brian Abbey, "Lab-based x-ray tomography of a cochlear implant using energy discriminating detectors for metal artefact reduction," Proc. SPIE 10456, Nanophotonics Australasia 2017, 104565S (2 January 2018); https://doi.org/10.1117/12.2283320