Presentation + Paper
1 March 2019 Cone-beam CT statistical reconstruction with a model for fluence modulation and electronic readout noise
Author Affiliations +
Abstract
Purpose: Cone-beam CT (CBCT) systems with a flat-panel detector (FPD) have advanced in a variety of specialty diagnostic imaging scenarios, with fluence modulation and multiple-gain detectors playing important roles in extending dynamic range and improving image quality. We present a penalized weighted least-squares (PWLS) reconstruction approach with a noise model that includes the effects of fluence modulation and electronic readout noise, and we show preliminary results that tests the concept with a CBCT head scanner prototype. Methods: Statistical weights in PWLS were modified using a realistic noise model for the FPD that considers factors such as system blur and spatially varying electronic noise in multiple-gain readout detectors (PWLSe). A spatially varying gain term was then introduced in the calculation of statistical weights to account for the change in quantum noise due to fluence modulation (e.g. bowtie filter) (PWLS∗). The methods were tested in phantom experiments involving an elliptical phantom specially designed to stress dual-gain readout, and a water phantom and an anthropomorphic head phantom to quantify improvements in noise-resolution characteristics for the new PWLS methods (PWLS𝑒 and PWLS∗, and combined PWLS∗e). The proposed methods were further tested using a high-quality, low-dose CBCT head scanner prototype in a clinical study involving patients with head injury. Results: Preliminary results show that the PWLSe method demonstrated superior noise-resolution tradeoffs compared to conventional PWLS, with variance reduced by ~15-25% at matched resolution of 0.65 mm edge-spread-function (ESF) width. Clinical studies confirmed these findings, with variance reduced by ~15% in peripheral regions of the head without loss in spatial resolution, improving visual image quality in detection of peridural hemorrhage. A bowtie filter and polyenergetic gain correction improved image uniformity, and early results demonstrated that the proposed PWLS∗ method showed a ~40% reduction in variance compared to conventional PWLS when used with a bowtie filter. Conclusion: A more accurate noise model incorporated in PWLS statistical weights to account for fluence modulation and electronic readout noise reduces image noise and improves soft-tissue imaging performance in CBCT for clinical applications requiring a high degree of contrast resolution.
Conference Presentation
© (2019) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
P. Wu, A. Sisniega, J. W. Stayman, W. Zbijewski, D. Foos, X. Wang, N. Aygun M.D., R. Stevens M.D., and J. H. Siewerdsen "Cone-beam CT statistical reconstruction with a model for fluence modulation and electronic readout noise", Proc. SPIE 10948, Medical Imaging 2019: Physics of Medical Imaging, 1094814 (1 March 2019); https://doi.org/10.1117/12.2513417
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation

Head

Imaging systems

Sensors

Modulators

Image filtering

Scanners

RELATED CONTENT


Back to Top