17 January 1990 Picosecond Streak Camera Measurements Of Short X-Ray Pulses
M. M. Murnane, H. C. Kapteyn, R. W. Falcone
Author Affiliations +
Abstract
A commercial x-ray streak camera was modified to obtain an impulse response of 2.5 psec. It was used to measure the x-ray emission from a plasma produced by a sub-picosecond laser. An upper limit of 1.5 psec was obtained for the x-ray pulse duration over the 50 eV to the 1 keV region of the spectrum.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
M. M. Murnane, H. C. Kapteyn, and R. W. Falcone "Picosecond Streak Camera Measurements Of Short X-Ray Pulses", Proc. SPIE 1155, Ultrahigh Speed and High Speed Photography, Photonics, and Videography '89: Seventh in a Series, (17 January 1990); https://doi.org/10.1117/12.962462