8 July 2021 Short-wavelength Raman spectroscopy for in-situ water analysis: a feasibility study
Andrea Sterzi, Olga Sambalova, Urs Schneider, Davide Bleiner, Andreas Borgschulte
Author Affiliations +
Proceedings Volume 11886, International Conference on X-Ray Lasers 2020; 118860Y (2021) https://doi.org/10.1117/12.2596045
Event: XVII International Conference on X-Ray Lasers, 2020, Online Only
Abstract
Short-wavelength (λ < 160 nm) Raman spectroscopy offers an advantage of a generally higher sensitivity than Raman spectroscopy in the visible range. An application with high potential is its use for environmental water analysis targeting archetypal compounds that are present in industrial and urban sewage water. However, this application is feasible only if specific performance benchmarks are met. We validate the applicability of a simple and cost-effective deep-UV Raman spectrometer (λexc = 236.5 nm). The analysis brings to the fore that the experimentally derived detection limits the studied archetypal compounds are to high by several orders of magnitude. We outline potential further development and associated limitations. These are the deterioration of the analysed species by photolysis as a consequence of the high photon energy and intensity, and the self absorption of the UV radiation. These effects are explained and partially corrected along a simple mathematical model from which a general limit of detection is estimated.

1.

INTRODUCTION

Clean water is an essential and one of the most important aspects of the United Nations sustainable development goals.1 However, water pollution is diminishing the accessibility to clean water worldwide. Apart from long known fertilizer-derived compounds, such as nitrate 00052_PSISDG11886_118860Y_page_1_1.jpg and nitrite 00052_PSISDG11886_118860Y_page_1_2.jpg ions,2, 3 new threats are posed by the appearance of so-called Contaminants of Emerging Concern (CECs), which are difficult to remove from effluent streams. Among them, antidepressant agents Citalopram (CIT), Venlafaxine (VEN) and Carbamazepine (CBZ) and a widely used corrosion inhibitor, Benzotriazole, are in the spotlight.4 Several studies report their concentration levels in waste water in the range of tens of μg/l, leading to serious threats for the aquatic fauna and plants.58 In the case of nitrate ions, the maximum reported contamination levels is higher, approx. ≤ 50 mg/l, according to the World Health Organization (WHO).911 Even more harmful are nitrite ions, which can be formed via photolysis process under UV illumination.11–14 These compounds in waste water are normally measured via liquid chromatography - mass spectrometry (LC-MS).57 The technique provides high sensitivity (beyond μg/l) and chemical selectivity.15 However, the process has to be performed in a chemical laboratory and requires intermediate steps (e.g. sample preparation and separation of the elemental components). In addition, the instrumentation cost and dimension renders this technique unsuitable for on-line detection.

Optical spectroscopy may be a valid on-line alternative. Optical vibrational spectroscopy has high potential, because the spectra resemble a molecular fingerprint. Infrared gas analysis is known to detect trace impurities down to ppb level.16 Infrared spectroscopy is not applicable to water analysis because of the large absorption in this wavelength range. An optical water analysis method must use the water window from around 1 μm to 200 nm,17 a criterion, which is met by Raman spectroscopy. However, despite its high versatility, it suffers from an intrinsically low sensitivity. In fact, the intensity of the Raman scattered light I is several orders of magnitude lower than the excitation light I0, being approximately one Raman photon scattered for every 10–6 incident photons. Another substantial limiting factor is fluorescence, which is often dominant in organic samples and overlaps with the Raman bands.18 Several approaches have been proposed to enhance the signal while limiting fluorescence, e.g. taking advantage of new technologies in the field of laser sources, time-resolved detection, optical components, devices, sensors and plasmonic substrates.1821 Summarizing, Raman spectroscopy in the deep-UV energy range fulfils the above given criteria. Its main advantage is the confinement of Raman Stokes lines in the region up to 4000 cm–1 (⇒ λ ≃ 260 nm), well separated from the typical fluorescence emission (λ > 260 nm).22, 23 The Raman signal, I, is inversely proportional to the forth power of the wavelength 00052_PSISDG11886_118860Y_page_2_2.jpg, hence it is markedly enhanced by shortening the latter. Further signal enhancement can be reached, if the photon energy is resonant with a specific electronic transition, the so-called resonant Raman spectroscopy (RRS), which is very likely to occur in the UV range (UV-RRS). Based on this selective enhancement mechanism, UV-RRS has been exploited to study several molecular species showing electronic absorption bands in the UV spectral range, e.g. nucleic acids, proteins and drugs.24, 25 Recently, this method proved its applicability also in the field of water analysis on example of antibiotics in solutions and urine.26 This pioneering study by C. Domes et al. reports the detection by deep-UV RRS of Moxiflacin, Meropenem and Peperacilin in aqueous media with a limit of detection (LOD) in the order of 5 mg/l. Nitrate and nitrite ions in waste water have been more extensively studied, A. Ianoul et al.27 reached a LOD of about 1 mg/l (14 μM) for 00052_PSISDG11886_118860Y_page_2_3.jpg and 00052_PSISDG11886_118860Y_page_2_4.jpg. Later on, several works reported Raman detection of nitrite28, 29 while focusing on the UV light induced photo-degradation, e.g. photolysis of nitrate to form nitrite ions, another “side-effect” correlated to the use of highly energetic photons.1214, 30, 31

Despite such remarkable results, the application of deep-UV Raman as a standard analytical method to perform routine water analysis is still limited. Several technical factors restrict its use. First, the mere generation of deep-UV light requires customized high power laser sources such as Ti:Sapphire oscillators, Nd:YAG (4th and 5th harmonics at 266 and 213 nm), CW intracavity frequency-doubled argon or kripton ion lasers (lines at 257, 248, 244, 229 nm), Raman-shifted third harmonics in gas from Nd:YAG (204 nm) or diode-pumped solid state (DPSS) Nd:YVO4 (213 nm).26, 3234 Their cost and complexity is also often accompanied with a lack of mobility. Alternatively, hollow cathode lasers (e.g. NeCu pulsed laser emitting at 248.5 nm) are more affordable and have recently found applications in aerospace, being used for performing in-situ minerals investigation.35 However, their lifetime is relatively short and their peculiar annular beam profile may impede a proper beam focusing. Tunable light sources covering the deep-UV range exist in the form of tunable lasers31, 36 and UV synchrotron radiation,37 enabling measurement of analytes in the resonance condition. However in both cases they cannot be implemented in view of developing a portable device. Furthermore, the collection of the Raman signal from a multi-wavelength light source, i.e. the separation of the Rayleigh from the Raman light, is even more challenging.31 Additional deep-UV range effects potentially include optical self-absorption, photo-decomposition of the sample and resonances, practically affecting the measured spectrum. Those “intrinsic-processes”, which are seldom present in standard Raman spectroscopy, while boosting the sensitivity (resonance enhancement) increase the complexity of the measurement and thus impede an extensive application.

In this work we evaluate a compact and cost-effective deep-UV Raman spectrometer for in-situ environmental analysis, i.e. detection of archetypal contaminants in aqueous environment at mg/l concentrations (sketched in Fig. 1). We investigate advantages and limitations of the quasi-resonance approach on different pollutant categories: antidepressant agents, corrosion inhibitor and fertilizer-derived compounds. Specifically, the following analytes are measured: Citalopram, Venlafaxin, Carbamazepine, Benzotriazole in solutions, nitrate and nitrite ions in deionized (pure) and environmental (river) water. We provide a reliable LOD and asses effects such as self-absorption, fluorescence and photolysis, which selectively affect the acquired spectra. The deep-UV Raman spectrometer built in our lab is based on recent technological advances in deep-UV micro-chip laser and optical components. A direct comparison between visible (commercial Raman, λexc = 532 nm) and deep-UV (λexc = 236.5 nm) was made for all samples under investigation. The impact of the self-absorption, i.e. an induced non-linear dependence between the Raman intensity and the analyte concentration, is investigated. From this, the effective laser beam absorption length through the sample is estimated. Although the strong attenuation of laser and scattered light common for strong UV absorbers, such as benzotriazole, hampers the analysis, we show how the effect can be used to estimate the overall molar concentration of even strongly UV-absorbing compounds. This important analysis parameter (attenuation) is already used to perform on-line analysis in sewage water plants. As Raman sensitivity of compounds is often not known in advance and is low, determined by the scattering cross sections, the simultaneous acquisition of Raman and attenuation parameters paves the way for the automatic (computerized) evaluation of multi-component systems.

Figure 1.

Deep-UV Raman spectroscopy for monitoring water quality and its potential applications.

00052_PSISDG11886_118860Y_page_2_1.jpg

2.

METHODOLOGY

2.1

Deep-UV Raman setup

This study is based on the prototype deep-UV Raman spectrometer described in Ref.31 The system is further reduced in dimensions and cost and is adapted for measurements of liquid samples in 180° backscattering geometry (Fig. 2). A commercial, passively q-switched microchip laser (Pulselas-P-236 by Alphalas) emitting 5 mW @λexc = 236 nm, with 10 KHz repetition rate is used as a light source. A single element (BS/filter in Fig. 2) reflects the incident light wavelength, while transmitting the Stokes line. Thereby, the laser line is removed. A single aluminum 90° off-axis parabolic mirror is used to focus the beam and collect the scattered light.31 The complete instrument with the compact light source fits on a 45 x 30 cm breadboard. The advantages of the chosen excitation wavelength include a spectral separation of the fluorescence and Raman signals and availability of the optical components.23 In particular, commercial Rayleigh Raman filters (e.g., from Semrock) are available for the chosen wavelength and not for shorter excitation wavelength, e.g. Nd:YAG based 213 nm. The signal loss due to the absorption by optical components is negligible. The effective laser beam power at the sample position is reduced to less than 40% (from 5 to 3 mW). Due to its simplicity, the presented optical scheme is adaptable to further modifications without requiring the complete re-design of the set-up.

Figure 2.

Simplified optical scheme of the deep-UV Raman spectrometer. The optical components include UV-mirrors (M), a diverging lens (L), off axis parabolic mirrors (P) and a beam splitter/edge filter (BS/filter). The laser beam is shown in blue while the scattered light is shown in purple.

00052_PSISDG11886_118860Y_page_3_1.jpg

2.2

Raman spectroscopy in the visible energy range

Visible Raman spectra were measured by meas of a continuous wave laser (Ocean Optics, 100 mW nominal power, 40 mW effective power at the focal point, λexc = 532 nm) and a Raman probe (Bruker) connected by an optical fiber. All Raman spectra were acquired with the integration time of 20 seconds and 100 averages.

The dark spectrum (camera noise) was subtracted with no additional smoothing procedure. As a threshold to perform peak identification during the spectra evaluation, we assumed a signal to noise ratio equal to 3.

2.3

Ion Chromatography and UV absorption spectroscopy

As a validation method, Ion Chromatography (IC) was adopted to determine the nitrate content in the sample withdrawn from the river. The measurements were performed on a Metrohm IC 882 instrument with a ASUPP 5-250 separation column. Certified standard reference materials were used. UV absorption measurements were performed on a UV spectrometer (Varian UV scan 50 BIO).

2.4

Samples

Pharmaceuticals and benzotriazole were purchased from Sigma-Aldrich, specifically: Citalopram hydrobromide (C20H21FN2O·HBr, molecular weight = 405.3 g/mol); Venlafaxine hydrochloride (C17H27NO2·HCl, molecular weight = 313.8 g/mol); Carbamazepine (C15H12N2O, molecular weight = 236.27 g/mol), benzotriazole (C6H5N3, molecular weight = 119.1 g/mol). The structure of each molecule is reported in Fig. S1 Supplemental Material. Solutions were prepared using deionized water and diluted to required concentration. Nitrate solutions were prepared in deionized water and an environmental sample was taken from a local river (Chriesbach, Dübendorf Switzerland). Samples were stored at 5 degrees Celsius and were measured in quartz cuvettes.

3.

THEORY

3.1

Excitation and relaxation processes in the deep-UV spectral range

The interaction of a Raman-active molecule with the incident photon depends on the laser wavelength. The physical origin of such wavelength-dependent behaviour is briefly schematically represented in an extended Jablonsky diagram (Fig. 3). Upon non-resonant irradiation the excitation frequency v0 is far from inducing any electronic transition, hence the molecule is excited from the ground state (GS) to a virtual state (Fig. 3A).24 Radiation is emitted back at the same frequency v0 (Rayleigh elastic scattering) and via inelastic scattering at a shifted frequency ν± v0 (Stokes and Anti-Stokes inelastic scattering), which corresponds to the vibrational or rotational energies. In Fig. 3, for simplicity, only the the Stokes inelastic scattering (νν0) is represented. This phenomenon is known as Standard Raman (or non-Resonant Raman) and virtually occurs at any incident photon wavelength.38

Figure 3.

Extended Jablonsky diagram showing main possible quantum electronic transitions occurring at different excitation wavelengths. (A) Standard Raman in the visible range, (B) Resonant Raman (C) Photon absorption and time delayed (clock) fluorescence emission (D) Photolysis generating new compound exciting new Raman emissions. The broken line separates the system before and after the photolysis. Second order processes and the probability of each electronic transition are omitted for simplicity.

00052_PSISDG11886_118860Y_page_5_1.jpg

When the photon energy of the excitation source is equal to the energy of an electronic transition, the molecule is excited from the GS to a real electronic state (Fig. 3B). In such resonance condition, the scattered signal can be enhanced of several order of magnitude (up to 108 times higher) when compared to standard Raman scattering.22 The enhancement is due to the resonance-increased probability of excitation compared to an excitation to a virtual state.25 The signal is also enhanced in quasi-resonance condition, and increases as the photon energy approaches the electronic transition. The emission of Stokes and anti-Stokes light is the fastest relaxation channel with competing emission originating from fluorescence (Fig. 3C). Fluorescence takes place on longer time scale than Raman scattering (nanosecond timescale) and often obstructs the weaker Raman signal.22 Additionally, if the molecule undergoes a process of photolysis, molecular bonds are broken (Fig. 3D). This leads to the formation of new compounds, which then act as Raman-active light scattering molecules.

The majority of molecular species exhibit absorption bands in the UV and deep-UV spectral range,25 hence the scenario depicted in cases (B,C,D) is expected when deep-UV light is used to excite the specimen. We notice that the molecules under investigation do not exhibit absorption bands precisely at the excitation wavelength (236.5 nm; Figs.S5 and S6). Therefore, our method of choice is quasi-resonant deep-UV Raman Spectroscopy.

3.2

Raman intensity and self-absorption

The sensitivity of Raman analysis is mainly determined by the Raman scattering cross section, σ. The scattering is much stronger in resonant Raman, as σ increases proportionally to the square of the molar absorptivity, ε, of the absorber at the resonant frequency ν0:39

00052_PSISDG11886_118860Y_page_4_1.jpg

The values of σ can be as large as 10–22cm–2·mol–1·sr−1, which is roughly 108 times higher than in non-resonant Raman (≈ 10−30cm−2·mol−1·sr−124). Practically, effective sensitivity depends also on various parameters: the collection optics, scattering geometry, excitation wavelength, laser power, etc. In short-wavelength Raman the factor determining Raman sensitivity is the phenomenon of differential self-absorption. Briefly, the excitation light is attenuated by passing through the sample (self-absorption), affecting the collected Raman signal. The scattered light travelling back through the sample is also self-absorbed, but the single Raman lines are differently attenuated due to the dependency of ε (molar absorptivity) on the frequency of the scattered radiation (differential self-absorption).3941 Several theoretical and optical models have been proposed to describe the intensity of Raman bands accounting for experimental conditions and considering the differential self-absorption phenomenon (mostly analyte-related).35, 3941 It is not the scope of this work to introduce a new model or discuss in detail the existing ones, but rather to evaluate, in a simplified way, the impact of differential self-absorption on the measured Raman spectra (we refer to it from here as self-absorption) by adapting the model introduced by Shiver et al.40, 41

The intensity of a specific Raman band i, quantified in counts during the measurement process, can be generally expressed by:

00052_PSISDG11886_118860Y_page_5_2.jpg

where k is a coefficient accounting for the experimental conditions, i.e. laser Intensity (here defined as W/cm2, illuminated volume V, collection aperture and detector efficiency, σi is the scattering cross section and Ni is the number of analyte molecules per unit of volume. In order to quantify Ii, the intensity of the Raman scattered light at a specific wavelenght, Ii,R, is estimated as follows. During the measurement process the Raman light is converted by the spectrometer, through a conversion factor ks:

00052_PSISDG11886_118860Y_page_5_3.jpg

We define V as the interacting volume of the focused laser beam with the liquid sample. V can be roughly approximated by a portion L of a Gaussian beam propagating along the z direction (Fig. 4).

Figure 4.

Illuminated Gaussian-like volume in back-scattering geometry and self-absorption. Laser and scattered light propagates from the origin O to the point (x’, y’, z’). L is the depth of focus.

00052_PSISDG11886_118860Y_page_6_1.jpg

The following approximations are considered: 1) the scattered light, which may originate outside of the interacting volume is neglected and 2) the detector collects homogeneously the radiation scattered from different point inside the interacting volume. The infinitesimal scattering element dV at the position (x’,y’,z’), stems for a generic sampling point which is excited by the laser and which scatters light backwards. The laser light distribution function, Iλ is attenuated along the z direction according to the Beer Lambert’s law and expressed by:

00052_PSISDG11886_118860Y_page_6_2.jpg

where the nominal intensity of the laser is I0 and α (mm−1) is the absorption coefficient at the excitation wavelength λ. The Raman scattered light is also attenuated according to the Beer Lambert’s law, and the loss term is 00052_PSISDG11886_118860Y_page_6_3.jpg, where β is the absorption coefficient at the Raman band wavelength. The solid scattering angle Ω is relatively small in our setup and each scattering point dV scatters over a negligible angle , hence cosθ ≈ 1. By defining IΩ as its distribution function, the fraction of total scattered light is expressed by kΩ = ∫ IΩdΩ over the solid angle Ω. The overall expression for Ii,R, accounting for all the introduced terms is:

00052_PSISDG11886_118860Y_page_6_4.jpg

The terms σi Ni do not depend on the spatial coordinates and P (Watt) is the laser power without any absorption loss at a surface centered at (x,y) along z direction, estimated as the integral of the laser nominal intensity P = ∫∫ I0(x,y)dxdy. By incorporating the constant ks of eq.(3), we obtain the expression for the detected intensity:

00052_PSISDG11886_118860Y_page_6_5.jpg

The integral is calculated over the length l, namely effective absorbance length. The solution and final expression for Ii is then:

00052_PSISDG11886_118860Y_page_6_6.jpg

This equation was originally formulated to evaluate self-absorption in mixtures containing resonantly active components.41 Here, we make use of it by considering the solution as a mixture of resonant (analyte) and non-resonant (water) components. In non-resonant condition (α+β) is small (<1). The term 00052_PSISDG11886_118860Y_page_6_7.jpg 1, so in eq.(7) the intensity Ii,RNi, hence proportional to the analyte concentration. The length l, in absence of absorption, coincides with L as the collected light is scattered from the entire illuminated volume V. Conversely, in resonance (or pre-resonance) (α + β) > 1 and the exponential 00052_PSISDG11886_118860Y_page_6_8.jpg << 1, meaning that the beam is extensively absorbed along the probed volume. In this case, 00052_PSISDG11886_118860Y_page_12_2.jpg is ≈ 1 and the second bracket of eq.(7) is constant, as both Ni and the denominator are proportional to the concentration c (α = εα·c and εα is the molar absorption coefficient). A strong attenuation of the non-resonant component, i.e. a water Raman mode, is expected. In this respect we define the inhibition coefficient Ki as the ratio of the inhibited and non-inhibited water mode Iwater ÷ I0,water. This term, as derived from eq.(7), results in the following relation:

00052_PSISDG11886_118860Y_page_6_9.jpg

Water peak intensities are taken from Raman spectra while α and β are obtained from the UV absorption measurements through α = Abs÷10 mm·ln(10), where Abs is the measured Absorbance. The term l is estimated by graphically solving eq.(8). It is worth noticing that neglecting the effective laser intensity distribution and the spectrometer clipping effect represent a strong approximation.35

4.

EXPERIMENTAL RESULTS

4.1

Antidepressant agents

Solutions of CIT, VEN and CBZ were prepared at different concentrations and measured both by deep-UV and visible light (Fig. 5). Deep-UV spectrum of CIT presents two peaks: (1) 1236 cm−1 and (2) 2232 cm−1, assigned to the aromatic C-F stretching vibrations and the triple-bonded C-N stretching mode, respectively.42 The water-related mode appears at about 1620 cm−1. The visible spectrum presents a huge fluorescence background, which prevents any reliable peak identification. The estimated LOD for the deep-UV Raman is 5 mg/l (12 μM). VEN deep-UV spectra have a slightly smaller signal to noise ratio than the CIT ones. A Raman mode is clearly identified at 1280 cm−1 and assigned to a O-CH3 stretching mode.43 The visible range Raman spectrum presents a weaker fluorescence background than in case of CIT, however any clear peak identification is still perturbed by the overlap of the signals. The estimated LOD, according to a concentration series, is 6 mg/l (20 μM).

Figure 5.

Deep-UV and visible Raman spectra of CIT, VEN, CBZ at different concentrations. Main Raman modes are marked and assigned in the text. The left and right axes refer to the deep-UV and visible Raman measurements, respectively. The constant background is arbitrary and here used to increase readability of the spectra.

00052_PSISDG11886_118860Y_page_8_1.jpg

CBZ specrum is more complex, as several vibrational modes are detected. Main modes are assigned as follows: (1) (C-H) in plane bending mode at 1118 cm−1, (2) (C-H) bending mode at 1160 cm−1, (3) (C-C) stretching modes at 1219 cm−1, (4) (C-H) bending and (C-C) stretching modes 1491 cm−1, (5) (C-C) stretching mode 1566 cm−1 and (6) (C-C) double bond stretching mode 1622 cm−1.44, 45 The background intensity of the deep-UV spectra decrease more evidently relative to the other two pharmaceuticals, as the fluorescence observed in visible Raman is weaker. The estimated LOD of CBZ is 12.5 mg/l (50 μM).

As expected, all measured compounds present no fluorescence-related background in deep-UV Raman.18, 22 However, another factor, the analyte concentration, affects the background intensity, as shown prominently in the CBZ case. The dependence between the band intensity and concentration is not strictly linear for CIT and VEN, and is almost constant for CBZ (Fig. 6). At the same time, a strong attenuation of the O-H water modes at 3440 cm−1 is observed (Figs. S2, S3 and S4, Supplemental Material). Addressing such effects to self-absorption, we apply eq.(8) to estimate the absorption length l. UV absorption at the laser wavelength and at λ = 257 nm (O-H water mode wavelength) was measured (Fig. S5, Supplemental Material) for the following solutions: CIT (100 mg/l), VEN (50 mg/l), CBZ (33 mg/l). The estimated l (mm) values and all the coefficients are listed in Table S1 (Supplemental Material). The strongest inhibition of the water mode is observed in CBZ, where the highest absorption is also measured. The estimated l = 9.8 mm is as long as the cuvette (10 mm). This physically means that the beam is extensively absorbed over the entire illuminated volume V41 explaining the observed non-linearity that originates from absorption (Fig. 6). CIT presents the lowest absorptivity, the highest ki ratio and l = 6.7 mm.

Figure 6.

Intensity of CIT(2) @ 2232 cm−1, VEN(1) @1280 cm−1 and CBZ(5) @1566 cm−1 Raman modes as a function of the concentration. The noise level is marked, (approximately 5 counts arbitrary unit).

00052_PSISDG11886_118860Y_page_9_1.jpg

4.2

Nitrate and nitrite

Compared to visible Raman spectra, the deep-UV Raman of a 100 mg/l NaNO3 solution (Fig. 7 top), exhibits a clearly visible 00052_PSISDG11886_118860Y_page_7_1.jpg peak at 1050 cm−1, which corresponds to the ν symmetric stretching mode. The related LOD, as demonstrated in our previous study,31 is approximately 2 mg/l and it is achievable by simply increasing the integration time from 20 to 60 seconds. Notably, a broader Raman mode at around 1330 cm−1 is assigned to a 00052_PSISDG11886_118860Y_page_7_2.jpg vibration.30, 31 As discussed in our previous work, for λexc = 236.5 nm a quantum yield ϕ = 0.05 was estimated, hence a contribution of nitrite within the LOD is not surprising. The visible spectrum has lower signal to noise ratio. Above the fluorescence background only water-related bands at 1620 cm−1 and about 2100 cm−1 are detectable, with only a minor, barely visible nitrate band at 1050 cm−1. Nitrate was clearly detectable also in the samples taken from an aquatic ecosystem (Fig. 7 bottom), including a 00052_PSISDG11886_118860Y_page_7_3.jpg peak indicative of 00052_PSISDG11886_118860Y_page_7_4.jpg photolysis. Using the intensity of the ν nitrate peak in the lab-made solution as a reference, an estimated 35 mg/l 00052_PSISDG11886_118860Y_page_7_5.jpg is present in the river. These results agree with the Ion Chromatography (IC) analysis adopted as a validation method. The latter estimated a 25 ± 0.04 mg/l 00052_PSISDG11886_118860Y_page_7_6.jpg concentration, with no presence of nitrite (at a concentration higher than 0.1 mg/l). The fluorescence background of visible Raman is especially strong in the river water spectrum, more than a factor of ten compared to the pure nitrate solution. We ascribe it to the presence of dissolved organic matter.46 Therefore, the visible Raman, unlike its deep-UV alternative, is unsuitable for the detection of environmentally-relevant nitrate concentrations, especially in the samples taken directly from aquatic ecosystems. It can be noted, that the O-H water peak is slightly suppressed compared to the Raman spectra of the pharmaceuticals. By applying the same procedure as outlined above, the absorption length for the river sample (Table S1, Supplemental Material) was evaluated. Notably, the beam is absorbed within a limited length (l = 2.4 mm), hence the intensity of the nitrate mode has to scale linearly with the concentration, as predicted in the section 3.2. UV Absorption spectra are plot in Fig. S6 (Supplemental Material). The relatively high molar absorptivity a is assigned to the presence of organic matter.

Figure 7.

Deep-UV and visible Raman measurements of: (top) 100 mg/l NaNO3 solution and (bottom) of the river water sample. The main features are marked by a red and black arrows for 00052_PSISDG11886_118860Y_page_9_3.jpg and 00052_PSISDG11886_118860Y_page_9_4.jpg, respectively.

00052_PSISDG11886_118860Y_page_9_2.jpg

4.3

Benzotriazole

Owing to its low sensitivity, visible Raman is not commonly used for detecting benzotriazole in solution at low concentrations (mg/l). On the other hand, the use of deep-UV light is hugely hindered by the ultraviolet absorption exhibited by this compound.8 In fact, the molecule presents two main absorption peaks at 259 and 278 nm.47 Aware of such limitations, we first evaluated visible Raman LOD by performing a high concentration series (from 500 mg/l to 10 g/l (Fig. S7, Supplemental Material). Despite the presence of fluorescence background between 500 and 2500 cm−1, main benzene ring modes are detectable at 783, 1010 and 1387 cm−1. As expected, the minimal concentration for reliably identifying those vibrational modes is high, approximately 1 g/l. The deep-UV Raman concentration series were performed on a different range, from 0.5 to 1000 mg/l. As before, no fluorescence overlap was present, however, unexpectedly, no benzotriazole related Raman modes were identified at any concentrations (see Fig. S8 Supplemental Material). Furthermore, above 1 g/l all Raman modes disappear and almost no signal is acquired. This effect of attenuation is most evident on the main water O-H modes at 3420 cm−1 (Fig. 8 top), with a strong attenuation in the deep-UV range and virtually none in the visible. This absorption effect is quantitatively and exponentially correlated to the analyte concentration (Fig. 8 bottom), with an appreciable drop above approximately 0.5 mg/l concentrations. Fitting the experimental data to the self absorption model discussed above yields the following function, which is based on eq.(8):

00052_PSISDG11886_118860Y_page_10_2.jpg

Figure 8.

(Top) Deep UV and visible measurement of deionized water and 10 mg/l benzotriazole concentration solution. (Bottom) Water Raman mode intensity (O-H stretching water mode at 3420 cm−1) plotted with respect to benzotriazole concentration and fitted through eq.(9). Inset: corresponding spectra (0.5 mg/l, 5 mg/l, 25 mg/l, 50 mg/l, 100 mg/l, 250 mg/l, 500 mg/l, 1000 mg/l concentration).

00052_PSISDG11886_118860Y_page_10_1.jpg

where the constant A = I0, x is the independent variable, C and B are free fit parameters. The intensity of the water peak is well fitted by the model (R2 = 0.99), (Fig. 8 bottom), providing a value of Cfit = 0.088. This term represents the main calibration parameter for estimating the analyte concentration, while B accounts for intensity of the signal background. We notice that few data points (from 50 to 250 mg/l) lie below the fitting curve and this can be ascribed to a variation of the laser intensity during the measurement process. The penetration depth, calculated for a 50 mg/l solution, is equal to 7.8 mm (Table S1, Supplemental Material), indicating an extensive beam absorption.

5.

FUTURE PERSPECTIVES

In this paper, various potent pollutant have been successfully measured, although the achieved LOD is much above their average concentration in urban waste water (tens of μg/l). Briefly, we summarize in this sections themain outcomes and our strategies to improve current limitations.

5.1

Photolysis

Nitrate was successfully detected in pure water (2 mg/l) and in aquatic ecosystem, and the LOD may be improved as discussed above. In particular the deep-UV Raman analysis of nitrate in aquatic ecosystems is promising, as dissolved organic matter often generates an overlapping fluorescence background, which hinders the visible or near-UV Raman analysis. Nevertheless, the photolysis of nitrate ions upon UV light illumination affects the reliability of the analysis. The photolysis effect is diminished by the reduction of the exposure time of the sample to UV light, e.g. by means of a circulating liquid cell that allows for a “constant flow” mode measurement.

5.2

Absorbance

Benzotriazole vibrational modes were not detected using deep-UV light at any concentration between 0.5 mg/l to 1 g/l, due to the self-absorption phenomenon. This effect highlights the constraint for a potential use of deep-UV Raman spectroscopy as an environmental analysis method. If a strong UV absorber is present in the specimen, the light may be completely attenuated by the UV absorbing compounds. However, using the Raman intensity of water as internal standard it is possible to estimate the concentration of the UV-absorbers in the solution even in the absence of Raman signal from the analyte. Pure water can be used for reference. This can be achieved in one set-up, e.g by using a liquid cell connected to both the solution and the pure water. The measured Raman intensities of detected pollutants can then be corrected using the model introduced in the theory section. The intensity change on the water signal is a qualitative indicator of the overall water purity comparable to already established UV-absorption sensors with the advantage of providing fundamental chemical information not accessible otherwise. In particular for automatic sensing applications, this quality factor is of high relevance. This is most easily be explained using an example. The Raman setup will have difficulties in detecting nitrate, if the solution also contains benzotriazol. As benzotriazol (at these concentrations) does not show Raman lines, an automatic system will report ‘clean water’. The quality factor using the water Raman signal does not provide a chemical analysis, but ensures that the Raman signal is reliable in case of no changes. Small changes can be used to correct the Raman analysis, while strong changes indicate that the Raman analysis is not reliable anymore.

5.3

Estimation of minimal limit of detection

The above given concrete examples may generalized to estimate the theoretically minimal limit of detection. We have to consider:

  • wavelength selection: the water window ends around 200 nm, the chosen wavelength of 236 nm is nearly ideal.17 At this wavelength, the absolute Raman crosssections for water and archetypal organic substances such as ethanol are similar.48 Although resonance enhancement may be possible, the wavelength dependence is negligible in the targeted range (deepUV).49

  • signal-to-noise ratio: peaks outside the water Raman transitions are detectable, if their signal is above the noise of background signal from water, which is estimated to be 100 to 1000 times lower than the Raman transitions (see, e.g., Figs. 8, 5), i.e., the signal intensity Is must be greater than

    00052_PSISDG11886_118860Y_page_11_1.jpg

    The ratio Is/Iwater is given the Raman crosssections times concentration, which are assumed to be equal (see above):

    00052_PSISDG11886_118860Y_page_11_2.jpg

    The minimal detectable concentration is thus simply

    00052_PSISDG11886_118860Y_page_11_3.jpg

    For environmental water analysis, a limit of detection in the order of μg/l ≃ cm = 10 −12 is envisioned. To reach this value, an SNR of 90 dB must be achieved.

  • Increasing SNR: various advances are required to achieve an SNR of 90 dB. This includes adopting collection optics of shorter focal length and higher numerical aperture (N.A.).50 Actually, in our setup the N.A. is relatively small (≈ 0.3) with respect to an average commercial Raman spectrometer. UV compatible lenses with high N.A. are commercially available. Taking as example (Edmund Optics) a fused-silica aspherical lens with N.A.=0.50 a significant increase in the intensity is expected. By assuming that the radiant flux collected by the optical component is ϕ ∝ (F/#)2 and being the 00052_PSISDG11886_118860Y_page_12_1.jpg, the collected light can be increased by at least 4 times. By reducing the focal length the beam waist decreases accordingly, minimizing l. Further signal to noise improvements are also achievable by using a highly sensitive spectrometer, such as an intensified CCD sensor (ICCD-based detector), increasing intensity by another factor of 10 to 1000. However, marked increase of the laser beam intensity is not advisable because of the associated photolysis effects. Although the SNR is only proportional to the square root of intensity increase, the envisioned limit of detection is achievable.Furthermore, the use of a reference cell would allow easy subtraction of the reference spectrum from the investigated one by alternating measurement of the two samples, thus further increasing the sensitivity of the analysis. Time-resolved approaches, e.g. time-gated Raman (TG Raman), would be a valid technique to be considered to reduce the fluorescence while increasing the SNR ration. During last years the diffusion of this technique has been boosted through different application fields due to remarkable improvements in the electronics, laser sources and sensing.51

6.

CONCLUSIONS

Opportunities and drawbacks of deep-UV Raman spectroscopy for monitoring water quality have been outlined by benchmarking a cost-effective and portable deep-UV Raman spectrometer. The outlined system is based on a commercial deep-UV laser source and inexpensive optical components. Potentially harmful compounds were measured at environmentally relevant concentrations in pure water and in aquatic environment, providing a reliable LOD for the technique. In particular, the impact of fluorescence, photolysis and self-absorption during the analysis process is thoroughly discussed. Such effects, can be considered as collateral since they invalidate the measurement by affecting the Raman spectra. Adopting a simple mathematical model, the self-absorption was described and the beam absorption depth quantified. In principle, by implementing the various measures including improvement of hardware as well as evaluation procedures, it is foreseeable that an LOD in the μg range is reachable, being an important benchmark in environmental water analysis. The perspective to carry out Raman spectrometry using narrow line X-ray pulses (see also Sects. 1-4 [52]) on electronic transition is attractive.

7.

ACKNOWLEDGMENTS

The project was partially funded by BAFU (Bundesamt für Umwelt). We thank Renato Figi and Jasmin Hubeli for IC measurements performed at Empa.

REFERENCES

[2] 

Singh, B. and Sekhon, G., “Nitrate pollution of groundwater from farm use of nitrogen fertilizers — a review,” Agriculture and Environment, 4 207 –225 (1979). https://doi.org/10.1016/0304-1131(79)90022-5 Google Scholar

[3] 

Doane, T. A. and Horwáth, W. R., “Spectrophotometric determination of nitrate with a single reagent,” Analytical Letters, 36 2713 –2722 (2003). https://doi.org/10.1081/AL-120024647 Google Scholar

[4] 

Trček, B., Žigon, D., Zidar, V. K., and Auersperger, P., “The fate of benzotriazole pollutants in an urban oxic intergranular aquifer,” Water Research, 131 264 –273 (2018). https://doi.org/10.1016/j.watres.2017.12.036 Google Scholar

[5] 

Lajeunesse, A., Smyth, S., Barclay, K., Sauvé, S., and Gagnon, C., “Distribution of antidepressant residues in wastewater and biosolids following different treatment processes by municipal wastewater treatment plants in canada,” Water Research, 46 5600 –5612 (2012). https://doi.org/10.1016/j.watres.2012.07.042 Google Scholar

[6] 

Giebultowicz, J. and Nalecz-Jawecki, G., “Occurrence of antidepressant residues in the sewage-impacted vistula and utrata rivers and in tap water in warsaw (poland),” Ecotoxicology and Environmental Safety, 104 103 –109 (2014). https://doi.org/10.1016/j.ecoenv.2014.02.020 Google Scholar

[7] 

Schultz, M. M., Furlong, E. T., Kolpin, D. W., Werner, S. L., Schoenfuss, H. L., Barber, L. B., Blazer, V. S., Norris, D. O., and Vajda, A. M., “Antidepressant pharmaceuticals in two u.s. effluent-impacted streams: Occurrence and fate in water and sediment, and selective uptake in fish neural tissue,” Environmental Science & Technology, 44 1918 –1925 (2010). https://doi.org/10.1021/es9022706 Google Scholar

[8] 

Molins-Delgado, D., Távora, J., Díaz-Cruz, M. S., and Barceló, D., “UV filters and benzotriazoles in urban aquatic ecosystems: The footprint of daily use products,” Science of The Total Environment, 601-602 975 –986 (2017). https://doi.org/10.1016/j.scitotenv.2017.05.176 Google Scholar

[9] 

Ward, M., Jones, R., Brender, J., de Kok, T., Weyer, P., Nolan, B., Villanueva, C., and van Breda, S., “Drinking water nitrate and human health: An updated review,” International Journal of Environmental Research and Public Health, 15 (1557), (2018). Google Scholar

[10] 

Wick, K., Heumesser, C., and Schmid, E., “Groundwater nitrate contamination: Factors and indicators,” Journal of Environmental Management, 111 178 –186 (2012). https://doi.org/10.1016/j.jenvman.2012.06.030 Google Scholar

[11] 

Assubaie, F. N., “Comparative study of methods used for the determination of nitrate and nitrite pollutants in well, ground water and soil in al-hassa area, saudi arabia,” Pakistan Journal of Biological Sciences, 7 508 –513 (2004). https://doi.org/10.3923/pjbs.2004.508.513 Google Scholar

[12] 

Goldstein, S. and Rabani, J., “Mechanism of nitrite formation by nitrate photolysis in aqueous solutions: the role of peroxynitrite, nitrogen dioxide, and hydroxyl radical,” Journal of the American Chemical Society, 129 10597 –10601 (2007). https://doi.org/10.1021/ja073609+ Google Scholar

[13] 

Meusinger, C., Berhanu, T. A., Erbland, J., Savarino, J., and Johnson, M. S., “Laboratory study of nitrate photolysis in antarctic snow. i. observed quantum yield, domain of photolysis, and secondary chemistry,” The Journal of Chemical Physics, 140 (244305), (2014). Google Scholar

[14] 

Sharpless, C. M. and Linden, K. G., “UV photolysis of nitrate: effects of natural organic matter and dissolved inorganic carbon and implications for UV water disinfection,” Environmental Science & Technology, 35 2949 –2955 (2001). https://doi.org/10.1021/es002043l Google Scholar

[15] 

Ludovici, M., Ialongo, C., and Camera, E., “Principles, current applications, and future perspectives of liquid chromatography-mass spectrometry in clinical chemistry,” Liquid Chromatography, 727 –751 Elsevier(2017). https://doi.org/10.1016/B978-0-12-805392-8.00022-0 Google Scholar

[16] 

Hundt, P. M., Tuzson, B., Aseev, O., Liu, C., Scheidegger, P., Looser, H., Kapsalidis, F., Shahmohammadi, M., Faist, J., and Emmenegger, L., “Multi-species trace gas sensing with dual-wavelength qcls,” Applied Physics B: Lasers and Optics, 124 (6), 108 (2018). https://doi.org/10.1007/s00340-018-6977-y Google Scholar

[17] 

Hale, G. M. and Querry, M. R., “Optical constants of water in the 200-nm to 200-μm wavelength region,” Appl. Opt., 12 555 –563 (1973). https://doi.org/10.1364/AO.12.000555 Google Scholar

[18] 

Kostamovaara, J., Tenhunen, J., Kögler, M., Nissinen, I., Nissinen, J., and Keränen, P., “Fluorescence suppression in raman spectroscopy using a time-gated CMOS SPAD,” Optics Express, 21 (31632), (2013). Google Scholar

[19] 

Bruzas, I., Lum, W., Gorunmez, Z., and Sagle, L., “Advances in surface-enhanced raman spectroscopy (SERS) substrates for lipid and protein characterization: sensing and beyond,” The Analyst, 143 (17), 3990 –4008 (2018). https://doi.org/10.1039/C8AN00606G Google Scholar

[20] 

Demirel, G., Usta, H., Yilmaz, M., Celik, M., Alidagi, H. A., and Buyukserin, F., “Surface-enhanced raman spectroscopy (SERS): an adventure from plasmonic metals to organic semiconductors as SERS platforms,” Journal of Materials Chemistry C, 6 (20), 5314 –5335 (2018). https://doi.org/10.1039/C8TC01168K Google Scholar

[21] 

Tolles, W. M., Nibler, J. W., McDonald, J. R., and Harvey, A. B., “A review of the theory and application of coherent anti-stokes raman spectroscopy (CARS),” Applied Spectroscopy, 31 253 –271 (1977). https://doi.org/10.1366/000370277774463625 Google Scholar

[22] 

Asher, S. and Johnson, C., “Raman spectroscopy of a coal liquid shows that fluorescence interference is minimized with ultraviolet excitation,” Science, 225 311 –313 (1984). https://doi.org/10.1126/science.6740313 Google Scholar

[23] 

Troyanova-Wood, M. A., Petrov, G. I., and Yakovlev, V. V., “Simple and inexpensive instrument for deep-UV raman spectroscopy,” Journal of Raman Spectroscopy, 44 1789 –1791 (2013). https://doi.org/10.1002/jrs.4394 Google Scholar

[24] 

Asher, S. A., “UV resonance raman spectroscopy for analytical, physical, and biophysical chemistry. part 1,” Analytical Chemistry, 65 59A –66A (1993). https://doi.org/10.1021/ac00052a715 Google Scholar

[25] 

Efremov, E. V., Ariese, F., and Gooijer, C., “Achievements in resonance raman spectroscopy,” Analytica Chimica Acta, 606 119 –134 (2008). https://doi.org/10.1016/j.aca.2007.11.006 Google Scholar

[26] 

Domes, C., Domes, R., Popp, J., Pletz, M. W., and Frosch, T., “Ultrasensitive detection of antiseptic antibiotics in aqueous media and human urine using deep UV resonance raman spectroscopy,” Analytical Chemistry, 89 9997 –10003 (2017). https://doi.org/10.1021/acs.analchem.7b02422 Google Scholar

[27] 

Ianoul, A., Coleman, T., and Asher, S. A., “Uv resonance raman spectroscopic detection of nitrate and nitrite in wastewater treatment processes,” Analytical Chemistry, 74 1458 –1461 (2002). https://doi.org/10.1021/ac010863q Google Scholar

[28] 

Furuya, N., Matsuyuki, A., Higuchi, S., and Tanaka, S., “Determination of nitrate ion in waste and treated waters by laser raman spectrometry,” Water Research, 13 (4), 371 –374 (1979). https://doi.org/10.1016/0043-1354(79)90025-3 Google Scholar

[29] 

Gajaraj, S., Fan, C., Lin, M., and Hu, Z., “Quantitative detection of nitrate in water and wastewater by surface-enhanced raman spectroscopy,” Environmental Monitoring and Assessment, 185 5673 –5681 (2012). https://doi.org/10.1007/s10661-012-2975-4 Google Scholar

[30] 

Asher, S. A., Tuschel, D. D., Vargson, T. A., Wang, L., and Geib, S. J., “Solid state and solution nitrate photochemistry: Photochemical evolution of the solid state lattice,” The Journal of Physical Chemistry A, 115 4279 –4287 (2011). https://doi.org/10.1021/jp200406q Google Scholar

[31] 

Sterzi, A., Schneider, U., Sambalova, O., Bleiner, D., and Borgschulte, A., “Tunable deep-UV raman spectroscopy reveals nitrate photolysis,” UV and Higher Energy Photonics: From Materials to Applications 2019, SPIE(Google Scholar

[32] 

Bykov, S. V., Mao, M., Gares, K. L., and Ashera, S. A., “Compact solid-state 213 nm laser enables standoff deep ultraviolet raman spectrometer: Measurements of nitrate photochemistry,” Applied Spectroscopy, 69 895 –901 (2015). https://doi.org/10.1366/15-07960 Google Scholar

[33] 

Lednev, I. K., Karnoup, A. S., Sparrow, M. C., and Asher, S. A., “-helix peptide folding and unfolding activation barriers: a nanosecond UV resonance raman study,” Journal of the American Chemical Society, 121 8074 –8086 (1999). https://doi.org/10.1021/ja991382f Google Scholar

[34] 

Gares, K. L., Hufziger, K. T., Bykov, S. V., and Asher, S. A., “Review of explosive detection methodologies and the emergence of standoff deep UV resonance raman,” Journal of Raman Spectroscopy, 47 124 –141 (2016). https://doi.org/10.1002/jrs.v47.1 Google Scholar

[35] 

Hollis, J. R., Rheingold, D., Bhartia, R., and Beegle, L. W., “An optical model for quantitative raman microspectroscopy,” Applied Spectroscopy, 74 684 –700 (2020). https://doi.org/10.1177/0003702819895299 Google Scholar

[36] 

D’Amico, F., Saito, M., Bencivenga, F., Marsi, M., Gessini, A., Camisasca, G., Principi, E., Cucini, R., Di Fonzo, S., Battistoni, A., Giangrisostomi, E., and Masciovecchio, C., “Uv resonant raman scattering facility at elettra,” Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 703 33 –37 (2013). https://doi.org/10.1016/j.nima.2012.11.037 Google Scholar

[37] 

Masciovecchio, C., Bencivenga, F., and Gessin, A., “Water dynamics at the nanoscale,” Condensed Matter Physics, Google Scholar

[38] 

“Quantum mechanical theory of rayleigh and raman scattering,” The Raman Effect, 49 –84 John Wiley & Sons, Ltd.Google Scholar

[39] 

Hong, Z. and Asher, S. A., “Dependence of raman and resonance raman intensities on sample self-absorption,” Applied Spectroscopy, 69 75 –83 (2015). https://doi.org/10.1366/14-07531 Google Scholar

[40] 

Shriver, D. F. and Dunn, J. B. R., “The backscattering geometry for raman spectroscopy of colored materials,” Applied Spectroscopy, 28 319 –323 (1974). https://doi.org/10.1366/000370274774332371 Google Scholar

[41] 

Liu, C. and Berg, R. W., “Nonlinearity in intensity versus concentration dependence for the deep UV resonance raman spectra of toluene and heptane,” Applied Spectroscopy Reviews, 48 425 –437 (2013). https://doi.org/10.1080/05704928.2012.708888 Google Scholar

[42] 

Structure Determination of Organic Compounds, Springer, Berlin Heidelberg (2009). Google Scholar

[43] 

Bock, P., Nousiainen, P., Elder, T., Blaukopf, M., Amer, H., Zirbs, R., Potthast, A., and Gierlinger, N., “Infrared and raman spectra of lignin substructures: Dibenzodioxocin,” Journal of Raman Spectroscopy, (2020). https://doi.org/10.1002/jrs.v51.3 Google Scholar

[44] 

Czernicki, W. and Baranska, M., “Carbamazepine polymorphs: Theoretical and experimental vibrational spectroscopy studies,” Vibrational Spectroscopy, 65 12 –23 (2013). https://doi.org/10.1016/j.vibspec.2012.11.011 Google Scholar

[45] 

Zanchi, C., Lucotti, A., Tommasini, M., Trusso, S., de Grazia, U., Ciusani, E., and Ossi, P. M., “Laser tailored nanoparticle arrays to detect molecules at dilute concentration,” Applied Surface Science, 396 1866 –1874 (2017). https://doi.org/10.1016/j.apsusc.2016.03.147 Google Scholar

[46] 

Aziz, F., Ouazzani, N., Mandi, L, Assaad, A., Pontvianne, S., Poirot, H, and Pons, M.-N., “Characteriza-tion by fluorescence of dissolved organic matter in rural drinking water s torage tanks in morocco,” Journal of Water and Health, 16 321 –328 (2018). https://doi.org/10.2166/wh.2018.002 Google Scholar

[47] 

Li, Y, Wang, J., Guo, WGao, C., and Cheng, Z., “Determination of benzotriazole in industrial cooling water by fluorescence,” Instrumentation Science &Technology, 45 290 –300 (2016). https://doi.org/10.1080/10739149.2016.1220392 Google Scholar

[48] 

Abe, N., “Effects of hydrogen bonding on the raman intensities of methanol, ethanol and water,” Journal of Raman Spectroscopy, 7 (3), 161 –167 (1978). https://doi.org/10.1002/jrs.v7:3 Google Scholar

[49] 

Faris, G. W. and Copeland, R. A., “Wavelength dependence of the raman cross section for liquid water,” Appl. Opt., 36 2686 –2688 (1997). https://doi.org/10.1364/AO.36.002686 Google Scholar

[50] 

Kelly, J. F., Blake, T A., Bernacki, B. E., and Johnson, T J., “Design considerations for a portable raman probe spectrometer for field forensics,” International Journal of Spectroscopy, 2012 1 –15 (2012). https://doi.org/10.1155/2012/938407 Google Scholar

[51] 

Kögler, M. and Heilala, B., “Time-gated raman spectroscopy – a review,” Measurement Science and Technology, 32 Google Scholar

[52] 

D. Bleiner, “The Science and Technology of X-ray Lasers:,” in A 2020 Update Proc. SPIE, 1188602 (2021). Google Scholar
© (2021) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Andrea Sterzi, Olga Sambalova, Urs Schneider, Davide Bleiner, and Andreas Borgschulte "Short-wavelength Raman spectroscopy for in-situ water analysis: a feasibility study", Proc. SPIE 11886, International Conference on X-Ray Lasers 2020, 118860Y (8 July 2021); https://doi.org/10.1117/12.2596045
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
Back to Top