This study aims to make the implementation of the attenuation and backscattering study to be apply for skin tissue. Attenuation and backscatter coefficients were defined in the range 22 to 48 MHz based on the L38 transducer which has a central frequency of 30 MHz and a 60% bandwidth. The attenuation coefficient was computed using local attenuation estimation algorithms (i.e., spectral difference and spectral log difference methods). Using a reference phantom technique, we compensated the recorded backscatter spectra for system-independent measurements. This study shows the experimental and theoretical steps needed to implement a quantitative study of ultrasounds images in the skin.
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Benavides Lara, I. Demirkan, M. Tsoukas, and K. Avanaki "Implementing the quantitative study of ultrasound imaging of skin", Proc. SPIE 12352, Photonics in Dermatology and Plastic Surgery 2023, 123520H (14 March 2023); https://doi.org/10.1117/12.2651493