Poster + Paper
7 April 2023 Parameter selection for convex optimization time calibration for a 2-panel PET system
Gregory Romanchek, Shiva Abbaszadeh
Author Affiliations +
Conference Poster
Abstract
Coincidence timing calibration is fundamental to PET imaging. The electronics, cable lengths, and detector physics such as charge drift and depth dependence add to the measured time differences in coincidence sorting – increasing random rate, decreasing true rate, and degrading system performance. This work investigates the parameter selection for convex optimization (Ordinary Least Squares) for timing calibration. We test the correlation between commonly selected parameters and the experimentally measured coincidence time difference. Additionally, we test 127 nested models of a parameterized regression equation to identify the those which optimize MSE, BIC, and FWHM, respectively. In each of these models, the FWHM performance improved ~53%, though the value shifted from ~ 304 to 160 ns – far from ~ 10 ns FWHM CZT can achieve. The results point to the lack of a necessary parameter, such as trigger threshold level or temperature, or data which is too variable for the OLS optimization.
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gregory Romanchek and Shiva Abbaszadeh "Parameter selection for convex optimization time calibration for a 2-panel PET system", Proc. SPIE 12463, Medical Imaging 2023: Physics of Medical Imaging, 124633W (7 April 2023); https://doi.org/10.1117/12.2654423
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Positron emission tomography

Data modeling

Mathematical optimization

Performance modeling

Convex optimization

Education and training

Back to Top