1 April 2024 Dual-source photon-counting CT: impact of residual cross-scatter on quantitative spectral results
Leening P. Liu, Edgar Salazar, Pooyan Sahbaee, Harold I. Litt, Peter B. Noël
Author Affiliations +
Conference Poster
Abstract
Dual-source photon-counting CT combines the high temporal resolution and high pitch of dual-source CT with the material quantification capabilities of photon-counting CT. It, however, results in cross-scatter that increases in severity with increased patient size and collimation. This cross-scatter must be corrected to ensure the removal of scatter artifacts and improve quantitative accuracy. To evaluate residual cross-scatter of a first-generation dual-source photon-counting CT and the effect of phantom size, collimation, and radiation dose, a phantom was scanned in single- and dual-source modes with and without its extension ring at three collimations and three radiation doses. Virtual monoenergetic images (VMI) at 50 keV, VMI 150 keV, and iodine density maps were reconstructed to determine variation between acquisition parameters in single- and dual-source modes. Additionally, differences relative to single-source acquisitions and to singlesource and small collimation acquisitions were calculated to reflect residual cross-scatter with and without matched collimation. At VMI 50 keV, inserts exhibited accuracy and similar variation between single- and dual-source modes, averaging 5.4 ± 2.6 and 6.2 ± 2.5 HU, respectively, across phantom size, collimation, and radiation dose. Differences relative to single-source measured 5.1 ± 8.5 and 0.4 ± 4.2 HU while differences relative to single-source and small collimation acquisitions were 6.4 ± 10.8 HU and -0.5 ± 3.9 HU for VMI 50 and 150 keV, respectively. This minimal residual cross-scatter increases confidence in the quantitative accuracy of spectral results necessary for clinical applications of dual-source photon-counting CT with motion, such as cardiac imaging.
(2024) Published by SPIE. Downloading of the abstract is permitted for personal use only.
Leening P. Liu, Edgar Salazar, Pooyan Sahbaee, Harold I. Litt, and Peter B. Noël "Dual-source photon-counting CT: impact of residual cross-scatter on quantitative spectral results", Proc. SPIE 12925, Medical Imaging 2024: Physics of Medical Imaging, 129252N (1 April 2024); https://doi.org/10.1117/12.3006109
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Collimation

Iodine

Photon counting

Computed tomography

Image quality

Back to Top