|
1.INTRODUCTIONThe cutting position of cigarettes is one of the key factors that determine the stability and consistency of cigarette quality [1]. Firstly, the cutting position affects the processability of cut tobacco; for example, a high cutting position will lead to insufficient tensile strength and easy breakage. Secondly, it also affects the composition of cigarette smoke, and any deviation in the cutting position will affect the uniformity of the cigarette’s flavor and burning, thereby affecting the smoking experience. In addition, the accuracy of the cutting position is also related to the appearance quality of the cigarette. In general, a cutting position that is too low or too high will negatively impact the appearance quality of cigarettes, such as causing surface wrinkles or breakage of the cut tobacco, and this effect is more obvious in cigarettes with a smaller diameter [2-4]. Therefore, it is of vital importance to control the cutting position precisely to guarantee the quality of cigarettes and increase production efficiency. As a non-contact and high-precision measuring method, microwave technology has been widely used in cigarette cutting position measurement, such as the MW 4420 Microwave Moisture and Density Profile Measuring System by TEWS Elektronik. There have been many successful applications of these instruments in cigarette cutting position measurements [5-10]. However, the stability and accuracy of the microwave measuring system are affected by various factors, such as equipment aging, environmental changes, and measurement errors between different pieces of equipment, etc. Therefore, regular calibration of the instrument is necessary to ensure its measuring precision. The purpose of this study is to thoroughly investigate the calibration approach for cigarette cutting position measurement and validate its efficacy through practical application, thereby enhancing the accuracy of the microwave measuring system, increasing the level of automation in cigarette production, minimizing manual intervention, and improving production efficiency and product quality. This has significant practical implications for facilitating the technological advancement and productivity of the tobacco industry. 2.THEORY AND MODELSThe main calibration methods for measuring the cutting position of cigarettes are as follows: Conduct a statistical analysis of “dense-end part” and “normal dense part” data in relation to cigarette density, and calculate the density ratio between them, which serves as the design constraint parameter for the calibration kit. Design imitation cigarette strips of different fixed lengths, by simulating the density distribution of the cigarette strip, to create a complete set of calibration kits with varying dense-end lengths for “inner and outer discharge” cigarettes as the calibration standards. Calibrate the microwave density equipment using the calibration kit. Measure with the calibration kit to obtain an error value by comparing the true cutting position value with the calculated cutting position value. Set the optimal threshold and resolution based on the comparison between the density ratio obtained in step 1) and the cutting position error value, followed by iterative calculations to determine the current equipment’ s calibration value. 2.1Structure and density of calibration kitThe operating frequency of equipment based on microwave technology for measuring cigarette cutting positions is typically in the range of 2.0 to 3.0 GHz. Therefore, the design of the calibration kit must consider the material response within this frequency band. Table 1 shows the dielectric properties and density information for common plastic materials. Since the density of cigarettes is mostly in the range of 200 to 300 mg/cm³ (i.e., 0.2 to 0.3 g/cm³), which is significantly different from the densities of common plastics listed in Table 1, polypropylene materials with low density values (0.85 to 0.91 g/cm³) and moderate dielectric constants were selected. In addition to its stability and processability, polypropylene was chosen as the calibration material due to its suitable microwave response. Table 1.Electrical characteristics of plastics.
Based on the analysis of the differences in distribution position and density between the “dense-end part” and the “normal dense part” of cigarettes, the calibration kit structure shown in Figure 1 was designed. The white segment in the figure simulates the “normal dense part”, while the gray segment simulates the “dense-end part”. The length of the “dense-end part” varies between the left and right ends, mimicking the actual length difference of the “dense-end part” in real cigarette strips. In addition, this paper incorporates the density distribution characteristics of cigarettes, specifically the density ratio of the “dense-end part” to the “normal dense part” (8 ~ 15%). On this basis, and in accordance with the actual machining accuracy (± 0.1 mm), the processing was carried out. Figure 2 shows the density distribution diagram of the calibration kits, where different colors represent the density distribution of the calibration kits with varying dense-end lengths. The density distribution within the range of 1 to 11 mm reflects the corresponding length differences, and the “normal dense part” in the middle section exhibits similar density and length. Additionally, the density ratio of the “dense-end part” to the “normal dense part” falls within the design range, indicating that the density distribution range of the calibration kits is reasonable and suitable for calibrating the cutting position measurement. 2.2Calibration of evaluation parametersIn this study, two key calibration techniques are elaborated in depth: density measurement calibration and cutting position calculation calibration. In the domain of microwave density measurement and calibration, a calibration kit made from polypropylene material is selected as the calibration reference. This calibration kit exhibits stable density characteristics at room temperature (25°C), ensuring the accuracy and repeatability of measurements with an error less than 3 mg/cm³, thus meeting the high standard requirements for calibration. In the cutting position calibration procedure, this paper uses half the length difference of the “inner and outer row” dense-end parts as the key calibration index. Calibration kits with different specifications have varying dense-end part lengths, specifically 0 mm, 1 mm, 2 mm, 3 mm, and 4 mm, corresponding to sizes 0# to 4#. These length values establish the standard offset for the cutting position. During the calibration process, the cutting position of the calibration kit is precisely calculated and compared with the actual offset of the cutting position. The calibration result is considered acceptable when the deviation between the calculated outcome and the actual offset is less than 1 mm, and the repeatability of the calibration is less than 0.5 mm. 2.3Calibration methods2.3.1Environmental conditionsThe ambient temperature should be maintained at 25±2 °C, and the relative humidity should be between 40% and 70% RH. 2.3.2EquipmentMW-T Microwave Moisture and Density Profile Measuring System.The equipment needs to be preheated for ten minutes. 2.3.3MW-T density calibration procedureSet up a calibration brand according to the size of the calibration kits (length 54mm, circumference 24.5 mm). Select the default model and cigarette measurement type, and set the batch number to 8. Put the 0# calibration kit into sample pool, start the measurement, and record the measurement results for 8 times. Compare the measured density with the actual density of 0# calibration kit, calculate measurement error and density ratio. Calibrate the MW-T with density ratio as correction value, obtaining a calibration coefficient. Set the coefficient as a measurement parameter. 2.3.4Cigarettes cutting position measurement calibration procedureSet up a calibration brand according to the size of the calibration kits (length 54mm, circumference 24.5 mm). Select the default model and cutting position measurement type, and set the batch number to 10. Designate the “dense-end part” with arrowheads (i.e., as shown in Figure 2) as the “inner row” cigarette, and the section without arrowheads as the “outer row” cigarette. Sample the 0# calibration kit 10 times in the direction of the arrow. Reverse the calibration kit and sample it 10 more times, completing a cutting position measurement. Measure the remaining calibration kits in sequence using the above method. Set the iteration range from 2 to 4, the optimization threshold value to 0.5 mm, and the resolution to 0.01%, and import CPSA software for iteration to calculate the best parameters for cutting position calibration. 3.RESULTS AND ANALYSIS3.1Calibration of density measurement and evaluation of uncertainty3.1.1Density measurement calibration analysisAccording to the density calibration procedure outlined in Section 2.3.3, the density calibration results for the 0# calibration kit are presented in Table 2. The acatual density of the 0# calibration kit was confirmed by a tripartite measuring mechanism to be 380.84 mg/cm³. Prior to calibration, the maximum error between the density measurement values recorded in Table 2 and the actual density values reached 53.26 mg/cm³, while the minimum error was 51.02 mg/cm³, and the average error was 52.44 mg/cm³. The density calibration ratio is 1.16. Following calibration, the density of the 0# calibration kit was remeasured. The mean error between the measurement results and the actual density was reduced to 0.96 mg/cm³, with a repeatability of 1.54 mg/cm³, both meeting and exceeding the calibration accuracy requirement of less than 3 mg/cm³. Table 2.0# calibration kit measured density.
3.1.2Uncertainty evaluation measurement modelWhere, ΔD is the indication error of densimeter, in mg/cm3; D is the indication value of densimeter, in mg/cm3; DS is the constant density value of calibration kit at 25°C, in mg/cm3. 3.1.4Source and evaluation of standard uncertainty3.1.4.1Standard uncertainty assessment of DThe standard uncertainty [11,12] u1(D) was introduced for repeated measurements of 0 # calibration kit by MW-T, with the density values were 380.72 mg/cm3, 381.81 mg/cm3, 382.41 mg/cm3, 378.72 mg/cm3, 378.75 mg/cm3, 379.38 mg/cm3, 380.94 mg/cm3, 382.25 mg/cm3, 380.81 mg/cm3, 380.64 mg/cm3 for 10 times. The standard deviation Sn of the density values was 1.26 mg/cm3. Referring to JJG 42-2011 Verification Regulation of Working Glass Hydrometers, by repeating the measurement 10 times, the standard uncertainty u1(D) is calculated as: In addition, the standard uncertainty u2(D) is introduced from the MW-T resolution d (0.01 mg/cm3), which is calculated by the following formula: 3.2Standard uncertainty evaluation of DSThe standard uncertainty of DS is primarily introduced by the density of 0# calibration kit. The density uncertainty of the calibration kit is not more than U = 1 mg/cm3, with a coverage factor K = 2, then: 3.3Calibration and analysis of cigarettes cutting position3.3.1Analysis of cutting position calibrationAfter density calibration of the MW-T, the calibration was performed according to the cigarette cutting position calibration procedure outlined in Section 2.3.3. The measurement data of the calibration kits were imported into CPSA software for iterative analysis, with the iteration range set from 2 to 4 and a resolution of 0.01%. A total of 201 iterations were conducted. Taking 0.5 mm as the threshold for iterative optimization, the optimal parameter for cutting position calibration was finally determined to be 2.42. After testing, the average error of each calibration kit was 0.16 mm, which meets the calibration accuracy requirement that the difference between the calculated cutting position and the actual offset is less than or equal to 1 mm. Table 3 displays the results and errors of cutting position calculation for each calibration kit. The data in the table indicate that the calculated cutting position error of each calibration kit after calibration is less than 1 mm, and the standard deviation is 0.23 mm. In summary, the method of cutting position calibration using calibration kits meets the accuracy requirements for calibration. Table 3.Results of calibration kits offset.
3.2Calibration repeatabilityAccording to the data in Table 4, the results of calibrating each calibration kit ten times were obtained. The standard deviations were all less than 0.5 mm. The coefficient of variation was less than 0.2 for all kits except for the relatively large 0# calibration kit, indicating that all indicators have met the evaluation criteria for calibration repeatability. Table 4.Repeatability of cutting position calibration.
Note: The average value in the table represents the average of calibrated results obtained from ten repetitions; the standard deviation is the standard deviation of calibrated results obtained from ten repetitions; the coefficient of variation is the ratio of standard deviation to the mean value. 3.3Calibration stabilityFor the calibration kits, the MW-T was used to perform calibration measurements continuously for five months under the same ambient temperature and humidity (25 ± 2 °C, 40-70% RH), and the results were presented in Table 5. The results indicate that after long-term calibration tests, the standard deviation of calibrated offset of each calibration kit is less than 0.5 mm. This suggests that the calibration kits and the calibration method employed exhibit good stability and are suitable for long-term calibration applications. Table 5.Stability of cutting position calibration.
Note: The calibration results in the table represent the average of the calibration values obtained from three calibration runs per month; the standard deviation is the standard deviation of the average measurements over five months. 3.4Calibration applicabilityIn order to verify the applicability of the calibration kit for cutting position calibration across multiple MW-T systems, four MW-T systems were selected as experimental equipment in this study. Firstly, each piece of equipment was calibrated for density and cutting position using the 0# calibration kit, following the steps outlined in Section 2.3. Then, the cutting position of the calibration kits was measured using each equipment. The optimal calibration parameters and average errors after calibration are presented in Table 6. According to the table, there is a certain degree of inter-device variability between different pieces of equipment. Even when the same group of calibration kits is used for calibration, the optimal calibration parameters for each equipment differ. However, the average measurement error of the calibration kit after calibration is less than 1 mm, which indicates that the method has good universal applicability for cutting position calibration across different equipment. Table 6.Multi-equipment cutting position measurements.
4.CONCLUSIONSDue to the current lack of an effective and feasible calibration method for cigarette cutting position using the microwave method in the industry, this paper statistically analyzed the density distribution of cigarettes and comprehensively considered the proportion of the dense-end length between the inner and outer row cigarettes. Based on this analysis, a set of calibration kits was designed. The MW-T was calibrated and its uncertainty was evaluated using this set of calibration kits. The results indicate that the average error between the density value obtained by the microwave method and the true density value of the calibration kits is 0.96 mg/cm³, with a repeatability of 1.54 mg/cm³, both of which meet the calibration accuracy requirement of less than 3 mg/cm³. The measurement uncertainty for the calibration of the MW-T using the 0# standard part is: U = 1.30 mg/cm³, K = 2. Additionally, the repeatability and stability tests demonstrate that the calibration kits and the calibration method can achieve long-term and stable efficient calibration of the equipment. The method is notable for its fast and convenient calibration process and ease of operation, providing an effective quality tool for the tobacco industry. The method will continue to be optimized to enhance the accuracy and stability of calibration, and to explore the possibility of integrating with other advanced technologies to meet the higher and more comprehensive quality control requirements of the tobacco industry. REFERENCESZhang Z,
“Cigarette Weight Control and Adjustment of Cigarette Dense-end Position[J],”
Industrial Control Computer, 23
(12), 110
–111
(2010). Google Scholar
Wang X, Wang Y, Men J, et al,
“Influence of Fine Cigarette Cutting Position on Cigarette Quality Index [J],”
Anhui Agricultural Science Bulletin, 25
(01), 93
–94
(2019). Google Scholar
Deng H.,
“Study on the Influence of Cigarette’s Dense end Position of Cigarette Machines [J],”
Scientific and Technological Innovation, 2017
(35), 64
–65 Google Scholar
Zou Q, Chen R, Zhao Y, et al,
“Characterization Method and Control Technique of Cut Tobacco Distribution in Slim Cigarette [J],”
Tobacco Science & Technology, 52
(05), 87
–93
(2019). Google Scholar
Guan X, Zhu B, Zhang Q, et al,
“Analysis of Physical Quality Changes in Domestic Slim Cigarettes [J],”
Journal of China Tobacco Journal, 30
(04), 1
–14
(2024). Google Scholar
Yu S, Huang J, Zhang Y, et al,
“Difference Analysis of Microwave Moisture and Density Distribution Measuring Instrument and Oven to Determine Moisture Content of Finished Cigarette [J],”
Light Industry Science and Technology, 38
(05), 81
–83
(2022). Google Scholar
Zhao C, Ma T, Zeng Z, et al,
“Study on the Relationship Between the Weight of Cut Tobacco at the End of Cigarette and the Density Difference Between inside and outside of the Compacted end of Cigarette [J],”
Light Industry Science and Technology, 36
(05), 117
–118
(2020). Google Scholar
Wang K, Wang J, Fang L,
“Influence of Cigarette Density on Quality of Cigarette Mainstream Smoke [J],”
Hubei Agricultural Sciences, 58
(17), 81
–84
(2019). Google Scholar
Wang H, Ning N, Liu H,
“Application of Cigarette Density and Moisture Detector in Pressure Drop Stability Test of Filter Rod [J],”
Modern Agricultural Science and Technology, 22 26
–27
(2011). Google Scholar
Shen X, Li H, Wang R, et al,
“Effects of Particle Size Distribution of Cut Tobacco on Cigarette Weight and Density [J],”
Journal of China Tobacco Journal, 15
(06), 23
–26
(2009). Google Scholar
Jiang T, Guo X, Dong J, et al,
“Study on the Calibration Method and Uncertainty Evaluation of the Thermogravimetric Moisture Meter [J],”
Brand and Standardization, 03 248
–250
(2024). Google Scholar
JJG 42-2011, Verification Regulation of Working Glass Hydrometers [S], Google Scholar
|