Paper
16 June 1997 Digital fluoroscopy using the optically coupled CCD: preliminary investigations
Author Affiliations +
Proceedings Volume 2976, Biomedical Sensing, Imaging, and Tracking Technologies II; (1997) https://doi.org/10.1117/12.275534
Event: BiOS '97, Part of Photonics West, 1997, San Jose, CA, United States
Abstract
This paper reports our preliminary investigation of developing an optically coupled CCD x-ray imaging system for digital x-ray fluoroscopy. The limitations of an image intensifier, TV camera based fluoroscopy are discussed. The contrast sensitivity and contrast-detail deductibility of a lens coupled CCD prototype was compared with an image- intensifier, TV camera based fluoroscopy. The results indicated that a well designed coupled CCD can be x-ray quantum noise limited, and it offers better contrast and resolution than the investigated image-intensifier, TV camera fluoroscopy systems.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hong Liu, Jian Xu, Gary E. Halama, and James A. McAdoo "Digital fluoroscopy using the optically coupled CCD: preliminary investigations", Proc. SPIE 2976, Biomedical Sensing, Imaging, and Tracking Technologies II, (16 June 1997); https://doi.org/10.1117/12.275534