Paper
2 May 1997 Contrast resolution study of CdZnTe detectors for medical imaging
Bindu Pillai, George C. Giakos, Amlan Dasgupta, Samir Chowdhury, Srinivasan Vedantham, P. Ghotra, J. Odogba, Victor A. Vega-Lozada, Ravi K. Guntupalli, Sankararaman Suryanarayanan, Robert J. Endorf, Anthony M. Passalaqua, William J. Davros
Author Affiliations +
Abstract
The detected signal and noise contributions were measured and related to the radiation exposure and tube current tube setting. Furthermore, the detector contrast has been experimentally determined. The experimental results indicate that Cd1-xZnxTe detectors have high detector contrast resolution. Therefore, they appear to be very attractive for x-ray digital imaging applications.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bindu Pillai, George C. Giakos, Amlan Dasgupta, Samir Chowdhury, Srinivasan Vedantham, P. Ghotra, J. Odogba, Victor A. Vega-Lozada, Ravi K. Guntupalli, Sankararaman Suryanarayanan, Robert J. Endorf, Anthony M. Passalaqua, and William J. Davros "Contrast resolution study of CdZnTe detectors for medical imaging", Proc. SPIE 3032, Medical Imaging 1997: Physics of Medical Imaging, (2 May 1997); https://doi.org/10.1117/12.274015
Lens.org Logo