2 May 1997 Ellipsoid scan: chest cone-beam CT with a large ellipsoidal view field using a 16-in. x-ray image intensifier
Rika Baba, Ken Ueda, Hironori Ueki, Keiji Umetani
Author Affiliations +
Abstract
We propose a cone-beam computed tomography (CT) system that has a large field of view and high image quality, and that can be applied to the chest imaging. This system uses a 16- inch x-ray image intensifier and a television camera. To enlarge the field of view, an ellipsoid-scan sequence is used, which requires the subject to be moved parallel to the transaxial plane during scanning. The contrast resolution is increased by applying a television camera with high sensitivity and non-linear efficiency. We tested this system using a stationary apparatus, and obtained an isotropic 3D image of a chest phantom which had 0.64 mm voxels and covered both lungs. This system attained better spatial resolution for coronal images and its contrast resolution in the transaxial images was only 3 to 7 times larger than conventional spiral-scan CT at the same x-ray dose. This system is promising for ling cancer screening, precise diagnosis and surgery.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Rika Baba, Ken Ueda, Hironori Ueki, and Keiji Umetani "Ellipsoid scan: chest cone-beam CT with a large ellipsoidal view field using a 16-in. x-ray image intensifier", Proc. SPIE 3032, Medical Imaging 1997: Physics of Medical Imaging, (2 May 1997); https://doi.org/10.1117/12.274002
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-ray computed tomography

X-rays

X-ray imaging

Sensors

Cameras

Chest

Image resolution

Back to Top