16 April 1997 Comparative evaluation of conventional and zero-crossover rare-earth screen-film systems in pediatric chest radiography
Thomas C. Fearon, Daniela Dumitru-Buna, David C. Kushner, Dorothy Bulas, Carlos Sivit
Author Affiliations +
Abstract
This study compares the image quality of conventional and zero-crossover symmetric and asymmetric rare-earth screen film systems for the specific imaging task of chest radiography in pediatric patients by means of evaluation of the visualization of anatomic landmarks. Advances in anticrossover technology in double emulsion film have made it possible to design each emulsion layer in a double- emulsion film as if it were a distinct system. Chest radiographs of 120 pediatric patients performed with one of six screen-film systems (two conventional rare-earth, two zero crossover symmetric and two zero crossover asymmetric screen-film systems) were evaluated for image quality by three radiologist observers. The evaluation was based on the visualization of anatomic landmarks and the subjective evaluation of the physical properties of each film. InSight VHC Thoracic Imaging system with IPC film and Pediatric Insight screens with Insight Pediatric film were ranked highest by the radiologist observers both with respect to the visualization of anatomic landmarks and the subjective evaluation of the physical properties of the systems. The difference between these two systems was not statistically significant (p less than 0.05), however, the ranking of these systems is statistically different (p less than 0.05) relative to the remaining four screen-film systems. Zero- crossover screen-film systems have been shown to provide improved image performance for the specific task of pediatric chest radiography based upon the evaluation of the visualization of anatomic landmarks relative to conventional systems evaluated in this study.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Thomas C. Fearon, Daniela Dumitru-Buna, David C. Kushner, Dorothy Bulas, and Carlos Sivit "Comparative evaluation of conventional and zero-crossover rare-earth screen-film systems in pediatric chest radiography", Proc. SPIE 3036, Medical Imaging 1997: Image Perception, (16 April 1997); https://doi.org/10.1117/12.271299
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Imaging systems

Visualization

Radiography

Chest

Image quality

Lung

Image visualization

RELATED CONTENT

Image-display optimization using clinical history
Proceedings of SPIE (May 01 1991)
Efficacy Of SER In Chest Diagnosis
Proceedings of SPIE (September 19 1985)
Motion blur detection in radiographs
Proceedings of SPIE (March 11 2008)
The Effect Of Pixel Size On The Detection Rate Of...
Proceedings of SPIE (August 28 1984)

Back to Top