Paper
6 June 2002 A reference tristimulus colorimeter
Author Affiliations +
Proceedings Volume 4421, 9th Congress of the International Colour Association; (2002) https://doi.org/10.1117/12.464724
Event: 9th Congress of the International Color Association, 2001, Rochester, NY, United States
Abstract
A reference tristimulus colorimeter has been developed at NIST with a transmission-type silicon trap detector (1) and four temperature-controlled filter packages to realize the Commission Internationale de l'Eclairage (CIE) x(λ), y(λ) and z(λ) color matching functions (2). Instead of lamp standards, high accuracy detector standards are used for the colorimeter calibration. A detector-based calibration procedure is being suggested for tristimulus colorimeters wehre the absolute spectral responsivity of the tristimulus channels is determined. Then, color (spectral) correct and peak (amplitude) normalization are applied to minimize uncertainties caused by the imperfect realizations of the CIE functions. As a result of the corrections, the chromaticity coordinates of stable light sources with different spectral power distributions can be measured with uncertainties less than 0.0005 (k=1).
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
George P. Eppeldauer "A reference tristimulus colorimeter", Proc. SPIE 4421, 9th Congress of the International Colour Association, (6 June 2002); https://doi.org/10.1117/12.464724