Paper
18 December 2001 High-resolution x-ray spectroscopy using a large format GaAs array
Didier D. E. Martin, Alan Owens, Christian Erd, S. Andersson, Anthony J. Peacock, Hans Andersson, V. Lamas, Seppo Arvo Anter Nenonen, N. Haack
Author Affiliations +
Abstract
We report X-ray measurements on a prototype 1.1 cm2, 32x32 GaAs pixel array produced to assess the technological feasibility of making large area, near Fano-limited arrays which operate near room temperature. Measurements were carried out on four widely separated pixels bot in our laboratories and using monochromatic X-ray pencil beams at the HASYLAB synchrotron research facility in Hamburg, Germany. The pixels were found to be very uniform both in their energy and spatial responses. For example, typical energy resolutions of 300 eV at 10.5 keV, rising to ~560 eV at 60 keV were achieved. The corresponding resolutions measured under full-area illumination were found to be the same within statistics, indicating uniform crystallinity and stoichiometry. Likewise, by raster scanning at 15 keV, 15x15 micrometers 2 beam across each of the pixels, the gain uniformity across the pixels (and by implication the entire array) was determined to be statistically flat.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Didier D. E. Martin, Alan Owens, Christian Erd, S. Andersson, Anthony J. Peacock, Hans Andersson, V. Lamas, Seppo Arvo Anter Nenonen, and N. Haack "High-resolution x-ray spectroscopy using a large format GaAs array", Proc. SPIE 4507, Hard X-Ray and Gamma-Ray Detector Physics III, (18 December 2001); https://doi.org/10.1117/12.450753
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

X-rays

Gallium arsenide

Image resolution

Spectroscopy

Amplifiers

Solids

RELATED CONTENT

The wide field imager for IXO status and future...
Proceedings of SPIE (July 29 2010)
The x-ray microcalorimeter spectrometer onboard Athena
Proceedings of SPIE (September 17 2012)
MCP readout for the AXAF-I grating spectrometer
Proceedings of SPIE (December 16 1993)
Progress on GaAs cryogenic readout circuits for SISCAM
Proceedings of SPIE (June 27 2006)

Back to Top