5 June 2003 Compact high-speed computed radiography (CR) system using a linear CCD with a large-area phtodiode (PD) and dual transfer lines
Satoshi Arakawa, Hiroaki Yasuda, Takao Kuwabara, Hideki Suzuki, Toshiaki Suzuki, Tatsuya Hagiwara
Author Affiliations +
Abstract
We performed an image quality simulation for the line scan system, which realizes a compact and high-speed Computed Radiography (CR) system. The line scan system uses a line light source and a linear CCD sensor. In this system, the emitted light must be efficiently focused onto the CCD sensor to detect the emitted light as much as possible. To realize the effective light detection, we analyzed the spread of the light in the photostimulable phosphor layer. We also estimated the image quality based on X-ray absorption, the amount of emitted light, light collecting efficiency and electric noise. It clarified the image quality is affected strongly by such factors as the spread of the PSL, the size of photo diodes of the CCD sensor and electric noise.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Satoshi Arakawa, Hiroaki Yasuda, Takao Kuwabara, Hideki Suzuki, Toshiaki Suzuki, and Tatsuya Hagiwara "Compact high-speed computed radiography (CR) system using a linear CCD with a large-area phtodiode (PD) and dual transfer lines", Proc. SPIE 5030, Medical Imaging 2003: Physics of Medical Imaging, (5 June 2003); https://doi.org/10.1117/12.479988
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
CCD image sensors

Image quality

Light scattering

Chromium

X-rays

Scattering

Computing systems

Back to Top