29 September 2004 Ground calibration of x-ray CCD detectors with charge injection for the X-ray imaging spectrometer on Astro-E2
Author Affiliations +
Abstract
The X-ray Imaging Spectrometer (XIS) on the Astro-E2 mission contains X-ray CCDs fabricated at MIT Lincoln Laboratory. We present results from CCD subassembly calibration of these detectors. Absolute detection efficiency and spectral resolution in the 0.3-10 keV band are presented for four front-illuminated and two back-illuminated flight candidate devices. In addition, we characterize the charge injection capabilities of these devices. The quantum efficiency measurements have been used to model the various dead layers in the gate structure and to constrain the depletion depth of the detectors. Noise levels of 2 electrons, RMS, and spectral resolution better than 50 eV (FWHM) at 525 eV have been achieved. The small but measurable charge transfer inefficiency (less than 1e-5 per transfer at 5.9 keV) of these devices has been characterized as a function of X-ray energy.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Beverly LaMarr, Mark W. Bautz, Steven E. Kissel, Gregory Y. Prigozhin, Kiyoshi Hayashida, Takeshi Go Tsuru, and Hironori Matsumoto "Ground calibration of x-ray CCD detectors with charge injection for the X-ray imaging spectrometer on Astro-E2", Proc. SPIE 5501, High-Energy Detectors in Astronomy, (29 September 2004); https://doi.org/10.1117/12.553204
Lens.org Logo
CITATIONS
Cited by 14 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

X-rays

Charge-coupled devices

X-ray imaging

Spectral resolution

Quantum efficiency

Calibration

RELATED CONTENT


Back to Top