20 April 2005 Estimating surface radiation dose from multidetector CT: cylindrical phantoms, anthropomorphic phantoms, and Monte Carlo simulations
J. J. DeMarco, C. H. Cagnon, D. D. Cody, D. M. Stevens, C. H. McCollough, J. O'Daniel, M. F. McNitt-Gray
Author Affiliations +
Abstract
Multidetector CT (MDCT) systems offer larger coverage and wider z-axis beams, resulting in larger cone angles. One impact on radiation dose is that while radiation profiles at isocenter are constant when contiguous axial scans are performed, the increased beam divergence from the larger cone angle results in significant surface dose variation. The purpose of this work was to measure the magnitude of this effect. Contiguous axial scans were acquired using an MDCT for two sizes of cylindrical phantoms and an anthropomorphic phantom. Film dosimetry and/or radiation detector measurements were performed on the surface of each phantom. Detailed mathematical models were developed for the MDCT scanner and all phantoms. Monte Carlo simulations of contiguous axial scans were performed for each phantom model. From cylindrical phantoms, film dosimetry at the surface showed differences between peak and valley that reached 50%. From the anthropomorphic phantom, measured values ranged from 7.9 to 16.2 mGy at the phantom surface. Monte Carlo simulations demonstrated these variations in both cylindrical and anthropomorphic phantoms. The magnitude of variation was also related to object size. Even when contiguous axial scans are performed on MDCT, surface radiation profiles show considerable variation. This variation will increase as MDCT cone angles increase and when non-contiguous scans (e.g. pitch > 1) are acquired. The variation is also a function of object size. While average surface doses may remain constant, peak doses may increase, which may be significant for radiation sensitive organs at or near the surface (e.g. breast, thyroid).
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. J. DeMarco, C. H. Cagnon, D. D. Cody, D. M. Stevens, C. H. McCollough, J. O'Daniel, and M. F. McNitt-Gray "Estimating surface radiation dose from multidetector CT: cylindrical phantoms, anthropomorphic phantoms, and Monte Carlo simulations", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); https://doi.org/10.1117/12.595933
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Monte Carlo methods

Field effect transistors

Scanners

Sensors

X-ray computed tomography

Head

Natural surfaces

RELATED CONTENT


Back to Top