7 September 2006 Cone beam reconstruction algorithm for PET-VCT
J. Hsieh, A. Lonn, M. Nyka
Author Affiliations +
Abstract
A typical X-ray CT scanner has a 50cm reconstruction field-of-view (FOV) which is significantly smaller than the 70cm FOV typically used for PET. When CT images are used for attenuation correction in PET, significant artifacts result when portions of the scanned object extend beyond 50 cm. In this paper, we present a reconstruction algorithm to estimate the missing projection samples so that a significantly larger reconstruction FOV can be achieved. The estimation is based on the observation that the total attenuation is a continuous function over views. The initial estimate of the projection distribution makes use of the property that the projection profile does not change abruptly as a function of the projection angle. When combined with the projection values at the location of truncation, the size and location of water cylinders that best fit these parameters are calculated. Additional adjustments are made to ensure the total attenuation of the expanded projection samples matches closely to the estimated total missing projection. The estimated projection is further blended with the mirror image of the measured projections near the location of truncation. Various phantom experiments were conducted to evaluate the proposed algorithm. Both axial and helical scans were collected on a 64-slice scanner. Results show that the proposed algorithm can restore the fidelity of the reconstruction for the portion of the object inside 50 cm FOV, and provide an adequate impression of the object outside 50cm.
© (2006) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J. Hsieh, A. Lonn, and M. Nyka "Cone beam reconstruction algorithm for PET-VCT", Proc. SPIE 6318, Developments in X-Ray Tomography V, 63180H (7 September 2006); https://doi.org/10.1117/12.681209
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Signal attenuation

Reconstruction algorithms

Positron emission tomography

Error analysis

Scanners

X-ray computed tomography

Sensors

RELATED CONTENT

The reason of why dynamic dual energy CT is better...
Proceedings of SPIE (October 17 2022)
Neural MLAA for PET-enabled dual-energy CT imaging
Proceedings of SPIE (February 15 2021)
Comparison of scatter correction methods for CBCT
Proceedings of SPIE (March 02 2006)
A solution to arc correction in cylindrical PET scanner
Proceedings of SPIE (September 13 2005)
Dynamic CT simulation for minimal invasive surgery
Proceedings of SPIE (May 06 2004)
Signal statistics in x-ray computed tomography
Proceedings of SPIE (May 03 2002)

Back to Top