9 March 2010 Strain correction in interleaved strain-encoded (SENC) cardiac MR
Abdallah G. Motaal, Nael F. Osman
Author Affiliations +
Abstract
The strain encoding (SENC) technique directly encodes regional strain of the heart into the acquired MR images and produces two images with two different tunings so that longitudinal strain, on the short-axis view, or circumferential strain on the long-axis view, are measured. Interleaving acquisition is used to shorten the acquisition time of the two tuned images by 50%, but it suffers from errors in the strain calculations due to inter-tunings motion of the heart. In this work, we propose a method to correct for the inter-tunings motion by estimating the motion-induced shift in the spatial frequency of the encoding pattern, which depends on the strain rate. Numerical data was generated to test the proposed method and real images of human subjects were used for validation. The proposed method corrected the measured strain values so they became nearly identical to the original ones. The results show an improvement in strain calculations so as to relax the imaging constraints on spatial and temporal resolutions and improve image quality.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Abdallah G. Motaal and Nael F. Osman "Strain correction in interleaved strain-encoded (SENC) cardiac MR", Proc. SPIE 7626, Medical Imaging 2010: Biomedical Applications in Molecular, Structural, and Functional Imaging, 762602 (9 March 2010); https://doi.org/10.1117/12.841777
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Magnetic resonance imaging

Temporal resolution

Heart

Tissues

Computer programming

Medicine

Motion measurement

Back to Top