3 March 2012 Feasibility study to demonstrate cardiac imaging using fast kVp switching dual-energy computed tomography: phantom study
Author Affiliations +
Abstract
Dual-energy computed tomography is a novel imaging tool that has the potential to reduce beam hardening artifacts and enhance material separation over conventional imaging techniques. Dual-energy acquisitions can be performed by using a fast kVp technology to switch between acquiring adjacent projections at two distinct x-ray spectra (80 and 140 kVp). These datasets can be used to further compute material density and monochromatic images for better material separation and beam hardening reduction by virtue of the projection domain process. The purpose of this study was to evaluate the feasibility of using dual-energy in cardiac imaging for myocardial perfusion detection and coronary artery lumen visualization. Data was acquired on a heart phantom, which consisted of the chambers and aorta filled with Iodine density solution (500 HU @ 120 kVp), a defect region between the aorta and chamber (40 HU @ 120 kVp), two Iodinefilled vessels (400 HU @ 120 kVp) of different diameters with high attenuation (hydroxyapatite) plaques (HAP), and with a 30-cm water equivalent body ring around the phantom. Prospective ECG-gated single-energy and prospective ECG-gated dual-energy imaging was performed. Results showed that the generated monochromatic images had minimal beam hardening artifacts which improved the accuracy and detection of the myocardial defect region. Material density images were useful in differentiating and quantifying the actual size of the plaque and coronary artery lumen. Overall, this study shows that dual-energy cardiac imaging will be a valuable tool for cardiac applications.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Priti Madhav, Yasuhiro Imai, Suresh Narayanan, Sandeep Dutta, Naveen Chandra, and Jiang Hsieh "Feasibility study to demonstrate cardiac imaging using fast kVp switching dual-energy computed tomography: phantom study", Proc. SPIE 8313, Medical Imaging 2012: Physics of Medical Imaging, 83132Y (3 March 2012); https://doi.org/10.1117/12.911526
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Iodine

X-ray computed tomography

Cardiac imaging

Computed tomography

Dual energy imaging

Arteries

Heart

Back to Top