Paper
25 September 2012 HST/WFC3 UVIS detectors: radiation damage effects and mitigation
S. M. Baggett, K. Noeske, J. Anderson, J. W. MacKenty, L. Petro
Author Affiliations +
Abstract
Devices in low Earth orbit are particularly susceptible to the cumulative effects of radiation damage and the Hubble Space Telescope Wide Field Camera 3 (HST/WFC3) UVIS detectors, installed on HST in May 2009, are no exception. Such damage not only generates new hot pixels but also generates charge traps which degrade the charge transfer efficiency (CTE), causing a loss in source flux as well as a systematic shift in the object centroid as the trapped charge is slowly released during readout. Based on an analysis of internal and external monitoring data, we provide an overview of the consequences of the ~3 years of radiation damage to the WFC3 CCD cameras. The advantages and disadvantages of available mitigation options are discussed, including use of the WFC3 post-flash and charge injection modes now available to observers, and the status of an empirical pixel-based correction similar to the one adopted for the HST Advanced Camera for Surveys (ACS).
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. M. Baggett, K. Noeske, J. Anderson, J. W. MacKenty, and L. Petro "HST/WFC3 UVIS detectors: radiation damage effects and mitigation", Proc. SPIE 8453, High Energy, Optical, and Infrared Detectors for Astronomy V, 845336 (25 September 2012); https://doi.org/10.1117/12.926901
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Radiation effects

Sensors

Stars

Amplifiers

Signal attenuation

Photometry

Calibration

RELATED CONTENT


Back to Top