6 March 2013 Multi-energy performance of a research prototype CT scanner with small-pixel counting detector
S. Kappler, A. Henning, B. Krauss, F. Schoeck, K. Stierstorfer, T. Weidinger, T. Flohr
Author Affiliations +
Proceedings Volume 8668, Medical Imaging 2013: Physics of Medical Imaging; 86680O (2013) https://doi.org/10.1117/12.2006747
Event: SPIE Medical Imaging, 2013, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
We have investigated the multi-energy performance of our most recent prototype CT scanner with CdTe-based counting detector. With its small pixel pitch of 225 μm this device is prepared for the high X-ray fluxes occurring in clinical CT. Each of these pixels is equipped with two adjustable counters. The ASIC architecture of the detector allows configuration of the counter thresholds in chess patterns, enabling data acquisition in up to four energy bins. We have studied the material separation capability of counting CT with respect to potential clinical applications. Therefore we have analyzed contrast and noise properties in material decomposed CT images using up to four base materials. We have studied contrast agents containing iodine, gadolinium, or gold, and the body-like materials calcium, fat, and water. We describe the mathematical framework used in this work and demonstrate the general multi-energy capability of counting CT with simulations and experimental data from our prototype scanner. To prove the clinical relevance of our studies we compare the results to those obtained with well-established dual-kVp techniques recorded at same patient dose and with identical image sharpness.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. Kappler, A. Henning, B. Krauss, F. Schoeck, K. Stierstorfer, T. Weidinger, and T. Flohr "Multi-energy performance of a research prototype CT scanner with small-pixel counting detector", Proc. SPIE 8668, Medical Imaging 2013: Physics of Medical Imaging, 86680O (6 March 2013); https://doi.org/10.1117/12.2006747
Lens.org Logo
CITATIONS
Cited by 35 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

X-ray computed tomography

Scanners

Calcium

Gadolinium

Iodine

Prototyping

Back to Top