Paper
26 September 2013 The effect of gain variation in micro-channel plates on gamma-ray energy resolution
Author Affiliations +
Abstract
A Monte Carlo simulation of micro-channel plate (MCP) with particular interest in its effect on energy resolution performance is presented. Important physical processes occurring in MCP channels are described and modeled, including secondary electron (SE) yield, SE emission, and primary electron reflection. The effects causing channel saturation are also introduced. A two dimensional Monte Carlo simulation is implemented under the assumption of unsaturated channel. Simulation results about basic MCP performances and especially gain and energy resolution performances are presented and analyzed. It’s found that energy resolution as an intrinsic property of MCP cannot be improved simply by adjusting system parameters; however it can be improved by increasing input signal or number of photoelectrons (PEs) in the context of image intensifier. An initial experiment with BazookaSPECT detector and CsI(Tl) scintillator is performed to validate and correlate with the simulation results and good agreement is achieved.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ling Han, Harrison H. Barrett, H. Bradford Barber, and Lars R. Furenlid "The effect of gain variation in micro-channel plates on gamma-ray energy resolution", Proc. SPIE 8853, Medical Applications of Radiation Detectors III, 88530D (26 September 2013); https://doi.org/10.1117/12.2030076
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Microchannel plates

Monte Carlo methods

Selenium

Image intensifiers

Sensors

Scintillators

Image resolution

Back to Top