The intercomparison experiment for RadNet is necessary and was conducted in Huailai Remote Sensing Station from 5th Aug to 10th Aug in 2012. Time series observations from various sensors were collected and analyzed. The maximum relative differences among sensors of UVR, SWR, LWR, PAR, and LST are 4.83%, 5.3%, 3.71%, 11%, and 0.54%, respectively. Sensor/parameter differences indeed exist and are considerable large for PAR, SWR, UVR, and LWR, which cannot be ignored. The linear normalization and quadratic polynomial normalization perform similar for CUV5/UVR, PQS1/PAR, CNR4/SWR, and SI-111/LST. As for CNR4/LWR, quadratic polynomial normalization show higher accuracy than linear normalization, especially in node2, node4, and node5. Thus, the LWR measured by CNR4 is proved to be nonlinear, and should be normalized with quadratic polynomial coefficients for higher precision. |