1 September 2006 Noncontact backscatter-mode near-infrared time-resolved imaging system: preliminary study for functional brain mapping
Ichiro Sase, Akira Takatsuki, Junji Seki, Toshio Yanagida, Akitoshi Seiyama
Author Affiliations +
Abstract
To improve the spatial resolution and to obtain the depth information of absorbers buried in highly scattering material, we developed a noncontact backscatter-mode near-infrared time-resolved imaging system (noncontact B-TRIS) that is intended for functional human brain mapping. It consists of mode-locked Ti-sapphire lasers as light sources and a charge-coupled device camera equipped with a time-resolved intensifier as a detector. The system was tested with a white polyacetal phantom as a light-scattering medium and black polyacetal particles as absorbers. Illumination and detection of light through an objective lens system (Φ=150 mm) enabled us to capture images from an area whose diameter is about 70 mm without coming into contact with it. The scattering and absorption coefficients of the white phantom obtained by B-TRIS were similar to those obtained by a conventional time-resolved spectroscopy. Although the imaged diameter of an absorber buried within a phantom was considerably larger than the actual diameter, the center position of the absorber coincided with the actual position with accuracy <2 mm. Furthermore, the depth information can be also detected by the noncontact B-TRIS. These results suggest a potential of noncontact B-TRIS for imaging cognitive human brain function.

1.

Introduction

Simultaneous measurements of a near-infrared imaging technique with other noninvasive functional neuroimaging techniques revealed that near-infrared optical measurement has become one of the important tools for understanding human brain activities.1, 2, 3 One significant feature of the optical method is that near-infrared spectroscopy (NIRS) can provide the redox state of the terminal mitochondrial enzyme cytochrome c oxidase noninvasively as well as changes in concentrations of deoxygenated and oxygenated hemoglobin, which occurs in the human cerebral cortex in response to neural activity.4 Another important feature is that the optical system does not restrict subjects in a constrictive imaging chamber during measurements, so the system could be applied to measurements of brain activations induced by various types of tasks, such as gait tasks,5 visual stimulation of infants,6 and metal tasks with a written test,7 as well as those of higher-order brain functions.8, 9, 10 Furthermore, the optical method is now applied to detect the fast responses linked with electrophysiological phenomena,11, 12, 13 which more directly reflects the neuronal activity in the human cerebral cortex.

Most conventional NIRS imaging systems utilize several pairs of optical fibers for the irradiation and detection of light to provide surface-projection images of an object. However, those multiple fiber systems have the following disadvantages:14, 15 (1) spatial resolution and sensitivity depend on the location of the optical fibers, and (2) the signal-to-noise (S/N) ratio is affected by the attachment or angle of the fiber tip against the skin (or the subject). As NIRS is increasingly applied to clinical and cognitive neuroscience, these issues have come to be pointed out more often. Furthermore, it is important to improve spatiotemporal resolution of NIRS to obtain more “precise functional mapping” of the brain by expanding the use of the features of light, such as wavelength, polarization, intensity, and traveling time. Thereby, we expect a new insight into unexplained brain functions such as the “sensory motor paradox” during tactile stimulation,16 pain examinations,17 and imagery experiments,18 which are occasionally difficult to detect by conventional brain mapping techniques such as functional magnetic resonance imaging (fMRI) and position emission tomography.

Recently, various attempts have been made to improve spatial resolution and/or quantification of NIRS signals using fiber optical systems19, 20, 21 and noncontact systems 22, 23, 24, 25, 26 based on continuous wave19, 23, 24 and time20, 21, 25 or frequency22, 26 domain methods.

Advantages of the time-resolved optical measurement system have been demonstrated by many fundamental experiments trying to evaluate the contrast-to-noise ratio (i.e., sensitivity) by varying inclusion depth20 or source to detector distance21 and the resolving power of optical tomographic schemes using fast-transmitted photons of ultrashort pulses27 and by trying to calculate path length for adopting the modified Lambert-Beer equation using the time-of-flight value.28

In the present study, we have developed a noncontact type of backscatter-mode time-resolved imaging system (noncontact B-TRIS) to overcome the problems mentioned before. We have assembled a reflective imaging system and demonstrated its merits when detecting and mapping light-absorbing particles buried inside the scattering media. There are several merits of noncontact B-TRIS:

  • (1) Direct visualization of absorbers at their actual positions, without any complicated calculations or interpolations.

  • (2) Improvement of spatial resolution resulting from liberation from the restriction of fiber positioning.

  • (3) Noncontact detection (without attachment of optical fibers to the skin).

  • (4) Detection of depth information by time window (detection of Z position and improvement of S/N ratio by volume limitation).

We tested the above four items with noncontact B-TRIS using a polyacetal phantom. In this paper, we discuss the development of this technology to apply it to human brain mapping in comparison with the conventional multichannel NIRS.

2.

Materials and Methods

2.1.

Camera System

A schematic drawing of a reflective-mode time-resolved optical imaging system is shown in Fig. 1 . Two sets of mode-locked Ti-sapphire lasers (Tsunami, Spectra-Physics Lasers, USA), which are tunable to conventionally used wavelengths to measure brain activity (e.g., 780, 805, or 830nm ; 80MHz , 10 to 16ps , 800 to 1100mW ), were used as light sources. Pulsing of the 730-nm laser was synchronized with that of the 830-nm laser by a lock-to-clock system (Lock-to-clock, Spectra-Physics Lasers). Stabilities of laser pulsing were monitored by a photodiode (PD). Both lasers are guided into separate optical fiber switchers (F-SM19, Piezosystem Jena GmbH, Germany) through objective lenses. One laser pulse within 1 sec was incident on one optical fiber, and it was switched to the other optical fibers one by one by an optical fiber switcher controlled by a personal computer (PC). Each optical fiber output was introduced into a set of optical elements that consisted of a ball lens, iris, and convex lens, which collimate output beams [Fig. 2b ]. Laser powers were reduced to 1/100 by neutral density (ND) filters placed immediately after the laser output. Then, the laser beams were led to the optical switching system. Each laser power at the sample surface was also adjusted to 100μW by the output iris. Pulse width at the sample was around 100ps (initial laser pulse widths were about 10 to 15ps ).

Fig. 1

Block diagram of noncontact B-TRIS system.

054006_1_032605jbo1.jpg

Fig. 2

Noncontact illumination system. (a) Schematic drawing of the optical system around the sample. (b) Laser illumination assembly. (c) Photographic image of entire optical assembly.

054006_1_032605jbo2.jpg

Images were captured by a charge-coupled device (CCD) camera (Imager3, LaVision GmbH, Germany) equipped with a time-resolved intensifier (Picostar, La Vision). Laser output triggered a high-voltage gate pulse for the time-resolved intensifier. Timing of image acquisition and laser illumination was controlled by the PC. Acquired images were transferred to the PC independently from laser pulse timing (80MHz10 to 100Hz) and analyzed by a separate software (Image Pro Plus, Media Cybernetics, USA).

2.2.

Noncontact Illumination System

To achieve noncontact illumination, we have prepared a light illumination assembly [Figs. 2a, 2b, 2c] encircling an objective lens [ f=150mm , diameter (ϕ)=150mm , borosilicate crown glass (BK7), antireflection (AR) coated]. A schematic drawing of the optical system around the sample is shown in Fig. 2a. The sample in the drawing is a large phantom block and the signal from the phantom is imaged by the objective lens through a large aperture on the upright optical bench. The sample image is relayed by a combination of the objective lens, and two sets of camera lenses (Nikkor 50mm F1.2S, Nikon, Japan). The upright optical bench board holds multiple laser illuminating assemblies shown in Fig. 2b. The laser beam emanating from the optical fiber is collimated by the ball lens and focused on the sample surface by a convex lens and iris to a spot with a diameter less than 1mm . Time lags of the laser output between each optical fiber caused by the optical switcher were between 25 and 200ps . However, fibers with the closest output times were selected so that the time difference was maintained at less than 50ps for every experiment. A photograph of the optical assembly is shown in Fig. 2c. Orange fibers are the incident optical fibers from a fiber switcher. In the present study, the distance between the phantom surface and the detector (and the convex lens for illumination) depends on the focal length of the objective lens (f=150mm) , and the distance between the phantom surface and CCD was about 700mm .

3.

Results

In the present study, we evaluated a noncontact B-TRIS to measure optical properties of scattering media and resolving power for mapping light-absorbing particles buried inside the scattering media in the reflectance mode for the purpose of applying this technology to the human brain mapping.

3.1.

Measurement of Optical Properties of Phantom Sample

For the phantom, we have used white polyacetal as a light-scattering medium and black polyacetal particles for absorbers. Temporal profiles of light intensity reflected by the large bulk phantom (Fig. 3 ) were measured by noncontact B-TRIS. Plots of the light intensity (solid lines) and the fitting curve using Patterson’s half infinite model29 are shown in Fig. 3. Their optical properties were obtained by the present system (noncontact B-TRIS) and the conventional fiber-based time-resolved system (TRS-10, Hamamatsu Photonics, Hamamatsu, Japan)30 and are shown in Table 1 . Although there is difference in the wavelength used here, 780nm for noncontact B-TRIS and 761 and 795nm for TRS-10, the scattering coefficient (μs) obtained by both instruments gave similar results. On the other hand, absorption coefficient (μa) showed about twofold difference between these instruments. This probably indicates that the order 103mm1 is the sensitivity limit of μa for B-TRIS and/or TRS-10. However, optical properties of the human brain tissue (cortex) are μs =1.00mm1 and μa0.02mm1 .31 The value of μa of the human cortex is about 20times larger than that of the present phantom sample. Thus, the B-TRIS and TRS-10 may be practical to measure the optical properties of the human brain tissue.

Fig. 3

Temporal profiles of light intensity at 780nm reflected by large bulk phantom (solid lines) and fitting curves (dashed lines) using Patterson’s half infinite model (Ref. 29) assuming refractive index (n) of the phantom was 1.48. Measurements were performed every 25ps . Light intensities, detected at distance (ρ) of 12, 15, 21, and 26mm from light illumination spot, were integrated over 3×3mm2 of the detection area. Estimated optical properties of the phantom are shown in Table 1.

054006_1_032605jbo3.jpg

Table 1

Optical properties obtained with near-infrared time-resolved spectroscopy.

B-TRISTRS-10
Distance between Input and Output μs′ 1 (mm−1) μa 2 (×10−3mm−1) μs′ 1 (mm−1) μa 2 (×10−3mm−1)
15mm 1.011.36nd3nd
21mm 1.100.98ndnd
26mm 0.960.95ndnd
30mm ndnd1.000.59
40mm ndnd0.970.61

1 μs′ , scattering coefficient.

2 μa , absorption coefficient. Optical properties of a bulk white polyacetal were measured with a noncontact B-TRIS instrument and conventional time-resolved spectroscopy [TRS-10, Hamamatsu Photonics, Japan (Ref. 30)]. The values for noncontact B-TRIS show the results obtained at 780nm , and the values for TRS-10 show the mean value of the results obtained at 761 and 795nm .

3nd: not determined (data for distances longer than 30mm could not be obtained by B-TRIS due to low signal intensity and low S/N ratio, and the data for distances shorter than 26mm could not be obtained by TRS-10 due to strong reflected light intensity greater than the sensitivity of the detector).

3.2.

Time-Resolved Imaging in Scattering Media

The captured images and calculation procedures for detecting small absorber particles buried inside the polyacetal phantom are illustrated in Figs. 4a, 4b, 4c, 4d . The illumination spots marked with filled red circles and the position of the absorber marked with an open black circle, where illuminating laser spots were placed hexagonally to cover the imaging area of about 90mm in diameter, are illustrated in Fig. 4e [see also Fig. 2a].

Fig. 4

Detection of black absorber. (a) Images without an absorber; bright spots denote illuminating positions. (b) Images with an absorber. (c) Images subtracted (b) from (a); bright spot predicts the position of absorber. (d) Integrated image of seven images in (c). (e) Illumination spots (red circles) illustrated on the image (d); position of the absorber is shown by a black open circle. Wavelength of 805nm , which corresponds to an isosbestic point of oxygenated and deoxygenated hemoglobin, was used to capture images.

054006_1_032605jbo4.jpg

Backscattered images from the phantom without an absorber are shown in Fig. 4a, and the same images with an absorber buried at a depth of 17.5mm are shown in Fig. 4b. The absorber is 10mm in diameter (φ) and 5mm thick ( t ). A pulse laser was illuminated sequentially to seven different spots [see light spot in images in Figs. 4a and 4b], and images were obtained with a 30-ms CCD integration time and a 600-ps time window (1800 to 2400ps after the laser illumination, see Fig. 8 ) of the gated intensifier. Images in Fig. 4c were obtained by subtracting images of Fig. 4b from those of Fig. 4a, where specularly reflected light, which appeared as bright spots in Figs. 4a and 4b, were removed by subtraction and only diffusely reflected lights were obtained in Fig. 4c. An integrated image of the seven images in Fig. 4c is shown in Fig. 4d. As referred to in Fig. 4e, Fig. 4c indicates decreases in the reflected signal (“shadow,” i.e., light blue and yellow area) just above the absorber (except for numbers 2 and 3, whose illuminating spots are too far from the absorber). The center of the shadow is less than 2mm from the center of the absorber. Those correspondences occurred not only for the integrated image [Fig. 4d] but also for each subtracted image [Fig. 4c].

Fig. 8

Changes in normalized light intensity obtained at different depths of the absorber. Configuration of incident optical fibers and an absorber was the same as that shown in Fig. 4e, but the depth of absorber was changed. Wavelength of 805nm was used, and each plot was obtained with a time window of 600ps . The data points indicate the light intensity integrated over the region of 3×3mm2 on the phantom surface at a distance of 40mm from the incident light spot. The absorber was buried in the middle of incident and detection points.

054006_1_032605jbo8.jpg

In the present study, the time window was selected to be 600ps (1800 to 2400ps after the illumination of the pulse laser), which gave a broadening of the diameter of the shadow [full width at half maximum (FWHM)=25mm ] 2.5 times larger than the actual diameter of the absorber (10mm) . If earlier arriving photons are captured, the backscattered light from the superficial area may become dominant and its intensity may increase. As a result, the signal from deeply buried absorbers (cf. Fig. 8) is lost.

3.3.

Spatial Resolution of B-TRIS

For noncontact B-TRIS, the sensitivity to the target object depends on the geometry of the illuminator, but the detected position of the target object is probably unaffected by the position or geometry of the detector (Fig. 4), because the detection area of B-TRIS is the whole area (i.e., 90-mm diameter of field of view using ϕ=150-mm objective lens). The spatial configuration of light illumination and detection of B-TRIS are illustrated in Fig. 5. Figure 5a shows the position of three absorbers buried in the phantom. The three absorbers buried at a depth of 20mm in the phantom. Figure 5b shows the region of light illumination (closed circles) and the region of detection (slashed mark region). Images obtained using noncontact B-TRIS, in which individual shadows of three absorbers can be visualized clearly, are shown in Figs. 5c and 5d [Fig. 5c, pseudo colored image; Fig. 5d, black and white image]. Furthermore, the position and shape of shadows obtained by noncontact B-TRIS did not change even when the relative positions of sample and light illumination were changed (data not shown).

Fig. 5

(a) Position of absorbers buried at the depth of 20mm in the phantom. (b) Illustration of light illumination position and receiving areas of noncontact B-TRIS. Closed circles indicate the region of illumination and the slash mark indicates the region of detection. (c) Pseudo color image obtained with B-TRIS. (d) Grayscale image of (c). Wavelength of 805nm was used. Time window of noncontact B-TRIS was 600ps (between 1800 and 2400ps ).

054006_1_032605jbo5.jpg

Further, to evaluate the spatial resolution of noncontact B-TRIS quantitatively, we placed two absorbers in the phantom at the depth of 20mm (inset of Fig. 6 ). The profiles of the reflected light intensity along the line connecting the centers of absorbers are shown in Fig. 6. The time window was set between 1800 and 2400ps , where the reflected light intensity is at its maximum at a depth of 20mm (see Fig. 8). One can clearly notice the two separate peaks when the distance of two absorbers is longer than 20mm . However, two absorbers could be discriminated from a two-dimensional image even when the gap was 15mm .

Fig. 6

Profiles of the reflected light intensity along the line connecting centers of the absorbers. Inset is an illustration of two absorbers placed in the phantom at the depth of 20mm . Wavelength of 805nm was used, and the time window of noncontact B-TRIS was 600ps (between 1800 and 2400ps ).

054006_1_032605jbo6.jpg

3.4.

Sensitivity Depending on the Sample Position

The uniformity of the sensitivity throughout the target area obtained by noncontact B-TRIS is shown in Fig. 7 . We measured peak intensity changes in backscattered light from a single absorber that moved across the field of view, as shown with arrows (trace A and B) in the inset of Fig. 7, where the illumination positions are shown with closed circles. The sensitivity of B-TRIS is relatively flat near the center and it gradually decreases as the target position approaches the edge of the target area. Furthermore, the position of the absorber measured by B-TRIS coincided the actual position with accuracy <2mm as described before.

Fig. 7

Top, illustration of illumination (black circles) and detection (white plane) positions of B-TRIS, and traces of absorber (A and B). Bottom, changes in reflected light intensity accompanying the movement of absorber. Experimental conditions were the same as those of Fig. 5.

054006_1_032605jbo7.jpg

3.5.

Depth Measurement

Normalized temporal changes in light intensity depending on the depth of the absorbing samples are shown in Fig. 8 . As the depth of the absorber increased, the backscattered light path and the time of flight also increased. The peak value of the backscattered light intensity decreases exponentially and the time-of-flight profile is broadened with increasing absorber depth. This result indicates that the present noncontact B-TRIS provides not only the two-dimensional distribution of the absorbers but also their depth information.

4.

Discussion

Time-resolved optical imaging methods are powerful for extracting not only temporal but also spatial profiles of target objects. There have been many reports applying time-resolved methods to capture ballistic or banana-shaped propagated light through scattering media by discriminating the specific component of scattered light. As described before, advantages of the time-resolved optical measurement system have been demonstrated by many fundamental experiments and human applications: the higher resolving power of optical tomographic schemes using fast-transmitted photons of ultrashort pulses,27 better depth sensitivity than continuous wave systems,20, 21, 32 and the ability to acquire an effective path length using the time-of-flight value.28 Recently, non-contact imaging systems using near-infrared light and fluorescent probe have been developed to detect cancer tissues.22, 24, 26 In the present study, we have developed a noncontact reflective-mode time- resolved imaging system to visualize light absorbing particles placed inside the scattering medium. Our aim is focused on future applications of this system to detect intrinsic signals, such as oxygenated and deoxygenated hemoglobin, for functional mapping in the human head. At present, however, there are several problems to be solved for practical application to functional human brain mapping, such as (1) movement of subject is restricted within a focus area of the detecting objective lens, and (2) the hair is highly critical for injection and detection of light, thus the measurement area is restricted to the hairless region such as the forehead. However, one important merit of this system was that we could actually visualize the shadow of each absorbing particle right above its location, without any interpolation or complicated calculations. Therefore, by extracting a suitable portion of the photons using the time-resolved method, local changes in absorption at a deeper region can be projected perpendicularly onto the surface by taking the shortest path through the scattered medium. This would reduce the annoyance of remapping and interpolation of the data obtained using the source-detector (S-D) imaging system. Although S-D imaging systems are currently used to detect stimulus-induced brain activation in human subjects, the system provides accurate data only when the activation area is larger than the optical pixel size. It is restricted by the S-D distance as well as their positioning.

Another merit of the time-resolved method is the ability to map time-of-flight data of the target sample. Time-of-flight profiles would allow us to know the depth information of the object and allow us to determine absolute concentration changes of absorber such as hemoglobin.

5.

Conclusion

We reported the construction of noncontact B-TRIS and its application to in vitro experiments. Improvement of spatial resolution and detection of absorbing materials buried in the scattering medium in the reflection mode was shown. Further fundamental experiments using more practical models are required, because the brain has layered structures, which are more complicated than plain phantoms, and blood vessels running near the surface would make interpretation of the data more complicated. We, however, hope that the B-TRIS will become a powerful tool for clinical and cognitive neuroscience research in human subjects.

Acknowledgments

This research was supported by the Breakthrough 21 Project of the Ministry of Posts and Telecommunications of Japan, and in part by Grants-in-Aid from the Nissan Science Foundation of Japan. The authors thank Dr. H. Eda for helpful comments and suggestions.

References

1. 

M. Moosmann, P. Ritter, I. Krastel, A. Brink, S. Thees, F. Blankenburg, B. Taskin, H. Obrig, and A. Villringer, “Correlates of alpha rhythm in functional magnetic resonance imaging and near infrared spectroscopy,” Neuroimage, 20 145 –158 (2003). 1053-8119 Google Scholar

2. 

B. M. Mackert, G. Wubbeler, S. Leistner, K. Uludag, H. Obrig, A. Villringer, L. Trahms, and G. Curio, “Neurovascular coupling analyzed non-invasively in the human brain,” NeuroReport, 19 63 –66 (2004). 0959-4965 Google Scholar

3. 

A. Seiyama, J. Seki, H. C. Tanabe, I. Sase, A. Takatsuki, S. Miyauchi, H. Eda, S. Hayashi, T. Imaruoka, T. Iwakura, and T. Yanagida, “Circulatory basis of fMRI signals: Relationship between changes in the hemodynamic parameters and BOLD signal intensity,” Neuroimage, 21 1204 –1214 (2004). 1053-8119 Google Scholar

4. 

K. Uludag, J. Steinbrink, M. Kohl-Bareis, R. Wenzel, A. Villringer, and H. Obrig, “Cytochrome-c-oxidase redox changes during visual stimulation measured by near-infrared spectroscopy cannot be explained by a mere cross talk artefact,” Neuroimage, 22 109 –119 (2004). 1053-8119 Google Scholar

5. 

I. Miyai, H. C. Tanabe, I. Sase, H. Eda, I. Oda, I. Konishi, Y. Tsunazawa, T. Suzuki, T. Yanagida, and K. Kubota, “Cortical mapping of gait in humans: A near-infrared spectroscopic topography study,” Neuroimage, 14 1186 –1192 (2001). 1053-8119 Google Scholar

6. 

T. Kusaka, K. Kawada, K. Okubo, K. Nagano, M. Namba, H. Okada, T. Imai, K. Isobe, and S. Itoh, “Noninvasive optical imaging in the visual cortex in young infants,” Hum. Brain Mapp, 22 122 –132 (2004). 1065-9471 Google Scholar

7. 

T. Higashi, Y. Sone, K. Ogawa, Y. T. Kitamura, K. Saiki, S. Sagawa, T. Yanagida, and A. Seiyama, “Changes in regional cerebral blood volume in frontal cortex during mental work with and without caffeine intake: Functional monitoring using near-infrared spectroscopy,” J. Biomed. Opt., 9 788 –793 (2004). https://doi.org/10.1117/1.1755233 1083-3668 Google Scholar

8. 

M. N. J. Herrmann, A. C. Ehlis, and A. J. Fallgatter, “Frontal activation during a verbal-fluency task as measured by near-infrared spectroscopy,” Brain Res. Bull., 30 51 –56 (2003). 0361-9230 Google Scholar

9. 

T. Suto, M. Fukuda, M. Ito, T. Uehara, and M. Mikuni, “Multichannel near-infrared spectroscopy in depression and schizophrenia: Cognitive brain activation study,” Biol. Psychiatry, 55 501 –511 (2004). 0006-3223 Google Scholar

10. 

A. Watanabe and T. Kato, “Cerebrovascular response to cognitive tasks in patients with schizophrenia measured by near-infrared spectroscopy,” Schizophr Bull., 30 435 –444 (2004). 0586-7614 Google Scholar

11. 

J. Steinbrink, M. Kohl, H. Obrig, G. Curio, F. Syre, F. Thomas, H. Wabnitz, H. Rinneberg, and A. Villringer, “Somatosensory evoled fast optical intensity changes detected noninvasively in the adult human head,” Neurosci. Lett., 291 105 –108 (2000). 0304-3940 Google Scholar

12. 

M. Wolf, U. Wolf, J. H. Choi, R. Gupta, L. P. Safonova, L. A. Paunescu, A. Michalos, and E. Gratton, “Functional frequency-domain near-infrared spectroscopy detects fast neuronal signal in the motor cortex,” Neuroimage, 17 1868 –1875 (2002). 1053-8119 Google Scholar

13. 

M. A. Franceschini and D. A. Boas, “Noninvasive measurement of neuronal activity with near-infrared optical imaging,” Neuroimage, 21 372 –386 (2004). 1053-8119 Google Scholar

14. 

D. A. Boas, T. Gaudette, G. Strangman, X. Cheng, J. J. A. Marota, and J. B. Mandeville, “The accuracy of near infrared spectroscopy and imaging during focal changes in cerebral hemodynamic,” Neuroimage, 13 76 –90 (2001). 1053-8119 Google Scholar

15. 

H. Obrig and A. Villringer, “Beyond the visible—Imaging the human brain with light,” J. Cereb. Blood Flow Metab., 23 1 –18 (2003). https://doi.org/10.1097/00004647-200301000-00001 0271-678X Google Scholar

16. 

E. Paulesu, R. S. J. Frackowiak, and G. Bottini, “Maps of somatosensory systems,” Human Brain Function, 183 –242 Academic Press, San Diego (1997). Google Scholar

17. 

P. Petrovic, K. M. Petersson, P. Hansson, and M. Ingvar, “A regression analysis study of the primary somatosensory cortex during pain,” Neuroimage, 16 1142 –1150 (2002). 1053-8119 Google Scholar

18. 

J. T. E. Richardson, Imagery, Psychology Press, London (1999). Google Scholar

19. 

A. D. Boas, A. M. Dale, and M. A. Franceschini, “Diffuse optical imaging of brain activation: Approaches to optimizing image sensitivity, resolution, and accuracy,” Neuroimage, 23 5275 –5288 (2004). 1053-8119 Google Scholar

20. 

J. Selb, J. J. Stott, M. A. Franceschini, A. G. Sorenson, and D. A. Boas, “Improved sensitivity to cerebral dynamics during brain activation with a time-gated optical system: Analytical model and experimental validation,” J. Biomed. Opt., 10 (1), 011013 (2005). https://doi.org/10.1117/1.1852553 1083-3668 Google Scholar

21. 

B. Montcel, R. Chabrier, and P. Poulet, “Detection of cortical activation with timeresolved diffuse optical methods,” Appl. Opt., 44 1942 –1947 (2005). https://doi.org/10.1364/AO.44.001942 0003-6935 Google Scholar

22. 

A. B. Thompson, D. J. Hawrysz, and E. M. Sevick-Muraca, “Near-infrared contrast-enhanced imaging with area illumination and area detection: The forward imaging problem,” Appl. Opt., 42 4125 –4136 (2003). 0003-6935 Google Scholar

23. 

R. B. Schulz, J. Ripoll, and V. Ntziachristos, “Noncontact optical tomography of turbid media,” Opt. Lett., 28 1701 –1703 (2003). 0146-9592 Google Scholar

24. 

R. B. Schulz, J. Ripoll, and V. Ntziachristos, “Experimental fluorescence tomography of tissues with noncontact measurements,” IEEE Trans. Med. Imaging, 23 492 –500 (2004). https://doi.org/10.1109/TMI.2004.825633 0278-0062 Google Scholar

25. 

G. M. Turner, G. Zacharakis, A. Soubret, J. Ripoll, and V. Ntziachristos, “Complete-angle projection diffuse optical tomography by use of early photons,” Opt. Lett., 30 409 –411 (2005). https://doi.org/10.1364/OL.30.000409 0146-9592 Google Scholar

26. 

R. Roy, A. B. Thompson, A. Godavarty, and E. M. Sevick-Muraca, “Tomographic fluorescence imaging in tissue phantoms: A novel reconstruction algorithm and imaging geometry,” IEEE Trans. Med. Imaging, 24 137 –154 (2005). https://doi.org/10.1109/TMI.2004.839359 0278-0062 Google Scholar

27. 

H. Eda, I. Oda, Y. Ito, Y. Wada, Y. Oikawa, Y. Tsunazawa, M. Takada, Y. Tsuchiya, Y. Yamashita, M. Oda, A. Sassaroli, Y. Yamada, and M. Tamura, “Multi-channel time-resolved optical tomographic imaging system,” Rev. Sci. Instrum., 70 3595 –3602 (1999). https://doi.org/10.1063/1.1149965 0034-6748 Google Scholar

28. 

D. T. Delpy, M. Cope, P. van der Zee, S. Arridge, S. Wray, and J. Wyatt, “Estimation of optical pathlength through tissue from direct time flight measurement,” Phys. Med. Biol., 33 1433 –1442 (1988). https://doi.org/10.1088/0031-9155/33/12/008 0031-9155 Google Scholar

29. 

M. S. Patterson, B. Chance, and B. C. Wilson, “Time resolved reflectance and transmittance for the non-invasive measurement of tissue optical properties,” Appl. Opt., 28 2331 –2336 (1989). 0003-6935 Google Scholar

30. 

M. Oda, Y. Yamashita, T. Nakano, A. Suzuki, K. Shimizu, I. Hirano, F. Shimomura, E. Ohmae, T. Suzuki, and Y. Tsuchiya, “Nearinfrared time-resolved spectroscopy system for tissue oxygenation monitor,” Proc. SPIE, 3597 611 –616 (1999). https://doi.org/10.1117/12.356809 0277-786X Google Scholar

31. 

F. Bevilacqua, D. Piguet, P. Marquet, J. D. Gross, B. J. Tromberg, and C. Depeursinge, “In vivo local determination of tissue optical properties: Applications to human brain,” Appl. Opt., 38 4939 –4950 (1999). 0003-6935 Google Scholar

32. 

A. Liebert, H. Wabnitz, J. Steinbrink, H. Obrig, M. Moller, R. Macdonald, A. Villringer, and H. Rinneberg, “Time-resolved multidistance near-infrared spectroscopy of the adult head: Intracerebral and extracerebral absorption changes from moments of distribution of time of flight of photons,” Appl. Opt., 43 3037 –3047 (2004). https://doi.org/10.1364/AO.43.003037 0003-6935 Google Scholar
©(2006) Society of Photo-Optical Instrumentation Engineers (SPIE)
Ichiro Sase, Akira Takatsuki, Junji Seki, Toshio Yanagida, and Akitoshi Seiyama "Noncontact backscatter-mode near-infrared time-resolved imaging system: preliminary study for functional brain mapping," Journal of Biomedical Optics 11(5), 054006 (1 September 2006). https://doi.org/10.1117/1.2363359
Published: 1 September 2006
Lens.org Logo
CITATIONS
Cited by 17 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Imaging systems

Picosecond phenomena

Brain mapping

Pulsed laser operation

Brain

Optical fibers

Neodymium

Back to Top