1.Introduction1.1.BackgroundIt is generally accepted that the majority of acute coronary events is precipitated by the rupture of a vulnerable atherosclerotic plaque in the coronary system, and subsequent thrombogenesis.1, 2, 3 The thin-cap fibroatheroma is currently hypothesized to be the most likely class of arterial wall pathology to constitute a vulnerable plaque.4, 5 The key to plaque vulnerability is still elusive,6 even though recent technological advances in intravascular imaging technology have enabled the collection of a wealth of data on unstable atherosclerosis in all its stages of development,7 both in clinical and in ex vivo settings. It appears very likely that combined information on physiological, anatomical, chemical, and mechanical parameters8, 9, 10 is required for a reliable assessment of the proneness of a specific lesion to rupture. Some of these parameters may be accessible through intravascular imaging methods.11, 12, 13, 14, 15, 16, 17, 18, 19 In addition, plaque type and morphology prior to intervention influence the long-term procedural outcome significantly.20 Among the parameters that influence plaque vulnerability are the thickness of the fibrous cap overlying the necrotic core, inflammation, intraplaque hemorrhage, and composition.9, 21 Data on plaque composition and stability can complement the image, and can aid decisions regarding whether a particular section of coronary artery should be treated and how. The advent of optical coherence tomography (OCT) for cardiovascular applications has boosted the resolving power of intravascular imaging to a level where imaging the thin cap of a vulnerable plaque is possible.22 Insight into the physiology of a plaque is complementary to the structural information offered by the OCT grayscale image. In intravascular ultrasound (IVUS) imaging, parametric imaging techniques have been developed that quantify properties of the underlying tissue in the image. Examples of this approach are IVUS elastography, characterizing local elasticity,11, 12, 23, 24 and spectral analysis methods for tissue identification.25, 26, 27 In intravascular OCT, published research into parametric imaging methods so far has been very limited. Recent publications have highlighted the possibility of tissue identification in OCT images. Yabushita 28 published a qualitative image classification scheme, which has become the de facto standard, distinguishing fibrous, calcified, and lipid-rich tissues. A comparison with later work by Kume 29 reveals, however, that the interpretation of OCT images based on qualitative criteria can be ambiguous in some cases. In this study, we attempt to resolve some of the ambiguity associated with qualitative OCT tissue classification by developing a parametric imaging technique that images the optical extinction coefficient (sum of scattering and absorption: ) in atherosclerotic plaques in situ. A few papers have been published recently30, 31, 32 that targeted the optical attenuation coefficient for arterial tissue identification with OCT. These studies were performed on longitudinally sectioned arteries, imaged with a scanning-stage type OCT setup. Xu 33 published a study in which both the backscatter coefficient and the extinction coefficient were measured on transverse sections of paraffin-embedded, intact human coronaries. The latter study can be interpreted as a quantitative basis for the qualitative scheme of Yabushita 28 For this work, we used an intracoronary, catheter-based OCT system, enabling us to image the optical attenuation in situ and even in vivo. 1.2.OCT Signal ModelThe expectation value of the detected OCT signal can be modeled using a single scattering model that incorporates the axial point spread function due to the coupling of the emitted and backscattered intensity to the Gaussian mode profile of a single-mode fiber34 (SMF). In this equation, the OCT signal is modeled by a Lambert-Beer exponential decay curve, multiplied by the coupling factor. The expectation value operator denotes the ensemble average over many realizations of the speckle generated by the beam in the tissue. Here is the longitudinal point spread function (PSF) for an SMF-based OCT system, governed by the position of the beam waist and the Rayleigh length . For analysis of Fourier-domain OCT (FD-OCT), Eq. 1 models the amplitude reflectivity.In FD-OCT, one must also account for the spectral coherence of the source.35 In this study, we use a swept-source system for in vivo imaging. A Gaussian coherence function can be assumed in the frequency domain during the linear scan. The finite width of introduces an extra factor (the Fourier transform of ), describing the signal roll-off with depth. In Eq. 3, is the center of the scan, which has been moved to positive by frequency shifting,36 and is the half width of the roll-off function. Our object is the attenuation coefficient , appearing in the exponent of Eq. 1. The scale factor is the locally available intensity multiplied by the backscattering coefficient , . The backscatter efficiency is a tissue property that can be measured independently in homogeneous media.33 The local intensity is equal to the source intensity incident on the vessel wall diminished by the attenuation along the path from the lumen border to the imaged depth, . As the OCT image traces provide only one intensity measurement for every depth, it is not possible to extract two depth-dependent parameters and from the analysis. Equation 1 describes the OCT signal in a homogeneous medium. The vessel wall, and biological tissue in general, are heterogeneous structures. An OCT A-line usually samples more than one tissue type. Hence, the signal intensity must be fitted in windows of a length that is unknown a priori.31 Here, variations in may confound the analysis of . In addition, the presence of speckle means that the signal-to-noise ratio (SNR) in the image is inherently low,37, 38 and several frames must be averaged to reduce speckle. Averaging is complicated by artifacts due to cardiac motion, and by nonuniform catheter rotation, which can usually be corrected.39, 40 In this paper, we present a method for intracoronary imaging of the optical extinction coefficient . The principle is demonstrated in human coronary arteries ex vivo. The optical attenuation maps are compared with histopathology to identify the correspondence between tissue type and extinction coefficient. After validation, we recorded suitable intracoronary OCT sequence in patients and applied the algorithm to the data. Analysis was successful and the resulting optical attenuation maps both corroborated and complemented the qualitative interpretations of plaque type. 2.Materials and Methods2.1.Coronary Artery SpecimensWe examined 14 human coronary arteries, acquired during autopsy at the Department of Pathology of the Erasmus Medical Center (MC), from 14 human hearts (57% men, 12 left anterior descending coronary arteries, 2 right coronary arteries, mean age 64). Inclusion criteria were age and noncardiac death. Permission to use autopsy material for scientific study was obtained from the relatives; this study was approved by the Medical Ethics Committee of the Erasmus MC. The artery segments were excised from the heart and varied in length between 5 and . In these vessels, 65 sites were identified (up to 5 per segment) for imaging and further histologic processing. All vessels except one were observed to be atherosclerotic based on gross pathology. During excision of the arteries, side branches were closed with sutures. OCT imaging was performed within postmortem. 2.2.Imaging Protocol In VitroOCT imaging was performed with an M2CV time-domain OCT system, and ImageWire 200 catheters (Lightlab Imaging Inc., Westford, Massachusetts). A vessel cross section was imaged by an IR light beam, central wavelength , that was swept along the vessel wall by the rotating fiber inside the catheter sheath. The OCT system had an axial resolution of and a tangential resolution of in the focus. The imaging depth was . Each frame consisted of , corresponding to per pixel. Imaging was performed at a rate of (fps). The artery under investigation was placed in a water bath filled with saline solution32 at . Both ends were mounted on plastic sheaths, and the vessel was pressurized to using a water column system filled with saline. The vessel was inspected longitudinally by moving the imaging catheter along the lumen. Sites of interest were marked with a needle, visible in the OCT images. At these sites, a stationary sequence of 40 to 50 frames was recorded and stored in polar coordinates in stacked TIFF (tagged image file format). After removal from the water tank, the needles were replaced by a color-coded suture. 2.3.HistopathologyAfter imaging, the artery sections were pressure fixed at in formaldehyde for at room temperature, and subsequently stored in formaldehyde at for further processing. Vessels were partially decalcified for in formic acid.41 After fixation and decalcification, sutures marking the imaged cross sections were replaced by ink dots. The tissue was embedded in paraffin and serially sectioned for histological staining. Each imaged cross section was stained with hematoxylin-eosin (H&E), picrosirius red, elastic van Gieson (EvG), and immunohistochemical stain CD68 for macrophage identification. Histological cross sections were characterized by two pathologists blinded for the OCT results. Characterization was done by making a map of all the cross sections, called cartoon histology, with color coding for different types of tissue. We distinguished fibrous tissue, including intimal smooth muscle cells (SMCs), necrotic core in early or advanced stage, hemorrhage, and dense calcium. In case of disagreement, the two pathologists reevaluated the slides and reached a consensus diagnosis. A summary of the color coding used in the cartoon histology is given in Table 1 . Table 1Color coding used for tissue type in the cartoon histology.
2.4.Imaging Protocol In VivoQuantitative analysis of intracoronary OCT imagery is complicated by motion and other artifacts, resulting from catheter eccentricity or blood remnants in the lumen. The influence of cardiac motion can be mitigated by increasing the imaging speed. The diastolic filling phase, where coronary motion is minimal, lasts about , measured from the end of the T wave in the electrocardiogram (ECG) to the beginning of atrial systole, marked by the start of the P wave. At a frame rate of , each heart cycle will have 15 to 20 frames in which the effects of cardiac motion are limited, and enable averaging of those frames. In vivo imaging was performed in the catheterization laboratory of the Thorax Center of the Erasmus MC (Rotterdam, The Netherlands) using an optical frequency domain imaging (OFDI) system,42, 43 built at the Wellman Center for Photomedicine (Boston, Massachusetts). This swept-source system had a center wavelength of and an axial resolution of . The frame rate was , and the ranging depth in tissue is about . The resulting images were sampled at per frame. Patients undergoing percutaneous coronary intervention (PCI) were enrolled in this study, having given informed consent. In the OFDI pullback and x-ray angiography data, lesions of interest were identified for optical attenuation imaging. The catheter imaging tip was then positioned at such a site, and a stationary (i.e., no pullback) sequence was recorded. The blood was cleared from the artery by flushing with 100% iodixanol 370 (Visipaque ™, GE Healthcare, Cork, Ireland) at , either manually or using an injection pump (Mark-V Pro Vis, Medrad Inc., Indianola, Pennsylvania) at . Recording was started as soon as clearing was achieved, and stopped after about five heart cycles; flushing was also stopped at this time. An ECG was recorded synchronously with the OCT acquisition at a sampling rate equal to the A-line frequency . In the acquired data, we selected the frames corresponding to the cardiac phase between the T and P waves for further analysis. The imaged plaques were classified2, 28 by two experienced OCT readers for comparison with the assessment based on optical attenuation. 2.5.Catheter CharacterizationIndividual catheters of the type used in this study may have nonidentical focal length and Rayleigh length . It is essential to characterize the catheters before attempting to extract quantitative data from the OCT image. Catheter beam parameters were determined by recording the OCT signal in a series of Intralipid (Fresenius Kabi, ’s-Hertogenbosch, The Netherlands) dilutions (base solution 20% w/w; diluted 1:4 to 1:128) in water. Every recording consisted of about 15000 A-lines. These were averaged to reduce speckle. An -dependent (catheter-independent) intensity offset was measured in water, with negligible attenuation and backscattering. This offset was subtracted from the averaged OCT traces. To reduce the dynamic range of the fitted signal and reduce sensitivity to noise, we linearized Eq. 1 by logarithmic compression: The second term in Eq. 4, representing the catheter-specific PSF, is calculated with Eq. 2 for a matrix of combinations, with and , both with increments, and added to the offset-corrected measured intensity profile. We minimized the root mean square (rms) difference between the fitted line and measurement, compensated for roll-off if applicable, see Eq. 3. All catheters yielded a unique, paraboloid minimum. The values of and at the minimum of the cost function represent the best estimate of the beam parameters.2.6.Data AnalysisAll data analysis was performed in MATLAB R2007b (The Mathworks, Natick, Massachusetts). Frames from the OCT movies were aligned using our nonlinear rotation correction algorithm,39, 40 and then averaged for speckle reduction. This mean OCT image was used as the basis for analysis of the optical attenuation coefficients. If there was evidence of remaining motion artifacts (e.g., blurring in the averaged frame) in the corrected sequence, the data were discarded. The intensity traces were linearized in the manner of Eq. 4, using the beam parameters specific for the catheter that was used for imaging the vessel. Starting at the lumen boundary, the log-compressed data was least-squares fitted with a linear model, in a window of variable length . With a linear optimization, there is a unique optimum for each set of data, and results are independent of the initial guess required for an iterative nonlinear model. Speed is another advantage.We extracted from the data, using a fitting procedure optimized to search for homogeneous tissue regions. The fitting window was extended until an inhomogeneity was encountered that degraded the result. Searching for the longest window maximizes the accuracy of the fitted attenuation coefficient, and reduces sensitivity to intensity fluctuations that result from inhomogeneities. A schematic representing the data flow in the fitting procedure is shown in Fig. 1 . Using an adaptive threshold criterion, the lumen boundary is determined in the mean image. A linear least-squares (lsq) fit is performed in a window of minimum length , corresponding to (px). Results with are discarded. The cost function (fxn) is defined as the root of mean square (rms) difference between the measured OCT trace [Eq. 5] and the model . The window is iteratively extended until a decrease in fit quality is detected. The optimum values (for smallest ) are stored, and the window is moved forward. This procedure is repeated until the window encounters the end of the A-line. All parameters in the fitting procedure were optimized in a simulation study.44 This algorithm results in an estimate of for every A-line. The attenuation coefficients are color coded on a linear scale and plotted in the regular gray-scale OCT image as an overlay or side by side. The in vitro attenuation image is then compared with the cartoon histology to identify a correspondence between tissue type and . As histological tissue slices are much thinner than the accuracy of the needle marker there is an unavoidable sampling error.45, 46, 47 Imaged cross sections that were obviously mismatched with histology, based on anatomical features, were removed from the data set before analysis. 3.Results3.1.In Vitro ImagingWe analyzed OCT data of 39 lesions. Other sites had either too much catheter motion, preventing the averaging of the data, or the match between the OCT data and histology was inaccurate. Results for different lesion types are shown in Figs. 2, 3, 4 . Figure 2 displays data for a lesion containing a large calcification and an advanced necrotic core. Low attenuation is found in areas with fibrous tissue and calcification. The necrotic core exhibits a high attenuation coefficient, and stands out much more clearly in the attenuation image than in the gray-scale OCT. The adventitia also has a high attenuation, but can be distinguished from pathological features by its morphology and location: the high attenuation region occurs outside the media and is circumferential. A concentric fibrous lesion shows up as a homogenous, circumferential, low-attenuation region in Fig. 3. Again, the attenuation coefficient rises in the adventitial region. A more complex plaque type is analyzed in Fig. 4. The OCT image is very heterogeneous, and this is reflected by the morphology in the histology. The cartoon histology shows intraplaque hemorrhage and necrotic core, embedded in a mostly fibrous matrix. The green and orange boxes highlight areas with high attenuation coefficient. The green box contains some necrotic core, which is in accordance with the example in Fig. 2. The orange box, however, is only fibrous tissue and SMC. The CD68 stain reveals strong topical macrophage infiltration in both these areas. Macrophages, being strong optical scatterers, attenuate the OCT beam. An example of intimal xanthoma is shown in Fig. 5 . Macrophages in a lesion that is mostly fibrous otherwise cause strong attenuation. The limited size and composition of our data set does not permit formulation of a statistically sound, quantitative classification scheme for automatic identification of necrotic core and macrophages, or plaque type in general. We can, however, formulate tentative criteria based on the appearance of certain pathologies in our in vitro optical attenuation data. The characteristics we observed are summarized in Table 2 . Table 2Appearance of vessel wall condition in the attenuation image.
3.2.In Vivo ImagingStationary OCT sequences with ECG registration were obtained from nine sites in four patients. Of these, eight sites had OFDI data with small enough catheter motion that the residual shifts could be corrected and the frames could be averaged in at least one heart cycle. Of the imaged plaques, five were classified as fibroatheroma, and three were fibrous lesions. Figure 6 presents optical attenuation imaging of atherosclerotic lesions in vivo in two plaques of different types. The measured optical attenuation reproduces accurately if a single site is imaged repeatedly. In Fig. 7 , we show attenuation data from a necrotic-core coronary lesion in vivo, imaged over three consecutive heart cycles. The features observed in the plaque, as well as the absolute values of the attenuation coefficient, appear very similar in the different images. A quantitative estimate of the reproducibility is not straightforward as the data from different heart cycles are not exactly aligned angularly and radially. The characteristics noted in Table 2 were also observed in vivo, as evidenced by Fig. 6. Plaque identification by application of these criteria agreed with the interpretation by two OCT readers for seven out of eight cases. Interpretation of the last one (a fibrous plaque by gray scale; significant necrotic core according to optical attenuation) was complicated by inhomogeneity in both the gray-scale and attenuation images. 4.Discussion4.1.Study ResultsWe compared the optical attenuation coefficient, measured by OCT, with the composition of atherosclerotic lesions, based on histology. High optical attenuation— —is associated with two important markers of atherosclerotic plaque vulnerability: necrotic core and macrophage infiltration. Other common plaque tissue types, fibrous and calcific, were found to have a low attenuation coefficient— . The in vitro measurements were done on intact excised human coronary arteries, with a commercially available, clinically approved OCT system and catheters. We demonstrated that the analysis can be performed, with similar results, on patient data. The application in interventional procedures is easy, quick, and does not cause any patient discomfort. It requires the recording of a stationary sequence of about five heart cycles at a spot of interest, and about of extra flush medium for clearing. Automated, intraprocedural, real-time attenuation analysis should be possible, if the OCT catheters are well-characterized before imaging. The only steps that currently require user attention are the selection of end-diastolic frames for processing, and a check to ensure the lumen contour detection did not produce any errors. Data processing is computationally light: analysis of a cross section takes (depending on frame size) in MATLAB on an office PC, without any custom optimization. This study is the first to report in situ tissue typing of atherosclerotic lesions in intact human coronary arteries by means of the optical attenuation coefficient. It is also the first to demonstrate the in vivo feasibility of such an approach, and to assess the diagnostic potential of this parameter for plaque type in a clinical setting. One of the main difficulties in OCT-based plaque characterization is the distinction between necrotic core and calcified lesions, as the difference is only in border definition.28, 29 In optical attenuation, there is a marked contrast between the two plaque components. The analyzed attenuation coefficients for the different tissue types are similar in vivo compared to in vitro. Given that the two data sets were acquired with fundamentally different OCT systems, and different catheters, this inspires confidence in the data. It also suggests that the tissue optical properties did not change significantly after excision of the arteries at autopsy and storage in saline during the experiment. 4.2.Comparison with LiteratureOur results are in general agreement with those presented by Xu 33 Using a benchtop OCT system, they measured low for fibrous and calcific tissue and high for necrotic core. Macrophage infiltration was not identified in that study. The scale is also comparable, taking uncertainty margins into account and reading the “lipid” category of Ref. 33 as our category “necrotic:” necrotic core has . Xu report a value of the fibrous and calcific, , that is slightly larger than the one we found . This may be a result of the fixation procedure applied in Ref. 33. Other literature reports both higher30 and lower31 values for necrotic core attenuation. Our data do confirm the observation33 that lipid-rich tissue attenuates more strongly than other atherosclerotic tissue types, or healthy vascular tissue. 4.3.Image Artifacts and Study LimitationsWe were unable to determine the backscatter coefficient from our data because of tissue heterogeneity and the lack of an intensity calibration ( in Sec. 1.2) that is constant for all lines and frames. We put in a significant effort to reduce the sensitivity of the fitting algorithm to sudden steps in intensity that result from a drop in at a sharp interface between different tissues. This approach has been successful, by and large, but evidence of artifacts, resulting from changes in being interpreted as large , remains in the data. An example can be seen in Fig. 2, where fibrous and calcified tissue are adjacent. Both fibrous and calcified tissue have a low , but the transition from high (fibrous) to low (calcified) produces a spot of high fitted near 3 o’clock. This illustrates that the attenuation display should always be interpreted in conjunction with the gray-scale OCT image. The model we used for analysis is a single-scattering model, assuming straight light paths to and from the scattering site where backreflection occurs, and not incorporating interfaces between different tissues. In reality, specular interface reflections at the boundary between different tissues—with differing refractive indices32—may produce a strong signal locally, which cannot be accounted for by Eq. 1. Likewise, there is evidence of multiple scattering, causing a higher signal than predicted by the model, depending on beam parameters and tissue optical properties.48, 49 More sophisticated optical modeling50 could reduce these artifacts, but will substantially complicate the analysis and increase dependence on parameter guesses. Two aspects of the comparison with histology merit discussion. The first is in-plane (as opposed to longitudinal, pointed out in Sec. 2.6) image registration: the histologic processing inevitably deforms the tissue. This complicates automated correlation and one-to-one mapping between tissue type and optical attenuation. Second, density variations are apparent in the histology slides, particularly in fibrous tissue. Certain areas seem to be more tightly packed than others. While in some cases this may be an artifact due to histologic processing, these regions can often be identified in the OCT images as well, thus representing real tissue inhomogeneities. They will undoubtedly reflect in the attenuation coefficient. Our standard—tissue type, color coded in the cartoon histology—does not represent these density variations. This is a source of variation that our analysis cannot account for. Likewise, for the in vivo data, the plaque classification is more coarse-grained than our image analysis. It qualifies an entire cross section as a certain plaque type and cannot account for possible variability within a plaque that will be apparent in our data. Routine clinical use of the tissue characterization method is limited in its present implementation by the frame averaging used for speckle reduction. Multiple frame processing is incompatible with imaging of a coronary artery in a pullback, where each frame samples a different vascular cross section. The required amount of speckle reduction depends on the minimum acceptable SNR for the fit results to be meaningful. An unpolarized OCT image theoretically has37 ; we observe in our in vitro data and in vivo. Averaging independent speckles increases the SNR as . We found heuristically that, with the present algorithm, about 30 frames are required in the in vitro data, while 15 to 20 frames suffice in vivo. The difference may be attributed to residual cardiac motion, assisting speckle decorrelation between frames. In the future, we will investigate more sophisticated speckle reduction and data processing techniques, such as frequency compounding in FD-OCT data or digital filtering,51, 52, 53 to enable comprehensive imaging of the optical attenuation in a coronary artery. 5.ConclusionsWe developed a framework for imaging the optical attenuation coefficient in human coronary arteries using intravascular OCT. In vitro studies demonstrate that is a suitable parameter to distinguish between different tissue types in atherosclerotic lesions. Two markers of plaque vulnerability, the presence of necrotic core and macrophage infiltration, can be identified as areas of high attenuation . Fibrous and calcified tissue have . The analysis was successfully applied to OCT images acquired in patients. Optical-attenuation-based plaque classification, using the criteria developed with the in vitro analysis, corroborated the plaque classification based on the gray-scale OCT. ReferencesE. Falk,
P. K. Shah, and
V. Fuster,
“Coronary plaque disruption,”
Circulation, 92
(3), 657
–671
(1995). 0009-7322 Google Scholar
R. Virmani,
F. D. Kolodgie,
A. P. Burke,
A. Farb, and
S. M. Schwartz,
“Lessons from sudden coronary death—a comprehensive morphological classification scheme for atherosclerotic lesions,”
Arterioscler., Thromb., Vasc. Biol., 20
(5), 1262
–1275
(2000). 1079-5642 Google Scholar
E. Falk,
“Plaque rupture with severe pre-existing stenosis precipitating coronary thrombosis: characteristics of coronary atherosclerotic plaques underlying fatal occlusive thrombi,”
Br. Heart J., 50
(2), 127
–134
(1983). https://doi.org/10.1136/hrt.50.2.127 0007-0769 Google Scholar
F. D. Kolodgie,
A. P. Burke,
A. Farb,
H. K. Gold,
J. Y. Yuan,
J. Narula,
A. V. Finn, and
R. Virmani,
“The thin-cap fibroatheroma, a type of vulnerable plaque—the major precursor lesion to acute coronary syndromes,”
Curr. Opin. Cardiol., 16
(5), 285
–292
(2001). https://doi.org/10.1097/00001573-200109000-00006 0268-4705 Google Scholar
J. A. Schaar,
J. E. Muller,
E. Falk,
R. Virmani,
V. Fuster,
P. W. Serruys,
A. Colombo,
C. Stefanadis,
W. S. Casscells,
P. R. Moreno,
A. Maseri, and
A. F. W. van der Steen,
“Terminology for high-risk and vulnerable coronary artery plaques,”
Eur. Heart J., 25
(12), 1077
–1082
(2004). https://doi.org/10.1016/j.ehj.2004.01.002 0195-668X Google Scholar
P. W. Serruys,
H. M. Garcia-Garcia, and
E. Regar,
“From postmortem characterization to the in vivo detection of thin-capped fibroatheromas: the missing link toward percutaneous treatment: what if Diogenes would have found what he was looking for?,”
J. Am. Coll. Cardiol., 50
(10), 950
–952
(2007). https://doi.org/10.1016/j.jacc.2007.04.085 0735-1097 Google Scholar
S. Waxman,
F. Ishibashi, and
J. E. Muller,
“Detection and treatment of vulnerable plaques and vulnerable patients—novel approaches to prevention of coronary events,”
Circulation, 114
(22), 2390
–2411
(2006). https://doi.org/10.1161/CIRCULATIONAHA.105.540013 0009-7322 Google Scholar
P. D. Richardson,
M. J. Davies, and
G. V. Born,
“Influence of plaque configuration and stress distribution on fissuring of coronary atherosclerotic plaques,”
Lancet, 334
(8669), 941
–944
(1989). https://doi.org/10.1016/S0140-6736(89)90953-7 0140-6736 Google Scholar
R. Virmani,
F. D. Kolodgie,
A. P. Burke,
A. V. Finn,
H. K. Gold,
T. N. Tulenko,
S. P. Wrenn, and
J. Narula,
“Atherosclerotic plaque progression and vulnerability to rupture—angiogenesis as a source of intraplaque hemorrhage,”
Arterioscler., Thromb., Vasc. Biol., 25
(10), 2054
–2061
(2005). https://doi.org/10.1161/01.ATV.0000178991.71605.18 1079-5642 Google Scholar
E. Falk,
“Pathogenesis of atherosclerosis,”
J. Am. Coll. Cardiol., 47
(8), C7
–C12
(2006). https://doi.org/10.1016/j.jacc.2005.09.068 0735-1097 Google Scholar
J. A. Schaar,
C. L. de Korte,
F. Mastik,
C. Strijder,
G. Pasterkamp,
E. Boersma,
P. W. Serruys, and
A. F. W. van der Steen,
“Characterizing vulnerable plaque features with intravascular elastography,”
Circulation, 108
(21), 2636
–2641
(2003). https://doi.org/10.1161/01.CIR.0000097067.96619.1F 0009-7322 Google Scholar
C. L. de Korte,
M. J. Sierevogel,
F. Mastik,
C. Strijder,
J. A. Schaar,
E. Velema,
G. Pasterkamp,
P. W. Serruys, and
A. F. W. van der Steen,
“Identification of atherosclerotic plaque components with intravascular ultrasound elastography in vivo: a Yucatan pig study,”
Circulation, 105
(14), 1627
–1630
(2002). https://doi.org/10.1161/01.CIR.0000014988.66572.2E 0009-7322 Google Scholar
D. E. Goertz,
M. E. Frijlink,
D. Tempel,
V. Bhagwandas,
A. Gisolf,
R. Krams,
N. de Jong, and
A. F. W. van der Steen,
“Subharmonic contrast intravascular ultrasound for vasa vasorum imaging,”
Ultrasound Med. Biol., 33
(12), 1859
–1872
(2007). https://doi.org/10.1016/j.ultrasmedbio.2007.05.023 0301-5629 Google Scholar
D. E. Goertz,
M. E. Frijlink,
D. Tempel,
L. C. A. van Damme,
R. Krams,
J. A. Schaar,
F. J. ten Cate,
P. W. Serruys,
N. de Jong, and
A. F. W. van der Steen,
“Contrast harmonic intravascular ultrasound—a feasibility study for vasa vasorum imaging,”
Invest. Radiol., 41
(8), 631
–638
(2006). https://doi.org/10.1097/01.rli.0000229773.11715.da 0020-9996 Google Scholar
P. R. Moreno,
R. A. Lodder,
K. R. Purushothaman,
W. E. Charash,
W. N. O’Connor, and
J. E. Muller,
“Detection of lipid pool, thin fibrous cap, and inflammatory cells in human aortic atherosclerotic plaques by near-infrared spectroscopy,”
Circulation, 105
(8), 923
–927
(2002). https://doi.org/10.1161/hc0802.104291 0009-7322 Google Scholar
B. D. MacNeill,
I. K. Jang,
B. E. Bouma,
N. Iftimia,
M. Takano,
H. Yabushita,
M. Shishkov,
C. R. Kaufman,
S. L. Houser,
H. T. Aretz,
D. DeJoseph,
E. F. Halpern, and
G. J. Tearney,
“Focal and multi-focal plaque distributions in patients with macrophage acute and stable presentations of coronary artery disease,”
J. Am. Coll. Cardiol., 44
(5), 972
–979
(2004). https://doi.org/10.1016/j.jacc.2004.05.066 0735-1097 Google Scholar
M. Takano,
I. K. Jang,
S. Inami,
M. Yamamoto,
D. Murakami,
K. Okamatsu,
K. Seimiya,
T. Ohba, and
K. Mizuno,
“In vivo comparison of optical coherence tomography and angioscopy for the evaluation of coronary plaque characteristics,”
Am. J. Cardiol., 101
(4), 471
–476
(2008). https://doi.org/10.1016/j.amjcard.2007.09.106 0002-9149 Google Scholar
O. C. Raffel,
F. M. Merchant,
G. J. Tearney,
S. Chia,
D. D. Gauthier,
E. Pomerantsev,
K. Mizuno,
B. E. Bouma, and
I.-K. Jang,
“In vivo association between positive coronary artery remodelling and coronary plaque characteristics assessed by intravascular optical coherence tomography,”
Eur. Heart J., 29
(14), 1721
–1728
(2008). https://doi.org/10.1093/eurheartj/ehn286 0195-668X Google Scholar
P. Barlis,
P. W. Serruys,
N. Gonzalo,
W. J. van der Giessen,
P. J. de Jaegere, and
E. Regar,
“Assessment of culprit and remote coronary narrowings using optical coherence tomography with long-term outcomes,”
Am. J. Cardiol., 102
(4), 391
–395
(2008). https://doi.org/10.1016/j.amjcard.2008.03.071 0002-9149 Google Scholar
T. Kubo,
T. Imanishi,
H. Kitabata,
A. Kuroi,
S. Ueno,
T. Yamano,
T. Tanimoto,
Y. Matsuo,
T. Masho,
S. Takarada,
A. Tanaka,
N. Nakamura,
M. Mizukoshi,
Y. Tomobuchi, and
T. Akasaka,
“Comparison of vascular response after sirolimus-eluting stent implantation between patients with unstable and stable angina pectoris, a serial optical coherence tomography study,”
J. Am. Coll. Cardiol. Imaging, 1
(4), 475
–484
(2008). https://doi.org/10.1016/j.jcmg.2008.03.012 Google Scholar
R. Virmani,
A. P. Burke,
A. Farb, and
F. D. Kolodgie,
“Pathology of the vulnerable plaque,”
J. Am. Coll. Cardiol., 47
(8), C13
–C18
(2006). https://doi.org/10.1016/j.jacc.2005.10.065 0735-1097 Google Scholar
T. Kubo,
T. Imanishi,
S. Takarada,
A. Kuroi,
S. Ueno,
T. Yamano,
T. Tanimoto,
Y. Matsuo,
T. Masho,
H. Kitabata,
K. Tsuda,
Y. Tomobuchi, and
T. Akasaka,
“Assessment of culprit lesion morphology in acute myocardial infarction—ability of optical coherence tomography compared with intravascular ultrasound and coronary angioscopy,”
J. Am. Coll. Cardiol., 50
(10), 933
–939
(2007). https://doi.org/10.1016/j.jacc.2007.04.082 0735-1097 Google Scholar
B. M. Shapo,
J. R. Crowe,
A. R. Skovoroda,
M. J. Eberle,
N. A. Cohn, and
M. O’Donnell,
“Displacement and strain imaging of coronary arteries with intraluminal ultrasound,”
IEEE Trans. Ultrason. Ferroelectr. Freq. Control, 43
(2), 234
–246
(1996). https://doi.org/10.1109/58.485949 0885-3010 Google Scholar
E. I. Cespedes,
C. L. de Korte, and
A. F. W. van der Steen,
“Intraluminal ultrasonic palpation, assessment of local and cross-sectional tissue stiffness,”
Ultrasound Med. Biol., 26
(3), 385
–396
(2000). https://doi.org/10.1016/S0301-5629(99)00169-6 0301-5629 Google Scholar
A. Nair,
B. D. Kuban,
N. Obuchowski, and
D. G. Vince,
“Assessing spectral algorithms to predict atherosclerotic plaque composition with normalized and raw intravascular ultrasound data,”
Ultrasound Med. Biol., 27
(10), 1319
–1331
(2001). https://doi.org/10.1016/S0301-5629(01)00436-7 0301-5629 Google Scholar
A. Nair,
B. D. Kuban,
E. M. Tuzcu,
P. Schoenhagen,
S. E. Nissen, and
D. G. Vince,
“Coronary plaque classification with intravascular ultrasound radiofrequency data analysis,”
Circulation, 106
(17), 2200
–2206
(2002). https://doi.org/10.1161/01.CIR.0000035654.18341.5E 0009-7322 Google Scholar
A. Katouzian,
S. Sathyanarayana,
B. Baseri,
E. E. Konofagou, and
S. G. Carlier,
“Challenges in atherosclerotic plaque characterization with intravascular ultrasound (IVUS): from data collection to classification,”
IEEE Trans. Inf. Technol. Biomed., 12
(3), 315
–327
(2008). https://doi.org/10.1109/TITB.2007.912352 1089-7771 Google Scholar
H. Yabushita,
B. E. Bouma,
S. L. Houser,
T. Aretz,
I. K. Jang,
K. H. Schlendorf,
C. R. Kauffman,
M. Shishkov,
D. H. Kang,
E. F. Halpern, and
G. J. Tearney,
“Characterization of human atherosclerosis by optical coherence tomography,”
Circulation, 106
(13), 1640
–1645
(2002). https://doi.org/10.1161/01.CIR.0000029927.92825.F6 0009-7322 Google Scholar
T. Kume,
T. Akasaka,
T. Kawamoto,
N. Watanabe,
E. Toyota,
Y. Neishi,
R. Sukmawan,
Y. Sadahira, and
K. Yoshida,
“Assessment of coronary arterial plaque by optical coherence tomography,”
Am. J. Cardiol., 97
(8), 1172
–5
(2006). https://doi.org/10.1016/j.amjcard.2005.11.035 0002-9149 Google Scholar
D. Levitz,
L. Thrane,
M. Frosz,
P. Andersen,
C. Andersen,
J. Valanciunaite,
J. Swartling,
S. Andersson-Engels, and
P. Hansen,
“Determination of optical scattering properties of highly-scattering media in optical coherence tomography images,”
Opt. Express, 12
(2), 249
–259
(2004). https://doi.org/10.1364/OPEX.12.000249 1094-4087 Google Scholar
F. J. van der Meer,
D. J. Faber,
D. M. Baraznji Sassoon,
M. C. Aalders,
G. Pasterkamp, and
T. G. van Leeuwen,
“Localized measurement of optical attenuation coefficients of atherosclerotic plaque constituents by quantitative optical coherence tomography,”
IEEE Trans. Med. Imaging, 24
(10), 1369
–76
(2005). https://doi.org/10.1109/TMI.2005.854297 0278-0062 Google Scholar
F. J. van der Meer,
D. J. Faber,
I. Cilesiz,
M. J. C. van Gemert, and
T. G. van Leeuwen,
“Temperature-dependent optical properties of individual vascular wall components measured by optical coherence tomography,”
J. Biomed. Opt., 11
(4), 041120
(2006). https://doi.org/10.1117/1.2333613 1083-3668 Google Scholar
C. Xu,
J. M. Schmitt,
S. G. Carlier, and
R. Virmani,
“Characterization of atherosclerosis plaques by measuring both backscattering and attenuation coefficients in optical coherence tomography,”
J. Biomed. Opt., 13
(3), 034003
(2008). https://doi.org/10.1117/1.2927464 1083-3668 Google Scholar
T. G. van Leeuwen,
D. J. Faber, and
M. C. Aalders,
“Measurement of the axial point spread function in scattering media using single-mode fiber-based optical coherence tomography,”
IEEE J. Sel. Top. Quantum Electron., 9
(2), 227
–233
(2003). https://doi.org/10.1109/JSTQE.2003.813299 1077-260X Google Scholar
R. Leitgeb,
C. K. Hitzenberger, and
A. F. Fercher,
“Performance of Fourier domain vs. time domain optical coherence tomography,”
Opt. Express, 11
(8), 889
–894
(2003). 1094-4087 Google Scholar
S. H. Yun,
G. J. Tearney,
J. F. de Boer, and
B. E. Bouma,
“Removing the depth-degeneracy in optical frequency domain imaging with frequency shifting,”
Opt. Express, 12
(20), 4822
–4828
(2004). https://doi.org/10.1364/OPEX.12.004822 1094-4087 Google Scholar
J. M. Schmitt,
S. H. Xiang, and
K. M. Yung,
“Speckle in optical coherence tomography,”
J. Biomed. Opt., 4
(1), 95
–105
(1999). https://doi.org/10.1117/1.429925 1083-3668 Google Scholar
J. Rogowska and
M. E. Brezinski,
“Evaluation of the adaptive speckle suppression filter for coronary optical coherence tomography imaging,”
IEEE Trans. Med. Imaging, 19
(12), 1261
–1266
(2000). https://doi.org/10.1109/42.897820 0278-0062 Google Scholar
G. van Soest,
J. G. Bosch, and
A. F. W. van der Steen,
“Azimuthal registration of image sequences affected by nonuniform rotation distortion,”
IEEE Trans. Inf. Technol. Biomed., 12
(3), 348
–355
(2008). https://doi.org/10.1109/TITB.2007.908000 1089-7771 Google Scholar
G. van Soest,
J. G. Bosch, and
A. F. W. van der Steen,
“Alignment of intravascular optical coherence tomography movies affected by non-uniform rotation distortion,”
Proc. SPIE, 6847 684721
–8
(2008). https://doi.org/10.1117/12.761106 0277-786X Google Scholar
G. J. Friedrich,
N. Y. Moes,
V. A. Muhlberger,
C. Gabl,
G. Mikuz,
D. Hausmann,
P. J. Fitzgerald, and
P. G. Yock,
“Detection of intralesional calcium by intracoronary ultrasound depends on the histologic pattern,”
Am. Heart J., 128
(3), 435
–441
(1994). https://doi.org/10.1016/0002-8703(94)90614-9 0002-8703 Google Scholar
S. H. Yun,
G. J. Tearney,
B. J. Vakoc,
M. Shishkov,
W. Y. Oh,
A. E. Desjardins,
M. J. Suter,
R. C. Chan,
J. A. Evans,
I.-K. Jang,
N. S. Nishioka,
J. F. de Boer, and
B. E. Bouma,
“Comprehensive volumetric optical microscopy in vivo,”
Nat. Med., 12
(12), 1429
–1433
(2007). https://doi.org/10.1038/nm1450 1078-8956 Google Scholar
G. J. Tearney,
S. Waxman,
M. Shishkov,
B. J. Vakoc,
M. J. Suter,
M. I. Freilich,
A. E. Desjardins,
W.-Y. Oh,
L. A. Bartlett,
M. Rosenberg, and
B. E. Bouma,
“Three-dimensional coronary artery microscopy by intracoronary optical frequency domain imaging,”
J. Am. Coll. Cardiol. Imaging, 1
(6), 752
–761
(2008). https://doi.org/10.1016/j.jcmg.2008.06.007 Google Scholar
G. van Soest,
T. P. M. Goderie,
S. van Noorden, and
A. F. W. van der Steen,
“Algorithm optimization for quantitative analysis of intravascular optical coherence tomography data,”
Proc. SPIE, 7161 71612C
–8
(2009). 0277-786X Google Scholar
J. Rieber,
O. Meissner,
G. Babaryka,
S. Reim,
M. Oswald,
A. Koenig,
T. M. Schiele,
M. Shapiro,
K. Theisen,
M. F. Reiser,
V. Klauss, and
U. Hoffmann,
“Diagnostic accuracy of optical coherence tomography and intravascular ultrasound for the detection and characterization of atherosclerotic plaque composition in ex-vivo coronary specimens: a comparison with histology,”
Coron Artery Dis., 17
(5), 425
–430
(2006). https://doi.org/10.1097/00019501-200608000-00005 0954-6928 Google Scholar
O. A. Meissner,
J. Rieber,
G. Babaryka,
M. Oswald,
S. Reim,
U. Siebert,
T. Redel,
M. Reiser, and
U. Mueller-Lisse,
“Intravascular optical coherence tomography, comparison with histopathology in atherosclerotic peripheral artery specimens,”
J. Vasc. Interv. Radiol., 17
(2), 343
–349
(2006). 1051-0443 Google Scholar
N. Bruining,
S. Verheye,
M. Knaapen,
P. Somers,
J. Roelandt,
E. Regar,
I. Heller,
S. de Winter,
J. Ligthart,
G. Van Langenhove,
P. J. de Feyter,
P. W. Serruys, and
R. Hamers,
“Three-dimensional and quantitative analysis of atherosclerotic plaque composition by automated differential echogenicty,”
Catheter. Cardiovasc. Interv., 70
(7), 968
–978
(2007). https://doi.org/10.1002/ccd.21310 Google Scholar
J. M. Schmitt,
A. Knüttel,
M. J. Yadlowsky, and
M. A. Eckhaus,
“Optical-coherence tomography of a dense tissue—statistics of attenuation and backscattering,”
Phys. Med. Biol., 39
(10), 1705
–1720
(1994). https://doi.org/10.1088/0031-9155/39/10/013 0031-9155 Google Scholar
M. J. Yadlowsky,
J. M. Schmitt, and
R. F. Bonner,
“Multiple scattering in optical coherence tomography,”
Appl. Opt., 34
(25), 5699
–5707
(1995). https://doi.org/10.1364/AO.34.005699 0003-6935 Google Scholar
L. Thrane,
M. H. Frosz,
T. M. Jørgensen,
A. Tycho,
H. T. Yura, and
P. E. Andersen,
“Extraction of optical scattering parameters and attenuationcompensation in optical coherencetomography images of multilayered tissue structures,”
Opt. Lett., 29
(14), 1641
–1643
(2004). https://doi.org/10.1364/OL.29.001641 0146-9592 Google Scholar
M. Pircher,
E. Gotzinger,
R. Leitgeb,
A. F. Fercher, and
C. K. Hitzenberger,
“Speckle reduction in optical coherence tomography by frequency compounding,”
J. Biomed. Opt., 8
(3), 565
–569
(2003). https://doi.org/10.1117/1.1578087 1083-3668 Google Scholar
D. C. Adler,
T. H. Ko, and
J. G. Fujimoto,
“Speckle reduction in optical coherence tomography images by use of a spatially adaptive wavelet filter,”
Opt. Lett., 29
(24), 2878
–2880
(2004). https://doi.org/10.1364/OL.29.002878 0146-9592 Google Scholar
A. Ozcan,
A. Bilenca,
A. E. Desjardins,
B. E. Bouma, and
G. J. Tearney,
“Speckle reduction in optical coherence tomography images using digital filtering,”
J. Opt. Soc. Am. A, 24
(7), 1901
–1910
(2007). https://doi.org/10.1364/JOSAA.24.001901 0740-3232 Google Scholar
|