25 April 2014 Consideration of dynamic photothermal effect for evaluation of scanning light sources in optical devices using pulsed source criteria
Author Affiliations +
Abstract
Quantitative evaluation of the potential radiation hazards of scanning light sources in medical optical devices is critical. Currently, point scanning light sources of continuous radiation are treated as pulsed sources, where the dwell time at each point is equal to the pulse duration. This study compares the photothermal effects from scanning light and pulsed sources using numerical calculation for scanning without restricting aperture and with various spot sizes. The calculation results show that the thermal damage threshold of scanning source not restricted by measurement aperture does not significantly differ from that of pulsed source. Temporal temperature response and size-dependent photothermal effect also confirm the similarity between scanning and pulsed sources.

1.

Introduction

Many optical medical devices use scanning technology and the number of such devices is increasing. Continuous wave (CW) light sources that are scanned (henceforth called scanning sources) have unique characteristics, which distinguish them from stationary CW or pulsed light sources: scanning sources are spatially dynamic but temporally stationary, whereas pulsed sources are spatially stationary but temporally dynamic. Commonly used guidance documents and safety standards suggest evaluating scanning light sources using pulsed source criteria. Details of pulsed source criteria can be found in guidance documents, for example ISO 15004-2:2007.1 In a previous report using a melanin granule lattice model (MGLM), it was demonstrated that the photothermal effect of scanning source is different from that of pulsed source within restricting aperture.2 However, the difference reported in the previous study was at maximum 25%, which was not large enough to invalidate consensus use of pulsed source criteria for evaluating scanning source. Also, it was noted in another study that dwell-time and overlap of the adjacent exposure points (EPs) must be considered for worst-case scenario analysis.3

A number of questions still remain before it is definitively concluded that photothermal effects from scanning and pulsed sources are comparably equivalent:

  • 1. Previous results comparing photothermal effects from scanning and pulsed sources were confined to the retinal region within a restricting (or measurement) aperture. Actual scanning devices do not operate under such apertures, thus photothermal effects without apertures must be compared.

  • 2. The mechanism for producing a different photothermal response from scanning and pulsed sources is the heat storage and propagation in the tissue. Thus, different scanning spot sizes must be considered.

  • 3. Scanning sources resemble repetitive pulses, rather than single pulse. The additivity of thermal effect depends on temporal temperature characteristics of the scanning source, which must be considered.

  • 4. In addition to temporal temperature characteristics, the temporal change in thermal damage threshold must be considered.

  • 5. Local heat buildup may cause local hotspots in scanning pattern. For scanning devices with two-dimensional (2-D) scanning patterns, the entire scanning area must be considered to find hotspots caused by local heat buildup.

In this study, questions (1)–(4) will be addressed using the MGLM. Question (5) needs rigorous computational calculation, and thus will be discussed qualitatively without numerical calculation results. The main goal of this study is to find the difference (or similarity) between the thermal damage thresholds of linear scanning source and pulsed source. If the thermal damage threshold of scanning source is not significantly different from that of pulsed source, pulsed source criteria can be applied to the evaluation of photothermal effect from scanning source. Actual values of damage thresholds slightly vary depending on the calculation parameters used and will not be discussed in this study because this study aims for numerical calculation of relative differences in thermal damage thresholds between scanning and pulsed sources.

2.

Method

The MGLM is a modified version of melanin granule model4 using uniformly distributed melanosomes in retinal-pigmented epithelium and has been proven to be useful for studying photothermal damage of retinal tissue within its limits.2 The results using MGLM are similar to those using other computational models especially when the pulse duration or dwell-time is longer than 10 μs and shorter than 1 s.2,4

In this study, MGLM was adapted for calculating temperature and thermal damage threshold from scanning sources. Temperature at a given location (r) and time (t) can be analytically calculated using temperature function T(r,t), which depends on the thermal diffusivity of the tissue (D), thermal conductivity (k), granule radius (a), heat capacity (cp), absorption coefficient of melanosome (αm), and melanosome number density (ρm). Parameters for calculations are summarized in Table 1.2 The radiant exposure at the cornea (Ic) was set to 10μJ/cm2, and the retinal radiant exposure (I0) was approximated by multiplying Ic by a factor of 105.4 Each melanosome shown as filled circles in Fig. 1 has radius (a) 1.0 μm, and the melanosomes were placed in close proximity with each other. Melanosome number density (ρm) is the inverse of the volume containing one melanosome, which is a cube with 2.0-μm sides. A total of five layers of melanosomes without gap between each layer were used in this study. More details can be found elsewhere.2

Table 1

Summary of parameters for calculation.

VariableSymbolValue
Thermal conductivityκ5.0323×103J/cm/K/s
Tissue densityρ1.0g/cm3
Melanosome radiusaμm
Heat capacitycp4.186J/g/K
Melanin absorption coefficientαm2000/cm at 532 nm
Melanosome densityρm1.25×1011/cm3
Frequency factorlnA228.22
Activation energyEa627.6kJ/mole

Fig. 1

Illustration of melanosomes (gray circles) and scanning light source. Irradiated melanosomes are marked with darker gray circles.

JBO_19_4_045004_f001.png

The Arrhenius integral calculations were performed on locations z=1.5μm above the top melanosome layer, to avoid excessively high temperature inside the melanosome itself while obtaining lowest damage thresholds possible. Also, different from the previously reported results, restricting aperture of size D has been removed in this study. The calculation without restricting aperture can be simply implemented by enlarging the size of the restricting aperture within which scanning light source is moving. Figure 1 illustrates laser spot sizes (dx,dy) and scanning lengths (Dx,Dy) used for the calculation. Scanning lengths in x- and y-directions, Dx and Dy, can be considered as restricting aperture with same dimensions. Also, the spot size d was varied from 30 μm, which was the only value used in the previous study. Square tophat source profiles were used in this study due to its advantage over circular spots: Circular and Gaussian source profiles, although they are realistic profiles of actual optical devices, exaggerate geometrical effect from the source itself.2 Calculations were performed for different scanning speeds which correspond to 0.1, 1, 10, and 100 ms of dwell time. All dwell-times are well within the range where MGLM produces calculation results that are comparable to those found using other numerical methods.

3.

Numerical Calculations

3.1.

Photothermal Damage without the Aperture

As shown in Fig. 1, when a scanning source with size dx and dy is moving in x-direction, the range of melanosomes that are irradiated by the scanning source can be specified by Dx and Dy. Setting Dy equal to dy, the scanning is restricted to a linear scan. When Dx equals to dx, the scanning irradiation is restricted only within the aperture of size equal to the source spot size. When Dx is larger than dx, the scanning source is irradiated over a wide range of retinal tissue. Figure 2 shows the damage thresholds for 30×30μm2 tophat laser spot scanned for different Dx. The damage thresholds for scanning speeds corresponding to 0.1-, 1-, and 10-ms dwell times are almost identical to those of scanning laser with (Dx=30μm) and without aperture (Dx>30μm). This is in agreement with the previous findings that the damage thresholds for scanning sources do not differ from those for pulsed sources,2 and it further confirms that the discrepancy remains insignificant even without restricting aperture. The damage threshold decreased from 12.55×105J/cm2 to 12.07×105J/cm2 for the 100-ms dwell time, which is a slightly greater decrease compared with shorter dwell times. The longer the dwell time, the more thermal energy is deposited on the irradiated volume of tissue, and thus requires less energy to induce photothermal damage. However, a 4% decrease in damage threshold for 100 ms is still not significant. Considering the fact that a 100-ms dwell time corresponds to a scanning speed of 0.3mm/s for a 30-μm beam spot, which represents a lower limit on the scanning speed for practical imagers, consideration of a larger decrease in damage threshold for dwell time larger than 100 ms would not be practical. As a matter of fact, some modern fast scanners used in medical imaging devices can achieve frame rate as high as 60 frames-per-second, which produce 1000pixels×1000pixels images in total 1×1mm2 image sizes. The dwell time of such devices is as short as 500 ns, at which photothermal damage from scanning and pulsed sources are equivalent.

Fig. 2

Damage threshold of scanning source calculated using melanin granule lattice model (MGLM) for different scan lengths with 30-μm laser spot.

JBO_19_4_045004_f002.png

3.2.

Spot-Size Dependence

As described in the previous section, a small difference in thermal damage thresholds between scanning and pulsed sources exist and is mainly caused by the dynamic dissipation of moving heat generated by scanning light source. The greater the heat stored in the tissue, the more the effect of dynamic heat dissipation contributes toward thermal damage. Previous calculation results used a 30-μm spot size because this is considered to be the smallest achievable focal spot size at the retina without incorporating adaptive optics.4 Larger spot sizes must be considered because not all ophthalmic devices generate diffraction-limited spot on the retina, and they may generate a larger discrepancy in thermal damage thresholds between scanning and pulsed sources. For numerical calculation, larger dx (>30μm) was used and more dramatic change in damage thresholds were observed from increased spot size.

Figure 3 shows the damage thresholds for square laser spot of size dx (=dy) scanned over square aperture of size Dx (=Dy). The damage thresholds for 0.1- and 1-ms dwell-time were almost identical regardless of dx and Dx. However, damage threshold decreased from 12.55×105 to 7.42×105J/cm2 for 100-ms dwell time as the dx (=Dx) increased from 30 to 120 μm. This represents a 40% decrease in damage threshold. The damage threshold curve in Fig. 3 can be approximately fitted to (1/dx) curve. The (1/dx) dependency of the thermal damage threshold is similar to what was observed for spot-size dependency of retinal irradiations.5,6 The spot-size dependence study of scanning source also suggests that thermal damage threshold of larger scanning spots demonstrate similar tendency in spot-size dependence of pulsed source.

Fig. 3

Damage threshold of scanning source calculated using MGLM for different spot sizes.

JBO_19_4_045004_f003.png

3.3.

Temporal Thermal Characteristics

Scanning sources resemble repetitive pulses, rather than single pulse. It is particularly true for scanning sources in imaging medical devices, such as those used in scanning laser ophthalmoscopy and optical coherence tomography. For repetitive pulse irradiations of N pulses, the correction factor N1/4 was adopted in many consensus standard documents. Clark et al.7 recently performed a comprehensive study on the multiple-pulse thermal damage thresholds of the retina and pointed out the importance of interacting pulses for temperature rise and thermal damage threshold. For short pulses with the pulse duration in the range of 100 μs, the temperature from a single pulse does not reach steady state, thus the damage threshold not only depends on N but also depends on the duty cycle. When the duty cycle is high, the pulses are similar to a single long pulse with the total pulse duration approximated by total-on-time (TOT). When the duty cycle is low, the pulses are noninteracting thus the temperature returns to the initial value before each new pulse arrives, giving the highest damage threshold. Multiple-frame scanning irradiation resembles repetitive pulse irradiation, because multiple repetitive optical exposures are delivered to a certain location. It is very important to know whether temporal temperature change from scanning irradiation is similar to that of pulsed irradiation.

Figure 4 shows the numerical calculation results on normalized temperature rise versus normalized time for a 30-μm square spot both for scanning and pulsed sources. As can be seen in Fig. 4(a), the temporal temperature rise of a scanning source is quite different from that of a pulsed source shown in Fig. 4(b) for certain dwell times. In scanning, the normalized time origin (t=0) indicates the moment where laser spot started to irradiate a certain location. Negative normalized time means that the laser spot did not reach the specific location yet. The scanning light source itself is not turned on and off like a pulsed source, thus the temperature at time origin starts to rise before the light source reaches the region. The temporal temperature characteristic is similar for scanning and pulsed sources with a 0.1-ms duration, whereas it is dramatically different for the 100-ms duration. Although the temperature reaches steady state for a 100-ms pulsed source, it widely spreads out into negative-normalized time range for scanning source. A similar calculation was performed for larger spot sizes. The larger the irradiated area, the longer it takes for the temperature to reach steady state. Figure 5 shows normalized temperature rise versus normalized time for 300-μm square spot both for scanning and pulsed sources. Scanning source showed temporal temperature rise similar to that of a 30-μm spot, as can be seen in Fig. 5(a). However, pulsed source exhibited quite different response compared with a 30-μm spot, showing no steady state in Fig. 5(b).

Fig. 4

Normalized temperature rise as a function of normalized time for 30-μm square source spot for (a) scanning and (b) pulsed sources. Temperature is normalized to the peak temperature, and time is normalized to the pulse duration.

JBO_19_4_045004_f004.png

Fig. 5

Normalized temperature rise as a function of normalized time for 300-μm square source spot for (a) scanning and (b) pulsed sources. Temperature is normalized to the peak temperature, and time is normalized to the pulse duration.

JBO_19_4_045004_f005.png

Numerical calculation of temporal temperature response of scanning irradiations exhibited almost no steady state even for long dwell time of tau~100ms. Thermal additivity of longer pulses is the key mechanism of reduced thermal damage threshold of repetitive pulses, and thermal additivity is stronger when the temperature reaches steady state within single pulse.7 Lack of steady state temperature for scanning irradiation means that thermal additivity is negligible, thus thermal damage threshold may not be lower than that of repetitive-pulsed irradiation. The limits of such noninteracting source and the CW source coincide at the pulse duration of TOT.7

3.4.

Additivity of Thermal Effect

Evaluation of photochemical effects for pulsed sources mainly rely on total energy (dose) delivered to the tissue. Careful measurement or calculation of radiant exposure of scanning source with consideration of overlap and dwell time variation will suffice for evaluating photochemical effect. Photothermal effect from repetitive-pulsed irradiation cannot be evaluated by only measuring (or calculating) total radiant exposure because deposited heat will not dissipate into another tissue volume immediately. In addition to the temporal temperature characteristics considered in the previous section, additivity of thermal effect needs to be considered for photothermal effect. For pulsed sources, additivity of thermal effect, thus additivity of Arrhenius integral, has already been discussed in great detail in others’ work.7 For scanning sources, spatial dependence must also be considered because scanning sources are basically moving heat sources.

Shown in Fig. 6(a) is an illustration of a linear scanning source with scan length L and spot size d. Location 1 and Location 2 indicate the mid-point and turning-point of scanning. One period (T) of scanning is the time duration for the source to leave Location 2 and to return back to Location 2. Dwell time (t) is the time duration for the source to irradiate the local tissue area, which can be obtained by dividing spot size (d) by scanning speed (v). Assuming that the scanning speed (v) is constant, there are two extreme schemes for scanning exposure, both of which are illustrated in Fig. 6(b):

  • 1. At Location 1, scanning exposure is delivered every T/2 with each dwell time of t. Duty cycle of this repetitive exposure is t/(T/2)=2t/T.

  • 2. At Location 2, scanning exposure is delivered every T with each dwell time of 2t. Duty cycle of this repetitive exposure is 2t/T.

Fig. 6

(a) Illustration of linear scanning source of scanning length L and spot size d. Location 1 and Location 2 indicate the mid-point and turning-point of scanning, respectively. Solid line represents the laser scanning path (for the eye); (b) time-lapsed intensity measured at Location 1 and Location 2.

JBO_19_4_045004_f006.png

The main differences are: the period of (1) is half that of (2), and the dwell time of (2) is twice as long as (1). Case (1) imposes a higher probability that the next exposure is delivered before the temperature returns to original, whereas case (2) imposes a higher probability to reach damage threshold due to the longer dwell time.

Practical scanning devices of recent design, especially imaging devices, typically have at least 500pixels×500pixels of scanning dimension. This implies that one scan length (L) is at least 500 times larger than that of one spot size (d). At constant scanning speed, this also implies that T is 1000 times larger than t because T=2L/v=2(500d)/v=1000t. For both (1) and (2), the duty cycle is small enough so that each exposure can be treated independently. It is more probable that the longer exposure time of (2) will produce a larger Arrhenius integral value than the shorter exposure interval of (1) will give additivity in the Arrhenius integral. Sinusoidal scanning generates a dwell time at Location 2 much longer than 2t, thus evaluating the photothermal effect at turning points seems to be a reasonable worst case scenario analysis.

The above statement may not be valid when the scanning length is very small compared with the spot size, although such scanning devices are not practical. Additivity of thermal effect for such special scanning devices can be analyzed using the Arrhenius integral. For a pulsed (or scanning) irradiation of total duration τtot, the Arrhenius integral is

Eq. (1)

Ω(τ)=0τtotAe[EaRT(τ)]dτ,
where A is a material parameter (frequency factor), Ea is the activation energy, T(τ) is the temperature at time τ, and R is the universal gas constant. If this integral reaches unity during or shortly after irradiation, then thermal damage to the skin has occurred. When there is repetitive independent pulse irradiation with each Arrhenius integral less than unity, then this type of repetitive pulse irradiation will never exceed the damage threshold. However, if repetitive pulse irradiation is not independent, Eq. (1) can be expanded into multiple individual integrals:

Eq. (2)

Ω(τ)=0τtotAe[EaRT(τ)]dτ=0τ1+τ2++τnAe[EaRT(τ)]dτ=0τ1Ae[EaRT1(τ)]dτ+0τ2Ae[EaRT2(τ)]dτ++0τnAe[EaRTn(τ)]dτ.

Adaptation of different temperature functions T1(τ),,Tn(τ) in the integral reflects the fact that the starting temperatures for individual pulses are different due to insufficient cooling between pulses. Arrhenius integrals for cases (1) and (2) can be expressed using Eq. (2):

Eq. (3)

Ω(1)(τ)=1n0tAe[EaRTi(τ)]dτ,

Eq. (4)

Ω(2)(τ)=1n/202tAe[EaRTi(τ)]dτ.

For such special case of scanning irradiations, measurement (or calculation) of worst-case-scenario radiant exposure is to be performed at Location 1 of Fig. 6(a) if Eq. (3) is larger than Eq. (4), or at Location 2 if Eq. (4) is larger than Eq. (3).

3.5.

Local Heat Buildup

In 2-D scanning, heat is not only dissipated in x-direction dominantly, but also is dissipated in y-direction. Also, overlap of EPs greatly differs depending on the scanning scheme. Local hotspot may exist due to local heat buildup in y-directional moving heat source, and it requires rigorous computational power to calculate such 2-D heat distribution and dissipation from 2-D scanning device. This is an area for future investigation.

4.

Conclusion

Using a numerical method, MGLM, the photothermal effect from scanning and pulsed sources was calculated and compared. The results confirmed previous investigations, which produced similar thermal damage thresholds for both scanning and pulsed sources restricted in measurement aperture. These results also showed that the thresholds for both types of sources are not significantly different when the source is scanned without any restricting aperture. Various spot-sizes were also considered, and the calculated thermal damage thresholds demonstrated inverse dependence on spot size, which is similar to the pulsed source case. Temporal temperature characteristics showed that scanning sources do not reach steady state in temperature, thus each exposure can be treated independently. Such noninteracting pulses can be regarded as single long pulses using TOT pulse duration for thermal damage. Analysis of additivity of thermal effect at the mid-point and turning-point of scanning suggests that the worst case scenario (higher photothermal damage) is observed at the turning-point, unless the scanning length is impractically small compared with the spot size. Further study is needed to calculate the heat buildup from a 2-D scanning source and to address effect of overlap of EPs.

Acknowledgments

The author thanks Drs. Dexiu Shi, Jeeseong Hwang, Bruce Stuck, Robert Landry, and David Sliney for valuable discussions. This work was supported in part by Critical Path Initiative (CP 21 and CP 6) of US Food and Drug Administration. The mention of commercial products, their sources, or their use in connection with material reported herein is not to be construed as either an actual or implied endorsement of such products by the Department of Health and Human Services.

References

1. 

(2007). Google Scholar

2. 

D.-H. Kim, “Using a melanin granule lattice model to study the thermal effects of pulsed and scanning light irradiations through a measurement aperture,” J. Biomed. Opt., 16 (12), 125002 (2011). http://dx.doi.org/10.1117/1.3656746 JBOPFO 1083-3668 Google Scholar

3. 

D.-H. Kim, “Evaluation of phototoxicity from scanning biophotonic devices,” Proc. SPIE, 7894 78940H (2011). http://dx.doi.org/10.1117/12.892349 PSISDG 0277-786X Google Scholar

4. 

C. R. Thompsonet al., “Melanin granule model for laser-induced thermal damage in the retina,” Bull. Math. Biol., 58 (3), 513 –553 (1996). http://dx.doi.org/10.1007/BF02460595 BMTBAP 0092-8240 Google Scholar

5. 

K. Schulmeisteret al., “Ex vivo and computer model study on retinal thermal laser-induced damage in the visible wavelength range,” J. Biomed. Opt., 13 (5), 054038 (2008). http://dx.doi.org/10.1117/1.2982526 JBOPFO 1083-3668 Google Scholar

6. 

K. Schulmeisteret al., “Review of thresholds and recommendations for revised exposure limits for laser and optical radiation for thermally induced retinal injury,” Health Phys., 100 (2), 210 –220 (2011). http://dx.doi.org/10.1097/HP.0b013e3181ea51e3 HLTPAO 0017-9078 Google Scholar

7. 

C. D. Clark IIIW. J. MarshallR. J. Thomas, “Theoretical analysis of multiple-pulse thermal damage thresholds of the retina,” J. Laser Appl., 25 (1), 012005 (2013). http://dx.doi.org/10.2351/1.4788755 JLAPEN 1042-346X Google Scholar

Biography

Do-Hyun Kim finished his PhD in solid state physics in 2000 at Seoul National University in the Republic of Korea. He finished his second PhD in electrical engineering with biomedical optics as the research subject at Johns Hopkins University in 2006. He has been performing research on biomedical optics with special interest in optical radiation safety and optical microscopy at the US Food and Drug Administration since 2006. He is a senior member of SPIE.

© 2014 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2014/$25.00 © 2014 SPIE
Do-Hyun Kim "Consideration of dynamic photothermal effect for evaluation of scanning light sources in optical devices using pulsed source criteria," Journal of Biomedical Optics 19(4), 045004 (25 April 2014). https://doi.org/10.1117/1.JBO.19.4.045004
Published: 25 April 2014
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Laser damage threshold

Photothermal effect

Light sources

Optical components

Thermal effects

Tissue optics

Tissues

Back to Top